MOLECULAR MU-20L Laboratory Ultrapure Water Machine Instruction Manual
Molecular molar ∑H2O
Moore Ultrapure Water Machine MU-20L
Install the user manual
Owner Manuals

Think in Water
This manual is mainly intended to help users correctly use and maintain our company's "Molecular" equipment, ensuring optimal operation and extending its service life. Based on a scientific approach, users are advised to read this manual carefully before using the "Molecular" device.
Operation panel display code description:
Code 1--- Low inlet water pressure prevents normal water production
Code 2--- Pure water tank pure water level is lower than the low level switch, causing the machine to draw water
Code 3--- No Cards
Code 4--- illegal card
Code 5--- When the service life expires, the consumables need to be replaced
Code 6--- Communication failure, check if cables or hardware are damaged
Code 7--- High Voltage Protection
Code 8--- Consumables are about to expire and need to be replaced promptly
Friendly reminder: When replacing the pretreatment filter element, to reduce the amount of residual water in the pretreatment filter bottle, first set the pure water machine to water generation or RO flushing mode, then close the tap water inlet valve. When the low-pressure switch alarms (the control panel shows low inlet pressure), turn off the main power switch and then replace the filter cartridge.
Important reminder:
1) This machine is a custom machine, specifically designed for water use in temperature-humidity chambers and salt spray tanks. Do not use it for other purposes! If you really need to use it for other purposes, please consult our technical hotline: 18964105168
2) Please replace the consumables of this equipment regularly according to actual usage conditions (see attached consumable replacement instructions and methods below) to ensure the equipment is in optimal operating condition.
3) Please order high-quality consumables specifically for this machine from our company. If users purchase other consumables from the market without authorization and encounter equipment malfunctions, we reserve the right to refuse repair and service!
4) After training by our company's professional installation engineers, this machine should be operated and maintained by dedicated personnel.
5) For safety, please turn off the power and tap water when not in use!
6) To ensure safety, it is recommended to install a cofferdam around the pure water equipment (10-15cm high), preferably with a drainage floor drain to prevent laboratory safety issues in case of a water leak. Especially when this machine is working at night without anyone around.
Table of Contents
| 1 Equipment Diagram | 3 |
| 2 Introduction to the basic principles of the equipment | 4 |
| 3. Operating conditions of the Moore pure water processor | 5 |
| 4 Installation steps for water purifier equipment | 5 |
| 4.1 Installation Requirements | 5 |
| 4.2 Installation | 5 |
| 5. Standard procedures for equipment operation and maintenance | 7 |
| 6 Steps for replacing consumables | 11 |
| 7 Determination and Methods for Consumable Replacement | 12 |
| 8. Diagnosis and handling of fault phenomena | 12 |
| Appendix: Basic Knowledge of Ultrapure Water | 14 |
1 Equipment Diagram
1.1 Equipment appearance

1.2 Equipment Operation Panel

2 Introduction to the basic principles of the equipment
PP fiber filter cartridge: Unique loose outside and tight inside, effectively removes suspended solids, sediment, and particulate impurities above 5 microns or 1 micron (determined by the pore size of the PP fiber filter element) from the water source; High-efficiency activated carbon filter cartridge: Possesses strong adsorption and filtration performance, effectively removing impurities such as discoloration, odors, and organic matter from distilled water.
Activated carbon filtration: Removes chlorine by chemical adsorption, and soluble organic matter by organic adsorption. Because reverse osmosis membranes are sensitive to chlorine and soluble organic compounds, carbon columns are often placed in front of the RO membrane to remove these substances.
