HUXI HW-200G/TG microplate incubator instruction manual

HUXI HW-200G/TG microplate incubator operation manual, introducing temperature time setting, operation stop operation and microplate fixing method, covering fault handling and maintenance suggestions.

Microplate Incubator User Manual

Preface

Thanks to the purchase of the microplate incubator. This user manual includes instrument functions and operating procedures. To ensure correct use, please read the manual carefully before operating the instrument. Please keep the manual safe so you can read it quickly when you encounter problems.

Unboxing inspection

When users open the instrument box for the first time, please check the instrument and accessories against the packing list. If you find any errors, incomplete parts, or abnormalities, please contact the seller or manufacturer.

Table of Contents

Important Notes.... 3

1. Important safety operation information.... 3

2. Safety.... 3

3. Instrument maintenance.... 4

4. After-sales Service.... 4

Chapter 1 Introduction.... 5

Chapter 2 Characteristics.... 6

1. Normal working conditions.... 6

2. Basic Parameters and Performance.... 6

Chapter 3 Basic Operation Instructions.... 7

1. Structural Diagram.... 7

2. Operation panel diagram.... 8

Chapter 4 Operation Guide.... 9

1. Temperature and time settings.... 9

2. Operation and stopping operations.... 9

3. Installation and fixation of microvia plates.... 10

Chapter 5 Fault Analysis and Handling.... 11

Note.... 12

It is important to clarify

1. Important safety operation information

Users need a complete understanding of how the instrument works before safely operating it. Users should carefully read this manual before operating the instrument.

Safety warning

No one is allowed to operate the instrument before reading the manual. If the instructions are not followed, the heat generated during operation may cause severe burns and may result in electric shock accidents. Please carefully read the following safety tips and guidelines, and implement all precautions within them.

2. Safety

During all processes of operating, maintaining, and repairing this instrument, the following basic safety precautions must be followed. Failure to comply with these measures or warnings noted elsewhere in this manual may affect the protection provided by the instrument and the expected scope of its use.

Safety warnings

This instrument is a Class I Type B standard equipment compliant with GB9706.1 standards. This instrument is intended for indoor use.

Operation warning

Please read this operation manual carefully before operating this device; otherwise, it may cause personal injury. Only qualified inspectors trained in how to install and use electrical equipment can operate this instrument.

Maintenance warning

Operators should not attempt to open or repair the instrument, as doing so will disqualify you from warranty coverage and may also result in electric shock. If repairs are needed, our company will be responsible for the repairs.

To prevent electric shock accidents, the instrument's input power cable must be reliably grounded. This instrument uses a three-core grounding plug, with the third pin serving as the grounding pin, which should be used with a grounded power socket.

Before connecting the power supply, ensure the voltage matches the instrument's required voltage. And ensure that the rated load of the power socket is not less than the instrument's requirements.

Power cord safety

If the power cord is damaged, it must be replaced. When replacement, the power cord must be replaced with a power cord of the same type and specification. Do not press anything on the power cord during use. Do not place power cords where people are walking.

Always hold the power cord when plugging or unplugging. When plugging in, make sure the plug is fully inserted into the socket; do not forcibly pull the power cord when unplugging.

Placement requirements

This instrument should be placed in a place with low humidity, minimal dust, away from water sources, and away from direct sunlight and strong light. The room should be well-ventilated, free from corrosive gases or strong magnetic field interference, and kept away from heating, stoves, and all other heat sources. Do not place the instrument in damp or dusty areas.

The opening on this instrument is designed for ventilation. To avoid overheating, be sure not to block or cover these ventilation holes. When multiple instruments are used simultaneously, the distance between each instrument should be no less than 30cm.

Stop using it

When stopping operation, turn off the power. If the instrument will not be used for a long time, unplug it and cover the instrument with a soft cloth or plastic paper to prevent dust from entering.

In the following situations, immediately unplug the instrument's power cord from the outlet and contact the supplier or have a trained technician handle it:

Exception handling
  • Liquid spills into the inside of the instrument;

  • The instrument must be exposed to rain or water;

  • The instrument is not functioning properly, especially if any unusual sounds or odors appear;

  • Instrument dropped or casing damaged;

  • The instrument's functions have changed significantly.

3. Instrument maintenance

This instrument should be regularly cleaned with a clean, soft cloth dampened with a small amount of anhydrous alcohol.

If there are stains on the surface of this instrument, clean it with a soft cloth dipped in cleaning paste.

4. After-sales service

a) Warranty content

Within one month from the date of delivery, our company will be responsible for replacing any faults caused by defects in materials or manufacturing.

This instrument is covered for faults caused by material and manufacturing defects within 12 months from the date of delivery. During the warranty period, our company will selectively repair or replace any instruments found to be defective. Products under warranty must be sent by the user to a repair department designated by our company. The shipping cost for sending the instrument from the user to the maintenance department shall be borne by the user. Our company will cover the shipping costs for returning the instrument to the user.

