| Range-pH | 0.00 to 14.00 pH |
| Resolution -pH | 0.01 pH |
| Accuracy @25 ℃/77 ° F-pH | ±0.05 pH |
| Calibrated -pH | Acidity measurement calibrate: Manual calibration acidity pH buffer calibrate: two points automatic identification calibrate, built-in standard points: 4.01, 7.01, 10.01 pH |
| Resolution -ORP | 1 mV |
| ' '25oC/77oF-ORP' | ± 5 mV |
| Calibrate | Manual calibration |
| Accuracy - temperature | ±1℃(±1.8ºF) |
| Temperature compensation | Automatic temperature compensation -5.0 to 105.0 ℃ (23.0 to 221.0 ° F) |
| Pump Control | Automatic and manual modes; Flow rate adjustable from 0.5 to 3.5 L/h |
| recording function | All data.csv files are transferred via USB flash drive to automatically record pH, ORP and temperature measurements, GLP and events including alarms, errors and power failures, store data for 60 days, Data Logging interval of 10 seconds |
| alert | Triggered when 5 consecutive readings exceed/fall below the set value |
| password protection | Settings, calibrate and log call option features are all password protected by user settings |
| connect | USB |
| GLP | Automatic recording of acidity pH/oxidation reduction ORP electrode calibrated, date and time information |
| alert output | SPDT 5A/230 VAC activated by pH/ORP/temperature optional alarm conditions |
| analog output | 4 to 20 mA output configuration; Output Impedance ≤ 500 ohms Accuracy < 0.5% FS, up to 50 V relative to isolation current |
| auxiliary input | Low level in acid/alkali tanks (open contact) Low level in chlorine tanks (touchpoint open); hold input (touchpoint open) |
| Electrode | Composite electrode HI1036-1802 |
| Power | AC100-240V |
| Power | 10VA |
| Applicable environment | 0-50°C |
| Size/Weight | 245 x 188 x 55 mm(73 mm) 1700 g (60 oz) |