Reverse osmosis membrane: Osmosis is when water flows from a low concentration to the higher concentration side through a semipermeable membrane. If a high-pressure pump is used to provide pressure to a high-concentration solution with a greater pressure difference than the osmotic pressure, water molecules will be forced through the semipermeable membrane to the side of the lower concentration; this step is called reverse osmosis. Reverse osmosis can filter out the vast majority of pollutants, 90%-99%. Due to its outstanding purification efficiency, reverse osmosis is a highly effective technology for water purification systems. Because reverse osmosis can remove most contaminants, it is often used as a pretreatment method, significantly extending the service life of the deionization column. Water treated with reverse osmosis is high-quality prepure water, suitable for routine use in many laboratories. Using a reverse osmosis membrane originally from the United States, it can effectively remove various salt ions, organic matter, bacteria, and viruses from water, with a total salt removal rate exceeding 99%, and can remove organic matter with molecular weights greater than 300 daltons.
Purification column: Uses American ultrapure water-specific ion exchange resin and special flow path design, offering high ultra-purification efficiency and effectively removing trace ions from water.
The machine's operation control uses an intelligent microcomputer, with touch buttons, operational status display, and online water quality monitoring, fully realizing human-machine interaction.
3. Operating conditions of the Moore pure water processor
| Operating voltage | AC220V ± 10%/50Hz, with reliable grounding |
| Power | 250W |
| Raw water temperature | 5-40°C |
| Tap water pressure | 0.15-0.25MPa (if it exceeds 0.25 MPa, customers need to take pressure reduction measures; otherwise, overpressure damage and leakage may easily occur; If the temperature drops below 0.1 MPa, equipment alarms will be triggered and shutdowns will occur, requiring additional booster pumps. |
| Tap water TDS requirements | TDS ≤ 200ppm. If this value is exceeded, please consult our technical or after-sales engineers to add enhanced pretreatment or higher-end pure water machines; otherwise, the output water quality may be substandard |
| Work environment | The installation site should ensure a well-ventilated and dry environment, with a temperature between 5-30°C; Especially in northern regions, in winter, prevent the instrument from freezing and damage; freezing of water inside the instrument can cause complete damage; Do not place the device in environments with high acidity |
4 Installation steps for water purifier equipment
4.1 Installation Requirements
4.1.1 Equipment should be installed in a well-ventilated and dry environment whenever possible.
4.1.2 The specifications of the inlet and drain pipes should meet the matching interface requirements for the purifier and be within 2 meters of the main unit.
4.1.3 Equipment must be placed stably, reliably, and correctly positioned.
★ 4.1.4 Drainage ditches or trenches, as well as cofferdams, should be provided next to equipment.
4.1.5 The system power supply and wiring must meet design requirements.
4.2 Installation
4.2.1 Check that the machine's packing list matches the actual parts;
4.2.2 Please disconnect water and power sources;
4.2.3 Connect the source water tee connector in the accessory to the customer's water supply point, then use a 2-inch white PE pipe to connect the external pretreatment inlet to the quick-plug outlet of the source water tee valve with valve, and finally connect the external pretreatment outlet to the "raw water port" at the rear of the machine;
4.2.4 Connect a 2-inch white PE pipe from the "wastewater outlet" at the rear of the machine into the drainage pipe (trough), inserting it to a length of about 0.5 meters;
4.2.5 Connect a 2-inch white PE pipe from the "pure water outlet" at the rear of the machine to the pure water outlet at the bottom of the pure water tank;
4.2.6 Connect a 2-inch white PE pipe from the "return port" at the rear of the machine to the return port at the bottom of the pure water tank;
4.2.7 Insert the "signal cable" connector at the top of the water tank into the signal cable interface on the back of the machine, then insert and tighten it;
