TENLIN HX-0515 Low temperature thermostatic circulator - 5 ℃/ +/- 0.05 ℃ 15L
TENLIN HX-0515 low temperature thermostatic circulator Temperature range -5-100 ℃, resolution 0.01 ℃, Fluctuation +/- 0.05 ℃, HX-0515 low temperature thermostatic circulator has the advantages of fast cooling speed and strong temperature Stability.
TenLin-
Shangtian Jingyi DHC-0520N high-low temperature thermostatic tank, heating and cooling Water Bath -5~ 160 ℃/20L
$ 1886.00 -
LABTOP OBC220EB open bath thermostatic circulator
Login -
NELE DC-0550 Thermostatic Bath Vertical laboratory Thermostatic Water Bath 50L/-5 ℃ +/- 0.05 ℃
$ 2246.00 -
NELE DC-4020-II Temperature Cycling Unit 20L/-40 ℃
$ 4137.00 -
Shangtian Jingyi DHC-2020N (0.01) High-low temperature Thermostatic Bath Cyclic refrigeration Thermostatic Bath -20~160 ℃/20L
$ 2232.00
Features
1. Refrigeration system: It adopts low-noise air-cooled hermetic compressor unit, which has the advantages of fast refrigeration speed and strong temperature stability.
2. Control system: new special software research and development of self-made temperature control technology, equipped with high-standard Pt100 and all imported electronic components.
3. Circulation system: It is composed of internal and external circulation. During the external circulation, the constant temperature liquid in the tank can be introduced outside, and a second constant temperature field can be established. It can also be used as a cold source to guide the refrigerated liquid in the tank to the external experimental container to cool down. When circulating, the temperature of the liquid in the tank can be very uniform and stable.
4. Protection system: Refrigeration has overheating and overcurrent protection. The control system has functions such as over-temperature chirp alarm, upper and lower limits of temperature can be set for over-temperature alarm temperature, and the load can be automatically cut off when over-temperature.
5. High-end large LCD control instrument, easy to operate, strong temperature stability, PID can be automatically searched or manually adjusted.
6. The accurate temperature correction function can reach 0.01℃, and the temperature fluctuation can reach up to ±0.02℃~±0.05℃ (depending on the model).
external structure

Figure 1: Outline structure of constant temperature bath

Figure 2: Bath control panel
Model parameters
| model | temperature range | power | tank volume |
| HX-08 | 0-100℃ | 1.6kw | 8L |
| HX-0508 | -5-100℃ | 1.6kw | |
| HX-1008 | -10-100℃ | 1.7kw | |
| HX-2008 | -20-100℃ | 1.8kw | |
| HX-3008 | -30-100℃ | 2kw | |
| HX-4008 | -40-100℃ | 2.3kw | |
| HX-010 | 0-100℃ | 1.8kw | 10L |
| HX-0510 | -5-100℃ | 1.9kw | |
| HX-1010 | -10-100℃ | 2kw | |
| HX-2010 | -20-100℃ | 2.1kw | |
| HX-3010 | -30-100℃ | 2.4kw | |
| HX-4010 | -40-100℃ | 2.8kw | |
| HX-015 | 0-100℃ | 2.1kw | 15L |
| HX-0515 | -5-100℃ | 2.1kw | |
| HX-1015 | -10-100℃ | 2.2kw | |
| HX-2015 | -20-100℃ | 2.3kw | |
| HX-3015 | -30-100℃ | 2.5kw | |
| HX-4015 | -40-100℃ | 3kw | |
| HX-020 | 0-100℃ | 2.3kw | 20L |
| HX-0520 | -5-100℃ | 2.4kw | |
| HX-1020 | -10-100℃ | 2.5kw | |
| HX-2020 | -20-100℃ | 2.6kw | |
| HX-3020 | -30-100℃ | 2.8kw | |
| HX-4020 | -40-100℃ | 3.2kw | |
| HX-030 | 0-100℃ | 2.3kw | 30L |
| HX-0530 | -5-100℃ | 2.4kw | |
| HX-1030 | -10-100℃ | 2.5kw | |
| HX-2030 | -20-100℃ | 2.7kw | |
| HX-3030 | -30-100℃ | 3kw | |
| HX-4030 | -40-100℃ | 3.5kw | |
| Other parameters | |||
| temperature resolution | 0.01°C | ||
| temperature fluctuation | ±0.05℃ | ||
| pump flow | 13L/min | ||
| lift | 3m | ||
TenLin HX-0515 Low temperature thermostatic circulatorSpecifications
| LIST | VALUE |
|---|---|
| Temperature range | -5~100℃ |
| temperature resolution | 0.01℃ |
| Temperature fluctuation | ±0.05℃ |
| Dimensions | 320*370*765mm |
| power | 2.1kw |
| Tank Volume | 15L |
| Pump flow | 13L/min |
| pump head | 3m |
| Circular mode | Internal/external circulation |
TenLin HX-0515 Low temperature thermostatic circulator Packing list
host X1, cover X1, manual X1, certificate X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveJJF 2019-2022
- ActiveJJF 1030-2023
- ActiveGB/T 26809-2011
- ActiveJJF 1030-2010
- ActiveGB/T 26808-2011


HX-0515