Reader's Guide
This manual is compiled based on manufacturer's data. Due to equipment batch changes, the content may not fully match every physical unit. Please refer to the actual operation and on-device labels.
Copyright belongs to the manufacturer. For internal reference of purchasers only. If safety parameters in the manual differ from the physical unit, please suspend operation and consult our engineer for confirmation.
Disclaimer: This manual is for reference only. In case of any discrepancy, the physical unit shall prevail. Our company assumes no liability for any consequences arising from operations based on this manual.
SANFA GHP series light chamber instruction manual
1. Scope of Application
This product is the ideal equipment for cell tissue growth cultivation, seed germination, seedling cultivation, insect and animal breeding, microbial antibiotic cultivation, and preservation. It is especially suitable for production and research in fields such as bioengineering, medical research, agricultural science, aquaculture, and animal husbandry for constant temperature and light experimental cultivation.
2. Technical Indicators
| Model | GHP-160 | GHP-250 | GHP-300 | GHP-400 | GHP-500 |
|---|---|---|---|---|---|
| Volume | 160L | 250L | 300L | 400L | 500L |
| Refrigerants | Fluorine-free | ||||
| Ambient temperature | 5~35°C | ||||
| Temperature range | 10-60°C with light, 5-60°C without light | ||||
| Temperature uniformity | ±1°C | ||||
| Temperature fluctuations | ±1°C | ||||
| Illuminance | Universal light 0~6000LX adjustable in five levels, strong light 0~20000LX adjustable percentage | ||||
| Control methods | Intelligent microcomputer control | ||||
| Power supply | ~220V 50HZ | ||||
| Maximum power consumption | 700W | 1000W | 1200W | 1500W | 2500W |
| Note | |||||
3. Overview of the structure
The outer shell of this product is made of high-quality steel plates with a powdered surface. The chamber uses stainless steel plate, which is durable and corrosion-resistant. The outer door is equipped with an observation window and double-layer insulated glass doors for easy observation of temperature tests inside the chamber.
Uses smart microcomputer-controlled touch switches for easy operation. Multiple parameters of the lighting box (including time, temperature, and illuminance) can be set to simulate natural climates. Inside the box, axial fans are used for forced convection, ensuring uniform temperature and good performance. The lighting uses T4 fluorescent tubes, which are energy-saving, durable, and easy to maintain.
4. Panel display
(Photo 1)
5. Indicator lights and button descriptions
Temperature and parameter settings
After powering on, the system enters the main monitoring screen and presses the "Query/Shift" key to look up the set values. Self-tuning function: In compressor normally on and balanced mode, if temperature and humidity control is not ideal, start the temperature and humidity self-tuning. Press and hold the "Decrease" button to activate the temperature and PID auto-tuning.
1. Run parameter editing
On the main interface, press the "Set" key to enter the menu. Use the arrow keys to select "Run Parameters," then press the "Set" key to enter the editing interface. After each parameter is modified, press the "Set" key to move to the next parameter. If the workgroup is set to that group, it will run accordingly. If set to 0 for the current period, it will mark the last program segment and exit editing. In day-night mode, the two-segment ending is fixed. The second paragraph is set to 0, indicating that this group operates in fixed operating mode. Press and hold the "Settings" button to return to the main monitoring screen.
