HUXI HB-20 Corrosion Resistant Diaphragm Pump Instruction Manual
1. Product Introduction
The HG series diaphragm vacuum pump is a new type of diaphragm vacuum pump independently developed and produced. It can operate continuously without oil, with low noise, high work efficiency, and long service life. The product range is rich with different applications, including pumps that serve as solvent filtration devices that combine vacuum pumps and compressors, high-vacuum pumps paired with rotary evaporators, and pumps with both positive and negative pressure filters, greatly reducing laboratory procurement costs. The main advantages are: stable vacuum, adjustable pressure, compact size, light weight, easy to use, oil-free, clean and pollution-free, capable of long-lasting stable operation, continuous operation for 2 hours, simple maintenance and maintenance, and self-operation. Mainly used in vacuum distillation, reaction reactors, vacuum filtration, vacuum concentration, vacuum drying, and for compressing and converting gases.
2. Technical Parameters
| Model | HG-30/HG-30F | HG-30L/HG-30LF | HG-30D/HG-30DF | HG-60DL/HG-60DLF |
|---|---|---|---|---|
| Pumping speed (L/min) | 30 | 30 | 30 | 60 |
| Ultimate Pressure (MPa) | >0.085 | >0.085 | >0.095 | >0.085 |
| (150mbar) | (150mbar) | (50mbar) | (150mbar) | |
| Positive pressure (Psi) | â | 40 | â | 40 |
| Grade | 1 | 1 | 2 | 2 |
| Motor Power (W) | 180 | 180 | 180 | 180 |
| Dimensions LĂWĂH (mm) | 250Ă135Ă210 | 250Ă135Ă210 | 315Ă135Ă210 | 315Ă135Ă210 |
| Weight | 9 | 9 | 13 | 13 |
| Diaphragm material | Imported corrosion-resistant rubber | Imported corrosion-resistant rubber | Imported corrosion-resistant rubber | Imported corrosion-resistant rubber |
| Noise (dB) | <50 | <50 | <50 | <50 |
| Function | Negative pressure | Dual-use positive and negative pressure type | Negative pressure | Dual-use positive and negative pressure type |
Instructions: Please read this manual carefully before using this product. Only qualified engineers or electricians may install and commission this product.
3. Product usage
This product is only for air extraction.
Do not allow corrosive liquids or particulate materials into the pump. Water vapor, oil pollutants, or other liquids must be filtered out first.
The operating environment temperature must not exceed 55°.
Do not use this product at high altitudes, as low air pressure at high altitudes may affect the product.
This product is a precision instrument that protects the product from dirt and moisture.
4. Installation
Before installation or repair, please disconnect the power supply; otherwise, electric shock may occur or cause injury or death.
To avoid electric shock, do not use this product in environments that come into contact with water or other liquids. In case of rain or snow, be sure to protect the machine.
Motor wiring should be configured according to local power regulations.
Check whether the local voltage matches the local device (see label). If the voltage is different, use a transformer.
During or after use, do not touch the motor casing to avoid burning your hands.
This product does not require lubrication.
The size of the connecting rubber hose must match the air inlet.
Install the silencer we provide at the air outlet.
5. Placement
To reduce noise and vibration, the product can be placed on a shockproof device placed on a horizontal tabletop, which helps reduce resonance.
Ensure the product's ventilation openings are unobstructed; otherwise, it will affect heat dissipation.
Connect the power supply and turn on the power switch.
6. Handling of abnormal situations
Disconnect the power.
Do not disassemble without disconnecting the power.
Wait for the pump to cool down.
Refer to common fault handling.
Contact manufacturers or distributors for more assistance.
The motors are equipped with overheat protectors, ensuring the product will not be damaged by overheating or overload.
7. Product maintenance
After use, disconnect the power to ensure the product is insulated.
Release the pressure inside the pump.
Filter materials are especially prone to dirt in dirty environments, so please replace them regularly.
8. Common fault handling
Before repair, always disconnect the power supply. If the motor cannot run or starts running slowly under pressure, turn off the switch and disconnect the power. Ensure the power supply matches the power indicated on the nameplate. Check the plugs, wires, and switches to see if there are signs of aging or falling off. If the pump overheats or the noise is too loud, stop operation immediately and carry out repairs.
Note: We are not responsible for pumps that have been self-modified.
| Possible reasons | Low pressure | Low vacuum level | Excessive noise | Overheating | Unable to start |
|---|---|---|---|---|---|
| The valve plate is dirty | â | â | |||
| Valve plates are damaged | â | â | |||
| Diaphragm damage | â | â | |||
| Low voltage | â | â | â | â | |
| Voltage mismatch | â | â | â | ||
| Started under pressure | â | ||||
| The pressure regulator valve is too loose | â | â | â | ||
| Rubber hose leaking air | â | â | â |
9. Maintenance and Maintenance Steps
Please maintain regularly. Dirty filter material or muffler can cause poor air circulation and cause overheating and excessive noise, so regularly check the filter material inside the pump head. If necessary, clean the air chamber with alcohol. Do not use organic compounds, acids, corrosive, or flammable solvents for cleaning. Lubricate any parts to avoid affecting the pump's service life.
