YIHENG BPHJS series High and low temperature (alternating) Damping heat testing chamber Instruction manual
The matters recorded here are extremely important and must be strictly observed.
1. Safety Tips
Danger (may result in serious property loss or personal injury)
This product must be reliably grounded (do not use neutral or centerline wires as ground) to prevent electric shock.
Before use, it is essential to confirm that the voltage of the power supply meets the product requirements; installation should only be carried out by qualified personnel.
Products should use independent power outlets, and ensure that plugs and sockets are properly grounded.
It is not allowed to unplug or plug in the power cord arbitrarily without turning off the power switch during operation.
It is not allowed to arbitrarily extend or cut the product's power cord.
Do not place flammable, explosive, volatile, or corrosive substances during drying or baking.
Do not touch motors or compressors exposed to high temperatures above 60°C to prevent burns.
Do not install this equipment in damp areas or places where water may splash, as reduced insulation may cause electric leakage and electric shock.
Unauthorized repairs are not allowed; repairs must be carried out by professionals commissioned by our company.
Warning (unauthorized repairs may result in property damage or personal injury; responsibility is borne by the driver)
You must fully read and understand the product manual before proceeding.
When unplugging the power cord, do not drag the power cord directly.
If any of the following situations occur, you must unplug the power cord of this product:
3.1 When replacing the fuse tube; 3.2 When the product is discontinued for a long time;
3.3 When a product malfunction is awaiting inspection and repair; 3.4 When moving the product.
Non-professional personnel may not disassemble or repair the refrigeration and electrical parts.
Note (otherwise, it may affect the product's lifespan and cause it to malfunction)
Products should be placed on a hard and solid surface to keep them level.
A certain amount of clearance should be left around the product (at least 500mm at the rear of the cabinet).
Products must be used under certain operating conditions. (See Chapter 3: Product Usage Conditions)
If you place a heating sample inside the box, please use an external power supply and do not use the device's own power supply directly.
This product needs to be used indoors at temperatures below 27°C.
2. Main Technical Parameters
| Model | BPHJS-060A BPHJS-060B BPHJS-060C | BPHJS-120A BPHJS-120B BPHJS-120C | BPHJS-250A BPHJS-250B BPHJS-250C | BPHJS-500A BPHJS-500B BPHJS-500C | BPHJS-1000A BPHJS-1000B BPHJS-1000C |
| Temperature range | A: -20°C~130°C; B: -40°C~130°C; C: -60°C~130°C | ||||
| Temperature fluctuation | ±0.5°C | ||||
| Temperature uniformity | ±2.0°C | ||||
| Humidity range | 30%RH~95%RH | ||||
| Humidity fluctuation | ±3.0%RH | ||||
| Power supply voltage | AC220 50HZ | A: AC220 50HZ B: AC380 50HZ C: AC380 50HZ | AC380 50HZ | ||
| Power | 3750 W 4150 W 5250 W | 5950 W 7050 W 9300 W | 6000 W 7100 W 9300 W | 6650 W 7850 W | 12.4KW 13.9KW 15.4KW |
| Shelf | 2 pieces | 3 pieces | |||
| The inner liner measures accordingly W×D×H(mm) | 400*380*450 | 500*400*600 | 600*600*700 | 800*700*900 | 1000*1000*1000 |
| External dimensions W×D×H(mm) | 650*1040*1650 | 700*1040*1750 | 800*1160*1850 | 1000*1210*1980 | 1530*1500*1870 |
3. Product usage conditions
Temperature: 15~35°C; Relative humidity: ≤85% RH; Atmospheric pressure: 86~106 KPa
No direct sunlight or other heat sources radiating directly, and no strong magnetic field nearby.
There is no strong airflow nearby, but if the surrounding air needs to be forced, airflow should not blow directly onto the cabinet; It should be installed in a well-ventilated area.
There is no high concentration of dust or corrosive substances nearby, and no strong vibrations nearby.
