HUITAI LHS-SC/HC Series Constant Temperature and Humidity Chamber Instruction Manual
Constant temperature and humidity incubation chamber
● Please make sure to deliver this user manual to the final operator!
● Before starting operations, please read and understand all the contents of the user manual. Our company is not responsible for any adverse consequences caused by misoperations.
● The contents of the user manual may be changed or improved in the future. The company will not notify you separately at that time. Thank you for your understanding.
● If customers notice any abnormalities during use, please contact our company promptly.
● Please read the user manual carefully for easy reference, and be sure to complete and send your warranty card (sent to our after-sales service center).
| Product models | |
| Serial number | |
| Manufacturing date |
Table of Contents
(1) Product Applications...... 1
(2) Structural Introduction...... 1
(3) Working Principle...... 1
(4) Technical Indicators...... 1
(5) Working Conditions...... 2
(6) Usage Instructions...... 2
(7) Precautions...... 5
(8) Common fault symptoms and handling methods...... 6
(9) Electrical Connection Diagram ...... 7
Safety Tips:
1. The constant temperature and humidity incubator must be effectively grounded
2. The externally supplied power supply must meet the requirements of the constant temperature and humidity incubator
3. Use plugs and sockets with good contact; the power must match
4. Do not arbitrarily extend or cut the power cord.
5. Do not heat flammable, explosive, volatile, or corrosive substances
6. Do not perform repairs without authorization. Repairs must be done by professional technicians
7 When replacing the fuse, the power cord must be unplugged
8. Must fully read and understand the product manual before proceeding
1. Product Applications
This product is an environmental testing device, suitable for various small-scale electrical tests, instruments, materials, and other damp heat tests. This test chamber adopts a currently reasonable structure and a stable, reliable temperature control method. Temperature and humidity are displayed directly, with an attractive appearance, easy operation, and high temperature and humidity control accuracy. It is an ideal device for constant temperature and humidity experiments
2. Structural Overview
1 Casters 2 Boxes 3 Magnetic seal strips 4 Sink rack 5 Sink 6 Hanging screws 7 Temperature control box
8 Side Sticker 9 Power Switch 10 Front Door 11 Downwind Cover 12 Power Cord Socket 13 Rear Seal Panel of the Cabinet
14 Water storage tank 15 Rear seal plate 16 Water outlet 17 Overflow outlet
This series of constant temperature and humidity boxes consists of the enclosure, temperature control device, heating and refrigeration system, humidification system, and gas circulation air duct. The box chamber is made of mirror-finished stainless steel stamping, with curved edges for easy cleaning. The casing is made of high-quality steel plate, with a spray-coated surface. The height of the mesh panel in the studio can be adjusted freely
Polyurethane foam board is filled between the studio and the cabinet with excellent thermal insulation, providing excellent thermal insulation. The temperature control device mainly consists of a temperature controller, a temperature sensor, and a humidity sensor. The temperature controller has functions such as over-temperature protection and power-off protection, as well as a timer function for gas circulation air ducts. This constant temperature and humidity chamber features a well-designed circulating air duct, maximizing the uniformity of temperature inside the chamber.
3. Working Principle
When the signals measured by temperature and humidity sensors are transmitted via cables to the main controller, the CPU inside the controller makes correct judgments, outputs heating or humidification signals, and controls the corresponding devices to reach the set value. This series of products adopts internationally advanced fuzzy PID control methods, offering high temperature control accuracy, minimal temperature and humidity fluctuations, and uniform temperature.
4. Technical Indicators
| Model | Constant temperature and humidity incubator | ||||
| LHS-100CH | LHS-100HC | LHS-150HC | LHS-250HC | LHS-300HC | |
| Temperature control range | RT+10~85°C | 0~60°C | |||
| Temperature resolution | 0.1°C | ||||
| Temperature fluctuation | ±1.0°C | High temperature ± 0.5°C, low temperature ± 1.0°C | |||
| Humidity control range | 80~95%RH | 60~85%RH | |||
| Humidity deviation | ±5~8%RH | ±3RH | |||
| Power supply voltage | 220V 50Hz | ||||
| Input power | 1100W | 1400W | 1750W | 2000W | 2500W |
| Inner pot dimensions (mm): W*D*H | 450*400*600 | 450*400*600 | 600*400*630 | 650*550*700 | 680*600*750 |
| Dimensions (mm) W*D*H | 580*550*1150 | 580*550*1150 | 7501*750*1180 | 800*900*1350 | 830*950*1400 |
| Volume | 100L | 100L | 150L | 250L | 300L |
| Carrying goods and supporting shelves | 2 pieces | 3 pieces | |||
| Timed range | 1-9999min | ||||
| Note | Performance parameter tests under no-load conditions, no strong magnetism, no vibration: ambient temperature 20°C, humidity 50% RH | ||||
Operating curve of the constant temperature and humidity chamber
5. Working conditions
1) Temperature: 15°C~35°C.