You can return the pump to the factory for repair, or follow the steps below for installation. Each machine includes the following components: one vacuum pressure gauge (-0.1Mpa), one connecting rubber hose (ÎŚ7mm*ÎŚ13mm*ÎŚ1000mm), one silencer, and one manual.
We can provide all the necessary accessories. Please contact agents or manufacturers for purchase.
Below is an example of the basic HG-30 pump (remember the locations of all accessories):
Disconnect the power. Warning: Always disconnect the power before repairs to avoid injury or casualties.
Release the pressure valve to release the gas inside the pump.
Remove the rubber hose.
Remove the pressure gauge, remember where it should be placed, and remove the four-way valve.
Unscrew the six hex socket screws at the pump head, open the gas chamber, and keep the screws in place.
There is a sealing ring in the groove on the back of the air chamber. If the sealing ring is damaged, please discard it; There is filter cotton inside the air chamber; if it gets dirty, please discard it.
Below the air chamber is a circular membrane cavity (with valve plates installed), which are pulled out.
Check whether the valve plates (black rubber and stainless steel) are aged or damaged. If they are damaged or too dirty, please replace or clean them. Just unscrew the screws.
Below the membrane cavity is a black membrane secured with a small cap. Unscrew the three screws on the small pressure cap, keep the screws safe, remove the pressure cap, and visually inspect the diaphragm for damage. If there is damage or severe aging, please replace it.
If needed, use water-based solvents to clean the pump head or use high-pressure gas to blow away dirt. Note: Do not use organic compounds, acids, or corrosive solvents for cleaning, and do not lubricate any parts, as this will reduce the pump's service life. Only use water-based solvents.
If you need to replace the diaphragm, use a new original diaphragm and place it on the raised connecting rod.
Install the diaphragm cavity where the valve plate has been replaced with a new one. Note: Do not cause any scratches or stick anything to the valve plate.
A new sealing ring is installed in the groove below the air chamber, and new filter cotton is placed.
Cover the air chamber and screw in the hex socket screw.
Install the four-way valve and install the pressure gauge according to the previously made marks.
Note: Any improvements to the HG series diaphragm vacuum pump will not be notified separately.
Warranty card
Under normal use, finished pumps enjoy a 12-month warranty service provided by our company or its authorized agents. To enjoy better service, if any quality issues are found within 30 days of purchase, please contact our company or an authorized distributor immediately. The warranty does not cover motors, diaphragms, and valve plates, as motors are not manufactured by our company; diaphragms and valve plates are consumables. We are not responsible for damage caused by improper operation, lack of maintenance, accidents, or transportation. The liability for compensation for the product will not exceed the value of the product itself. You have the authority to decide whether to refund or exchange.
We are not responsible for any direct or indirect personal injury resulting from the use of this product. We are not responsible for waste generated from the use of products, wasted labor hours, labor costs incurred, or any other indirect or direct losses to the enterprise or property, regardless of whether the operation is done according to the manual. If there are transportation issues or delays in delivery, regardless of whether it is due to personnel negligence, unclear responsibility, or courier issues, the right is to choose to exchange goods, resend goods, or issue a refund within the product price.
Explicitly refusing to repair products or accessories intended for special purposes or purposes. If you extend the warranty card without authorization, the consumer is fully responsible. Consumers decide for themselves the application scenarios of the product, such as whether to resell, whether it can be used in combination with other products, or for equipment designed, assembled, and produced by the consumer themselves. This warranty card can only be modified by authorized personnel, and a special written modification document must be signed.
User Information (Name):
Warranty card number:
Product Name: Diaphragm vacuum pump
Product Model:
Shipping Date:
Warranty period:
Customer Name and Seal:
Maintenance records
| Date | Fault description | Maintenance personnel |
|---|---|---|
- 1HUXI HB-40B Corrosion resistant diaphragm pump instruction manual
- 2HUXI MD200 series, dry Nitrogen blowing instrument instruction manual
- 3HUXI HY-1200 In situ hybridization instrument instruction manual
- 4HUXI HW-200G/TG microplate incubator instruction manual
- 5HUXI JB series digital display constant speed powerbasic stirrer instruction manual
- 6HUXI Vortex Series, Instruction Manual for Vortex Mixer
- 7HUXI HT-200 Series Multi-Tube Vortex Mixer Instruction Manual
- 8HUXI HW-400T Microplate Shaker Instruction Manual
- 9HUXI HT-10/HT-20 Track Shaker Instruction Manual
- 10HUXI HX-10 series mini thermostatic metal bath instruction manual