LABTOP icooler- 8005 low temperature chilled down liquid circulating machine -80 ℃/5L
Temperature range -80~ RT ℃, Temperature Fluctuation +/- 0.5 ℃, LED display accuracy 0.1 ℃, PID intelligent control, PT100 temperature sensor. 5L volume.
labtop-
NELE DL-1020 Circulating Chiller 20L/-10 ℃
$ 1607.00 -
CNSHP DLSB-100/30 Circulating Chiller -30 ℃/100L
$ 8793.00 -
HUXI HLX-2017 low temperature coolant Circulating Chiller 17L/-20~40 ℃
$ 2061.00 -
LEPTO icooler-2003 Low temperature chilled down liquid circulating machine -20 ℃/3L
Login -
CNSHP DLSB-6/20 Circulating Chiller -20 ℃/6L
$ 977.00
Introduction
iCooler low-temperature coolant Circulating Chiller, two-stage compressor refrigeration, maintenance-free pressure suction pump, can carry out external closed circulation, can also carry out external open circulation, the refrigeration system adopts proportional refrigeration technology, the system energy consumption is low, and the work is stable. It is suitable for a variety of applications that require cryogenic liquids and cold traps externally.
Features
Maintenance-free pressure/suction pump, resistant to fiber and metal particles, long service life, pump pressure and flow automatically adapt to external loads, effectively protect external equipment, especially glass parts.
The refrigeration function and the circulation function can be controlled independently or simultaneously.
Sealed Water Tank with intuitive level window for easy user observation of the liquid level, while the tank is equipped with a protective overflow port to prevent accidents.
Stable and reliable system: high-performance and high-quality spare parts, stable quality and reliable performance.
When the alarm is alarmed, the light and sound are reminded at the same time, and various alarm codes are directly displayed, which can directly determine the error point.
Pt100 temperature sensor for single- and two-point temperature calibration.
A variety of operating modes are available, and unexpected power failure can be automatically resumed, which can ensure the continuous supply of cooling circulating water.
Optional intelligent control function: a variety of interfaces (RS485, RS232), which can remotely control the start/stop of the circulating water pump and refrigeration system.
The input of 0-10V or 4-20mA analog signal can be selected, which can be used for the start and stop of the external intelligent device control instrument.
Waterproof keyboard design, high-brightness LED display.
Environmentally friendly refrigerant, no toxic pollution.
The maximum and minimum temperature settings of the machine can be defined.
The buzzer can be set to silent or audible.
The serial number of the device can be checked through the software.
Serial models
| Model | Temperature range | Cooling capacity | Water tank volume |
| MiniCooler | -5~RT | 200 | 1 |
| iCooler-2003 | -20~RT | 450 | 3 |
| iCooler-2006 | -20~RT | 700 | 4.5 |
| iCooler-4006 | -40~RT | 800 | 4.5 |
| iCooler-2009 | -20~RT | 1200 | 9 |
| iCooler-4009 | -40~RT | 1500 | 9 |
| iCooler-2017 | -20~RT | 2400 | 17 |
| iCooler-3017 | -30~RT | 2400 | 17 |
| iCooler-8005 | -80~RT | 400 | 5 |
| Units: °C, W, L/min, L | |||
labtop iCooler-8005 Low temperature chilled down liquid circulatorSpecifications
| LIST | VALUE |
|---|---|
| Temperature range | -80~RT℃ |
| Temperature Fluctuation | ±0.5℃ |
| Temperature display | LED |
| Temperature display Accuracy | 0.1℃ |
| Temperature Control mode | Smart PID |
| Temperature sensor | Pt100 |
| Cooling Capacity (20 ℃) | 400W |
| Maximum pump pressure | 0.2bar |
| Flow rate of pump | 22L/min |
| Pump connection interface | ID=8 |
| Tank Volume | 5L |
| Dimensions | 355*455*865mm |
| Instrument fixing method | floor type |
| Allow continuous operation | 100% |
| Environmental temperature | 5~32℃ |
| Allowable Relative Humidity | 80% |
| Pump type | Pressure/suction pump |
| Maximum suction of pump | 0.1bar |
labtop iCooler-8005 Low temperature chilled down liquid circulator Packing list
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- Thermostatic circulator external open bath application
- Technical characteristics and application of iCooler small cryogenic coolant Circulating Chiller


iCooler-8005