Print Picutre Similar products View price

Factory NBC-LC-100L Hot and cold shock Test Chamber DATASHEET

Brand Factory Model NBC-LC-100L
From China Name Hot and cold shock Test Chamber
Specifications
Test temperature impact range -40 ℃~ + 150 ℃ (optional programming setting)
High temperature box temperature range + 60 ℃~ + 180 ℃ (optional programming setting)
Cryogenic box temperature range 0 ℃~ -55 ℃ (programmable arbitrary setting) test temperature range -40---150 degrees
Temperature transition time 10 seconds (time can be specified)
Controller control Accuracy ±0.2℃
box uniformity ±2
Volume L 100L
shell material Sheet metal paint
Inner box material SUS #304 Thickened heat-resistant and cold-resistant stainless steel plate, first-class light panel
Thermal Insulation High-density glass wool and high-strength PU foam insulation materials have good thermal insulation effect
Box structure This equipment is divided into: high temperature zone, low temperature zone
Air supply circulation system Japanese multi-wing centrifugal Circulating Fan, Japanese motor whole group
heater U-shaped fin type 304 #seamless steel high-speed heating electric heater.
Refrigeration compressor Adopt imported Compressor unit
Refrigeration accessories Table bend center force evaporator, Mar fan, etc
Refrigeration method Adopt air-cooled or water-cooled heat dissipation (water towers, pipes, the buyer installs them by himself)
Sample holder Hanging Basket
Test area size 500 * 500 * 400 (width * depth * height) * 2
Exterior size approx. Width 1270x Depth 2020x Height 1820mm
Electrical accessories Domestic well-known brand electrical components
Temperature and humidity control instrument Programmable platform Controller, touch screen input, Chinese and English switching. Different temperature points can be arbitrarily set according to customer requirements
wiring method The company's electronic control wiring method is implemented in accordance with international electrical standards
Protection device Compressor overheating, overtemperature, overload, high pressure, low pressure, oil pressure, overcurrent, under-phase, power failure standby and other alarm warnings.
Power Supply AC380V  50~60Hz    18KW
Meet the standard GB2423 [1] .02 High temperature test method, GB2423 [1] .01 Low temperature test method.
Packing list
host X1, manual X1, certificate X1, warranty card X1
After-sales instructions

Warranty days: 365 days

1. For product quality problems caused by instruments and equipment not artificially, the manufacturer will provide free warranty services during the warranty period;

2. There is no warranty period for fragile and consumable parts, and if you find product quality problems after receiving the goods, please apply for return and exchange within 7 days.


[Note] For after-sales services outside the Chinese mainland, if the instrument requires repair, the shipping and possible associated taxes shall be borne by the sender. Generation address of the manual: http://en.nbchao.com/p/50097/Specification.html , which is only used by users who purchase commodities from the NBCHAO.
Created on : 2025/12/31