YOUKE YK3057 Leakage marking Tester, leakage resistance marking test
Test Voltage 100~600V adjustable, short circuit current 1A +/- 0.1A, using platinum electrode high temperature corrosion resistance, drop count accurate, stable and reliable.
YouKe-
Jingyi JY-30A automatic swirling/spinning oil mixer, intelligent positive and negative two-way swirling/spinning vertical Stirrer
$ 2480.00 -
MODERNER RYK heat source block, bottom diameter φ 100mm
Login -
XINGHUO ZTT-930M protruding road sign abrasion performance Tester
$ 750.00 -
YUZHOU ZKXQ-1 Automatic Vacuum Inlay Machine
$ 3957.00 -
Guohua DB-B2 baking machine, power: 200W
$ 368.00
Guideline
IEC60335, GB4706.1, IEC60884-1, GB2099.1, GB/T4207-2003, IEC60112.
Technical Description
This product is used in electrical products under the influence of humidity and impurity environment, which may cause insulation leakage between live parts of different polarities or between live parts and grounded metal parts. Discharge can corrode materials and even ignite causing fires. This test instrument is a destructive test for insulating materials that simulates the above situation. It is used to test and evaluate the relative tracking resistance of the insulator under the action of the electric field and impurity water under the specified Voltage. Suitable for electrical and electronic products, solid electrical insulation materials for household appliances and their products, such as: relay sockets, transfer switch covers, contactor bodies, etc.
YouKe YK3057 Leakage marking TesterSpecifications
| LIST | VALUE |
|---|---|
| In-box Volume | ≥ 0.5 cubic meters with Glass observation door |
| Electrical load | The test Voltage is adjustable between 100 and 600V, and when the short-circuit current is 1A +/- 0.1A, the Voltage drop should not exceed 10%. When the short-circuit leakage current in the test circuit is equal to or greater than 0.5A, the time is maintained for 2 seconds, and the relay operates., cut off the current, indicating that the test product is unqualified. |
| Bipolar force on the sample | Using a rectangular platinum electrode, the force of the two poles on the sample is 1.0N +/- 0.05N respectively |
| Dripping device | The drop height is adjustable from 30mm to 40mm, the drop size is 44 to 50 drops/cm3, and the drop time interval is 30 +/- 1 second |
| Product Features | The test chamber structure is made of stainless steel or brass, the electrode head is platinum metal, high temperature and corrosion resistance, accurate drop count, and stable and reliable Control system. |
| Power | AC 220V,50Hz |
YouKe YK3057 Leakage marking Tester Packing list
Host X1, measurement ruler X1, power cord X1, ammonium chloride X1, distillation water X2, exhaust pipe X1, exhaust pipe buckle X1, measuringflask X1, Glass dish X1, water absorber X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.