YIHENG BPHJS series High and low temperature (alternating) Damping heat testing chamber Instruction manual

YIHENG BPHJS series Temperature Humidity Cyclic Chamber manual covers safety instructions, technical parameters, operation methods and BC1300 controller operation instructions, suitable for temperature and Humidity environmental adaptability test of electronic and electrical products.

The matters recorded here are extremely important and must be strictly observed.

1. Safety Tips

Danger (may result in serious property loss or personal injury)

  1. This product must be reliably grounded (do not use neutral or centerline wires as ground) to prevent electric shock.

  2. Before use, it is essential to confirm that the voltage of the power supply meets the product requirements; installation should only be carried out by qualified personnel.

  3. Products should use independent power outlets, and ensure that plugs and sockets are properly grounded.

  4. It is not allowed to unplug or plug in the power cord arbitrarily without turning off the power switch during operation.

  5. It is not allowed to arbitrarily extend or cut the product's power cord.

  6. Do not place flammable, explosive, volatile, or corrosive substances during drying or baking.

  7. Do not touch motors or compressors exposed to high temperatures above 60°C to prevent burns.

  8. Do not install this equipment in damp areas or places where water may splash, as reduced insulation may cause electric leakage and electric shock.

  9. Unauthorized repairs are not allowed; repairs must be carried out by professionals commissioned by our company.

Warning (unauthorized repairs may result in property damage or personal injury; responsibility is borne by the driver)

  1. You must fully read and understand the product manual before proceeding.

  2. When unplugging the power cord, do not drag the power cord directly.

  3. If any of the following situations occur, you must unplug the power cord of this product:

    • 3.1 When replacing the fuse tube; 3.2 When the product is discontinued for a long time;

    • 3.3 When a product malfunction is awaiting inspection and repair; 3.4 When moving the product.

  4. Non-professional personnel may not disassemble or repair the refrigeration and electrical parts.

Note (otherwise, it may affect the product's lifespan and cause it to malfunction)

  1. Products should be placed on a hard and solid surface to keep them level.

  2. A certain amount of clearance should be left around the product (at least 500mm at the rear of the cabinet).

  3. Products must be used under certain operating conditions. (See Chapter 3: Product Usage Conditions)

  4. If you place a heating sample inside the box, please use an external power supply and do not use the device's own power supply directly.

  5. This product needs to be used indoors at temperatures below 27°C.

2. Main Technical Parameters

ModelBPHJS-060A
BPHJS-060B
BPHJS-060C
BPHJS-120A
BPHJS-120B
BPHJS-120C
BPHJS-250A
BPHJS-250B
BPHJS-250C
BPHJS-500A
BPHJS-500B
BPHJS-500C
BPHJS-1000A
BPHJS-1000B
BPHJS-1000C
Temperature rangeA: -20°C~130°C; B: -40°C~130°C; C: -60°C~130°C
Temperature fluctuation±0.5°C
Temperature uniformity±2.0°C
Humidity range30%RH~95%RH
Humidity fluctuation±3.0%RH
Power supply voltageAC220 50HZA: AC220 50HZ
B: AC380 50HZ
C: AC380 50HZ
AC380 50HZ
Power3750 W
4150 W
5250 W
5950 W
7050 W
9300 W
6000 W
7100 W
9300 W
6650 W
7850 W
12.4KW
13.9KW
15.4KW
Shelf2 pieces3 pieces
The inner liner measures accordingly
W×D×H(mm)
400*380*450500*400*600600*600*700800*700*9001000*1000*1000
External dimensions
W×D×H(mm)
650*1040*1650700*1040*1750800*1160*18501000*1210*19801530*1500*1870

3. Product usage conditions

  1. Temperature: 15~35°C; Relative humidity: ≤85% RH; Atmospheric pressure: 86~106 KPa

  2. No direct sunlight or other heat sources radiating directly, and no strong magnetic field nearby.

  3. There is no strong airflow nearby, but if the surrounding air needs to be forced, airflow should not blow directly onto the cabinet; It should be installed in a well-ventilated area.

  4. There is no high concentration of dust or corrosive substances nearby, and no strong vibrations nearby.

  5. To ensure normal operation and ease of operation, in addition to maintaining horizontal installation, the distance between the equipment and surrounding objects or walls should be: front ≥ 900mm, left, right, top, and rear ≥ 500mm.