To ensure normal operation and ease of operation, in addition to maintaining horizontal installation, the distance between the equipment and surrounding objects or walls should be: front ≥ 900mm, left, right, top, and rear ≥ 500mm.
4. Scope of Application
The high and low temperature alternating damp heat test chamber is suitable for performance reliability testing of industrial products. This test chamber has a wide temperature control range, and its performance indicators meet national standards and basic environmental test specifications for electrical and electronic products. It is suitable for GB2423.1-89 Test A: low-temperature test method; GB 2423.2-89 Test B: high-temperature test method requirements; GB/T 2423.3-93 Test C: Constant damp heat test method; GB/T 2423.4-93 Test D: Alternating damp heat test method.
5. Product Features
This equipment simulates the changing cycles of high and low temperatures and humidity in nature, and is widely used for adaptation tests of electronic and electrical product components to temperature and humidity environments during the process (especially changes in electrical and mechanical properties).
In terms of product design:
Automatic temperature regulation ensures precise temperature control.
Using a touchscreen adjustment device enables human-machine interaction.
The refrigeration system features a multi-zone operating system.
The water level is automatically controlled by a switch.
The equipment is equipped with protection functions such as over-temperature, overvoltage, phase loss, overload, and short circuit. Once in protection mode, it automatically shuts down related devices and provides prompt functions to ensure system safety and reliability.
The inner liner is made of high-quality stainless steel plate, which is easy to clean and aesthetically pleasing.
The refrigeration system compressor unit and main components all use imported brand products and feature multiple protections such as overpressure, overheating, overload, and phase loss, ensuring reliable operation and low noise.
The casing uses cold-rolled steel electrostatic powder coating, with a uniform, attractive, and elegant color tone.
The insulation material uses ultra-fine glass wool, providing excellent insulation.
This test chamber is equipped with two silicone rubber doors, which have excellent sealing performance.
To measure humidity, simply hang the middle section of the wet medical gauze onto the humidity sensor, and drag both ends of the gauze into the water in the sink.
12. Optional communication software RS485 interface for remote control.
6. Precautions for Use
The left side of this machine has a test hole. When conducting sample power-on tests, please pay attention to selecting the wire and insert insulation material after connection.
When the device is powered on for the first time and performing AI computation, do not open the cabinet door to avoid calculation errors; Lighting is to be turned off except when necessary.
Except for those allowed in the manual, do not modify internal parameters of the program controller.
Before low-temperature preparation, the studio should be dried at 60°C for about 1 hour.
Avoid running the compressor continuously → → on or off within 5 minutes.
On any cross-section perpendicular to the prevailing wind direction, the sum of the test load surfaces should not exceed one-third of the operating cross-section at that location.
During testing, unless absolutely necessary, do not open the door or enter the chamber, as this may cause personal injury and malfunction.
The door lock of the equipment box can only be opened from the outside; entry into the box requires someone to supervise.
If there is an alarm output, it must be eliminated; forced power on is not allowed.
When performing humidity tests, be sure to frequently check the water level in the storage tank (lower part of the tank). The water level in the tank should not be too low or have no water.
Only after carefully reading the attached documents of this equipment can you operate the equipment.
