SANFA GHP series light chamber instruction manual

SANFA GHP series light chamber instruction manual, covering temperature lighting control, parameter setting, program operation, fault handling and other operation instructions, suitable for biological foster experiments.
Table of Contents    
1. Scope of Application
2. Technical Indicators
3. Overview of the structure
4. Panel display
5. Indicator lights and button descriptions
6. Fault Handling
7. Wiring diagram
8. Packing list

1. Scope of Application

This product is the ideal equipment for cell tissue growth cultivation, seed germination, seedling cultivation, insect and animal breeding, microbial antibiotic cultivation, and preservation. It is especially suitable for production and research in fields such as bioengineering, medical research, agricultural science, aquaculture, and animal husbandry for constant temperature and light experimental cultivation.

2. Technical Indicators

ModelGHP-160GHP-250GHP-300GHP-400GHP-500
Volume160L250L300L400L500L
RefrigerantsFluorine-free
Ambient temperature5~35°C
Temperature range10-60°C with light, 5-60°C without light
Temperature uniformity±1°C
Temperature fluctuations±1°C
IlluminanceUniversal light 0~6000LX adjustable in five levels, strong light 0~20000LX adjustable percentage
Control methodsIntelligent microcomputer control
Power supply~220V 50HZ
Maximum power consumption700W1000W1200W1500W2500W
Note

3. Overview of the structure

The outer shell of this product is made of high-quality steel plates with a powdered surface. The chamber uses stainless steel plate, which is durable and corrosion-resistant. The outer door is equipped with an observation window and double-layer insulated glass doors for easy observation of temperature tests inside the chamber.

Uses smart microcomputer-controlled touch switches for easy operation. Multiple parameters of the lighting box (including time, temperature, and illuminance) can be set to simulate natural climates. Inside the box, axial fans are used for forced convection, ensuring uniform temperature and good performance. The lighting uses T4 fluorescent tubes, which are energy-saving, durable, and easy to maintain.

4. Panel display

Panel diagram      

(Photo 1)

5. Indicator lights and button descriptions

Temperature and parameter settings

After powering on, the system enters the main monitoring screen and presses the "Query/Shift" key to look up the set values. Self-tuning function: In compressor normally on and balanced mode, if temperature and humidity control is not ideal, start the temperature and humidity self-tuning. Press and hold the "Decrease" button to activate the temperature and PID auto-tuning.

1. Run parameter editing

On the main interface, press the "Set" key to enter the menu. Use the arrow keys to select "Run Parameters," then press the "Set" key to enter the editing interface. After each parameter is modified, press the "Set" key to move to the next parameter. If the workgroup is set to that group, it will run accordingly. If set to 0 for the current period, it will mark the last program segment and exit editing. In day-night mode, the two-segment ending is fixed. The second paragraph is set to 0, indicating that this group operates in fixed operating mode. Press and hold the "Settings" button to return to the main monitoring screen.

2. User parameters, temperature parameters, and system parameter settings

In the menu interface, select the corresponding directory and press the "Set" key to enter. System parameters require password entry. The parameters are as follows:

User parametersFunction descriptionRange (factory value)
Power-On Mode (PM)0 - Power-on to shut down, 1 - Power-on to operate0~1 (0)
Program Pattern (PROG)0, fixed value; Paragraphs 1, 990~1 (0)
Timing Method (TM)0 - Start timing, 1 - Temperature stabilization timing0~1 (1)
Timing unit (HM)0 minutes, 1 hour0~1 (0)
Working Group (GP)0-8 → 1-9 groups0~8 (0)
Day and Night (BH)0 - Normal, 1 - Day and night (day first), 2 - Night first0~1 (0)
Disinfection Time (SRT)Disinfection time each time0~6000min(0)
Temperature parametersFunction descriptionRange (factory value)
Alarm Value (AL)0 deviation alarm, 1 lower deviation alarm-40~100 (3.0)
Scale Zone (P)The smaller the P, the faster the response0~100.0 (6.0)
Integration Time (I)The smaller the I, the stronger the corrected static error0~3600 (240)
Differential Time (D)The larger the D, the stronger the overheating resistance0~3600 (240)
Points Limit (AR)Suppresses humidity overshooting0~200 (100)
Control Cycle (T)Control cycle output1~180s (6)
Turn on Refrigeration (CH)Switching start refrigeration0~10.0 (0.8)
Turn off the refrigeration (CL)Switching shut-off of refrigeration0~10.0 (0.8)
Heating Off (HC)Switching off heating0~10.0 (0.5)
Refrigeration Delay (CT)Compressor delay0~1800s (120)
Zero Correction (TB)Temperature correction near zero-99.9~+199.9(0)
Full Correction (TK)Correction near the set temperature-1999~+1999(0)
Lower Temperature Limit (TL)Measure the lower limit-40.0~100.0(0)
Temperature Upper Limit (TH)Upper limit of measurement-40.0~100.0(60)
Humidity parameters
Scale Zone (P)The smaller the P, the faster the response0~100.0(10)
Integration Time (I)The smaller the I, the stronger the corrective static deviation0~3600(300)
Differential Time (D)The larger D is, the better the overshoot0~3600(70)
Points Limit (AR)Integration limit0~200(100)
Control Cycle (T)Control cycle1~180s(3)
Enable Temperature (KT)Enable temperature deviations for humidity control0~10.0(1)
Humidification (HC)Regarding humidification deviation0~10.0(1)
Turn on dehumidification (CH)Start the dehumidification condition0~10.0(2)
Turn off dehumidification (CL)Moisture removal conditions are applied0~1800s(1)
Zero Correction (HB)Actual humidity value - current measured value-99.9~+99.9(0)
Full Degree Correction (HK)Humidity fullness correction-1999~+1999(0)