2. User parameters, temperature parameters, and system parameter settings
In the menu interface, select the corresponding directory and press the "Set" key to enter. System parameters require password entry. The parameters are as follows:
| User parameters | Function description | Range (factory value) |
|---|---|---|
| Power-On Mode (PM) | 0 - Power-on to shut down, 1 - Power-on to operate | 0~1 (0) |
| Program Pattern (PROG) | 0, fixed value; Paragraphs 1, 99 | 0~1 (0) |
| Timing Method (TM) | 0 - Start timing, 1 - Temperature stabilization timing | 0~1 (1) |
| Timing unit (HM) | 0 minutes, 1 hour | 0~1 (0) |
| Working Group (GP) | 0-8 → 1-9 groups | 0~8 (0) |
| Day and Night (BH) | 0 - Normal, 1 - Day and night (day first), 2 - Night first | 0~1 (0) |
| Disinfection Time (SRT) | Disinfection time each time | 0~6000min(0) |
| Temperature parameters | Function description | Range (factory value) |
|---|---|---|
| Alarm Value (AL) | 0 deviation alarm, 1 lower deviation alarm | -40~100 (3.0) |
| Scale Zone (P) | The smaller the P, the faster the response | 0~100.0 (6.0) |
| Integration Time (I) | The smaller the I, the stronger the corrected static error | 0~3600 (240) |
| Differential Time (D) | The larger the D, the stronger the overheating resistance | 0~3600 (240) |
| Points Limit (AR) | Suppresses humidity overshooting | 0~200 (100) |
| Control Cycle (T) | Control cycle output | 1~180s (6) |
| Turn on Refrigeration (CH) | Switching start refrigeration | 0~10.0 (0.8) |
| Turn off the refrigeration (CL) | Switching shut-off of refrigeration | 0~10.0 (0.8) |
| Heating Off (HC) | Switching off heating | 0~10.0 (0.5) |
| Refrigeration Delay (CT) | Compressor delay | 0~1800s (120) |
| Zero Correction (TB) | Temperature correction near zero | -99.9~+199.9(0) |
| Full Correction (TK) | Correction near the set temperature | -1999~+1999(0) |
| Lower Temperature Limit (TL) | Measure the lower limit | -40.0~100.0(0) |
| Temperature Upper Limit (TH) | Upper limit of measurement | -40.0~100.0(60) |
| Humidity parameters | ||
| Scale Zone (P) | The smaller the P, the faster the response | 0~100.0(10) |
| Integration Time (I) | The smaller the I, the stronger the corrective static deviation | 0~3600(300) |
| Differential Time (D) | The larger D is, the better the overshoot | 0~3600(70) |
| Points Limit (AR) | Integration limit | 0~200(100) |
| Control Cycle (T) | Control cycle | 1~180s(3) |
| Enable Temperature (KT) | Enable temperature deviations for humidity control | 0~10.0(1) |
| Humidification (HC) | Regarding humidification deviation | 0~10.0(1) |
| Turn on dehumidification (CH) | Start the dehumidification condition | 0~10.0(2) |
| Turn off dehumidification (CL) | Moisture removal conditions are applied | 0~1800s(1) |
| Zero Correction (HB) | Actual humidity value - current measured value | -99.9~+99.9(0) |
| Full Degree Correction (HK) | Humidity fullness correction | -1999~+1999(0) |
Print and store
| Parameter name | Function description | Range (factory value) |
|---|---|---|
| Storage intervals | USB storage spacing | 0~9999 (0) |
| Print interval | Printer intervals | 0~9999 (0) |
| Units of time | Printing time units | 0~1 (0) |
| Export data | USB drive export | 0~1 (0) |
| Printer | Printer type | 0~9999 (0) |
| Print the curve | 0 data / 1 curve | 0~1 (0) |
| Set the print | Whether a set value is printed | 0~1 (1) |
| Paper width | Printer paper width | 0~9999 (0) |
System parameters Press and hold the 'Set' key, set the password to 12, and press the 'Set' key to enter.
| Parameter name | Function description | Range (factory value) |
|---|---|---|
| Emergency Reporting (AM) | Deviations from top to bottom, etc | 0~4(0) |
| Temperature Unit (CF) | Celsius/Fahrenheit | 0~1(0) |
| Gate Enable (DR) | None/closed/disconnect is effective | 0~2(0) |
| Water Level Enabled (WT) | No/closed/disconnect | 0~2(0) |
| Decimal point (DP) | Decimal places | 0~1(1) |
| Cooling Temperature (CB) | Compressor normally open threshold | -20.0~60.0(0) |
| Dehumidification Temperature (JS) | Keep the dehumidifying compressor running constantly | 0.0~99.9(0) |
| Temperature Filtering (ERT) | Filter level | 0~9(1) |
| Moisture Filtering (ERH) | Filter level | 0~9(4) |
| Auxiliary Output (OF) | Temperature/humidity/hydration | 0~4(0) |
| Auxiliary Temperature (OT) | Auxiliary output temperature | -40.0~99.9(60) |
| Auxiliary Humidity (OH) | Auxiliary output humidity | 0.0~99.9(99.9) |
| Parameter name | Function description | Range (factory value) |
|---|---|---|
| Dual compressor (CM) | Switch / Defrost mode | 0~9999h(0) |
| Defrosting interval 1 (PT1) | Set < 10°C | 0~9999h(0) |
| Defrosting time 1 (T1) | Set < 10°C | 0~9999s(0) |
| Defrosting interval 2 (PT2) | 10~25°C | 0~9999h(0) |
| Defrosting Time 2 (T2) | 10~25°C | 0~9999s(0) |
| Defrosting interval 3 (PT3) | >25°C | 0~9999h(0) |
| Defrosting Time 3 (T3) | >25°C | 0~9999s(0) |
Additional system parameters Press and hold the 'Set' key, set the password to 166, and press the 'Set' key to enter.
