HANNA HI934 Karl Fischer Coulomb Method (KF) Potentiometric titrator
Moisture content ranging from 1 ppm to 5%, resolution 0.1 ppm, dynamics with 3 speed settings, statistical mean of results, standard deviation, powerful built-in algorithms for termination criteria based on fixed mV endpoints or absolute/relative drift.
HANNA-
LEICI ZDJ-4A Potentiometric titrator
$ 2892.00 -
LEICI ZDJ-4B Potentiometric titrator
$ 3917.00 -
LEICI ZDJ-5B AutomaticTitrator
$ 6734.00 -
YueFeng ZD-2A Automatic Potentiometric titrator 0~14.00pH
$ 405.00 -
LEICI ZDCl-2 Potentiometric titrator
$ 3533.00
Introduction
The HI934 is a Karl Fischer coulomb titrator with high accuracy, flexibility and repeatability. This titrator is designed to perform titrations for a variety of applications, allowing the user to obtain good results and high-speed analysis. The moisture content range analyzed by HI934 is 1ppm to 5%. This powerful titrator efficiently monitors KF reactions, detects endpoints, and performs all necessary calculations and plots.
Features
Small footprint, requiring minimal bench space;
Housing made of sturdy, chemical-resistant plastic;
Powerful built-in algorithms for termination criteria based on fixed mV endpoints or absolute/relative drift;
Sample analysis averages and statistics;
Minimize water vapor ingress by sealing the solvent system;
Balance interface for automatic weighing;
Supports 100 titration methods;
User-defined reports;
Warning and error messages are clearly displayed.
HANNA HI934 Potentiometric titratorSpecifications
| LIST | VALUE |
|---|---|
| Measurement range | 1 ppm to 5% |
| resolution | 0.1 ppm |
| unit of result | %,ppm,mg/g,μg/g,mg,μg,mg/ml,μg/ml,ppt,mgBr/100g,gBr/100g,mgBr,gBr |
| Sample category | Liquid or solid (externally dissolved or extracted) |
| pre-titration condition | automatic |
| Background drift correction | Automatic or user-selectable |
| endpoint metric | Fixed mV continuous, relative drift stop or absolute drift stop |
| Measurement | Dynamic with 3 speed settings |
| result statistics | Mean, standard deviation |
| Titration container type | Borosilicate Glass with Standard Cone Glass Joints |
| operation Volume | 100 - 200 ml |
| Titration vessel diaphragm | silicone rubber |
| Diaphragm cap thread | GL-18 |
| reagent port | Standard taper 19 |
| Electrode type/connection | HI76330 |
| Glass connection | Standard taper 14/20 |
| Polarization current | 1,2,5,? 10μA |
| Voltage range | 2 - 1100 mV |
| Voltage resolution | 0.1 mV |
| Accuracy @25 ℃ | ±0.1% |
| Generator electrode model | Diaphragm or no diaphragm |
| Detection electrode type | automatic |
| Electrical connection | 5-Pin Connector with detachable cable |
| Glass connection | Standard taper 29/12 |
| Maximum current | 400 mA |
| current control | Automatic or fixed (400mA) |
| Type of mixing equipment | Magnetic, Electronic Regulation, Digital Stirrer |
| Stirring speed | 200 - 2000 RPM |
| Stirring resolution | 100 RPM |
| External Stirrer | 4-Pin mini-DIN connection allows control of external mixing equipment |
| Type of reagent processing system | A sealing system with integrated diaphragm air pump |
| Type of drying agent | molecular sieve |
| Bottle thread type | GL-45 |
| Glass connection | Standard taper 19 (using adapter provided) |
| Reagent/Waste Tube | PTFE |
| show | 5.7 "Graphic Color Display with Backlight |
| peripheral equipment | PC (USB Standard B); Flash Drive (USB Standard A); Analytical Balance (DB-9 socket); Printer (DB-25 socket); Keyboard (6-pin Mini-DIN) |
| language | English |
| Power Consumption | 100-240 VAC,50/60 Hz / 0.5 Amps |
| Plate material | ABS/PC and stainless steel |
| keyboard | Polyester brazing |
| Operating conditions | 10 - 40 0C;80% RH |
| Preserving the environment | -20 - 70 0C;95% RH |
| Size/Weight | 315 x 205 x 375 mm / 4.3kg |
HANNA HI934 Potentiometric titrator Packing list
machine X1, manual X1, certificate X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveGB/T 3653.1-2024
- ActiveGB/T 43788-2024
- ActiveSN/T 2261-2009
- ActiveHB 20056.4-2011
- ActiveYB/T 159.3-2015
- ActiveYB/T 505.4-2007
- Active 178-2014
- ActiveYS/T 710.1-2009
- Active 5110.2-2000
- Active 0995-2019
- Active 0980-2019
- ActiveSH/T 0688-2000
- ActiveHG/T 3838-2008
- ActiveYC/T 314-2009
- ActiveSH/T 1767-2008
- Active 0836-2010
- Active 5092.4-2001
- Active
- ActiveSN/T 4951-2017
- Active 6060-2022