HUXI HT-200 Series Multi-Tube Vortex Mixer Instruction Manual

HUXI HT-200 series multi-tube vortex Mixer operation manual, covering instrument structure, parameter setting, running stop, pointing and pulse functions, supporting a variety of tuberacks, processing up to 50 samples, suitable for biological, medical and other fields.

Product manual for the multi-tube vortex mixer

Contents Directory

Quality Assurance 01

[1] Instrument Introduction 02

[2] Product Features 03

1. Normal working conditions 03

2. Basic Parameters and Performance 03

[3] Operation Instructions 04

1. Institutional Diagram 04

2. Operation Panel 05

[4] Operation Guide 07

1. Speed and Timer Settings 07

2. Stop Operation Function 07

3. Jog function 08

[5] Faults and Analysis 09

Welcome to the user manual for the "Multi-tube Vortex Mixer." Users should carefully read this manual before using this instrument, follow the instructions for operation, and understand all precautions.

How to get help

If you encounter any issues or need assistance during installation and use, please contact the manufacturer/supplier's after-sales service department promptly.

Please prepare the following materials:

  • Product serial number

  • Warranty card

  • Description of the problematic phenomenon

  • The methods and operational steps you took to solve the problem

  • Your phone, fax, and email addresses are all available for contact

Quality assurance

According to the manufacturer's quality assurance terms, when using the instrument under normal operating conditions and operating methods specified in this manual, the warranty period for the instrument is 12 months (from the date of purchase). If the instrument's performance deterioration or damage occurs due to incorrect installation and operation, unauthorized disassembly and maintenance, or other violations of the operating terms specified in the manual, repairs cannot be performed according to this quality guarantee. If you encounter any issues related to the warranty terms, please contact the manufacturer/supplier.

1. Instrument Introduction

The multi-tube vortex mixer uses brushless DC motors and microcomputer control technology. The unique knob operation mode is simple and easy to use. By changing different test tubes to secure the sponge, it can mix and cultivate various commonly used test tubes, with up to 50 samples mixed at once. Suitable for fields such as biotechnology, microbiology, and medical analysis.

This product has the following features:

  • One-touch knob operation mode, simple and easy to use.

  • Set any timer within 1min ~ 99h59min, and an automatic prompt sound will be emitted after the operation ends.

  • Multiple sponge test tube racks are available, offering a wide range of applications.

  • Soft start ensures uniform acceleration, effectively preventing sample splashing.

  • Driven by a brushless DC motor, it offers precise speed, long lifespan, and maintenance-free operation.

  • It can process up to 50 test samples at once, making experiments efficient and fast.

  • The control panel is simple, precisely controlled by a microprocessor, and the LED displays speed and time in real time.

  • User-friendly programming with built-in three operating modes: timing, continuous, and jog

2. Product Features

1. Normal working conditions

  • Operating temperature: 0°C-50°C

  • Relative humidity: ≤70%

  • Power supply: AC100-240V 50/60Hz

2. Basic parameters and performance

ModelHT-200
Rotational speed200~2500rpm
Amplitude4mm horizontal rotation
Time setting1min ~ 99h59min
Maximum payload4.5kg
Input power70W
External dimensions410x285x430mm (L, W, H)
Sample sizeTest tube rack modelsTest tube rack specificationsApplicable test tube specifications (outer diameter mm)
MS200-T1050x Ø10mmØ9~Ø10mm
MS200-T1250x Ø12mmØ11~Ø13mm
MS200-T1550x Ø15mmØ14~Ø17mm
MS200-T2050x Ø20mmØ18~Ø21mm
MS200-T2515x Ø25mmØ24~Ø27mm
MS200-T2915x Ø29mmØ28~Ø32mm
Weight20kg

3. Operation instructions

This chapter mainly introduces the instrument's structure, the functions of the control panel, and preparations before powering it on. When using this instrument for the first time, you should familiarize yourself with the contents of this chapter before turning it on.

1. Structural schematic

Structural schematic diagram 1            
Structural schematic diagram 2            

2. Operation panel

Operation panel diagram            
Key namesDefinition
Speed adjustment knobPreset speed adjustment keys: short press to adjust step values, left/right rotation to adjust speed
Time adjustment knobTimer value adjustment key: short press to adjust step values, left/right rotation to adjust the timer value
ShortPress the button, press and hold to start immediately, then release to stop
Short/StartStart/Stop button: short press to start, long press for more than 2 seconds to stop

3. Preparation before powering on

  1. Flatten the lower test tube sponge pad (2) into the vibrating platform (1).

  2. Insert the test tube into the sponge tube mounting frame (3) and place it entirely on the test tube sponge pad (2).

  3. Place the upper tray sponge pad (4) on top of the test tube.

  4. Attach the fixing sleeves at both ends of the test tube fixing plate (5) onto the body's fixing rail, gently press down the upper fixing plate to the sponge position, then infuse 3~5 millimeters to ensure firm contact between the sponge and the top of the test tube, then tighten the handwheel.

