| Inner Volume |
50L |
| Temperature and Humidity Control Instrument |
Programmable TEMP8226S imported from Korea, touch screen input, different temperature points can be arbitrarily set according to customer requirements |
| Box structure |
This equipment is divided into: test area, low temperature cold storage area, high temperature heat storage area |
| Liner size |
500*400*300mm |
| Dimensions |
1500*1900*1400mm |
| Temperature Impact Range |
-60 ℃~ + 150 ℃ (programmable to set arbitrarily) |
| Temperature range of regenerator |
0 ℃~ + 180 ℃ (programmable to set arbitrarily) |
| Temperature range of storage tank |
0 ℃~ -60 ℃ (programmable to set arbitrarily) |
| Controller control Accuracy |
±0.2℃ |
| Temperature Impact Rate |
Complete in 5 minutes |
| Enclosure Material |
SUS #304 stainless steel plate |
| Liner material |
Heat-resistant and cold-resistant stainless steel plate, first-class light panel |
| Insulation material |
High Density Glass Cotton and High Strength PU Foam Insulation Material |
| box uniformity |
±2℃ |
| Air supply circulation system |
Japanese multi-wing centrifugal Circulating Fan, Japanese motor whole group |
| Heating system |
U-shaped fin type 304 #seamless steel high-speed heating electric heater |
| Refrigeration system |
France taikang compressor |
| Refrigeration accessories |
Table bending center force evaporator, Mar fan, etc |
| Refrigeration system |
air-cooled |
| Test box load |
20KG |
| Protection device |
Compressor overheating, overtemperature, overload, high Voltage, low pressure, oil pressure, overcurrent, lack of phase, power failure standby and other alarm warnings |
| Power supply Voltage |
AC380V 50~60Hz 10~15KW |
| meet the standard |
GB2423 [1] .02 High temperature test method, GB2423 [1] .01 Low temperature test method |
| Control mode |
5.7-Inch color touchscreen |
| Control mode |
program |
| Sensor type |
PT100 Sensors Input |
| output method |
PID control output; (4-20mA) analog circuit output; 8-way switching signal auxiliary input; (basic configuration of 10 relays, optional 20 relays) |
| Measurement range |
-200.00 ℃ --300.00 ℃, error +/- 0.2 ℃ |
| program editing |
There are three types of programs, namely 2ZONE, 3ZONE, and STEP. |
| language |
Chinese and English can be switched |
| Controller Power Supply |
Rated Voltage: 85-265V AC, 50/60Hz, lithium battery for at least 10 years, withstand Voltage 2000V AC/1 minute |
| Controller external Dimensions |
184*144*94mm |
| Controller mounting opening size |
173 × 133 (mm) (L × W) |
| TFT resolution |
640 × 64k color |