HUXI HY-1200 In situ hybridization instrument instruction manual

HUXI HY-1200 in situ hybridization instrument operation manual, covering boot inspection, basic parameters, operation guide, system configuration, temperature calibrating and fault handling. Support three operating modes, high temperature control accuracy.

Preface

Thanks to the purchase of slide denaturation/hybridization instruments. This user manual includes instrument functions and operating procedures. To ensure correct use, please read the manual carefully before operating the instrument. Please keep the manual safe so you can read it quickly when you encounter problems.

Unboxing inspection

When users open the instrument packaging box for the first time, please check the instrument and accessories against the packing list. If you find any errors, incomplete or abnormal parts, please contact the seller or manufacturer.

Table of Contents

Important Notes ---- 3

1. Important Safety Operation Information ---- 3

2. Safety ---- 3

3. Instrument Maintenance ---- 4

4. After-Sales Service ---- 4

Chapter 1 Introduction ---- 5

1. Product Introduction and Features ---- 5

2. Normal working conditions ---- 5

3. Basic Parameters and Performance ---- 6

Chapter 2 Basic Operation Instructions ---- 7

1. Structural Diagram ---- 7

2. Humidity sponge installation ---- 8

Chapter 3 Operation Guide ---- 9

1. Boot up and load the system ---- 9

2. Main Menu Interface ---- 9

3. Program List Menu ---- 10

4. Program Run and Edit Menu ---- 10

5. Parameter Settings Menu ---- 12

6. Program Name Editing Menu ---- 12

7. System Configuration Menu ---- 13

8. Temperature Calibration Menu ---- 14

Chapter 4 Precautions ---- 16

Chapter 5 Fault Analysis and Handling ---- 17

Note ---- 18

Important note

1. Important safety operation information

Users need to have a complete understanding of how the instrument works before operating it safely. Please read this manual carefully before operating the instrument.

Safety warnings        

No one is allowed to operate the instrument before reading the manual. If not operated according to the instructions in the manual, the heat generated during operation may cause severe burns and may lead to electric shock accidents. Please carefully read the following safety tips and guidelines, and implement all precautions within them.

2. Safety

During all processes of operating, maintaining, and repairing this instrument, the following basic safety precautions must be followed. Failure to comply with these measures or warnings noted elsewhere in this manual may affect the protection provided by the instrument and the intended scope of its use.

Indoor use signage        

This instrument is intended for indoor use.

Reads the manual prompt        

Please read this operation manual carefully before operating this device; otherwise, it may cause personal injury. Only qualified inspectors trained in how to install and use electrical equipment can operate this instrument.

No disassembly reminder        

Operators should not attempt to open or repair the instrument, as doing so will disqualify you from warranty coverage and may also result in electric shock. If repairs are needed, our company will be responsible for the repairs.

To prevent electric shock accidents, the instrument's input power cable must be reliably grounded. This instrument uses a three-core grounding plug, with the third pin serving as the grounding pin, which should be used with a grounded power socket.

Power safety reminder        

Before connecting the power supply, ensure the voltage matches the instrument's required voltage. And ensure that the rated load of the power socket is not less than the instrument's requirements.

Power cord check prompt        

If the power cord is damaged, it must be replaced. When replacement, the power cord must be replaced with a power cord of the same type and specification. Do not press anything on the power cord during use. Do not place power cords where people are walking.

Always hold the power cord when plugging or unplugging. When plugging in, make sure the plug is fully inserted into the socket; do not forcibly pull the power cord when unplugging.

Place environment prompts        

This instrument should be placed in a place with low humidity, low dust, away from water sources, and away from direct sunlight and strong light. The interior should be well-ventilated, protected from corrosive gases or strong magnetic field interference, away from heating and all other heat sources. Place the instrument away from damp or dusty areas.

The opening at the bottom of this instrument is designed for ventilation to prevent overheating, and never block or cover these ventilation holes. When multiple instruments are used simultaneously, the distance between them should not be less than 30cm.

Power outage maintenance reminder        

When stopping operation, turn off the power. When not in use, unplug the power cord and cover the instrument with a soft cloth or plastic paper to prevent dust ingress.