4.2.8 Installation should be clean and tidy.
4.2.9 Plug the machine's power cord into a 220V socket that meets installation standards.
4.3 Equipment Installation Diagram

Schematic diagram of the piping connections at the rear of the main chassis
5. Standard procedures for equipment operation and maintenance
5.1 Moore Intelligent Ultrapure Water Processor Operation and Maintenance Standard Procedure
| Daily operation details | 1. Check that the tap water has sufficient pressure and that the valve is opened (pressure range: 0.15-0.25Mpa); 2. Connect the device's power plug to the 220V power supply (the main power button behind the instrument must be turned on). Our company logo name will be displayed on the screen. Press the start switch, and the screen will scroll to show the current time, circuit status, and more. The device automatically performs RO flushing, displaying "Power-on flushing"; After about 3 minutes, the equipment stops flushing and starts producing water, displaying "Water Production"; If the water level in the pure water tank has reached a high level, the device is in standby mode and displays "full"; 3. Water Drawing: Press the Level III or Level I water switch; if it shows "Level 3 Water Intake or Level 1 Water Draw," you can discharge pure or ultrapure water from the machine's water intake; Press the corresponding button again to stop the corresponding water intake. Grade 1 water can be set to fixed quality, fixed quantity, and timed water intake modes; 4 After work or when the device is not in use, press the start switch again. The LOGO name will appear on the screen. Turn off the device (the main power switch behind the instrument must be turned off), and unplug the power cord. 5 Operation Guide 5.1 Keyboard Instructions Keys have different key functions depending on the operating environment. Under normal working conditions, the following button functions are available: start, rinse, circulate, three-stage water draw, and first-stage water draw. In the query and settings environment, the following key functions are available: set, back, up, down, and modify. Settings key: Used to enter the settings menu, confirm and return key: return to the upper menu. Up key: used to move the cursor or switch pages in the same layer. Down key: used to move the cursor or switch pages in the same layer. Modify key: used for value modification. 5.2 Power on the machine, connect all components as instructed, and carefully confirm everything is correct. Then turn on the external power supply (it is strongly recommended to install an external main power switch). The display screen is in standby mode. 5.3 User Settings 5.4 Main Menu The main menu consists of two pages. Press the set button in standby mode to display the main menu as shown in Figure 5-1: |

1. User settings 2. Information query 3. Record query 4. System query
Figure 5-1
Press the modify button to enter the next main menu page, as shown in Figure 5-2
5. Time setting 6. Filter element cleaning alarm 7. RO membrane cleaning alarm 8. Purification column alarm
Figure 5-2
In the main menu interface, pressing the back button returns to the standby screen.
5.5 User Settings
In the menu interface, press down or up to point the cursor to the "User Settings" item, then click the Settings key to confirm to enter the user settings interface, as shown in Figure 5-3:
Water intake mode: Fixed 100L
Water intake for fixed quality: None
Language: Chinese
Figure 5-3
Press the Settings key to bring up the cursor. In the corresponding cursor indicator, press the Modify key to modify the settings.
There are three water intake modes: random, timed, and quantitative. When the cursor points to the water intake mode, pressing the modify button allows repeated selection. When selecting timed or quantitative mode, corresponding water intake time or volume parameters will be displayed. Users can press the down button to move the cursor (after reaching the bottom of the screen, press down to return to the starting position) to the corresponding value, then press the modify button to modify (only increases in one direction; if the value exceeds 9, it returns to 0).
Fixed quality water intake can be selected with or without a fixed quality water extraction value. When selected, the corresponding fixed water extraction value can be set. After completing the corresponding item settings, press the settings button to save the settings data. The cursor disappears, and pressing the back button returns to the main menu.
5.6 Information Query
On the main menu interface as shown in Figure 5-1, when the cursor points to information query, press the settings button to enter the information query page, which displays the current filter usage time, current water intake, current circuit status, and current alarm status.
5.7 Record Query
On the main menu interface as shown in Figure 5-1, when the cursor points to record query, press the Settings button to enter the secondary menu page. Record queries include water sampling records and alarm records. Move the cursor to the corresponding menu and press the settings button to query the corresponding records.