4. Scope of Application

The high and low temperature alternating damp heat test chamber is suitable for performance reliability testing of industrial products. This test chamber has a wide temperature control range, and its performance indicators meet national standards and basic environmental test specifications for electrical and electronic products. It is suitable for GB2423.1-89 Test A: low-temperature test method; GB 2423.2-89 Test B: high-temperature test method requirements; GB/T 2423.3-93 Test C: Constant damp heat test method; GB/T 2423.4-93 Test D: Alternating damp heat test method.

5. Product Features

This equipment simulates the changing cycles of high and low temperatures and humidity in nature, and is widely used for adaptation tests of electronic and electrical product components to temperature and humidity environments during the process (especially changes in electrical and mechanical properties).

In terms of product design:

  1. Automatic temperature regulation ensures precise temperature control.

  2. Using a touchscreen adjustment device enables human-machine interaction.

  3. The refrigeration system features a multi-zone operating system.

  4. The water level is automatically controlled by a switch.

  5. The equipment is equipped with protection functions such as over-temperature, overvoltage, phase loss, overload, and short circuit. Once in protection mode, it automatically shuts down related devices and provides prompt functions to ensure system safety and reliability.

  6. The inner liner is made of high-quality stainless steel plate, which is easy to clean and aesthetically pleasing.

  7. The refrigeration system compressor unit and main components all use imported brand products and feature multiple protections such as overpressure, overheating, overload, and phase loss, ensuring reliable operation and low noise.

  8. The casing uses cold-rolled steel electrostatic powder coating, with a uniform, attractive, and elegant color tone.

  9. The insulation material uses ultra-fine glass wool, providing excellent insulation.

  10. This test chamber is equipped with two silicone rubber doors, which have excellent sealing performance.

  11. To measure humidity, simply hang the middle section of the wet medical gauze onto the humidity sensor, and drag both ends of the gauze into the water in the sink.

Diagram of gauze installation        
  1. 12. Optional communication software RS485 interface for remote control.

6. Precautions for Use

  1. The left side of this machine has a test hole. When conducting sample power-on tests, please pay attention to selecting the wire and insert insulation material after connection.

  2. When the device is powered on for the first time and performing AI computation, do not open the cabinet door to avoid calculation errors; Lighting is to be turned off except when necessary.

  3. Except for those allowed in the manual, do not modify internal parameters of the program controller.

  4. Before low-temperature preparation, the studio should be dried at 60°C for about 1 hour.

  5. Avoid running the compressor continuously → → on or off within 5 minutes.

  6. On any cross-section perpendicular to the prevailing wind direction, the sum of the test load surfaces should not exceed one-third of the operating cross-section at that location.

  7. During testing, unless absolutely necessary, do not open the door or enter the chamber, as this may cause personal injury and malfunction.

  8. The door lock of the equipment box can only be opened from the outside; entry into the box requires someone to supervise.

  9. If there is an alarm output, it must be eliminated; forced power on is not allowed.

  10. When performing humidity tests, be sure to frequently check the water level in the storage tank (lower part of the tank). The water level in the tank should not be too low or have no water.

  11. Only after carefully reading the attached documents of this equipment can you operate the equipment.

7. Operating Methods

This machine is a semi-automatic device. The over-temperature setting is set (90°C) and does not need to be reset. If overheating will cut off the power, the circuit breaker must be disconnected first, then closed and restarted. This applies to overload and overpressure in the chiller. (For instrument operation instructions, please refer to the appendix)

