| Test item | Low concentration COD, high concentration COD |
| Measuring method of low concentration COD | Potassium dichromate method |
| Low concentration COD Measurement range | (0.000-150.0)mg/L |
| Low concentration COD error | ±8% |
| Repeatability of low concentration COD | ≤3% |
| Low concentration COD Stability | <6mg/L(20min) |
| Low concentration COD wavelength | 470nm |
| Low concentration COD digestion | Need to be digested |
| Measuring method of high concentration COD | Potassium dichromate method |
| Measurement range of high concentration COD | (150.0-1500) mg/L, can be expanded to 15000mg/L |
| High concentration COD error | ±8% |
| High concentration COD Repeatability | ≤3% |
| High concentration COD Stability | <6mg/L(20min) |
| High concentration COD wavelength | 620nm |
| High concentration COD needs to be digested | Need to be digested |
| Absorption Measurement range | (0.000~2.000)Abs |
| Absorption error of indication | +/- 0.01 or +/- 5% |
| Absorption Repeatability | ≤ 0.005 or 2% |
| Absorption Measurement range | (0.000~2.000)Abs |
| Wavelength measurement wavelength | 420nm、620nm |
| Wavelength Precision | ±4nm |
| Transmittance error of indication | ±2% |
| Transmittance Repeatability | ≤ 0.5% |
| data storage | A total of 2000 test results |
| Measurement cell | 10Mm Colorimetric dish, 16mm Colorimetric tube |
| Power Supply | Power adapter, input AC: (100~240) V, output DC24V |
| Dimensions (mm); Weight (kg) | 302×205×122,1.33 |