Print and store

Parameter nameFunction descriptionRange (factory value)
Storage intervalsUSB storage spacing0~9999 (0)
Print intervalPrinter intervals0~9999 (0)
Units of timePrinting time units0~1 (0)
Export dataUSB drive export0~1 (0)
PrinterPrinter type0~9999 (0)
Print the curve0 data / 1 curve0~1 (0)
Set the printWhether a set value is printed0~1 (1)
Paper widthPrinter paper width0~9999 (0)

System parameters Press and hold the 'Set' key, set the password to 12, and press the 'Set' key to enter.

Parameter nameFunction descriptionRange (factory value)
Emergency Reporting (AM)Deviations from top to bottom, etc0~4(0)
Temperature Unit (CF)Celsius/Fahrenheit0~1(0)
Gate Enable (DR)None/closed/disconnect is effective0~2(0)
Water Level Enabled (WT)No/closed/disconnect0~2(0)
Decimal point (DP)Decimal places0~1(1)
Cooling Temperature (CB)Compressor normally open threshold-20.0~60.0(0)
Dehumidification Temperature (JS)Keep the dehumidifying compressor running constantly0.0~99.9(0)
Temperature Filtering (ERT)Filter level0~9(1)
Moisture Filtering (ERH)Filter level0~9(4)
Auxiliary Output (OF)Temperature/humidity/hydration0~4(0)
Auxiliary Temperature (OT)Auxiliary output temperature-40.0~99.9(60)
Auxiliary Humidity (OH)Auxiliary output humidity0.0~99.9(99.9)
Parameter nameFunction descriptionRange (factory value)
Dual compressor (CM)Switch / Defrost mode0~9999h(0)
Defrosting interval 1 (PT1)Set < 10°C0~9999h(0)
Defrosting time 1 (T1)Set < 10°C0~9999s(0)
Defrosting interval 2 (PT2)10~25°C0~9999h(0)
Defrosting Time 2 (T2)10~25°C0~9999s(0)
Defrosting interval 3 (PT3)>25°C0~9999h(0)
Defrosting Time 3 (T3)>25°C0~9999s(0)

Additional system parameters Press and hold the 'Set' key, set the password to 166, and press the 'Set' key to enter.

Parameter nameFunction descriptionRange (factory value)
Language choiceChinese/English0~1(0)
Humidity activatedHumidity is present0~1(0)
Light onwhether there is or no illumination0~1(0)

Time setting

(a) Long press the "Query Shift" button to change Beijing Time. After modification, click "Set" to save.

(b) Press the up and down arrow keys to select the scheduled entry, move it to the designated position, then click "Set" to start editing, then click the "Query Shift" button. After editing, press "Set" to exit the editing mode. Long press "Settings" to return to the menu. When year-month-day is a valid value, the day option is disabled. Example: 20-06-15 08:30 4 Startup means the computer started at 8:30 on June 15, 2020; 20-00-00 19:30 Fri Shutdown means every Friday at 19:30 shutdown.

6. Fault Handling

MalfunctionThe reasonProcessing
The unit is not operatingPower outages; Fuse blows; Poor socket contactWaiting for incoming calls; Replace the fuse; Reinsert it tightly
Poor cooling performanceRefrigerant leakage / capillary tube blockage / fan not turningProfessional personnel for maintenance
Poor temperature performanceItems are too crowded, ambient temperature, number of door openings, fan not runningleaving open spaces; Ambient temperature 0-30°C; Reduce door opening; Replace the fan; Reset the parameters
No heatingHeating tubes/relays/thyristors are damagedReplace the corresponding parts
Operations are in disorder24C02 is damagedProfessionals are replaced
The button failedDamaged buttons / loose socketsReplace the panel/tighten the plug
No relay noiseOptocoupler/relay failureProfessional replacement
The humidification lamp is on but not humidifiedHumidifier cuts out or is damagedAdd water/replace
The casing is electrocutedNo grounding wire or poor groundingDeactivate and re-ground

7. Wiring diagram

Wiring diagram    

Note: Light Level 1: Light Level 1 works; Light Level 2: Light 2 operates; Light Level 3: Light 1 and 2 operate simultaneously. Compressor 2 outputs 220V/5A, requires an external contactor or solid-state relay.

8. Packing list

Serial numberCategoryNameUnitQuantityNote
1DocumentsUser manualportion1
2DocumentsCertificate of Conformityportion1
3DocumentsProduct warranty cardportion1
4Spare partsMelt coreOnly2
5Spare partsShelfBlock3

The items listed on this list match the actual items inside the box. Packer 2