| Parameter name | Function description | Range (factory value) |
|---|---|---|
| Language choice | Chinese/English | 0~1(0) |
| Humidity activated | Humidity is present | 0~1(0) |
| Light on | whether there is or no illumination | 0~1(0) |
Time setting
(a) Long press the "Query Shift" button to change Beijing Time. After modification, click "Set" to save.
(b) Press the up and down arrow keys to select the scheduled entry, move it to the designated position, then click "Set" to start editing, then click the "Query Shift" button. After editing, press "Set" to exit the editing mode. Long press "Settings" to return to the menu. When year-month-day is a valid value, the day option is disabled. Example: 20-06-15 08:30 4 Startup means the computer started at 8:30 on June 15, 2020; 20-00-00 19:30 Fri Shutdown means every Friday at 19:30 shutdown.
6. Fault Handling
| Malfunction | The reason | Processing |
|---|---|---|
| The unit is not operating | Power outages; Fuse blows; Poor socket contact | Waiting for incoming calls; Replace the fuse; Reinsert it tightly |
| Poor cooling performance | Refrigerant leakage / capillary tube blockage / fan not turning | Professional personnel for maintenance |
| Poor temperature performance | Items are too crowded, ambient temperature, number of door openings, fan not running | leaving open spaces; Ambient temperature 0-30°C; Reduce door opening; Replace the fan; Reset the parameters |
| No heating | Heating tubes/relays/thyristors are damaged | Replace the corresponding parts |
| Operations are in disorder | 24C02 is damaged | Professionals are replaced |
| The button failed | Damaged buttons / loose sockets | Replace the panel/tighten the plug |
| No relay noise | Optocoupler/relay failure | Professional replacement |
| The humidification lamp is on but not humidified | Humidifier cuts out or is damaged | Add water/replace |
| The casing is electrocuted | No grounding wire or poor grounding | Deactivate and re-ground |
7. Wiring diagram
Note: Light Level 1: Light Level 1 works; Light Level 2: Light 2 operates; Light Level 3: Light 1 and 2 operate simultaneously. Compressor 2 outputs 220V/5A, requires an external contactor or solid-state relay.
8. Packing list
| Serial number | Category | Name | Unit | Quantity | Note |
|---|---|---|---|---|---|
| 1 | Documents | User manual | portion | 1 | |
| 2 | Documents | Certificate of Conformity | portion | 1 | |
| 3 | Documents | Product warranty card | portion | 1 | |
| 4 | Spare parts | Melt core | Only | 2 | |
| 5 | Spare parts | Shelf | Block | 3 |
The items listed on this list match the actual items inside the box. Packer 2
- 1SANFA LHP-160E Constant Temperature and Humidity Incubator
- 2SANFA LGP series Artificial climate box instruction manual
- 3SANFA SHP-100/160/250/300/400/500 biochemistry cultivating box instruction manual
- 4SANFA SHP-E series biochemistry cultivating box instruction manual
- 5SANFA SYQX-II anaerobic incubator instruction manual
- 6NOBOTECH NPT-PH101/PH106 Waterproof Pen pH/ORP Meter Instruction Manual
- 7NOBOTECH NPT-CD501/CD506 Waterproof Pen Type Conductivity detector /TDS Instruction Manual
- 8NOBOTECH NPT-DO601 Pen Dissolved Oxygen Detector Instruction Manual
- 9NOBOTECH NPT-CLOZ801 Pen type residual chlorine/ozone detector instruction manual
- 10NOBOTECH PH-1186 Industrial Online PH/ORP Meter Instruction Manual