There are 20 water intake records, each displayed vertically or upward. Each record includes: current water intake time, current water extraction volume, and current water quality. See Figure 5-4
The first water sampling record
Time: 08-11-12 08:50
Water quality intake: 18.23M
Water Withdrawal Capacity: 020L
Figure 5-4
There are 20 alarm records in total, each displayed downward or upward. Each record includes: the time of the alarm, the total water intake at that time, the alert content at that time, and more. See Figure 5-5
Record of the first alarm
Time: 08-11-12 08:50
Total water withdrawal: 123,456.78L
Current alarm: Filter element
Figure 5-5
5.8 System Information Query
On the main menu interface as shown in Figure 5-1, when the cursor points to system query, press the Settings button to enter system settings information inquiry. There are three screens displayed. Press the down or back buttons to flip the screen, as shown in Figures 5-6, 5-7, 5-8:
Automatic flushing interval: 10 hours
Automatic rinse time: 2 minutes
Excessive discharge: 05μs/cm
Water quality maintenance: 10M
Figure 5-6
Primary water flow rate: 023.23 mL/min
Third-stage water flow rate: 025.45 mL/min
Filter lifespan: 600 hours
Figure 5-7
Conductive: D=200
Flow meter: C=10.25L/P
Thermometer: Yes, UV: None
TOC meter: None
Figure 5-8
The content displayed in system information queries cannot be modified and can only be set via the RS232 communication port on the backend computer.
5.9 Working Interface
Press the start button while in standby mode to enter the system interface. See Figure 5-9:
Current alarm: None
Automatic flushing
Primary water intake 18.19M
Water is drawn in three stages
Figures 5-9
The first line alternately displays system information such as circuit board status, time, temperature (if temperature is configured), and current alarm status.
The second line displays the water production status. Depending on the working state, it may show information such as full water, automatic flushing, manual flushing, water creation, and low inlet pressure.
When starting, the system powers on and automatically flushes for 2-3 minutes, then checks the water tank level. If the tank level is insufficient, the water generation process begins.
Press the flush button to manually flush the system. Press the flush button again to enter manual discharge mode. Press the flush button again. If the water tank is full, it will show full; otherwise, it will enter automatic flushing mode before water generation.
Press the three-stage or primary water intake button to activate the third or primary water function, then press the third or primary button again to stop the corresponding water intake function. When the primary water intake function is activated, the following situations may occur:
A. During the water intake function, press the first-stage button to enter automatic circulation. Only when the water quality reaches the system's set value can the first-stage water intake valve automatically open. Otherwise, the automatic cycle timeout (about 15 minutes) will clear the water fetch command.
B. After meeting condition A, three water intake methods can be set: timed, quantitative, or random. Timed and quantitative methods: the primary water intake valve automatically closes after reaching the user-set water intake time or volume.
6 Steps for replacing consumables
First, close the tap water inlet valve. When the low-pressure switch alarms, turn off the power switch. This reduces the residual water in the filter and facilitates consumable replacement. Then open the movable doors on both sides of the chassis to replace the equipment using simple tools:
6.1 Primary Preprocessor Filter Element: Use a belt wrench to loop the cylinder labeled with the first-stage preprocessor filter (PP filter) and rotate counterclockwise. Slowly loosen the first-stage preprocessor filter cartridge, remove the cylinder, remove the internal filter element, clean the cylinder (be careful not to remove the black rubber ring inside the cylinder), then reinstall the new first-stage preprocessor filter (PP) cartridge back into the cylinder, making sure the center hole of the PP filter element aligns with the cylinder's center hole, then install the cylinder onto the fixed end cap interface. Use a belt wrench to fit the drum, rotate clockwise to tighten it, refer to Diagrams 1 and 2.
6.2 Secondary and Stage 3 Preprocessor Filter Cartridges: The replacement of the Secondary (active activated carbon) and Stage 3 (fine filtration) preprocessor filter elements is the same as that of the Primary Preprocessor Filter (please note that the black rubber gasket on the activated carbon must not be removed).
