Introduction
This series of products of constant temperature and humidity testing machine is mainly used to simulate the harsh high temperature, low temperature or damp heat environment that may be encountered in the transportation, storage and use of electronic instruments, new materials, electricians, vehicles, metals, electronic products, aerospace materials, etc.; or in the environment of high temperature, low temperature and alternating humidity and heat, inspect the high temperature, cold resistance, damp heat resistance of materials, accessories or instruments and equipment, and the reduction of the damage life that may be caused.
Features
Cabinet structure
1. Inner box material: SUS304# stainless steel plate, which has the advantages of acid resistance, corrosion resistance and easy cleaning.
2. Enclosure Material: cold-rolled steel plate baking paint treatment.
3. Door handle: non-reaction, explosion-proof door handle, easy to operate, safe and reliable.
4. Thermal insulation system: ultra-fine glass fiber is used to fill the insulation area, and the connecting parts of the inner and outer boxes are made of non-metallic high-low temperature resistant materials to effectively reduce temperature conduction; The door seal is made of refined silicone rubber, so that there is no aging and hardening under high-low temperature conditions, and the sealing is more reliable; The humidification method is isolated humidification.
5. Moving and positioning device: high-quality fixable PU movable wheels are installed at the bottom of the machine, which can easily move the machine to the designated position and fix the casters.
6. Sample holder: SUS304# stainless steel punching and bending into a mesh, easy to use, adjustable spacing between material racks, two standard configurations, and uniform distribution of samples according to the load bearing of 25kg.
7. Test hole: open a test hole of Φ50mm, and attach the corresponding thermal insulation accessories and special sealing soft plugs
8. Observation window: double-layer vacuum strengthened glass, inner glued sheet conductive film heating and defrosting, clear observation of the test process.
Refrigeration system
1. The refrigeration unit adopts the Japanese Panasonic compressor, which is the leading manufacturer of the global sealed refrigeration compressor (the core of the refrigeration system) market, which has a relatively high degree of recognition in the market, and uses the American DuPont environmentally friendly refrigerant (R404a), and adopts the binary low-temperature circuit system design, which is convenient for different temperature ranges to use different compressors to work to increase the equipment.
2. The temperature and humidity control adopts P.I.D + S.S.R system with the same channel coordinated control, which has the function of automatic calculation, which can immediately correct the temperature and humidity change conditions, so that the temperature and humidity control is more accurate and stable.
3. The temperature and humidity are all controlled by high-precision controller, and the relative temperature and humidity performance is accurately set and displayed, and the resolution reaches 0.1°C
Heated humidification system
1. Nichrome alloy high-speed heating wire is used for heating
2. High temperature completely independent system
3. The humidity adopts stainless steel pipe, built-in nickel-chromium alloy high-speed heating wire and high-temperature magnesium oxide powder, which has fast reaction and long life
4. The heating and humidification systems are completely independent, and the humidity has an automatic water replenishment device
Control system
1. Precision liquid crystal display programmable temperature controller
2. Intelligent digital display PID microcomputer temperature control instrument
3. Touch setting, digital and direct display, humidity directly display percentage, temperature sensor using PT100 platinum resistance Tester
4. The temperature control adopts the thermal balance temperature control method
5. The programmable temperature controller with precision liquid crystal display can be used to do program tests, and there can be multiple groups of multi-segment programming operations, fixed values and programs are optional, and the time can be set arbitrarily
6. The main electrical components of the equipment, such as circuit breakers and AC contactors, are all under the brand of "CHINT", and other electrical accessories are made of well-known domestic brands
7. Convenient data processing: it can be connected to a printer or 232 communication interface, with a USB data transfer interface (U disk), displayed by a computer, and printed with temperature, humidity and time curve, which provides a strong guarantee for the storage and playback of data in the test process
Protection system
The overall equipment is over-temperature, the fan is overheated protection, the overall equipment is under-phase/reversed-phase, the refrigeration system is overloaded, the refrigeration unit is overpressurized, the overall equipment is timed, and there are other protections such as leakage, operation indication, and automatic shutdown after fault alarm.
Table of temperature and humidity ranges

Test criteria
1、GB2423.1-2008/IEC6008-2-1-2007
2、GB/T 2423.2-2008
3、GJB150.3A-2009
4、GJB150.4A-2009
5、GB2423.22-2008
6、GBT 2424.5-2006
7、GB/T 10589-2008
8、GB/T2423.3-2008(IEC68-2-3)
LUGONG ST-150LB Humidity ChamberSpecifications
| LIST | VALUE |
|---|---|
| Studio Volume | 150L |
| Interior size | 500*600*500mm |
| Dimension | 950mm×1520mm×1055mm(W×H×D) |
| Temperature range | -40℃~+150℃ |
| Temperature Resolution | 0.01℃ |
| Temperature deviation | ±0.5℃ |
| Temperature Uniformity | ≦ 2℃ |
| Temperature Fluctuation | ±0.5℃ |
| Rising and Cooling rate | Heating time: -40 ℃ → + 150 ℃ average 3 ℃/min, cooling time: 150 ℃ → -40 ℃ average 1 ℃/min |
| Humidification system | Stainless steel humidifier |
| Heating system | Nickel chromium alloy stainless steel heater |
| water supply system | Front-mounted water supply tank, fully automatic water supply system |
| communication interface | 1. R232/485 communication |
| setting method | Chinese/English interface, touch input |
| operation mode | Program method/valuation method |
| setting range | Temperature range 5 ℃ |
| Display resolution | Temperature: 0.01 ℃; Time: 0.1 min; |
| Power-off memory function | The power failure recovery mode can be set as: hot start/cold start/stop |
| input | platinum resistance |
| storage function | Standard configuration, recording test data Linear dispersion, can upload and download data, can expand storage Capacity |
| Software Operating Environment | IBM PC compatible machine, CPU above PII, memory above 128M, Simplified Chinese Windows2000 or Simplified Chinese WindowsXP operating system |
| Chassis shell | Cold rolled steel plate |
| Chamber material | SUS304 #stainless steel plate |
| Insulation material | 100Mm polyurethane rigid foam + 100mm high density fiberglass cotton |
| Standard configuration | φ 50 Test Hole, Lighting, Double-decked Glass Viewing Window |
| Weight | About 235kg |
| current | 25/11A |
| Rated power | 5.4kw |
| Power Supply | Single phase, 220V |
| Humidity resolution | 0.1%R.H |
| Deviation deviation | ±2%R.H |
| Humidity Uniformity | ±3%R.H |
| Humidity stability | ± 2%R.H |
LUGONG ST-150LB Humidity Chamber Packing list
Host X1, instruction manual X1, warranty card X1, equipment certificate X1, packing list X1, specimen rack tray X2, specimen rack support track hook X8
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.