Print Picutre Similar products View price

HUDA QD-E703-100KS40 Hot & Cold Shock climate chamber DATASHEET

Brand HUDA Model QD-E703-100KS40
From China Name Hot & Cold Shock climate chamber
Specifications
Inner Volume 100L
Liner size 500*450*450mm
Dimensions ≈1455*1790*1980mm
Weight 650㎏
Maximum Power 15KW
Maximum current 30A
Power supply conditions AC380V three-phase R, S, T, N + protective grounding
Allowable Voltage fluctuation range 10%
Allowable Frequency Range 50±0.5Hz
Protective ground resistance Less than 4Ω
Power Supply TN-S mode Power or TT mode
Operating Environment The user is required to configure the air or power switch of the corresponding Capacity for the equipment at the installation site, and the switch must independently control the use of the equipment
Functional requirements 1. (air) indoor space cooling and heating; 2. According to the temperature setting, the indoor space reaches the set temperature; 3. It is divided into three parts: heat storage area, cold storage area, and test area; 4. The three areas are independent, and the temperature is switched through the damper
Operating Conditions Ambient Temperature is 0~ + 35 ℃, Relative Humidity ≤ 85%
High Temperature Impact Range RT~170°C
Low Temperature Impact Range RT~-60°C
Temperature range of test area -40~150℃
Temperature Impact Range High temperature: 60 ℃~ 150 ℃; Low temperature: -10 ℃~ -40 ℃; Room temperature: Indoor temperature
Temperature Resolution Accuracy 0.1℃
Temperature Fluctuation ±1℃
Temperature Uniformity ±2.0℃
Temperature deviation ±1℃
Storage area Heating time RT+→170°C≈30min
Cooling time of cold storage area RT+→-60°C≈50min
Temperature recovery time 5 minutes
Liner material 304 #stainless steel 1.0mm thick
Enclosure Material Cold plate baking paint 1.5mm thick
Thermal insulation material Polyurethane foam + fiberglass cotton 100mm thick
Box door Single door, explosion-proof handle
Lead Hole Open the test hole of Φ50mm, and attach the corresponding thermal insulation accessories and special sealing soft plug
Control mode Taiwan, China Youyi LCD touch programmable Controller; main power switch, operation indicator, abnormal indicator, end indicator, defrosting indicator, power indicator, RS-232 communication interface
Low temperature zone mechanical room Evaporator, Circulating Fan, drainage device, cold storage sheet, heating device, circulating wind wheel, volute
High temperature zone machinery room Heating device, Circulating Fan, heat storage sheet, pressure equilibrating port, circulating wind wheel, volute
Distribution control cabinet Total power circuit breaker, Controller, switchboard, cooling fan, over-temperature protector, cylinder, pressure protector, liquid storage tank, water-cooled condenser, refrigeration unit
Heating system Using scale heat dissipation heating pipe P.I.D to control heating to achieve temperature equilibrate
Refrigeration system (Optional) Using Domestic (cn-zh) high-quality chilled down water tower
Chill down the water tower Water circulation cooling use, Flow rate 8 tons/hour
Test material rack Stainless steel SUS #304 grille rack 2 pieces (load-bearing ≤ 20kg/piece)
Control mode Taiwan, China Youyi Controller, control temperature
Number of programs available Maximum 120 groups, can be set to be used in connection between programs
Command, repeability Each command can be up to 9999 times
RS-232 Communication interface Can display Linear dispersion, data acquisition, can be used as monitoring and remote Control system
U disk memory card Configurable 2~ 32G memory card, can record test data for two years, bring your own.
Operation mode Program loop mode
Set range Adjust according to the temperature working interval of the equipment (upper limit + 5 ℃, lower limit -5 ℃)
Display resolution Temperature: 0.1 ℃; Time: 0.1min
input Thermocouple
Circulatory system Low noise multi-wing long axis motor
Heating system Imported nickel-chromium alloy electric heating device
Heating Control mode No touchpoint equal period pulse width modulation, SSR (solid state relay)
Control mode Anti-integral saturation PID, BTC equilibrate temperature control mode (temperature test equipment)
subsidiary function Fault alarm and cause, processing prompt function, power-off protection function, Upper/Lower limit temperature protection function, self-diagnosis and automatic stop function
Software Operating Environment IBM PC compatible machine, CPU above PII, memory above 128M, Simplified Chinese Windows2000 or Simplified Chinese WindowsXP operating system
Temperature measurement T-Thermocouple
Refrigeration system Duplex refrigeration system
Refrigeration system Low temperature compressor imported from Europe
Refrigeration system Water-cooled shell and tube condenser, chilled down water tower
Evaporator finned tube heat exchanger
Evaporative condenser The plate heat exchanger brazed by high temperature copper has small surface area, small energy loss and high heat exchange efficiency
Throttle device Thermal expansion valve, capillary tubing
Oil separator Separate the refrigerated oil droplets mixed with high temperature and high Voltage refrigerant gas discharged from the compressor and return to the compressor curved rail cavity for compressor lubrication and cooling
refrigerant Environmentally friendly refrigerant R404A/R23
Refrigeration system protection device Compressor overheating, compressor overcurrent, compressor overpressure, condensing fan overheating
Protection device Adjustable over-temperature protection, air conditioning channel limit over-temperature, fan, motor overheating relay
Protection device Total Power phase sequence and phase loss protection, leakage protection, overload and short circuit protection, Ambient Temperature protection
Environmental conditions Ambient Temperature should be kept within 0 ℃~ + 35 ℃, Relative Humidity: ≤ 85%
Storage environment Ambient Temperature of the equipment should be kept within 0 ℃~ + 45 ℃; when the Ambient Temperature is lower than 0 ℃, the water remaining in the equipment should be discharged cleanly (the equipment is stopped for a long time) to avoid the water in the pipeline freezing and damaging the pipeline
Installation site requirements The distance between the left and right side panels of the test chamber and the rear side panels shall not be less than 500mm, and the distance between the front and the wall shall not be less than 1000mm
Packing list
complete-built unit X1, manual X1, certificate of conformity X1, warranty card X1
After-sales instructions

Warranty days: 365 days

1. For product quality problems caused by non-human products of instruments and equipment, the manufacturer will provide free warranty services during the warranty period;

2. There is no warranty period for fragile and consumable parts, and if you find product quality problems after receiving the goods, please apply for return and exchange within 7 days, and you will not wait for the overdue ones;


[Note] For after-sales services outside the Chinese mainland, if the instrument requires repair, the shipping and possible associated taxes shall be borne by the sender. Generation address of the manual: http://en.nbchao.com/p/41404/Specification.html , which is only used by users who purchase commodities from the NBCHAO.