In the following situations, immediately unplug the instrument's power cord from the outlet, contact the supplier, or have a trained technician handle it:

Abnormal situation alert        
  • Liquid spills into the inside of the instrument;

  • The instrument must be exposed to rain or water;

  • The instrument is not functioning properly, especially if any unusual sounds or odors appear;

  • Instrument dropped or casing damaged;

  • The instrument's functions have changed significantly.

3. Instrument maintenance

This instrument should be regularly cleaned with a clean soft cloth dampened with a small amount of anhydrous alcohol to ensure good heat conduction and avoid contamination.

Maintenance and cleaning tips        

If there are stains on the surface of this instrument, clean it with a soft cloth dampened with clean water.

When cleaning the instrument, the power must be cut off.

The surface of the instrument must not be cleaned with corrosive cleaning agents.

4. After-sales service

a) Warranty content

Within one month from the date of delivery, our company will be responsible for replacing any faults caused by defects in materials or manufacturing.

This instrument is covered for 24 months from the date of delivery against faults caused by defects in materials and manufacturing. During the warranty period, our company will selectively repair or replace any instruments proven to be defective.

Products under warranty must be sent by the user to a repair department designated by our company. The shipping cost for sending the instrument from the user to the maintenance department shall be borne by the user. Our company will cover the shipping costs for returning the instrument to the user.

For repairs outside the warranty period, our company will charge an appropriate repair cost.

b) Warranty coverage

The above warranty is not suitable for damage caused by improper use or maintenance by the user, use under non-compliant conditions, unauthorized repairs, or modifications.

Chapter 1 Introduction

1. Product Introduction and Features

The in-situ hybridizer uses microprocessing technology combined with PID control to achieve denaturation and hybridization of 12 slides. The sealed top cover and humidifying components ensure a more ideal hybridization environment, integrating three operating modes: denaturation/hybridization, hybridization, and multi-step operation, suitable for various in-situ hybridization experiments.

This product has the following features:

  • Full touchscreen operation with a user-friendly human-machine interface.

  • Chinese and English menus available.

  • Supports power outage recovery function; if an unexpected power outage occurs during operation, it can automatically resume operation according to the original program after power is restored.

  • Supports automatic cooling function after operation ends.

  • Supports automatic preheating function.

  • The platform offers high temperature control accuracy and minimal temperature fluctuations.

  • Can process 12 slides simultaneously.

  • Supports storage for 105 custom programs.

  • It integrates three operation modes: degeneration & hybridization, hybridization, and multi-step operation.

2. Normal working conditions

Operating temperature: 5°C ~ 30°C

Relative humidity: ≤70%

Power supply: AC220V/50Hz

3. Basic parameters and performance

Parameters and performance
Temperature control rangeRoom temperature +5 °C ~ 100 °C
Temperature setting range0°C~100°C
Time setting1min ~ 99h59min
Temperature control accuracy≤ ± 1°C
Temperature uniformity≤ ± 1°C
Heating time*≤2 minutes (temperature rises from 37°C to 95°C)
Cooling Time*≤6 minutes (cools from 95°C to 45°C)
Sample size12 pieces
Input power350W
Fuse250V 3A Φ5×20
Dimensions (mm)440x220x120 mm
Weight (kg)4.5KG

Chapter 2 Basic Operation Instructions

1. Structural schematic diagram

Structural schematic diagram 1        
Structural schematic diagram 2        

2. Humidity sponge installation

A) Before each experiment, use a tool to add 30ml of distilled water to each of the two humidity sponges.

b) Then place the humidity sponge into the slot on the top cover, making sure the sponge is placed flat.

Humidity sponge installation        

Chapter 3 Operation Guide

1. Boot up and system loading

Plug in the power to the instrument, turn on the power switch on the back of the instrument, and the display screen will show the system self-check and load menu. Please be patient during this process, and the system will automatically enter the operation interface.

Boot screen        

2. Main menu interface

After self-check and loading are complete, enter the main menu.