7. Operating Methods
This machine is a semi-automatic device. The over-temperature setting is set (90°C) and does not need to be reset. If overheating will cut off the power, the circuit breaker must be disconnected first, then closed and restarted. This applies to overload and overpressure in the chiller. (For instrument operation instructions, please refer to the appendix)
8. Fault detection
| Fault phenomena | Analyze possible causes | How to handle it |
| The evaporator has frosted or the inner liner is frosted | Frequently open doors during low temperatures | Stop cooling and dry at 80°C to reduce the number of door openings. |
| Lead holes are not properly sealed | When using low temperatures, seal them with a rubber stopper inside, | |
| The door was tightly shut | Please close the door tightly or contact our service department | |
| The temperature wouldn't go down | Evaporator frost | After drying the box, conduct a new test |
| The ambient temperature is too high | Lower the ambient temperature | |
| Check if the refrigerant pressure is at zero | If you are at zero, please contact our service department | |
| Doors and observation windows may have condensation or excessive frost | Humidity rises too quickly | The equipment was fine, so we continued testing |
| The defrost heater is either not working or is broken | Please contact our service department | |
| The exterior of the device is damp | The humidity in the environment is too high | The equipment was fine, so we continued testing |
| Humidity continues to rise | Evaporator frost | After drying the box, conduct a new test |
| Smelling something off | The box body left with a strange odor | Clean the inside and outside of the box |
| Smelling something off | The sample had an unusual odor | The equipment was fine, so we continued testing |
| The humidity doesn't go up | Whether humidity mode is selected | Determine the humidity mode |
| The OUT light on the humidity meter is on; whether the humidity meter has DC24 output. | Replace the instrument or contact the service department | |
| The humidity meter has DC to SSR output, but SSR does not have a 220V output | Close the circuit breaker to check SSRAC220V output and ensure it is in place | |
| SSRAC220V has input, DC24V output, and no SSRAC220V output. | Replace SSR | |
| SSR has AC220V output, check if the humidifier is open circuited or has a cold connection | If the humidifier resistance is ∞, replace it | |
| The shallow water tank is dry, and the wet-bulb sensor has deviations in the water level alarm action | Check the waterways to keep them clear. The float should operate with a switching mechanism; if the float is not replaced, the sensor should be replaced | |
| The water level in the radiator is insufficient | Add water to the brim | |
| The humidity doesn't go up | The float in the main water tank was broken | The water level float should be on and off at different positions |
| Hearing a strange noise | Stirring the wind blades loosened | Please contact our service department |
| Condenser fan and compressor are loose | Retighten the fixing screws | |
| Poor uniformity | The air flow in the test chamber is poor | Improves airflow |
| Sample heating | Reduce sample distribution | |
| Evaporator frost forms, and the air duct ventilation is low | After drying the box, conduct a new test | |
| Water supply was not supplied | The upper tank float is broken | Replace the water supply float |
| The submersible pump is broken | Replace with a pump of the same model | |
| The door is hard to close | Clutter gets stuck in doors or hinges | Remove debris |
| Frost causes the door sealing strip to deform | After drying the box, conduct a new test | |
| The high temperature and humidity inside the chamber cause excessive internal pressure | The equipment is flawless | |
| The equipment door is hard to open | The pressure inside the chamber is too low | The equipment is flawless |
| If the door is not tightly closed, the sealing strip will frost and harden | After connecting the drying box, open the door | |
| Foreign objects are stuck in the hinge | Remove foreign objects | |
| The device cannot start | The device's input power supply does not meet the requirements | Please adjust the power supply so that the power reaches the device |
| The input power is not reaching the device | Please enter the correct power supply | |
| The ground wire is mistaken for the neutral wire | Please connect the correct neutral wire | |
| Missing features | Rule out missing phase of the incoming wire, such as one phase connected or a weak connection | |
