Qiwei ORP-2 Portable ORP Meter
Qiwei ORP-2 portable ORP meter is a high-precision instrument for laboratory measurement of mV value, with dual functions of MV temperature, standard ORP electrode and temperature electrode, ORP-2 portable ORP meter measuring range mV: (0~ ± 1999) mV, resolution 1mV, temperature range (0~60) ℃.
QiWei-
General TDSscan20 pen TDS meter, multi-range pen TDS meter
$ 150.00 -
TONGAO TR 6900PN Water Quality Analyzer for Total phosphorus/Total nitrogen/turbidity/suspended solids
$ 3149.00 -
LEICI bromine ion electrode calibrated solution, suitable for bromine ion composite electrode
$ 106.00 -
JINGQI WD9101P Precision Ammonia Nitrogen Tester with Printing Function
$ 652.00 -
HANNA HI764080 Digital dissolved Oxygen electrode (built-in temperature sensor)
$ 817.00
Applications
The digital display ORP meter is a high-precision instrument for measuring mV values in the laboratory. It has dual functions of MV temperature, standard ORP electrode and temperature electrode, fast response and good stability. The instrument is suitable for industrial and mining enterprises, scientific research, agriculture, medicine, teaching, environmental protection and other units to measure the electrode potential (mV value) of aqueous solution.
operation method
1. Standard reagent configuration method:
Add 60ML distilled water to one package of standard reagents, fully dissolve and stir, and it can be used after 30 minutes.
2. Instrument Calibration
Switch to the calibration mode, wash the electrode with distilled water and dry it. Put in the calibration solution 256mV, stir to make it fully contact, after the reading is stable, press the confirm key, and the calibration is completed.
3. Instrument measurement
Switch to the measurement mode, put the electrode into the solution to be tested, stir to make it fully contact, after the reading is stable, press the confirmation key, and the value displayed on the screen is the correct reading of the solution to be tested.
Electrode use and precautions
1. Storage of ORP electrode: After the electrode is cleaned, put it into a saturated solution of KCL.
2. Cleaning of the ORP electrode: If the platinum head of the electrode is found to be dirty and the measured value in the standard solution of PH=4 exceeds the standard value, it should be cleaned.
QiWei ORP-2 Portable ORP meterSpecifications
| LIST | VALUE |
|---|---|
| Measurement range | mV:(0~±1999)mV |
| resolution | 1mV |
| Measurement range | 0~60℃ |
| basic configuration | ORP electrode, [suitable for solution temperature (5~ 60) ℃ range measurement] battery; temperature electrode; calibration reagent |
| Measurement parameter | ORP |
QiWei ORP-2 Portable ORP meter Packing list
complete-built unit X1, manual X1, certificate of conformity X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.


ORP-2