8. Fault detection

Fault phenomenaAnalyze possible causesHow to handle it
The evaporator has frosted or the inner liner is frostedFrequently open doors during low temperaturesStop cooling and dry at 80°C to reduce the number of door openings.
Lead holes are not properly sealedWhen using low temperatures, seal them with a rubber stopper inside,
The door was tightly shutPlease close the door tightly or contact our service department
The temperature wouldn't go downEvaporator frostAfter drying the box, conduct a new test
The ambient temperature is too highLower the ambient temperature
Check if the refrigerant pressure is at zeroIf you are at zero, please contact our service department
Doors and observation windows may have condensation or excessive frostHumidity rises too quicklyThe equipment was fine, so we continued testing
The defrost heater is either not working or is brokenPlease contact our service department
The exterior of the device is dampThe humidity in the environment is too highThe equipment was fine, so we continued testing
Humidity continues to riseEvaporator frostAfter drying the box, conduct a new test
Smelling something offThe box body left with a strange odorClean the inside and outside of the box
Smelling something offThe sample had an unusual odorThe equipment was fine, so we continued testing
The humidity doesn't go upWhether humidity mode is selectedDetermine the humidity mode
The OUT light on the humidity meter is on; whether the humidity meter has DC24 output.Replace the instrument or contact the service department
The humidity meter has DC to SSR output, but SSR does not have a 220V outputClose the circuit breaker to check SSRAC220V output and ensure it is in place
SSRAC220V has input, DC24V output, and no SSRAC220V output.Replace SSR
SSR has AC220V output, check if the humidifier is open circuited or has a cold connectionIf the humidifier resistance is ∞, replace it
The shallow water tank is dry, and the wet-bulb sensor has deviations in the water level alarm actionCheck the waterways to keep them clear. The float should operate with a switching mechanism; if the float is not replaced, the sensor should be replaced
The water level in the radiator is insufficientAdd water to the brim
The humidity doesn't go upThe float in the main water tank was brokenThe water level float should be on and off at different positions
Hearing a strange noiseStirring the wind blades loosenedPlease contact our service department
Condenser fan and compressor are looseRetighten the fixing screws
Poor uniformityThe air flow in the test chamber is poorImproves airflow
Sample heatingReduce sample distribution
Evaporator frost forms, and the air duct ventilation is lowAfter drying the box, conduct a new test
Water supply was not suppliedThe upper tank float is brokenReplace the water supply float
The submersible pump is brokenReplace with a pump of the same model
The door is hard to closeClutter gets stuck in doors or hingesRemove debris
Frost causes the door sealing strip to deformAfter drying the box, conduct a new test
The high temperature and humidity inside the chamber cause excessive internal pressureThe equipment is flawless
The equipment door is hard to openThe pressure inside the chamber is too lowThe equipment is flawless
If the door is not tightly closed, the sealing strip will frost and hardenAfter connecting the drying box, open the door
Foreign objects are stuck in the hingeRemove foreign objects
The device cannot startThe device's input power supply does not meet the requirementsPlease adjust the power supply so that the power reaches the device
The input power is not reaching the devicePlease enter the correct power supply
The ground wire is mistaken for the neutral wirePlease connect the correct neutral wire
Missing featuresRule out missing phase of the incoming wire, such as one phase connected or a weak connection
The sequence relay is incorrectPlease swap any two wires of phase A, B, and C to each other
The phase sequence relay indicator light is off, three phases are connected, any two phases have been switched and the phase sequence cannot operate properly, indicating the phase sequence relay is faultyReplace the phase sequence relay of the same model
The temperature won't go upThe circuit breaker in the control circuit was not closedClose the circuit breaker for the control circuit
The overheating setting is too lowSet the overtemperature value correctly
The heating circuit breaker is not closedPlease close the heating circuit breaker
The instrument panel is set too lowPlease set according to test requirements
The instrument panel has an OUT light but no DC outputReplace the instrument of the same model
There is a DC24V voltage, but since there is input but no output, the solid-state relay fails to heat, but the heating wire has an open circuitReplace solid-state relays of the same model with heaters of the same model

The wind does not turnCheck and confirm that the fan resistance is too high or the capacitor is faulty; replace with the same model fan or capacitor

There is a problem with the sensorReplace with sensors of the same model
Overheating is too greatIncorrect setting of relevant instrument parametersCheck the instrument manual or contact our service department
Solid-state relay malfunction, SSR failureReplace solid-state relays of the same model
The temperature won't dropCheck whether the box fan is runningCheck and confirm that the fan resistance is too high or the capacitor is faulty; replace with the same model fan or capacitor
Check whether the cold machine is running, and check if the circuit breaker is closedIf the circuit breaks after closing the chiller and it is still running, please contact our service department
Check if the pressure gauge behind the electrical control cabinet is at zeroIf you are in zero position, please contact our service department
Whether there is heating outputIf there is heating output, is it a solid-state fault? If so, please replace it with a solid-state relay of the same model
Is it because the instrument parameters are confused?If so, please correct according to the instructions
Cold air conditioner failureThe cold machine overload is set according to the compressor's standard power supply at 1.2 timesSet and reset correctly, then restart and contact our service department for overload
Overpressure, ambient temperature too highNecessary ventilation measures should be taken to lower the ambient temperature
Overpressure, and the cold system has issuesPlease contact our services
Overpressure, condensing fan not operatingPlease contact our company services

If the technical parameters and product structure provided in this manual do not involve improvements related to the main performance of the product, no further notice will be given. Thank you for your understanding.

