Yoke P704 Portable touch screen pH/Conductivity Detector
Acidity meter + Conductivity Detector, portable, automatically recognizes 8 kinds of Conductivity Detector standard solutions.
Yoke-
BANTE DO electrolysis solution, standard solution
$ 114.00 -
Qiwei Y-616 Benchtop Ozone Detector
$ 276.00 -
LEICI potassium ion electrode calibrated solution, suitable for potassium ion composite electrode
$ 106.00 -
JINGQI WD9102P Total Phosphorus Tester with Printing Function
$ 1305.00 -
JENCO 6377MB multiparameter STIP-scan pH/ORP/conductivity/salinity /TDS/temperature with Bluetooth connection
$ 1077.00
Introduction
A multi-parameter measuring instrument for laboratory high-precision measurement, covering various electrochemical parameters such as pH and conductivity. It is suitable for industries such as petroleum, chemical industry, medicine, power plant and environmental protection, and has wider applications in the fields of education and scientific research.
One machine with multiple functions: built-in pH module, conductivity module
Test items: pH value/mV value/conductivity/resistivity/TDS/salinity/temperature
Features
1.7-inch color high-resolution (1024*600) capacitive touch screen, UI design, navigation operation experience, high sensitivity, complete parameter display, intuitive, convenient and simple to use. The host shell is made of PEEK material, which is resistant to high temperature, corrosion and aging.
2. Built-in ARM32-bit microprocessor chip, using digital filtering and slip technology to intelligently improve the response speed of the instrument and the accuracy of measurement data. The circuit board adopts SMT patch technology, which improves the reliability of product processing.
3. The main interface of the instrument can display pH value/mV value/conductivity/temperature value at the same time. Click each item module to view the detailed detection value of each detection item and perform calibration of each item.
4. Meet the GLP requirements of international standards, with intelligent functions such as automatic calibration, automatic temperature compensation, data storage, USB output, clock display, wireless printing, function setting and self-diagnosis information.
5. The instrument has a built-in Chinese and English bilingual operating system, humanized operation design, adjustable brightness of the instrument, and the temperature unit can be freely switched between ℃ and ℉.
6. The built-in large-capacity storage can store 2000 sets of measurement data, the data can be stored and transferred through the U disk, and the stored data can be opened by the general office software EXELL, which is very convenient for customers to edit.
7. The instrument can be equipped with a Bluetooth module, which supports wireless Bluetooth printing, which is convenient for users to operate.
8. Based on the Wechat applet, you can view the instrument data in real time even if you download the program on your mobile phone.
9. Scalability: Different pH electrodes can be easily and quickly replaced (matching electrodes can be selected for measuring different samples) to measure pH. Replace different electrodes (electrodes with different cell constants can be selected) to measure conductivity.
10. Automatically identify 8 kinds of conductivity standard solutions, and there are two series of standard solutions to choose from: European and American series and Chinese series. Automatically identify 15 kinds of pH standard buffer solutions, there are three series of standard buffer solutions to choose from: European and American series, NIST series and Chinese series.
11. It can intelligently judge the electrode status to ensure accurate use by customers.
12. Switchable conductivity, TDS, salinity and resistivity. TDS and salinity are calculated using polynomials to ensure the conversion accuracy of the full range.
13. Rechargeable lithium battery power supply, support power and mobile power charging.
14. Comply with IP54 dust-proof and splash-proof design.
Yoke P704 Touch screen portable 2-in-1Specifications
| LIST | VALUE |
|---|---|
| pH Measurement range | (-1.99 ~ 19.99)pH |
| pH resolution | 0.1/0.01pH |
| pH measurement Accuracy | Electric meter: ± 0.01pH, matching: ± 0.02pH |
| Input current | ≤1×10-12 A |
| input impedance | ≥ 3 × 10 (12) Oh |
| Stability | ±0.01pH/3h |
| Temperature compensation range | Manual/automatic (0~100.0) ℃ |
| mV range | -2000mV ~ 0 ~ 2000mV |
| mV resolution | 1mV |
| mVAccuracy | ±0.1%FS |
| Conductivity Detector Measurement range 1 | 0.00 ~ 20.00μS / cm 20.0 ~ 200.0μS / cm200 ~ 2000μS / cm |
| Conductivity Detector Measurement range 2 | (2.00~20.00)mS/cm、(20.0~200.0)mS/cm、(200~2000)mS/cm(K=10) |
| Resistivity Measurement range | 0 ~ 100MZ · cm |
| TDS range | 0~100g/L |
| Salinity range | 0~100ppt |
| resolution | 0.01/0.1/1μS/cm,0.01/0.1/1 mS/cm |
| Accuracy | Electric meter: ± 0.5% FS, matching: ± 1.0% FS |
| Temperature compensation range | (0~100) ℃ (manual/automatic) |
| Electrode constant | 0.1/1/10cm-1 |
| reference temperature | 25 ℃, 20 ℃ and 18 ℃ |
| data storage | 2000 groups |
| Save content | Measurement number, measurement value, temperature value, ATC or MTC status, measurement date, measurement time, Power : rechargeable lithium battery Power , 3.7V/1800mA rechargeable lithium battery |
| Power | Rechargeable lithium battery Power , 3.7V/1800mA rechargeable lithium battery |
| Communication interface | USB2.0, wireless Bluetooth (optional) |
| Dimensions | 160*190*70mm |
| Environment temperature | Environment temperature: 5~ 35 ℃ (0.01 grade), 15~ 30 ℃ (0.001 grade) |
| Environment humidity | ≤75% |
| Protection class | IP54 |
| Weight | 880g |
Yoke P704 Touch screen portable 2-in-1 Packing list
complete-built unit X1, manual X1, certificate of conformity X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- Tide Review: Conductivity Meter 76652-03
- Inventory of experimental instruments: Preparation of PVDF-coated modified lithium battery separators on polyolefin separators
- Preparation of GO-modified TFC-FO membrane support layers with different particle sizes by film coating machine
- What are the technical requirements for mica pearlescent pigments? How to prepare samples?
- Measurement method for electrophoretic coating by conductivity meter
- What factors affect the effect of electrophoretic coating?
- How to judge whether laboratory ultrapure water equipment needs maintenance?
- Understanding Conductivity Cell Constants
- Standards for coating test methods
- How to choose an industrial pH sensor?
- ActiveGB/T 8884-2017
- ActiveJJF 1565-2016
- ActiveJJG 2059-2014
- ActiveHJ 1147-2020
- ActiveGB/T 17650.2-2021
- ActiveGB/T 41079.2-2022
- ActiveGB/T 6908-2018
- ActiveGB/T 7977-2007
- ActiveGB/T 32697-2016
- ActiveGB/T 33328-2016
- ActiveISO 21318-2007
- ActiveISO 9944-1990


P704