Main menu interface        

Transformation & Hybrid Mode: Enter the program list in Transformation & Hybrid Mode and select the program you want to edit or run.

Hybrid Mode: Enter the list of hybrid mode programs and select the program you want to edit or run.

Multi-Mode Mode: Enter the multi-point mode program list and select the program you want to edit or run.

System Configuration: Enter the system configuration menu to set menu language, key sounds, calibration, preheating, and automatic power-off recovery functions according to actual needs.

The number behind the red thermometer indicates the real-time temperature value of the current heating platform.

3. Program list menu

Select different running modes (degeneration & hybridization, hybrid, multi-point) and enter the corresponding program list menu.

Hybrid operation mode

Program list interface        
  • A. Back key to return to the upper menu.

  • B. Press the program editing and loading button: read the selected program and enter its run and edit menu.

  • C. Delete the key, clear all information for the selected program, and restore it to its default values.

  • D. Select the program.

  • E. Select the program.

  • F. Program list.

4. Program Run and Edit menu

In the program run and edit menu mode, you can edit program parameters or run the program directly.

Program Run Editing Interface 1        
Program Run Editing Interface 2        

Modification & hybridization operation mode

Interface for transmutation hybridization operation        
  • A. Back key to return to the upper menu.

  • B. Parameter display/setting button: Click this button to modify temperature and timer values.

  • C. Start the button and click it to start the current program.

  • D. Stop button: Clicking it stops the running program (this key is only effective when the program is running).

  • E. Pause button: If the program is running and the countdown begins, pressing this button will pause the countdown, but the platform temperature will remain at the current state.

  • F. Program Name: Click this button to modify the current program name.

  • G. Platform temperature, which displays the real-time temperature value of the current platform.

  • H. System status, which displays the instrument's current operating status (running, stopping, pausing).

5. Parameter settings menu

This menu is used to set the program's temperature and timer values.

Parameter settings interface        
  • A. The back key returns to the upper menu, leaving no current parameter modifications.

  • B. Temperature value: After clicking, enter the temperature value using the numeric keypad.

  • C. Timer value: After clicking, enter the timer value using the numeric keypad. The time format is (hours:minutes:seconds).

  • D. Clean button, click to clear the temperature value. (If the temperature value is empty, the program will automatically skip this step.)

  • E. Delete key: Click to delete the input parameter value.

  • F. Numeric keypad, used to input temperature and timer values.

  • G. Save key: After clicking, saves the current parameter value and automatically returns to the upper menu.

6. Program Name Edit Menu

This menu is used to edit the program name.

Alphabet keyboard

Alphabet keyboard interface        
Numeric keypad interface        
  • A. The back key returns to the upper menu, leaving no current parameter modifications.

  • B. A/a, the key to switch case between uppercase and lowercase letters

  • C. Numeric keypad: Click to switch to numeric keypad.

  • D. Alphabet Keyboard: Click to switch to Alphabet Keyboard.

  • E. Delete key: Click to delete the entered character.

  • F. Save key: Click to save the current program name and automatically return to the upper menu.

7. System configuration menu

This menu is used to configure the system's auxiliary functions.

System configuration interface        
  • A. Back key to return to the upper menu.

  • B. Menu language: Switch system menu language (Chinese/English).

  • C. Preheat at 37°C*, and turn the preheating function on or off.

  • D. Power outage recovery*: Enable or disable the power-off recovery function.

  • E. Temperature calibration*, click to enter temperature calibration mode.

  • F. Key sound: the sound of turning the button on or off.

Introduction to the 37°C preheating function:

If this function is enabled, the heating platform temperature of the instrument remains constant at 37°C in all other states except for program operation.

If this function is turned off, the system will turn on the fan after each program runs to cool the heating platform temperature.

*Introduction to the Power Outage Recovery Function:

If this function is enabled, during program operation, an unexpected external power outage causes the instrument to interrupt operation. After external power is restored, the instrument will automatically resume the incomplete program based on the operating status during the previous unexpected power outage.

*Introduction to temperature calibration functions:

If a large temperature error is found on the heating platform during use, this function can be used to recalibrate the instrument.