If you find any shortcomings in this equipment during use, we sincerely hope you will contact our company. Your valuable feedback is the driving force behind our company's development.

Company logo        

Appendix: BC1300 controller operation instructions

1. Main screen at startup

Main screen at startup        

Image one

NumberOrderExplanation
(1)Date/TimeIndicates the current date/time
(2)Temperature displayIndicates the current temperature indication (PV).
(3)Humidity displayIndicates the current humidity indicator value (PV).
(4)Move the button on the operation screenDuring operation, buttons alternately display various colors
1Surveillance footageClick once to go to the stop screen for run.
2Operation settingsTap to the Functions and Setting Operation screen.
3Appointment settingsTap the time up to now and the scheduled start time screen
4Curve displayClick to view the program curve screen
5Program group settingsClick once to go to the program settings screen.
6Screen settingsTap to go to the display settings screen.

2. Operation stop screen

Scene two

NumberOrderExplanation
1Directory keyTap to go to the main screen (Screen 1)
2Temperature displayDisplays the current actual temperature
3Temperature set valueSet the target temperature value
4TimeDisplay the current time
5Humidity set valueSet the humidity target value
6Humidity displayDisplays the current actual humidity
7Run stop buttonOperation stops and switches between each other
8Switch buttonWhen stopped, click once to set the set value and stop screen (Screen 3)
In running mode, click once to set the set operation screen (Screen 4)

3. Curved images

Curved images        

Scene three

NumberOrderExplanation
1Temperature valuesDisplays the current actual temperature, represented by a red curve
2Humidity valueDisplays the current actual humidity, represented by a blue curve
3Temperature set valueDisplays the target temperature value, represented by a yellow curve
4Humidity set valueDisplays the humidity target value, represented by a green curve
5Temperature and humidity curveDisplays real-time recorded temperature and humidity curves

Runtime screen

Runtime screen        

Scene 4

NumberOrderExplanation
1Temperature outputDisplays the current temperature control output
2The PID group number of the running PIDDisplays the PID group number of the current runtime application.
3Humidity outputDisplays the current humidity control output
4Status indicator lightsON status is indicated in red, while OFF mode is shown in dark gray.
5Running timeDisplays total running time
6Temperature calculationWhen temperature fluctuations are large, temperature PID calculations can be performed
7Humidity calculationWhen humidity fluctuates greatly, humidity PID calculation can be performed

4. Operation settings screen

Operation settings screen        

Scene 5

NumberOrderExplanation
1Operating modesYou can choose between program mode and fixed value mode
2Power-off modeYou can choose to stop after power outage, cold start, or hot start
3Fuzzy controlYou can choose to move or not act
4Set the slopeThe slope of temperature and humidity changes per minute
5Scheduled operationWhether to use or not, when choosing to use, you can set a time and stop operation when the time is up
6Limit the directory buttonYou can choose to use or not use it, and you can set a password
7Fan outputAutomatic or manual; when manual, the fan can be set to high, medium, or low speed
8Lock the buttonWhen the button is locked, the font appears red and other buttons cannot be used

Note: After pressing the switch button five times on the screen, proceed to the dot matrix printer to operate the settings screen

5. Dot Matrix printer operation settings

Printer operation settings        

Scene six

NumberOrderExplanation
1Print actionYou can choose not to move or to act
2Temperature printingYou can choose to use it or not to use it; the upper and lower temperature printing limits can be set
3Print modeSelectable curve or text mode, status printing interval 0~255 days
4Printing cycleSet the print interval per line to 0~99.59 hours
5Humidity printingYou can choose to use it or not; humidity printing limits can be set
6Paper supply settingsSet the paper feed spacing to 1~255 rows. After modification, press the execute button
7Humidity calculationWhen humidity fluctuates greatly, humidity PID calculation can be performed

6. Appointment Settings

Appointment settings        

Scene 7

NumberOrderExplanation
1Now is the timeThe current time setting
2Schedule the start timeRecording the controller's future startup time
3Appointment buttonWhen the reservation button is in red font, it means the reservation start setup was successful

