German EPK MikroTest G6 McCourt coated platings Thickness Gauge, thickness measurement: 0~100μm
0~100μm thickness range; suitable for underwater thickness measurement, waterproof and explosion-proof; no need to calibrate, no power supply; EPK hot-selling mechanical Coating Thickness Gauge.
-
German EPK MiniTest 735 F5 iron-based Thickness Gauge, measuring thickness 0~5000μm
$ 1787.00 -
DONGRU DR320 Coating Thickness Gauge electromagnetic induction principle for magnetic metal substrate surface platings measurement
$ 304.00 -
LEEB leeb252 Coating Thickness Gauge magnetic Induction/eddy Current dual-use automatic identification substrate
Login -
Defelsko PosiTector 6000 FHXS3 high-precision Coating Thickness Gauge for hot or rough surfaces
$ 2113.00 -
List Magnetik Pocket TOP-CHECK FE-B Pocket platings Thickness Gauge
$ 1501.00
Introduction
The German EPK MikroTest G6 McCourt Coating Thickness Gauge accurately and unmistakably measures the thickness of paint and plating layers on steel.
The MIKROTEST Coating Thickness Gauge measures the thickness of all non-magnetic coatings on steel (e.g. lacquer, powder coat, plastic, zinc, copper, tin and nickel). Fast, accurate and non-destructive, MICROTEST has been the most widely used instrument for the automatic determination of coating thickness for more than 30 years. Germany's "know-how" shows that it has the highest level of magnetic cladding Thickness Gauge in terms of process technology and precision.
Apply standards
The EPK MikroTest G6 is compliant with DIN EN ISO2178, DIN 50982, ASTM B 499, E 376, ASTM D 1186, ASTM G12, ASTM B 530, BS5411, ISO2361.
Features
MIKROTEST OPERATES FULLY AUTOMATICALLY. It has the following very good characteristics when in use:
1) Automatic measurement will not occur misoperation;
2) Simple setup and measurement;
3)No calibration required;
4)No batteries or other power sources are required;
5)Automatically report the thickness reading;
6)With a non-destructive probe, a little bit of measurement;
7)The metal armor is suitable for frequent outdoor operation and use, and is protected from mechanical impact;
8)Resistant to mechanical shock, acid and solvent corrosion;
9)The balancing device eliminates the influence of gravity and can accurately measure in any direction and in the tube.
Pictures

Comparison of the MikroTest series
| Model | Measuring range | Readability | Minimum measuring area diameter | The minimum radius is convex and concave mm | Minimum thickness of the substrate | Suitable for: |
| MIKROTEST G 6 | 0-100μm | 1 μm or 5% reading | 20mm | 5mm;25mm | 0.5mm | Electroplating layer on steel, iron, lacquer, enamel, plastic, rubber layer |
| MIKROTEST F 6 | 0-1000μm | 5 μm or 5% reading | 30mm | 8mm;25mm | 0.5mm | |
| MIKROTEST S3 6 | 0.2-3mm | 5% reading | 30mm | 15mm;25mm | 1.0mm | |
| MIKROTEST S5 6 | 0.5-5mm | 5% reading | 50mm | 15mm;25mm | 1.0mm | |
| MIKROTEST S10 6 | 2.5-10mm | 5% reading | 50mm | 15mm;25mm | 2.0mm | |
| MIKROTEST S20 6 | 7.5-20mm | 5% reading | 100mm | 100mm;150mm | 7.0mm | |
| MIKROTEST NIFE50 6 | 0-50μm | 2 μm + 8% reading | 20mm | 10mm;25mm | 0.5mm | Nickel plating on iron |
| MIKROTEST NI50 | 0-50um | 1 μm or 5% reading | 15mm | 5mm;25mm | - | Nickel plating on a non-ferromagnetic metal substrate |
| MIKROTEST NI100 | 0-100um | 1 μm or 5% reading | 15mm | 5mm;25mm | - |
ElektroPhysik MikroTest G 6 McCourt platings Thickness GaugeSpecifications
| LIST | VALUE |
|---|---|
| Measurement principle | mechanical |
| Measurement range | 0~100μm |
| margin of error | ±5% |
| Minimum measurement surface | 20mm |
| Minimum radius of curvature | Convex 5mm; Concave 25mm |
| thinnest substrate | 0.5mm |
| probe | Integrated probe |
| Calibrated Certificate | none |
| Calibrated mode | No need to calibrate |
| Unit display | μm |
| statistical function | none |
| icon function | none |
| Interface | none |
| Data storage | none |
| Power Supply | No batteries or power required |
| Standards | DIN EN ISO2178, DIN 50982, ASTM B 499, E 376, ASTM D 1186, ASTM G12, ASTM B 530, BS5411, ISO2361 standards |
| Display screen | Dial |
| Operating temperature | -20~100℃ |
| Machine Weight | 560G (including instrument case) |
| Machine size | 215*55*29mm |
| Packaging Dimensions | 235*185*46mm |
ElektroPhysik MikroTest G 6 McCourt platings Thickness Gauge Packing list
MikroTest Thickness Gauge, Operating Instructions, Encasement Box
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- Tide Review: McCourt Coating Thickness Gauge MikroTest G 6
- What are the Thickness Gauges that can test the thickness of underwater coatings?
- Can the coating Thickness Gauge measure the thickness of pipes?
- McCourt Coating Thickness Gauge Learn about these advantages, you can't choose the wrong instrument!
- Precautions for the use of MiKroTest McCourt Mechanical Coating Thickness G6
- What are the coating Thickness Gauges suitable for measuring the thickness of the coating in explosion-proof occasions?
- Dry-film thickness measurement: magnetic and mechanical methods
- Precautions for the use of Mikrotest Thickness Gauges
- Key points for maintenance of coating Thickness Gauge
- German EPK company McCourt automatic cladding Thickness Gauge operating instructions
- ReplacedNF EN ISO 2808-2007
- ActiveGB/T 1727-2021
- ActiveHG/T 20720-2020
- ActiveISO 2808-2019
- ISO 2361-1982
- AbolishISO 2808-2007
- ActiveGB/T 13452.2-2008
- AbolishGB/T 11374-1989


ElektroPhysik
MikroTest G 6