2) Relative humidity: not exceeding 85% RH.
3) Power supply: AC220V frequency 50Hz±1Hz
4) No strong sunlight nearby, no corrosive gases, good ventilation, no strong vibration sources or strong electromagnetic fields
6. Instructions for Use
1. Schematic diagram of the constant temperature and humidity chamber panel
2. Operating steps
1. Introduction to buttons and indicator lights
1. Key: On the main interface, click to toggle between display measurement value and runtime value; Long press to enter the program settings interface. In the settings interface, click down to switch values, then long press to exit the settings interface
2. Key: On the main interface, click to switch between display set value and set time value. Enter the parameter settings interface. In the settings interface, switch the cursor position
3. Keys: On the main interface, click on the lighting switch, long press to manually defrost, in the settings interface, decrease value or switch parameters downward, flip down page;
4. Key: On the main interface, tap Backlight to toggle; In the settings interface, add values or switch parameters upward, then scroll up;
5. Press and hold the button for 2 seconds to run or stop the program
2. Interface Introduction
1. In normal mode, the instrument displays the main interface;
Temperature: 30.0°C Humidity: 70.0% Illuminance: Level 6 [Stop operation!] ] Duration: 10 Segments: 30 Hours: 100 [Stop running!] 】
2. On the main interface, click the key to enter the query interface. If you go 60 seconds without pressing a button, you will automatically return to the main interface;
Query set value: Temperature: 30.0°C Humidity: 70.0% Illuminance: Level 6 Query set value Period: 10 Number of stages: 30 Time: 100
3. On the main interface, press and hold the key for 4 seconds to enter the settings interface. On this interface, press and hold the key for 4 seconds to return to the main interface and save the set values; 60 seconds without pressing a button, automatically returns to the main interface, setting values without saving them;
Total cycle: 99 Total sections: 30 Reservation time: 120 Set temperature: 90.0°C Set humidity: 90.0% Set lighting: Level 6 Set time: 100
4. On the main interface, press and hold the key for 4 seconds to enter the parameter password interface; Different parameters correspond to different passwords. Enter the correct password, click the key to enter the corresponding parameter interface; If the password is entered incorrectly, press the key to return to the main interface;
User parameters: 00 Temperature parameters: 00 Humidity parameters: 00
5. In the parameter password interface, enter the correct password, click the key to enter the corresponding parameter interface; In this interface, press and hold the key for seconds to return to the main interface and save the set value; If you press a button for 60 seconds, it will automatically return to the main interface, and the set value will not be saved
Communication address: 1 Power-off protection: 1 Control method: 2
6. On the main interface, press and hold the and keys simultaneously for 4 seconds to enter the internal parameter password interface, enter the correct password, and click the key to enter the corresponding parameter interface; If the password is entered incorrectly, press the key to return to the main interface;
Sampling parameters: 00 Defrost parameters: 00 Internal parameters: 00 Ambient temperature: 00
3. Introduction to Use
Example: The equipment needs to start working after 120 minutes. It operates for 10 cycles, with 2 stages set for each cycle. The first stage requires 720 minutes of operation, with temperature controlled at 30°C, illuminance at level 6, and humidity controlled at 60%RH; The second stage lasts 1200 minutes, with temperature controlled at 10°C, illuminance at level 4, and humidity controlled at 50%RH.
