INESA SHANGFEN 757 UV-Visible Spectrophotometer
SHANGFEN 757 UV-Visible Spectrophotometer Proportional dual beam of light, monochromator Czerny-Turner, focal length 160mm, grating 1200 lines/mm, 757 UV-Visible Spectrophotometer can meet a wide range of applications from daily analysis to scientific research.
ShangFen-
JINGHUA 7600SCRT Dual beam of light UV-Visible Spectrophotometer Standard computer/printer
$ 5667.00 -
YueFeng 7230G (PC) Visible Spectrophotometer with pc software
$ 661.00 -
JINGHUA 756 UV-Visible Spectrophotometer 2nm Light spectrum Bandwidth 190-1100nm Wavelength range
$ 1781.00 -
JINGHUA 756PC UV-Visible Spectrophotometer Wavelength range: 190nm~ 1100nm
$ 3149.00 -
JINGHUA 723 Automatic visible spectrophotometer High performance, low stray light, automatic wavelength setting
$ 770.00
Introduction
The 757 UV/VIS Spectrophotometer can meet the needs of a wide range of applications, from routine analysis to scientific research. The biter adopts a professional proportional dual-beam optical path system and a low-noise circuit design, which has good reliability and stability.
Features
It adopts professional optical, mechanical and electrical design, and has excellent performance and cost performance.
It adopts 8-inch color touch LCD display, which is flexible and convenient to operate.
It adopts a proportional dual-beam optical path structure and is equipped with a high-performance "shining holographic light shed" with a low-stray light and high-resolution monochromator, which has excellent optical accuracy, measurement accuracy, reproducibility and stability.
Featuring a unique auto-adjust "0" to adjust to "100", an automatic 8-well sample holder.
It has various advanced functions such as kinetic time scanning, GOTOλ, linear regression, concentration direct reading, peak-to-valley detection, etc.
Using thermal plotter (optional), it can draw curves such as data printing, spectral scanning, fixed-wave long-term scanning, and linear regression.
With USB interface, it can directly interact with the PC through the standard UVwin8 ultraviolet spectroscopy software, which strengthens the processing functions of the instrument's detection data and scanning spectrum, realizes the massive storage of test documents, and also provides convenience for users' secondary development.
Professional power-off protection measures, which can memorize the measured data, scanning spectrum, regression equation and instrument correction value, etc., and realize the rapid initialization of the boot into the test state.
The instrument uses a high-energy, long-life flange to locate the neon lamp. When the neon lamp needs to be replaced, only two flange positioning screws need to be removed and installed to ensure that the lamp source is in the right position without cumbersome optical path adjustment.
other
Light source: 12V20W tungsten halogen lamp and long life neon lamp
Power supply Voltage: AC220V±22V, 50Hz±1Hz
Rated power: 200W
Matching
Thermal Plotting Printer
ShangFen 757 UV-Visible Spectrophotometer Specifications
| LIST | VALUE |
|---|---|
| Optical inspection system | Proportional double beam of light |
| monochromator | Czerny - Turner |
| focal length | 160mm |
| grating line number | 1200 gr/mm |
| Detector | Photocells, photovoltaic cells |
| Light spectrum Bandwidth | 2nm |
| wavelength setting | touchscreen input |
| Wavelength range | 190~1100nm |
| Wavelength Precision | ±0.5nm |
| Wavelength Repeatability | ≤0.2nm |
| Light source switching wavelength | 340nm |
| stray light | ≤ 0.1% (T) (at 220 nm, as Nal) (at 360 nm, as NaNOz) |
| Photometric range | 0.0 ~200.0%T、-0.301~4.000A、0.000 ~9999c |
| Photometric Precision | ±0.3%T;±0.002Abs (0 ~ 0.5A);±0.004Abs (0.5 ~ 1A) |
| Luminosity Repeatability | ≤0.15%T;0.001A(0~0.5A);0.002A(0.5~1A) |
| Noise level | 100%(T)≤0.2%(T);0%(T)≤0.1%(T) |
ShangFen 757 UV-Visible Spectrophotometer Packing list
machine X1, manual X1, certificate X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- What are the research methods for the properties of temperature-sensitive membranes?
- Contrast: Use a Spectrophotometer to determine the opacity of paints and coatings
- What is UV-Vis spectrophotometry used for and how does the measuring device work?
- Application of UV-Vis Spectrophotometer in Industry
- How to extend the service life of the aging Test Chamber?
- Principle, characteristics and application of ultraviolet Spectrophotometer
- The difference between photometers at different angles
- Why is a Spectrophotometer calibrated? What is the principle?
- ActiveJJF 1641-2017
- ActiveJC/T 1040-2020
- ActiveSH/T 0409-1992
- ActiveSL 84-1994
- ActiveGB/T 6150.3-2023
- ActiveGB/T 6150.1-2023
- AbolishGB/T 14571.4-2008
- ActiveGB/T 14571.4-2022
- ActiveGB/T 36084-2018
- AbolishHJ/T 201-2005
- ActiveGB/T 13217.1-2020
- AbolishGB/T 11175-2002


757