Calibration prompt        

The calibration process requires room temperature and the thermometer used for calibration, so we do not recommend users calibrate it themselves. If you find a large temperature error in the instrument, please contact the dealer or manufacturer and consult technical support before calibrating.

8. Temperature calibration menu

If a large temperature error is found on the heating platform during use, this function can be used to recalibrate the instrument.

The calibration process requires room temperature and the thermometer used for calibration, so we do not recommend users calibrate it themselves. If you find a large temperature error in the instrument, please contact the dealer or manufacturer and consult technical support before calibrating.

1. Calibration thermometer: accuracy ≤ 0.2°C.

2. Room temperature: ≤30°C.

3. Add 30ml of distilled water to each humidity sponge and install it into the lid.

4. Install the calibration thermometer on the surface of the heating platform. (Note: Installation should ensure a smooth installation and complete contact between the thermometer and the heating platform. Before installing the thermometer, you can add an appropriate amount of paraffin oil between the heating platform surface and the thermometer to ensure thermal conductivity.) After installing the thermometer, close the lid.

5. The calibration process takes about 1 hour. During this process, ensure the instrument does not lose power unexpectedly. If power is accidentally lost during calibration, all data for this calibration will be invalid.

6. Enter the system configuration menu, click the temperature calibration button to start temperature calibration.

Temperature calibration interface        
  • A. Use the back button to exit temperature calibration; all data for this calibration will be invalid.

  • B. Calibration temperature point, indicating the temperature point currently being calibrated (system default 4 calibration points: 40/60/80/100°C).

  • C. Real-time temperature value: real-time temperature value of the heating platform.

Please be patient during the calibration and heating process. Once the temperature stabilizes, the system will automatically enter calibration point correction mode.

Calibration and correction interface        
  • A. Correction confirmation: Confirm the current temperature correction value, then proceed to the next temperature calibration point.

  • B. Temperature Correction -: Enter the actual temperature reading displayed on the calibration thermometer.

  • C. Temperature Correction+: Enter the actual temperature reading displayed on the calibration thermometer.

  • D. Correction value, displays the corrected actual temperature value.

Once all four calibration points are corrected, the system automatically exits calibration mode.

Chapter 4 Precautions

  • Do not move the instrument during use.

  • During use, to ensure temperature control accuracy, keep the instrument away from external heat sources.

  • After use, first cut off the power, let it cool for 30 minutes, then use a dry, clean soft cloth to wipe off any moisture left around the heating module. Store in a dry, ventilated place.

  • Non-professionals are prohibited from opening the instrument casing. If internal dust removal is required, please contact the dealer or manufacturer.

  • The moisture card is made of high-temperature resistant sponge, which normally does not need to be replaced.

High-voltage warning        

There is high voltage inside the instrument; unauthorized disassembly and repair are prohibited!

Chapter 5 Fault Analysis and Handling

Fault phenomenaCause of failure
No response after powering onPower not connected (a)
No response after powering onFuse wire burnout (b)
No response after powering onSwitch failure (c)
Excessive temperature deviationSensor Damage (c)
Cannot heat up or cool downTemperature Control System Failure (c)
The buttons are ineffectiveTouchscreen damage (c)
System prompt: Overtemperature alertOvertemperature Warning (c)
System prompt: Platform temperature abnormalPlatform temperature sensor failure (c)
System prompt: Abnormal temperature value on the lidUpper cover temperature sensor failure (c)
System prompt: Platform ADC initialization failedPlatform Data Sensor Initialization Failure (c)
System prompt: Upper cover ADC initialization failedFailure to initialize the upper cover data sensor (c)
System prompt: Platform ADC data is abnormalPlatform Data Sensor Data Anomalies (c)
System prompt: Abnormal data from the ADC on the upper coverAbnormal data sensor data on the top cover (c)

a) Check the power cord connections to ensure the external power supply meets the requirements.

b) After replacing the fuse wire, restart the instrument. If the problem persists, please contact the supplier or manufacturer.

c) Check the power cord connections to ensure the external power supply meets the requirements.

Note