6.3 Unscrew the clip fixed to the integrated ultra-purification package mu6, remove the two PE pipes attached to the top of the purification package, remove the ultra-purification package, then connect the newly arrived ultra-purification packet to the original position and connect the PE tube.
6.4 Replacement of the Reverse Osmosis Membrane: Understand the inlet and outlet directions of the membrane's three interfaces, paying special attention to not confusing the pure water port with the wastewater outlet. Then loosen the clamp or single clamp of the fixing membrane, connect the newly arrived reverse osmosis membrane, and secure all corresponding interface tubes.
Note:
(1) Install in order and assign seats according to the assigned numbers.
(2) All components must be tightened and locked in place; otherwise, the equipment may leak during the water production process.
(3) After installation, connect the power to rinse and run. If any leaks are found, tighten the connectors promptly and confirm there are no leaks and that the water quality meets standards before use.

Figure 1: Usage method of a belt wrench

Figure 2: Belt wrench fits the drum
7 Determination and Methods for Consumable Replacement
| Name | Model specifications | Standard amount | Replacement interval | Determination method |
| Primary pretreatment filter element | Molcot 10 | 1 piece | 1-2 months | You can observe the filter through the cylinder body to see if it turns dark yellow and then replace |
| Secondary pretreatment filter element | Molgac10 | 1 piece | 4-6 months | If the local water quality is poor and there is a lot of drifting powder, the replacement time should be shortened |
| Third-stage pretreatment filter element | Molpp10 | 1 piece | 4-6 months | If the local water quality is poor and there is a lot of sediment and other particulate matter, the replacement time should be shortened |
| Reverse osmosis membranes | Molrofilm | 1 piece | About a year, replace all three at once | It can be tested for Class III water conductivity; if the conductivity > 10μs/cm, replacement is required |
| Integrated ultra-purification pack | Moluppack mu6 | 2 pieces | If the equipment control panel displays resistivity <10MΩ·cm, replacement is required; Or replace according to actual resistance requirements. | |
| Note: The actual service life of consumables depends on the quality of the influent water. The replacement interval varies depending on the quality of tap water. | ||||
8. Diagnosis and handling of fault phenomena
| Serial number | Common faults | Cause analysis | Solution |
| 1 | System protection appears on startup, device alarm triggers | 1. The inlet valve is not open 2. Tap water pressure below 0.1Mpa 3. Filter cartridge is clogged 4. The inlet solenoid valve is not fully open | 1. Open the water inlet valve 2. Adjust tap water pressure 3. Replace with a new filter element 4. Replace with a new inlet solenoid valve of the corresponding model |
| 2 | The main unit shows that when the water is full, the pure water tank is overflowing | 1. Foreign objects are stuck in the inlet solenoid valve 2. The overflow outlet on the pure water tank is not connected to the water tank | 1. Remove debris and replace the solenoid valve 2. Use white PE pipes to connect the overflow outlet to the water tank |
| 3 | After pressing the water draw button, no corresponding water intake text appears, and water cannot be drawn normally | 1. The pure water tank water level is lower than the low level switch 2. Loose liquid level signal line | 1. Wait until the water level in the water tank rises before taking it 2. Reconnect the signal cable |
| 4 | After pressing the water fetch button, the corresponding water fetch text appears, but normal water fetch is not possible | 1. Water intake solenoid valve is faulty 2. The smart controller is faulty 3. Air enters the system | 1. Replace with a new solenoid valve 2. Replace with a new smart controller 3. Place the water tank above the main unit to release air |
| 5 | Plug in the power cord, but the display screen shows no signs and cannot turn on | 1. The socket is dead 2. Fuse blowing 3. The gray cable on the smart controller is loose 4. The regulated power supply has a 220V input and no 24V output | 1. Replace with an electrical outlet 2. Replace the fuse 3. Reconnect the gray cable 4. Replace with a new regulated power supply of the corresponding model |
| 6 | Water production decreases | 1. Pretreatment filter cartridge blockage 2. RO membrane blockage 3. Weather temperature below 10°C | 1. Replace the pretreatment filter element 2. Replace the RO membrane 3. Once temperatures rise, water production will naturally increase |
| 7 | Effluent quality decreases | 1. Cycle time is too short 2. Consumables expire 3. The probe of the resistance or conductance meter is faulty | 1. Extend the cycle time before use 2. Replace consumables 3. Replace the corresponding probe |
| 8 | The machine does not produce water | 1. Incorrect piping connection 2. The machine won't start without power 3. Consumables expire 4. The booster pump does not operate 5. Smart controller is faulty | 1. Properly connect the pipelines 2. Replace circuit components 3. Replace consumables 4. Replace the booster pump 5. Replace the smart controller |
Appendix: Basic knowledge of ultrapure water
Classification of ultrapure water:
Class I water: testing time requires minimal interfering and maximum purity and precision, such as atomic absorption and trace element analysis.
