| Detect gas |
Hydrogen sulfide H2S |
| Range |
H2S(0~100ppm) |
| Resolution |
H2S(0.01ppm) |
| Sensors principle |
Electrochemistry |
| Sensors Life |
2-3 years |
| Sampling method |
Diffusion |
| Display mode |
2.4 inch 320X240 high definition color screen display, 5 button operation |
| Detection Accuracy |
≤±3%(F.S) |
| Linearity |
≤±2% |
| Repeatability |
≤±2% |
| Response time |
T90≤20s |
| recovery time |
≤30s |
| Power Supply |
DC3.6V |
| Alarm mode |
Sound and light alarm, vibration alarm, visual alarm, sound and light + vibration + visual alarm, turn off alarm optional |
| Battery Capacity |
3.6VDC, 4500mA large Capacity rechargeable polymer battery, with overcharge, overdischarge, overVoltage, overheating, short circuit protection function |
| Operating Environment |
Temperature -40 ℃~ + 60 ℃; Relative Humidity ≤ 0-99% RH (Built-in Filter can be used in high Humidity or high dust environment) |
| Sample temperature |
-40 ℃~ + 60 ℃, optional high temperature sampling cooling filtrate handle, can detect the flue gas concentration of 600 degrees or higher temperature |
| Temperature measurement |
-40 ℃~ + 120 ℃ (optional) Accuracy 0.5 ℃ |
| Humidity measurement |
0-100% RH (optional) Accuracy 3% RH |
| Data storage |
Standard 100,000 data Capacity, support local real-time view, delete or data export, free host computer communication software, storage function defaults to off state, can be set to open state, storage time interval can be set arbitrarily |
| Interface |
USB (charging and communication), optional RS485 or RS232 Communication interface |
| interface language |
Chinese or English can be set, the default Chinese interface |
| Explosion-proof sign |
Intrinsically safe |
| Explosion-proof sign |
Exia II CT4 Ga |
| Protection level |
IP65, dustproof, splash-proof |
| Dimensions |
160*70*28mm |
| Weight |
300g |