Search
    Cart(0)
      总价
      前往结算
      您的购物车是空的 : )
      Home> Rotational Viscometer> Fungilab> Fungilab V-pad L Rotational viscometer

      Fungilab V-pad L Rotational Viscometer 6 million cP

      Fungilab V-pad L Rotational Viscometer Measurement range 1-6000000cp, Speed range 0.01-250rpm, Repeatability 0.20%, V-pad L Rotational Viscometer has Self-check function and sound and light alarm reminder.

      Brand
      SKU: NB035125
      Spain Fungilab
      Stock
      Loading…
      Price
      $ 7657.00
      Promotion
      Service
        Service
        The delivery and service By NBchao.Com
        VS
        Genuinegoods
        • Fungilab V-pad L Rotational viscometerOverview

        • Specifications
        • Comments
        • FAQs
        V-pad L
        Rotational viscometer

        Data Display

        • Speed: rpm or S-1

        • Viscosity: cP, P, mPa.s, Pa.s, cSt, St, mm²/s

        • Percent of full scale: %

        • Sample temperature: ℃ or OF

        • Shear speed (SR): S-1

        • Shear stress (SS): N/m2, Pa, Dyne/m2

        • Density (customer input required): g/cm³ or Kg/m³

        • Note: Shear stress and shear rate can only be displayed when the coaxial rotor L2C+L3C is used.

        Viscosity measurement 

        • Dynamic viscosity (cp or mPas)

        • Kinematic viscosity (cSt) (before viscosity measurement, the customer needs to enter the sample density value in advance)

        program measurement

        • Multi-step program measurement mode can be used and edited;

        • Programs can be stored for quick recall later;

        • You can choose TTT (stop measuring when reaching a certain torque) and TTS (stop measuring when reaching a certain time period) measurement mode, multi-step measurement and slope measurement.

        data storage

        • All test data will be stored in a comprehensive database;

        • Can activate or cancel quality detection line (QCL) alarm and over-range alarm;

        • Users have access to program parameters for each measurement experiment.

        Customer customized 

        • Customers can create measurement speeds (0.01 to 250rpm);

        • Can activate or cancel quality detection line (QCL) alarm and over-range alarm;

        • Customers can customize the horizontal and vertical coordinates on the measurement curve.

        Self-checking function : Viscometer self-checking function, sound and light alarm reminder

        Range function : After selecting a combination of rotor and speed, the maximum range under this combination will be automatically calculated.

        Tablet control : All parameters can be modified via access via an 8-inch tablet, and the mobile app is available for free in the Google Play Store

        Language : Multiple languages ​​available

        QCL : Select and set up sample quality detection line

        IAM : Data measurement and upload times are adjustable

        Graph mode : displays experimental data with real-time graphs

        Report export : Test results are exported in PDF and CSV formats

        Calibration : Allows the user to enable viscosity and temperature calibration (this calibration will be reset to factory calibration)

        Offset : Adjust temperature by offset

        Heldal (only available in V-COMPACT): Helical viscosity measurement mode, suitable for measuring non-Newtonian fluid samples

        Accessories : Small sample adapter (APM) and low viscosity adapter (LCP) optional

        PPR : Rotor quick disassembly and assembly device

        Optional

        Small sample volume sample adapter APM, low viscosity sample adapter LCP, high viscosity sample adapter Heldal, standard silicone oil, Water Bath, universal adapter, PPR quick release system (V-PAD optional), double-layer temperature-controlled beaker (V-PAD optional) ), blade rotor (yield stress), cone-plate system.

        Parameter comparison

        Order numbermodelIncluding PPRMeasuring range(cp)
        VP100003V-PAD LVP1010031*-6000000
        VP100002V-PAD RVP10100225**-40000000
        VP100001V-PAD HVP101001200***-320000000
        Other parameters
        range of rotation0.01-250rpm

        * It is recommended to use an LCP adapter when the viscosity range is from 0.1 to 200RPM; the minimum viscosity measured without a matching LCP adapter is 15CP

        ** It is recommended that the rotation speed of R-1 rotor is 60rpm (higher rotation speed measurement will be unstable)

        *** The recommended rotation speed for R-1 rotor is 60rpm

        Resolution according to viscosity range measurement

        Viscosity valueresolution
        100cp0.01cp
        1.000cp0.1cp
        5.000cp1.0cp
        10.000cp5.0cp
        25.000cp10.0cp
        50.000cp25.0cp
        100.000cp100.0cp
        1000.000cp1000.0cp


        VS

        Fungilab V-pad L Rotational viscometerSpecifications

        LIST VALUE
        Measurement range 1~6000000cp
        Rotor speed 0.01~250rpm
        Repeatability ±0.2%F.S
        Measurement Accuracy ±1%F.S
        Temperature range -40~+300℃
        Temperature Resolution 0.1℃
        Temperature Accuracy ±0.3℃
        temperature sensor PT100, Class A
        Rotor material All rotors are made of AISI 316 stainless steel
        Power supply Voltage 90-240VAC,50 / 60HZ

        Fungilab V-pad L Rotational viscometer Packing list

        machine X1, manual X1, certificate X1, warranty card X1

        [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

        FAQ

        [Q]
        What are the differences between these two, and are they both suitable for testing pulp viscosity?
        [A]
        The two parameters are basically the same, both are 0~ 2600 rpm. The main difference is that its structure is different. EVO is a traditional Viscometer. If you need a computer to use software, you can control the instrument on the computer. Linear dispersion can be done for many years. V-pad is a host, no display screen, no buttons, it is a relatively novel design device, it uses pad to control measurement, Bluetooth connection, experimenters can walk around, there is a certain convenience, but the price is slightly more expensive.
        Fungilab
        Fungilab V-pad L Rotational Viscometer 6 million cP
        Service