For repairs outside the warranty period, our company will charge an appropriate repair cost.

b) Warranty coverage

The above warranty is not suitable for damage caused by improper use or maintenance by the user, use under non-compliant conditions, unauthorized repairs, or modifications.

Chapter 1 Introduction

Microplate incubators are formed by combining microprocessing technology with PID control methods. It features a compact size, light weight, and no pollution; Temperature and time LCD display, elegant and easy to operate. Mainly used for microplate labeling plates (96-well/384-well plates), 96-well cell culture plates, and other samples for cell culture and incubation at appropriate temperatures.

This product has the following features:

  • The LCD displays the system's operating status and parameter values in real time.

  • Uses high-quality switching power supplies, ensuring stable and reliable long-term operation of the entire machine.

  • Uses a one-touch knob for operation, simple and easy to use.

  • It has a power-off recovery function; when the external power supply is cut off and power is restored, the device can automatically resume operation according to the original set program.

  • Microprocessor control, good temperature control linearity, minimal fluctuations.

  • The cultivation time can be set arbitrarily within a range of 0~100 hours, and after the scheduled interval ends, an automatic prompt sound is emitted.

  • Double-layer heating system ensures stability during constant temperature cultivation.

Chapter 2: Characteristics

1. Normal working conditions

Operating temperature: 5°C ~ 30°C

Relative humidity: ≤70%

Power supply: AC220V

2. Basic parameters and performance

Parameters and performance
Temperature control rangeRoom temperature +5 °C ~ 80 °C
Time setting1min ~ 99h59min
Temperature control accuracy≤ ±0.5 °C
Display accuracy0.1 °C
Module temperature uniformity≤ ±0.5 °C
Heating time≤ 10 minutes (room temperature rises to 80°C)
Sample size2 microplates or culture plates
HeatingHeating film
Input power120W
Dimensions (mm)280×270×140
Weight (kg)4.0

Chapter 3 Basic Operation Instructions

This chapter mainly introduces the instrument's structure, control panel layout, and preparations before powering it on. When using this instrument for the first time, you should familiarize yourself with the contents of this chapter before turning it on.

1. Structural schematic

Structural schematic diagram 1    
   Structural schematic diagram 2

2. Operation panel diagram

Operation panel

PV: Represents the current system status parameters.

SV: Indicates the current user-set temperature/time parameters. sv 45.0 01:30

START: This indicates that the system is operating according to user-set parameters.

STOP: This indicates the system has stopped running and is in configuration mode.

KEEP: Indicates the user-set runtime as continuous runtime.

Operation panel diagram

Chapter 4: Operation Guide

1. Temperature and time settings

a) Turn on the power switch; the display will display the screen as shown, the instrument will enter initialization, accompanied by a "beep" sound.

System loading

b) The system enters parameter configuration mode. The PV displays the real-time temperature value of the heating platform (25.0°C), and STOP means the system has stopped heating and is in configuration mode. At this point, the cursor points to the temperature

c) At this point, a short press of the knob button will rotate the cursor position to the left and select the parameter values you want to adjust. Turn the knob clockwise/counterclockwise to adjust the value.

stop
37.0

2. Operation and stopping operations

a) After setting the parameters, press and hold the knob button (for more than 2 seconds). The system will emit two "beep" prompt sounds, and the system will start running. When the temperature rises to the preset value, the system starts counting down.

b) After the countdown ends, the system stops heating, the instrument begins to cool naturally, and the current heating platform temperature is displayed in real time. At this point, a short press of the knob button will return the system to configuration mode.

Operation schematic    

Note:In operating mode, press and hold the knob button (for more than 2 seconds) to automatically stop heating and return to configuration mode.

3. Installation and fixation of micro-hole plates

a) Open the cover, angle one end of the micro-hole plate slightly downward toward the spring, and press the spring piece tightly according to the arrow direction

Installation step a    

b) Press down the other end of the micro-hole plate in the direction of the arrow.

Installation step B    

c) After placing the microporous plate, press down slightly on the bottom surface of the microporous plate and contact the surface of the temperature-controlled plate.

Installation step c    

d) Fix another microporous plate using the above method.

Installation step d    

Chapter 5 Fault Analysis and Handling

Serial numberFault phenomenaCause analysisHandling methods
1The display screen does not light up after turning on the power switchPower is not connectedCheck the power supply and turn it on
The switch is damagedSwitch switch
OthersContact suppliers or manufacturers
2The temperature display seriously does not match the actual temperatureSensor damageContact suppliers or manufacturers
3The LCD screen displays the following message: "Fault Status" "PT-Error!"Sensor failureContact suppliers or manufacturers
4The LCD screen displays the following content: "Fault Status" "ADC-Error!"Data acquisition circuit failureContact suppliers or manufacturers
5The LCD screen displays the following content: "Fault Status" "Unknown!"Unknown errorsRestart the instrument or contact the manufacturer
6The module does not heat upTemperature sensor or heater damagedContact suppliers or manufacturers
7The knob keys are not workingThe knob keys are damagedContact suppliers or manufacturers

Note

The contents of this handbook are subject to change without prior notice. If you have any questions, please consult the supplier.