Press and hold the key for 4 seconds to enter the main settings interface;
Click the key to switch the cursor to the total period value, use the and keys to change the total period setting value to 10; then click the key to hover the total number of segments and change the total segment setting value to 2; click again to hover the scheduled time value and change the scheduled time to 120 minutes;
Then click the set key to enter the first setting interface. Hover the cursor on the temperature value and change the temperature setting to 30°C. Click the set key to set the humidity value, change the humidity setting value to 60% RH, then click the key to hover on the illuminance value and change the illuminance setting to level 6; Then press the key to hover the time value and change the time setting to 720 minutes;
Then click the key to enter the setting state in section 2, and follow the above method to sequentially change the set values for temperature, humidity, illuminance, and time to 10°C, 50% RH, Level 4; Finally, press and hold the key for 4 seconds to exit the settings interface, return to the main interface, and set up;
Press and hold the button for 2 seconds, and the controller will start timing according to the scheduled time. The running status will display: Scheduled timing... After 120 minutes, the program will start running the first segment;
After the program enters the first segment, if the timing mode in the internal parameters is 0, timing will start. The running status shows timing is starting... If the timing mode is 1, the running status will be displayed as: Running... When the temperature reaches the set value, timing begins; If the timing method is set to 2, timing starts when both temperature and humidity reach the set value;
After the first segment ends, it enters the second segment; after the second segment finishes, it enters the first segment of the second cycle, and so on; When the second segment of the 10th cycle finishes, the program finishes, all outputs are turned off, and the buzzer sounds (mute by pressing any key).
Note: When humidity is set to 0, both humidification and dehumidification are disabled. It can be used as a biochemical chamber
Parameter introduction
User parameters: Password 3
| Parameter name | Parameter and function descriptions | Range (factory parameters) |
| Mailing address | When connected to a computer, the communication address of the instrument should be used; when multiple instruments are connected to a computer, the communication address should not be duplicated; | 1~16(1) |
| Power-off protection | 0: Turn off the power-off protection function; 1: Enable power-off protection. After the instrument is powered off and restored, the program continues to operate according to the cycle, number of segments, and time during the power outage; | 0~1(0) |
| Control methods | 0: The system determines whether the compressor is normally open or disconnected based on the normally open temperature parameter + ambient temperature and normally open humidity; In disconnectable mode, the opening and closing points are automatically calculated based on ambient temperature; 1: The system determines whether the compressor is normally open or disconnected based on the normally open temperature parameters; In disconnectable mode, the opening and closing points are automatically calculated based on ambient temperature; 2: Compressor disconnect type, where the compressor controls cooling based on the set values for turning on and off cooling in the temperature parameters; Dehumidification is controlled according to the set values for turning dehumidification on and off in the humidity parameters; | 0~3(1) |
Temperature parameters: Password 4
| Parameter name | Parameter and function descriptions | Range (factory parameters) |
| Overtemperature alarm | When the temperature measurement > set temperature + over-temperature alarm, the instrument alarm light turns on, the heating protection relay is cut off, cooling is turned on, and a buzzer sounds; | 5.0~50.0 (20.0)°C |
| Zero position adjustment | Zero position adjustment = mercury thermometer reading - displayed temperature; | -99.9 ~ 99.9(0)°C |
| Full level adjustment | If the actual temperature deviates, the parameter value can be adjusted to correct it. Full degree adjustment = 1000 * (mercury thermometer reading - displayed temperature) / display temperature; | -999~999(0) |
| Scale band | Time proportional effect adjustment. | 0.0~90.0 (15.0) |
| Scoring time | Integration regulation. | (1~9999 seconds) 600 |
| Differential time | Differential action regulation. | (0~9999 seconds) 200 |
| Cycle time | Heating control cycle. | 0~60 (5) |
| Turn off the heating | When the measured temperature > = set temperature + heating is turned off, the heating output is turned off; | -50.0~50.0(2.0)°C |
| Normally open temperature | When the compressor control mode is set to 0, if the set temperature > = ambient temperature + normally open temperature, the compressor will automatically switch to disconnect mode and calculate the on/off point based on the ambient temperature; When the compressor control mode is 1, if the set temperature > = normally open temperature, the compressor automatically switches to disconnect mode and calculates the on/off points based on ambient temperature; | -50.0 ~ 50.0(50.0)°C |