Class II water: When tested with tolerable bacteria, such as preparing general reagents that do not require storage.
Grade III water: general water for washing, water for preparing high-purity pure water, water for preparing microbial culture media, etc.
Special water use: During operation, special contaminants must be removed, such as removing heat sources in tissue/cell cultures and trace organic matter in HPLC.
Storage of ultrapure water:
Storing purified water in a typical storage tank absorbs environmental impurities, causing water quality to deteriorate. Additionally, light can induce microbial growth of various impurities, and microorganisms can seriously affect downstream ultrapure systems. Therefore, the storage unit should be cleaned regularly with pure water.
Issues with ultrapure water detection:
Users easily point out common issues in measuring the conductivity of pure and ultrapure water: slow electrode response, errors, instability, flow rate effects, and short electrode service life. In fact, all of this is just a matter of appearance—helplessness and problem-solving. We have summarized the problems from different causes into four aspects:
Flow sensitivity: Different flow rates may cause resistance differences in ordinary measuring devices, and changes in flow rate will also affect the measured values.
Poor long-term stability: After a period of use, the electrode performance changes significantly.
Susceptibility to Interference: Various external disturbances and Jasenic disturbances from pure water flow also make measurement values unstable
25°C conversion: The requirements for (ultra) pure water pH measurement are very strict, with a narrow qualification range, so the measured value must be converted to 25°C.
Ions in ultrapure water are removed to trace levels. In a sealed state, low conductivity or specific resistance values can be detected. However, once ultrapure water is removed from the water intake, gases in the air, especially CO2, quickly dissolve into the ultrapure water, forming carbonate ions and increasing its conductivity. Therefore, the conductivity monitored in non-sealed conditions can only serve as a reference. Testing ultrapure water purity mainly relies on precision instruments such as online resistors or mass spectrometers, atomic absorption, and other instruments.
Expression of ultrapure water:
Due to the special nature of ultrapure water, in professional practice ultrapure water is generally expressed by resistance, while conductivity is usually expressed as the conductivity of water or solutions above 10 us/cm.
- 1YUEPING SH-10A Moisture Rapid Tester Instruction Manual
- 2GREELOY drying tower Y double tower instruction manual
- 3JIAHANG JL series thermostatic water bath instruction manual
- 4ANGNI AD200L-H/AD300L-H (Digital Display) High Speed Dispersive Mulser Instruction Manual
- 5ANGNI AD Series High Speed Dispersible Mulser Instruction Manual
- 6HUITAI LHS-SC/HC Series Constant Temperature and Humidity Chamber Instruction Manual
- 7HUITAI LRH liquid crystal biochemistry cultivating box instruction manual
- 8YUEPING NDJ-1 rotational viscometer operating instructions
- 9YUEPING DV-1, Digital Viscometer Instruction Manual
- 10YUEPING NDJ-4 rotational viscometer instruction manual