| The sequence relay is incorrect | Please swap any two wires of phase A, B, and C to each other | |
| The phase sequence relay indicator light is off, three phases are connected, any two phases have been switched and the phase sequence cannot operate properly, indicating the phase sequence relay is faulty | Replace the phase sequence relay of the same model | |
| The temperature won't go up | The circuit breaker in the control circuit was not closed | Close the circuit breaker for the control circuit |
| The overheating setting is too low | Set the overtemperature value correctly | |
| The heating circuit breaker is not closed | Please close the heating circuit breaker | |
| The instrument panel is set too low | Please set according to test requirements | |
| The instrument panel has an OUT light but no DC output | Replace the instrument of the same model | |
| There is a DC24V voltage, but since there is input but no output, the solid-state relay fails to heat, but the heating wire has an open circuit | Replace solid-state relays of the same model with heaters of the same model | |
| The wind does not turn | Check and confirm that the fan resistance is too high or the capacitor is faulty; replace with the same model fan or capacitor | |
| There is a problem with the sensor | Replace with sensors of the same model | |
| Overheating is too great | Incorrect setting of relevant instrument parameters | Check the instrument manual or contact our service department |
| Solid-state relay malfunction, SSR failure | Replace solid-state relays of the same model | |
| The temperature won't drop | Check whether the box fan is running | Check and confirm that the fan resistance is too high or the capacitor is faulty; replace with the same model fan or capacitor |
| Check whether the cold machine is running, and check if the circuit breaker is closed | If the circuit breaks after closing the chiller and it is still running, please contact our service department | |
| Check if the pressure gauge behind the electrical control cabinet is at zero | If you are in zero position, please contact our service department | |
| Whether there is heating output | If there is heating output, is it a solid-state fault? If so, please replace it with a solid-state relay of the same model | |
| Is it because the instrument parameters are confused? | If so, please correct according to the instructions | |
| Cold air conditioner failure | The cold machine overload is set according to the compressor's standard power supply at 1.2 times | Set and reset correctly, then restart and contact our service department for overload |
| Overpressure, ambient temperature too high | Necessary ventilation measures should be taken to lower the ambient temperature | |
| Overpressure, and the cold system has issues | Please contact our services | |
| Overpressure, condensing fan not operating | Please contact our company services |
If the technical parameters and product structure provided in this manual do not involve improvements related to the main performance of the product, no further notice will be given. Thank you for your understanding.
If you find any shortcomings in this equipment during use, we sincerely hope you will contact our company. Your valuable feedback is the driving force behind our company's development.
Appendix: BC1300 controller operation instructions
1. Main screen at startup
Image one
| Number | Order | Explanation |
| (1) | Date/Time | Indicates the current date/time |
| (2) | Temperature display | Indicates the current temperature indication (PV). |
| (3) | Humidity display | Indicates the current humidity indicator value (PV). |
| (4) | Move the button on the operation screen | During operation, buttons alternately display various colors |
| 1 | Surveillance footage | Click once to go to the stop screen for run. |
| 2 | Operation settings | Tap to the Functions and Setting Operation screen. |
| 3 | Appointment settings | Tap the time up to now and the scheduled start time screen |
| 4 | Curve display | Click to view the program curve screen |
| 5 | Program group settings | Click once to go to the program settings screen. |
| 6 | Screen settings | Tap to go to the display settings screen. |
2. Operation stop screen
Scene two
| Number | Order | Explanation |
| 1 | Directory key | Tap to go to the main screen (Screen 1) |
| 2 | Temperature display | Displays the current actual temperature |
| 3 | Temperature set value | Set the target temperature value |
| 4 | Time | Display the current time |
| 5 | Humidity set value | Set the humidity target value |
| 6 | Humidity display | Displays the current actual humidity |
| 7 | Run stop button | Operation stops and switches between each other |