| Turn on the cooling | When the compressor control mode is set to 2, the cooling activation point is: Measured temperature > = set temperature + Cooling start; | -20.0~20.0(0.5)°C |
| Turn off the refrigeration | When the compressor control mode is set to 2, the cooling shutdown point is: measured temperature < = set temperature + cooling off mode; (The cooling point should be less than the cooling point turned on.) | -20.0~20.0(0.2)°C |
| Cooling delay | Compressor delay protection time, interval time between startup and last stop > = delay time; | 0~60 (3)Min |
Humidity parameters: Code 5
| Parameter name | Parameter and function descriptions | Range (factory parameters) |
| Over-humidity alarm | When humidity measurement > = set humidity + over-humidity alarm, turn off the humidification protection relay. | 0.0~50.0 (20.0)%RH |
| Humidity at zero | Zero position adjustment = hygrometer reading - displayed humidity; | -99.9~99.9(0)%RH |
| Humidity is at full level | If actual humidity deviates, this parameter value can be adjusted for correction. Full degree adjustment = 1000 * (hygrometer reading - displayed humidity) / displayed humidity; | -999~999 (0) |
| Scale band | Time proportional effect adjustment. | 0.0~90.0 (15.0)%RH |
| Scoring time | Integration regulation. | (1~9999) 200 seconds |
| Differential time | Differential action regulation. | (0~9999) 30 seconds |
| Cycle time | Humidification control cycle. | 0~60 (5) seconds |
| Turn off and add humidification | When humidity measurement > = set humidity + humidification is turned off, the humidification output is turned off | -50.0~50.0(-2.0)%RH |
| Keep the humidity on at a constant level | When the compressor control mode is set to 0, if the humidity is set to < = normally open humidity, the compressor will automatically switch to the disconnecting mode. | 00.0~90.0(10.0)%RH |
| Turn on dehumidification | When the compressor control mode is set to 2, the dehumidification activation point is: Measure humidity > = set humidity + start dehumidification; | -20.0~20.0(5)%RH |
| Turn off dehumidification | When the compressor control mode is set to 2, the dehumidification shutdown point is: Measured humidity < = set humidity + dehumidification turned off; (The dehumidifier should be turned off less than the dehumidifier turned on) | -20.0~20.0(2)%RH |
| Low-temperature protection | When the measured temperature falls below the low-temperature protection, turn off humidification and dehumidification, and the alarm light will turn on. | -20.0~60.0(5.0)°C |
Defrost parameters: Password 9
| Parameter name | Parameter and function descriptions | Range (factory parameters) |
| Defrosting method | 0: Solenoid valve defrost method; During defrosting, the solenoid valve is open and the fan stops; 1: Heating defrosting method; During defrosting, the solenoid valve opens, heating outputs, the fan stops, and the compressor stops; | 0~1(0)°C |
| Interval one | Defrosting interval time at 0.0~9.0°C, unit hour | 0~200 (3) hours |
| Interval two | 9.1~15.0°C defrost interval, unit hour | 0~200 (3) hours |
| Interval three times | The defrosting interval at temperatures above 15.0°C, measured in hours | 0~200 (6) hours |
| Defrost one | Defrosting output time at 0.0~9.0°C, unit seconds | 0~300 (60) seconds |
| Defrost 2 | Defrosting output time at 9.1~15.0°C, unit seconds | 0~300 (60) seconds |
| Defrost three | Defrosting output time above 15.0°C, unit seconds | 0~300 (60) seconds |
| Manual defrosting | 0: Automatic defrosting when the compressor is normally open; 1: On the main interface, press and hold the button for 8 seconds to start defrosting. Defrosting time depends on the parameters: Defrost 1, Defrost 2, Defrost 3; | 0~1(0) |
Internal parameters: Password 101
| Parameter name | Parameter and function descriptions | Range (factory parameters) |
| Temperature limit | Set the upper temperature limit | 0.0~90.0 (60.0)°C |
| Lower temperature limit | Set the lower temperature limit | -40.0~0.0 (0.0)°C |
| Instrument type | 0: Artificial climate box; 1: Constant temperature and humidity chamber; 2: Light box | 0~2(0)°C |
| Display language | 0: Chinese display; 1: English display | 0~1(0) |
| Timing method | 0:0~9999 minutes; 1:0~99 hours: 0~59 minutes | 0~1(0) |
| Timing judgment | 0: Timing starts as soon as it runs; 1: Timing begins when the temperature reaches the set value; 2: Start timing when both temperature and humidity reach the set values | 0~2(0) |
| High temperature control | When the temperature setting > high temperature control, the compressor only turns on when an over-temperature alarm occurs | 0.0~99.9(50.0) |
| Humidity deviation | Reserve parameters | 0.0~99.9 (2.0) |
Ambient temperature: Password 18 (When the ambient temperature chip fails, the alarm indicator flashes)
7. Notes
The water tank of the constant temperature and humidity chamber should be filled with purified water; tap water is strictly prohibited.
It is strictly forbidden for the water level in the water tank to fall below the warning level during the operation of the constant temperature and humidity chamber
During transportation and handling, artificial climate incubators must not be placed upside down or at an angle greater than 45 degrees. After transport, they should be kept still for more than 5 hours before starting operation (consequences caused by failure to follow this procedure are borne by the incubator).