7. Curve display

Curve display        

Image eight

NumberOrderExplanation
1Program numberYou can view curves for different numbered programs
2Check the timeWhen viewing the program curve, you can select different times
3Program curveThe red curve represents temperature, and the blue curve represents humidity

Press the switch button on screen eight to enter screen nine

Curve shows 2        

Image 9

NumberOrderExplanation
1Zoom in or out keyWhen viewing the curve, you can zoom in or out of time
2PV archiveAfter opening the PV file, you can view the historical curve
3Shift keyWhen viewing the curve, you can move the cursor to view temperature and humidity values for different time periods

Press the switch key on screen nine to enter screen ten

Curve shows 3        

Scene 10

NumberOrderExplanation
1Record the cycleThe interval between recording PV curves can be selected from 1 second to 1 minute
2Record operationsAutomatic or manual
3Temperature curveSet the upper and lower limits of the PV temperature curve
4Humidity curveSet the upper and lower limits of the PV humidity curve

8. Program group settings

Press the switch key on screen 10 to enter screen 11

Program group settings        

Scene 11

NumberOrderExplanation
1Backup projectsBackup items you need can be downloaded or uploaded
2Transmission buttonAfter the project backup is complete, press the transfer button, which must be operated while the machine is off
Program group setting 2        

Scene 12

Program group setting 3        

Scene 13

NumberOrderExplanation
1GroupsProgram groups of 1~120 can be set
2Starting conditionsThe actual values of the conditions when the program group starts running, temperature SSP or humidity SSP
3Program group simulation curveDisplays the simulation curve of the edited program
4Section numberingDisplays editable program segments 1~99
5Temperature and humidity valuesDisplays the temperature and humidity values of the edited program group
6TimeDisplays the time of each segment in the edited program group
7Time signalDisplays the time signals for each segment of the edited program group
8Wait for the actionThe program group selects wind speeds for each stage of the fan at high, medium, and low speeds
9Shift keyWhen editing programs, press this key to flip pages forward or backward
10Insert and delete keysWhen editing a program, first click "Segment Order," then click the "Insert/Delete" key to edit it
Program group setting 4        

Scene 14

NumberOrderExplanation
1Group selectionProgram group selection
2Group loopsThe number of loops in a program group can be linked to other program groups
3Cycle of cyclesThe number of loops in each segment of the program
4Segment loop modeProgram sets in segmental loop mode, allowing you to choose to end SP or start SP
5The group ended modeWhen the entire program group finishes running, you can choose to stop, pause, or link to run
Program group settings 5        

Scene 15

NumberOrderExplanation
1Source of copyEnter an existing program group
2Copy the targetEnter the group number of the program you want to copy, start grouping to the last group, then click the copy key
3Used file informationAll edited program file messages will be displayed at this time
4Group deletedEnter the group number you want to delete, then click Delete or Delete All
Program group setting 6        

Scene 16

NumberOrderExplanation
1Waiting for the setupWait for the settings to use or not to use
2Waiting rangeThe program only starts timing when the temperature and humidity enter the waiting range
3Waiting timeWhen the temperature and humidity enter the waiting range, the waiting time begins
4Wait for action modeYou can choose program groups or wait segments

9. Screen settings

Screen settings        
NumberOrderExplanation
1Humidity displayYou can choose between automatic or manual mode
2Base colorYou can choose the background color of the controller
3Buzzer on/offButton to control the buzzer switch
4Backlight selectionBacklight duration can be set to 0~99 minutes
5LCD brightnessAdjust the controller brightness
6Internal memoryDisplays the internal storage status of the controller

Packing list

Serial numberCategoryNameUnitQuantityNote
1DocumentsUser manualportion1
2DocumentsCertificate of Conformityportion1
3DocumentsWarranty cardportion1
4Spare partsRubber stoppersOnly1
5Spare partsShelfBlock
See main technical parameters
6Spare partsHookPiece
Each partition is equipped with 4 panels
7Spare partsWater trayOnly1
8Spare partsMedical gauzeBlock2
9Optional accessoriesPrinting paperVolume1
10Optional accessoriesSoftwareSet1485 communication function

Optional accessoriesCommunication cables
1

The items listed on this order match the actual items inside the box

Packer: 2 Inspector: 1