YANTE YT-CA725 Dirt Count Tester paper Dirt Count Tester
Using 20W fluorescent light source and 360 ° swirl/spin table design, the standard dust map area is 0.05~ 5.0mm ², which meets the inspection requirements of GB/T1451-1989 and is easy to operate.
YanTeIntroduction
YT-ZCA725 Paper Dust Tester is manufactured according to the standard GB/T1451-1989 "Measurement Method of Paper and Cardboard Dust", which is professionally used to measure the dust of paper surface . With reasonable structure and convenient operation, it is widely used in the field of testing paper materials in universities, scientific research units, quality inspection departments , and paper printing and packaging production departments.
Installation and Precautions
This instrument cannot be used in flammable and explosive environments.
The laboratory should maintain a temperature of 23±1°C and a Relative Humidity of 50±5%RH. There should be no vibration, no electromagnetic interference, and slight air flow in the room. There should be no vibration source around the installation environment of the instrument.
YanTe YT-CA725 Dust MeterSpecifications
| LIST | VALUE |
|---|---|
| Light source | 20W fluorescent lamp |
| irradiation angle | 60° |
| Workbench size | 270 * 270Mm, swirling/spinning 360 ° |
| Standard dust chart area | 0.05~5.0mm² |
| Dimensions | 428*350*250mm |
| Weight | 12.5Kg |
| Power Supply | AC 200~240 (wide Voltage range), 50~60Hz |
| Sampling | 4 or more 250mm × 250mm samples |
| meet the standard | GB/T1451-1989 |
| Operating Environment | Temperature 23 +/- 1 ℃, Relative Humidity 50 +/- 5% RH |
| Workbench size | Effective area 0.0625 m2, can swirl/spin 360 ° |
YanTe YT-CA725 Dust Meter Packing list
host X1, manual X1, certificate X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- Dust meter detection core and quality control essence
- Principles and steps for dust determination of paper and paperboard
- Determination Method and Standardized Process of Paper and Pulp Dust
- Pulp - Determination of dust content
- What performance requirements should printing paper meet?
- ActiveQB/T 3524-1999
- ActiveQB/T 2693-2005
- ActiveQB/T 2249-1996
- ActiveGB/T 10335.5-2008
- ActiveGB/T 1541-2013
- AbolishGB/T 10335.3-2004


YT-CA725