| 8 | Switch button | When stopped, click once to set the set value and stop screen (Screen 3) In running mode, click once to set the set operation screen (Screen 4) |
3. Curved images
Scene three
| Number | Order | Explanation |
| 1 | Temperature values | Displays the current actual temperature, represented by a red curve |
| 2 | Humidity value | Displays the current actual humidity, represented by a blue curve |
| 3 | Temperature set value | Displays the target temperature value, represented by a yellow curve |
| 4 | Humidity set value | Displays the humidity target value, represented by a green curve |
| 5 | Temperature and humidity curve | Displays real-time recorded temperature and humidity curves |
Runtime screen
Scene 4
| Number | Order | Explanation |
| 1 | Temperature output | Displays the current temperature control output |
| 2 | The PID group number of the running PID | Displays the PID group number of the current runtime application. |
| 3 | Humidity output | Displays the current humidity control output |
| 4 | Status indicator lights | ON status is indicated in red, while OFF mode is shown in dark gray. |
| 5 | Running time | Displays total running time |
| 6 | Temperature calculation | When temperature fluctuations are large, temperature PID calculations can be performed |
| 7 | Humidity calculation | When humidity fluctuates greatly, humidity PID calculation can be performed |
4. Operation settings screen
Scene 5
| Number | Order | Explanation |
| 1 | Operating modes | You can choose between program mode and fixed value mode |
| 2 | Power-off mode | You can choose to stop after power outage, cold start, or hot start |
| 3 | Fuzzy control | You can choose to move or not act |
| 4 | Set the slope | The slope of temperature and humidity changes per minute |
| 5 | Scheduled operation | Whether to use or not, when choosing to use, you can set a time and stop operation when the time is up |
| 6 | Limit the directory button | You can choose to use or not use it, and you can set a password |
| 7 | Fan output | Automatic or manual; when manual, the fan can be set to high, medium, or low speed |
| 8 | Lock the button | When the button is locked, the font appears red and other buttons cannot be used |
Note: After pressing the switch button five times on the screen, proceed to the dot matrix printer to operate the settings screen
5. Dot Matrix printer operation settings
Scene six
| Number | Order | Explanation |
| 1 | Print action | You can choose not to move or to act |
| 2 | Temperature printing | You can choose to use it or not to use it; the upper and lower temperature printing limits can be set |
| 3 | Print mode | Selectable curve or text mode, status printing interval 0~255 days |
| 4 | Printing cycle | Set the print interval per line to 0~99.59 hours |
| 5 | Humidity printing | You can choose to use it or not; humidity printing limits can be set |
| 6 | Paper supply settings | Set the paper feed spacing to 1~255 rows. After modification, press the execute button |
| 7 | Humidity calculation | When humidity fluctuates greatly, humidity PID calculation can be performed |
6. Appointment Settings
Scene 7
| Number | Order | Explanation |
| 1 | Now is the time | The current time setting |
| 2 | Schedule the start time | Recording the controller's future startup time |
| 3 | Appointment button | When the reservation button is in red font, it means the reservation start setup was successful |
7. Curve display
Image eight
| Number | Order | Explanation |
| 1 | Program number | You can view curves for different numbered programs |
| 2 | Check the time | When viewing the program curve, you can select different times |
| 3 | Program curve | The red curve represents temperature, and the blue curve represents humidity |
Press the switch button on screen eight to enter screen nine
Image 9
| Number | Order | Explanation |
| 1 | Zoom in or out key | When viewing the curve, you can zoom in or out of time |
| 2 | PV archive | After opening the PV file, you can view the historical curve |
| 3 | Shift key | When viewing the curve, you can move the cursor to view temperature and humidity values for different time periods |
Press the switch key on screen nine to enter screen ten
Scene 10
| Number | Order | Explanation |
| 1 | Record the cycle | The interval between recording PV curves can be selected from 1 second to 1 minute |
| 2 | Record operations | Automatic or manual |
| 3 | Temperature curve | Set the upper and lower limits of the PV temperature curve |
| 4 | Humidity curve | Set the upper and lower limits of the PV humidity curve |
8. Program group settings
