FLOM IFBZ2002-UP Two-stage reverse osmosis (RO) ultra-pure water machine, reagent type 20L/H
FLOM reverse osmosis (RO) ultrapure water machine IFBZ2002-UP is a two-stage reverse osmosis (RO) laboratory ultrapure water machine, equipped with ultrapure water circulation program design, IFBZ2002-UP reverse osmosis (RO) ultrapure water machine has a water production capacity of 20L/H, using LCD touch screen.
FLOM-
BIO0021 Puro10 Laboratory Water Purification System, three-stage water 10L/H water storage tank 30L
$ 5515.00 -
BIO0021 Puro50 Laboratory Water Purification System, three-stage water 50L/H water storage tank 30L
$ 7999.00 -
TIANMA UPTM-B100 Vertical ultrapure water mechanism Water volume 100L/H
$ 2984.00 -
TIANMA UPTM-C40 Vertical ultrapure water mechanism Water volume 40L/H
$ 4246.00 -
TIANMA UPTM-A10 Experimental Benchtop Ultrapure Water System
$ 739.00
Characteristics:
1. LCD touch screen control systemThere is self-diagnosis of abnormal operation (raw water, pure water, ultrapure water quality and quantity), and the warning function has the function of displaying the service life of consumables and membranes and restoring the reminder function;
2. The unique membrane treatment technology and accessories in the United States can adapt to different municipal tap water in various places;
3. The automatic, manual and timed flushing function of the reverse osmosis system can effectively protect and prolong the service life of the RO membrane;
4. Fully enclosed management design, avoid secondary pollution, keep water quality stable, and one-key internal disinfection and pressure relief;
5. Automatic continuous water production, real-time operation status display, adjustable operating parameters;
6. The water production system and the circuit system are completely separated, which effectively eliminates potential safety hazards;
7. Chinese and English language switching function (factory default is Chinese), more in line with your language habits;
8. The equipment conforms to the national laboratory water standard (GB/T6682-2008), Chinese national standard for electronic grade ultrapure water (GB/T11446.1-1997), and the American test grade water (CAP, ASTM, NCCLS) standards;
9. The company has passed the EU CE certification and ISO9001-2008 national quality management system certification;
Applications:
Laboratories for universities, scientific research, food, petrochemical, chemical preparation, cosmetics, electronic power, etc
FLOM IFBZ2002-UP Laboratory ultrapure water machineSpecifications
| LIST | VALUE |
|---|---|
| Product name | IFBZ standard reagent type |
| reverse osmosis (RO) | double stage |
| Flow Rate | 20L/h |
| Inlet source | Municipal tap water (conductivity < 200PPM) |
| source water pressure | 0.15 Mpa~0.5Mpa |
| Inlet temperature | 5-40℃ |
| conductivity | 1-5μS/cm |
| resistivity of ultrapure water | ≥18.25MΩ.cm |
| Total organic carbon removal rate TOC | <10ppb |
| Ion removal rate of inorganic matter | >99% |
| bacterial removal rate | >99% |
| Endotoxin content | <0.003EU/ml |
| Soluble Silicon (SiO2) | <0.01mg/L |
| Resistance pool sensitivity constant | 0.01cm-1 |
| bacteria | <1CFU/ml |
| Flow velocity | 1-2L/min |
| desalination rate | ≥99% |
| Regular quantitative water intake | 1mL-10L, the value can be adjusted arbitrarily, the Accuracy error is 1% |
| Dimension | 500*500*600mm |
| Weight | 40kg |
FLOM IFBZ2002-UP Laboratory ultrapure water machine Packing list
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- Detailed purchase method of laboratory ultra-pure water machine [dry goods]
- What is the difference between "water intake flow" and "water intake flow rate" in laboratory ultrapure water machines?


IFBZ2002-UP