1. After the equipment is installed, if the ground is uneven, it should be leveled. Leave at least 300mm of space on the left and right sides and back of the equipment. If using the equipment at ambient temperatures above 30°C, it is recommended to strengthen ventilation measures behind the equipment and lower the ambient temperature (for example: installing air conditioning to keep the operating temperature around 25°C) to prevent compressor failure due to overheating
2. During normal operation, the equipment should be neatly arranged inside the box and must not be overcrowded. This should not affect air circulation inside the chamber to ensure uniform temperature
3 The equipment should be kept away from sources of electromagnetic interference, and its ground wire should be grounded
4 When the incubator chamber temperature approaches the set temperature, the heating indicator light flickers on and off, repeating multiple times, which is normal. Generally, about 1 hour after the measured temperature reaches the controlled temperature, the temperature inside the workroom enters a constant state
5. The inner wall of the cabinet and the surface of the equipment should be wiped regularly. First, keep them clean and increase the transparency of the glass; Do not use acids, alkalis, or other corrosive solutions to wipe the exterior surface
10. If the equipment will not be used for a long time, unplug the power cord and wipe off any accumulated water inside the cabinet. Run regularly (generally quarterly) according to usage conditions for 5 hours, set the temperature to 40°C, and open the door every 2 hours to release moisture. After treatment, unplug the power cord and store it to remove moisture from electrical parts and avoid damaging related components
11 If equipment malfunctions, professional personnel or the manufacturer should repair and resolve it
8. Common fault symptoms and handling methods
| Fault phenomena | Speculation on the cause of the fault | Troubleshooting methods |
| No power on (indicator light not on) | Power outlets have no power or poor plug contact | Repair |
| The box's power cord is broken or not properly plugged in | Repair, replug | |
| The fuse blows | If the motor burns out after replacement, check for short circuits (if the motor is faulty, it may have odors or black wires) or the casing (insulation resistance is zero). Repair it before restarting. | |
| Power switch is broken (or not turned on) | Replace the power switch | |
| The instrument panel shows no display | Wiring is detached or instrument panel is burned out | Reconnect the wires or replace the instruments |
| The instrument panel displays "----" | Sensor faulty or wiring end (detached) | Repair or replace PT100 (100 ohms at 0°C, 0.3 ohms/°C) |
| The temperature value exceeds the controlled range | Use appropriate temperature values within the specified range | |
| Temperature goes out of control and AL1 alarm light turns on | AL1 set to 0 or incorrect, bidirectional thyristor breakdown, thermostat faulty (AL1 alarm not lit) | Reset and replace (model: BTA16) with the thermostat |
| It does not heat up | Thermostat is broken (no power supply, HEAT light not working) | Replace the thermostat |
| The HEAT light is on | Switch to 3041/3061 flip-flops or BTA16 thyristors | |
| The HEAT light is on, but the bidirectional thyristor is not conducting | Replace bidirectional thyristor (Model: BTA16) | |
| A timer function is used, with temperature rising and then lowering | Make T1 ≠0 | |
| Large temperature control errors or static errors | Fan broken (not running) | Replace the fan |
| PT100 has poor contact, increasing resistance | Rewire | |
| Parameter (HL, SC) is set incorrectly | Reset | |
| PID and other parameters are set incorrectly | Readjust parameters | |
| Abnormal fan noise or noise greater than 70dB | The fan is faulty or the compressor fan is faulty | Replace the fan |
| After wiping, the air duct plate or blades are damaged | Repair |
9. Electrical Connection Diagram
PT100 HOUT GND VCC A+ B- gated water level

- 1HUITAI LRH liquid crystal biochemistry cultivating box instruction manual
- 2YUEPING NDJ-1 rotational viscometer operating instructions
- 3YUEPING DV-1, Digital Viscometer Instruction Manual
- 4YUEPING NDJ-4 rotational viscometer instruction manual
- 5YUEPING PZ-D-5 Liquid specific gravity Balance instruction manual
- 6YUEPING DC Low Temperature Circulating Bath
- 7YOKE VIS-721 visible spectrophotometer instruction manual
- 8YOKE 7230G visible spectrophotometer instruction manual
- 9YOKE UV754/UV754N UV-Visible Spectrophotometer Instruction Manual
- 10YOKE UV755B/UV756/UV759 UV spectrophotometer instruction manual