Note:

1. Do not use test tubes of different heights at the same time.

2. Test tubes must be covered (stoppered) and tightly tightened.

3. Test tubes are placed symmetrically on sponge test tube racks.

Diagram of preparation before powering on            

4. Operation Guide

1. Speed and timer settings

  1. When the power switch is turned on, the display screen will show the image shown in the picture, the instrument will enter initialization, accompanied by a "beep" sound.

  2. The system enters parameter configuration mode. The left window displays the current set speed, the right window shows the countdown time, and each window has a number flashing. The time separator in the right window does not flash.

  3. At this point, a short press of the knob button will rotate the flashing position to the left in a loop, selecting the digit you want to adjust. Turn the knob clockwise/counterclockwise to adjust the value.

  4. When the timer is less than 1 minute, it displays "00:00". Turn off the timer function and turn on the timer function. The system is manually stopped by the user.

2. Run and stop functions

  1. After setting the parameters, a short press of the Stop button will emit one "beep" sound from the system, signaling the system to start running.

  2. When the speed is adjusted to the preset value and the system starts countdowning, the time separator in the right window starts flashing.

  3. After the countdown ends, the system stops running. The system emits two "beep" prompt sounds, accompanied by three consecutive "beeps," and returns to configuration mode.

  4. Long pressing the button forcibly stops operation.

Runtime interface schematic 1            
Runtime interface schematic diagram 2            
Runtime interface schematic 3            
Runtime interface schematic 4            
Runtime interface schematic 5            
Operation interface schematic 6            

3. Jog function

Holding down the button will trigger one Short "beep" sound, signaling the system will start running. Release the button to stop the system jog mode.

Jogging function diagram            

Note:

When the time parameter is set to "00:00", it means the instrument continues to operate in timing mode.

After the timer is completed, the instrument will automatically stop and emit a "beep" prompt sound, then automatically enter configuration mode.

The left window value adjustment is controlled by the left knob, and the right window value adjustment is controlled by the right knob. When using the timing function, the maximum timing reaches 99 hours and 59 minutes. If the maximum timing time is exceeded, the system will continue running but then stop.

Non-professionals should never handle faults while powered on.

4. Pulse function settings

  1. Press and hold the knob on the left to enter the pulse settings interface. The left window displays the pause time, setting the range from 0 to 10 seconds; 0 means no pulse function. The right window displays the runtime, sets the range from 1s to 99min59s, and each window has a flashing number. The time separator in the right window does not flash.

Pulse function settings are shown in Figure 1            
  1. At this point, a short press of the knob button will rotate the flashing position to the left in a loop, selecting the digit you want to adjust. Turn the knob clockwise/counterclockwise to adjust the value.

Pulse function settings are shown in Figure 2            

Note: During operation, you can long-press the left knob to change the pulse setting parameters.

5. Fault Analysis and Handling

Fault phenomenaCause analysisHandling methods
The display screen does not light up after turning on the power switchPower is not connectedCheck the power supply and turn it on
The switch is damagedSwitch switch
OthersContact suppliers or manufacturers
After operation, the liquid has no vorticesThe speed setting is unreasonableReset the speed
The upper tray is pressed too looselyAdjust the sponge position and press again
The instrument vibrates loudlyMix evenly and place asymmetricallyPlace the mixture symmetrically
The displayed speed does not match the actual speedController failureContact suppliers or manufacturers
The left screen shows 8888Motor drive circuit failureContact suppliers or manufacturers
The knob doesn't workThe button is damagedContact suppliers or manufacturers
Excessive noiseThe handwheel wasn't tightenedTighten the left and right hand wheels
Loose components and other causesContact suppliers or manufacturers
  • Check the power cord connections to ensure the external power supply meets the requirements.

  • After replacing the fuse wire, restart the instrument. If the problem persists, please contact the supplier or manufacturer.

  • Restart the instrument. If the problem persists, contact the supplier or manufacturer.

Safety warning diagram            
There is high voltage inside the product; unauthorized disassembly is strictly prohibited!