| Lighting method | Reflection: D/8 (diffuse illumination, 8 ° direction reception); SCI/SCE measurement; includes UV/exclude UV measurement; Transmission: D/0 (diffuse illumination, 0 ° direction reception) SCI/SCE measurement; includes UV/exclude UV measurement; |
| Lighting method 2 | Haze (ASTM D1003); meets CIE No. 15, GB/T 3978, GB 2893, GB/T 18833, ISO7724/1, ASTM E1164, DIN5033 Teil7 |
| Integrating sphere size | Φ154mm |
| lighting source | 360~780Nm combined LED light source, 400nm cut-off light source, 420nm cut-off light source |
| spectroscopy | Concave grating spectroscopy |
| Sensor | 256 pixel dual array CMOS image sensor |
| Measurement Wavelength range | 360~780nm |
| wavelength interval | 10nm |
| Half bandwidth | 5nm |
| Reflectance Measurement Range | 0~200% |
| Measurement caliber | Reflection: φ 30mm/φ 25.4mm, φ 18mm/φ 15mm, φ 10mm/φ 8mm, φ 6mm/φ 4mm; Transmission: φ 25.4mm/φ 4mm |
| Light-containing method | Reflection SCI/SCE, Transmission SCI/SCE |
| Color space | CIE LAB,XYZ,Yxy,LCh,CIE LUV,Musell,s-RGB,HunterLab,βxy,DIN Lab99 |
| color difference formula | ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00, DINΔE99,ΔE(Hunter) |
| Other chroma metrics | WI (ASTM E313, CIE/ISO, AATCC, Hunter), YI (ASTM D1925, ASTM 313), homochromatic index Mt, colour fastness, color fastness, force, coverage |
| Other chroma metrics 2 | APHA/Hazen/Pt-Co (platinum cobalt index), Gardner index, 8 degree gloss, 555 shade classification, haze (ASTM D1003), Saybolt (Seibert index), ASTM D1500 color code, Chinese Pharmacopoeia color code |
| Viewing Angle | CIE 2°/10° |
| observation light source | D65,A,C,D50,D55,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12,CWF,DLF,TL83,TL84,TPL5,U30 |
| display mode | Light spectrum plots/data, sample chromatic values, color difference values/plots, chromaticity plots, Color simulation, pass/fail results, Color bias |
| Measurement time | Approximately 2.4s (approximately 5s for simultaneous SCI/SCE testing) |
| Sexual repeability | Spectroscopic reflectance: φ 25.4mm/SCI, standard deviation within 0.04%; chromatic value: φ 25.4mm/SCI, Delta E * ab within 0.01 (after instrument preheating correction, the average value of whiteboard 30 times at intervals of 5s measurement) |
| Sexual repeability 1 | Spectral transmittance: φ 25.4mm/SCI, standard deviation within 0.05%; chromatic value: φ 25.4mm/SCI, Delta E * ab within 0.02 (after instrument preheating correction, the average value of whiteboard 30 times at intervals of 5s measurement) |
| stand-alone difference | ΔE*ab<0.12(SCI) |
| Dimensions | 370*300*200mm |
| Weight | About 9.6kg |
| Power Supply | DC 24V, 3A power adapter Power Supply |
| Lifetime of lighting source | More than 3 million measurements in 5 years |
| Display | TFT true color 7inch, capacitive touch screen |
| Data Interface | USB, Bluetooth ® 5.0, print serial port |
| data storage | 5000 master standards, 40,000 samples (SCI/SCE counts as one data) |
| operating language | Simplified Chinese, Traditional Chinese, English, (customizable German, French, Spanish) |
| Operating environment | 0~40℃(32~104°F) |
| Storage environment | -20~50℃(-4~122°F) |
| optional accessories | Microprinters, instrument inverted test fixtures, culture dishes |
None