Press the switch key on screen 10 to enter screen 11
Scene 11
| Number | Order | Explanation |
| 1 | Backup projects | Backup items you need can be downloaded or uploaded |
| 2 | Transmission button | After the project backup is complete, press the transfer button, which must be operated while the machine is off |
Scene 12
Scene 13
| Number | Order | Explanation |
| 1 | Groups | Program groups of 1~120 can be set |
| 2 | Starting conditions | The actual values of the conditions when the program group starts running, temperature SSP or humidity SSP |
| 3 | Program group simulation curve | Displays the simulation curve of the edited program |
| 4 | Section numbering | Displays editable program segments 1~99 |
| 5 | Temperature and humidity values | Displays the temperature and humidity values of the edited program group |
| 6 | Time | Displays the time of each segment in the edited program group |
| 7 | Time signal | Displays the time signals for each segment of the edited program group |
| 8 | Wait for the action | The program group selects wind speeds for each stage of the fan at high, medium, and low speeds |
| 9 | Shift key | When editing programs, press this key to flip pages forward or backward |
| 10 | Insert and delete keys | When editing a program, first click "Segment Order," then click the "Insert/Delete" key to edit it |
Scene 14
| Number | Order | Explanation |
| 1 | Group selection | Program group selection |
| 2 | Group loops | The number of loops in a program group can be linked to other program groups |
| 3 | Cycle of cycles | The number of loops in each segment of the program |
| 4 | Segment loop mode | Program sets in segmental loop mode, allowing you to choose to end SP or start SP |
| 5 | The group ended mode | When the entire program group finishes running, you can choose to stop, pause, or link to run |
Scene 15
| Number | Order | Explanation |
| 1 | Source of copy | Enter an existing program group |
| 2 | Copy the target | Enter the group number of the program you want to copy, start grouping to the last group, then click the copy key |
| 3 | Used file information | All edited program file messages will be displayed at this time |
| 4 | Group deleted | Enter the group number you want to delete, then click Delete or Delete All |
Scene 16
| Number | Order | Explanation |
| 1 | Waiting for the setup | Wait for the settings to use or not to use |
| 2 | Waiting range | The program only starts timing when the temperature and humidity enter the waiting range |
| 3 | Waiting time | When the temperature and humidity enter the waiting range, the waiting time begins |
| 4 | Wait for action mode | You can choose program groups or wait segments |
9. Screen settings
| Number | Order | Explanation |
| 1 | Humidity display | You can choose between automatic or manual mode |
| 2 | Base color | You can choose the background color of the controller |
| 3 | Buzzer on/off | Button to control the buzzer switch |
| 4 | Backlight selection | Backlight duration can be set to 0~99 minutes |
| 5 | LCD brightness | Adjust the controller brightness |
| 6 | Internal memory | Displays the internal storage status of the controller |
Packing list
| Serial number | Category | Name | Unit | Quantity | Note |
| 1 | Documents | User manual | portion | 1 | |
| 2 | Documents | Certificate of Conformity | portion | 1 | |
| 3 | Documents | Warranty card | portion | 1 | |
| 4 | Spare parts | Rubber stoppers | Only | 1 | |
| 5 | Spare parts | Shelf | Block | See main technical parameters | |
| 6 | Spare parts | Hook | Piece | Each partition is equipped with 4 panels | |
| 7 | Spare parts | Water tray | Only | 1 | |
| 8 | Spare parts | Medical gauze | Block | 2 | |
| 9 | Optional accessories | Printing paper | Volume | 1 | |
| 10 | Optional accessories | Software | Set | 1 | 485 communication function |
| Optional accessories | Communication cables | 1 |
The items listed on this order match the actual items inside the box
Packer: 2 Inspector: 1
- 1YIHENG GHP series, waterproof series Electric heated incubators of constant temperature operation instructions
- 2YIHENG DHP-B series small microbial incubator operation instructions
- 3YIHENG DHP series Electric heated incubators of constant temperature operation instructions
- 4LEICI DZS-706 Multi-Parameter Water Quality Analyzer Instruction Manual
- 5YINGXU Zahn Series Zahn Cup Instruction Manual
- 6RDS Mayer Rods Use · Maintenance · FAQ Instructions
- 7LANDTEK SRT-6223 sand blasting shot peening Roughness Gauge
- 8Defelsko PosiPen Coating thickness gauge
- 9REDDRAGON TC-200 Coating thickness gauge instruction manual
- 10REDDRAGON MT-10 moisture meter instruction manual