TIMEPOWER TP3117 Desktop multiparameter analyzer COD ammonia nitrogen, total phosphorus, total nitrogen turbidity suspended solids
TIMEPOWER TP3117 desktop multiparameter analyzer can quickly determine COD, ammonia nitrogen, total phosphorus, total nitrogen, turbidity, SS (suspended solids) and other parameters in enterprise wastewater. Measurement range COD 5-10000 mg/L, ammonia nitrogen 0.01-50 mg/L, total phosphorus 0.02-20 mg/L, total nitrogen 0.5-100 mg/L, turbidity 0.5-1000 NTU, suspended solids 5-500 mg/L.
TIMEPOWERapply
The TP3117 multiparameter analyzer is a powerful, easy-to-use precision benchtop analytical instrument. It can quickly determine COD, ammonia nitrogen, total phosphorus, total nitrogen, turbidity, SS (suspended solids) and other parameters in enterprise wastewater. It is widely used in emergency monitoring, sewage treatment, chemical industry, wastewater, food, printing and dyeing and other industries wastewater detection, and can also be applied to wastewater research in scientific research units, colleges and universities.
principle
The instrument uses the principle of photoelectric colorimetry for measurement. According to the Lambert-Beer law: when a beam of monochromatic parallel light passes through a colored solution, a part of the light energy is absorbed by the solution, and if the thickness of the liquid layer remains unchanged, the degree of light energy absorbed is directly proportional to the concentration of colored substances in the solution.
Features
Large-screen LCD display, key operation, direct data reading, simple and time-saving operation.
It adopts imported cold light source, with good optical performance and a service life of up to 100,000 hours.
You can choose between standard or continuous measurement modes for speed and accuracy.
The digestion colorimetric tube is used to make the measurement simple, fast and safe.
Built-in printer for printing current data and historical measurement data.
Memory standard working curves and 96 user-defined curves, which can store 30,000 pieces of detection data. It has the functions of data storage and power-off protection to prevent data loss and facilitate the query of historical measurement data.
It is equipped with special detection reagents to reduce user operation as much as possible and make the detection more accurate.
The price of consumables is low and the amount is small.
It supports network port and USB interface, and can transmit data with PC or printer.
Precautions
1. In order to improve the accuracy of detection, the interaction of samples during testing should be reducedThe cap should be fixed to reduce errors caused by operation.
2. According to the concentration value of the measured sample, the corresponding standard curve should be selected.
3. Because the oxidant contains chloride ion masking agent, there may be a small amount of precipitation after digestionThe tube will not float up the sediment and cause deviation in the results.
TIMEPOWER TP3117 Desktop multiparameter analyzerSpecifications
| LIST | VALUE |
|---|---|
| Measurement range | COD 5~ 10000mg/L; ammonia nitrogen 0.01~ 50mg/L; total phosphorus 0.02~ 20mg/L, total nitrogen 0.5~ 100mg/L; turbidity 0.5~ 1000 NTU; suspended solids 5~ 500mg/L |
| Digestion environment | COD 165 ℃, 15min, total phosphorus 125 ℃, 30min, total nitrogen 125 ℃, 30min |
| error of indication | ≤±5% |
| Repeatability | ≤5% |
| anti-chlorine interference | ≤2000mg/L(COD) |
| Colorimetric method | Digestion Colorimetric Tube |
| Data storage | 30,000 records, 96 Linear dispersion (support user-defined calibration) |
| transmission method | USB cable, LAN, etc |
| printer | Built-in thermal printer |
| Operating interface | Chinese/English (optional) |
| Display | 5.0-Inch color LCD screen |
| Power Supply | AC(220±22)V;(50±0.5)Hz |
| Power | 5W |
| Environmental temperature | 5~40℃ |
| Working Relative Humidity | ≤ 85% non-condensing |
| Dimension | 348mm*250mm*145mm |
| Weight | 2.6Kg (main unit) |
TIMEPOWER TP3117 Desktop multiparameter analyzer Packing list
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- Turbidity simple operation process
- COD prefabricated reagent simple operation process (0-15000mg/L)
- Simple operation process of ammonia nitrogen reagent
- Simple operation process of total phosphorus reagent
- Simple operation process of total nitrogen reagent (0-100mg/L)
- ActiveGB/T 693-2024
- ActiveJJF 1700-2018
- ActiveJJF 1571-2016
- ActiveJJF 1424-2013
- ActiveJJF 1012-2019
- Active 126-2009
- ActiveHG/T 5669-2019
- ActiveSL/T 788.2-2019
- ActiveGB/T 42955-2023
- ActiveGB/T 11901-1989
- ActiveGB/T 29599-2013


TP3117