| Measurement range | 0.00 to 9.99、10.0 to 99.9、100 to 1200NTU |
| resolution | 0.01NTU (≤9.99NTU)、0.1NTU(≤99.9 NTU)、1NTU(≤1200NTU) |
| Measurement Accuracy | Reading +/- 2% + 0.05NTU@25 ℃/77 ° F |
| Measurement Repeatability | +/- 1% of the reading or 0.02 NTU, whichever is greater |
| minimum detection limit | <0.05NTU |
| Measuring method | Turbidity assay in accordance with USEPA Method 180.1 and Standard 2130B |
| Calibrated mode | Manual calibration, built-in standard calibrating points: < 0.1, 15, 100, 500NTU |
| Optical inspection system | Custom Dedicated Receiver & Light source System |
| data management | USB-RS232 Dual Interface, 200 sets of measurement Data storage |
| Power mode | 4X1.5VAA battery; 15 minutes without any operation, it will automatically shut down. |
| Operating Environment | 0 to 50 ℃ (32 to 122 ° F), RH max 95%, non-condensing |
| Size Weight | Host size: 224x87x77mm Host Weight: 512g |