NOBOTECH NPT-PH101/PH106 Waterproof Pen pH/ORP Meter Instruction Manual

NOBOTECH NPT-PH101/PH106 waterproof pen pH/ORP meter instruction manual, covering product characteristics, technical parameters, pH/ORP calibration, mode switching, temperature calibrating and maintenance instructions, intuitive and convenient operation.

Waterproof pen-type pH/ORP meter user manual

Product images        
Product image 2        
Product image 3        

Table of Contents

1. Introduction ....1

2. Product configuration and features ....1

(1) Standard Configuration:.... 1

(2) Features:.... 1

(3) Technical Parameters:.... 1

3. Function Introduction ....2

(1) Overall Structural Composition: .... 2

(2) Display Area Composition:.... 2

(3) Key function overview table

(4) pH calibration

(5) ORP calibration

(6) Mode switching

(7) Temperature calibration

(8) Backlight Adjustment (Large Screen Version)

4. Maintenance ....7

(1) Battery Replacement (Small Screen Version)

(2) Battery Charging (Large Screen Version)

(3) Electrode head replacement

(4) Electrode maintenance

(5) Frequently Asked Questions

1. Introduction

Thank you for using our pen-type product series. This product is a waterproof pen-type pH/ORP meter with patented technology. This product has been tested and calibrated before leaving the factory, so you can use it with confidence.

2. Product configuration and features

(1) Standard Configuration:

  • Pen-type pH/ORP meter

  • One user manual

  • One set of pH standard solution

  • One TYPE-C charging data cable (large screen version)

(2) Features:

  • Measurement range: pH 0.00~14.00, ORP -999mV~999mV, temperature: 0~60°C

  • LED segment LCD display (pH/ORP values and temperature values)

  • Numerical and temperature calibration functions

  • Equipped with a 32-bit high-performance processor, it offers high stability and low power consumption

  • Two groups of calibration solutions: (1) 6.86 and 4.00 (2) 6.86 and 9.18, switchable freely

  • Automatically recognizes standard liquid; calibration cannot be completed if the standard solution does not match

  • Freely switchable between pH/ORP modes, making pH/ORP electrode heads easy to replace

  • Automatically shut down after five minutes of no button operation

  • Low battery alert function

  • Ultra-long standby technology, theoretical standby time of over one year

  • Protection rating: IP65

(3) Technical Parameters:

  • Resolution: 0.01 pH/1mV, 0.1°C

  • Accuracy: ± 0.05 pH, ± 2mV, ± 0.3°C

  • Automatic temperature compensation: 0~60°C, temperature compensation: NTC 10K

  • Power cover

  • Operating conditions: ambient temperature 0~60°C, relative humidity ≤90%

  • Operating voltage: Large screen: rechargeable lithium battery 3.7V, small screen: LR44×4 button battery 6V

  • Dimensions: Overall size: 204mm×46mm; Large screens 25mm × 32mm, small screens 20mm × 28mm

  • Weight: Large screen version: 128g, Small screen version: 111g

3. Function Introduction

(1) Overall Structural Composition:

  1. Power cover

  2. Display area

  3. Operation area

  4. Waterproof cover

  5. Electrode head

Overall structural diagram        

ORP electrode head                pH electrode head            

ORP electrode head, pH electrode head

(2) Display Area Composition:

1. Main display area (pH/ORP measurement values) 2. Secondary display area: temperature measurement values 3. Calibration prompt 4. Battery level prompt 5. Lock prompt 6. Current mode

Display area icon        

(3) Key function overview table

Serial numberInterface/OperationKey 1Key 2Key 3
Short pressPress and holdShort pressPress and holdShort pressPress and hold
1Main interfaceLock the valueShut downNoneFactory Reset (hold down the MODE button first)Menu interfaceNone
2Select the actionReturn to the main interfaceShut downFlip downEnter/confirmFlip upwardNone
3Edit operationsReturn to the main interfaceShut downEditor position minus 1ConfirmAdd 1 to the editorial positionMove the edit bit
9Boot up and operateStart the machineStart the machineNoneNoneNoneNone

(4) pH calibration

Before calibration and when replacing the standard solution, wash the pH electrode tip in clean water, gently wipe dry with a clean tissue, then place the electrode tip into the standard solution to be calibrated and gently shake 3~5 times.

pH standard solutions are generally prepared by dissolving a packet of pH buffer in 250mL of purified or distilled water.

StepsOperationDisplay
1On the main interface, short press the MODE button to enter the menu, then use the MODE or CAL keys to select the corresponding menu item. Press and hold the CAL key to enter the interface.0 CAL calibration
2Use the MODE or CAL keys to select the calibration value of the secondary display area up and down. Place the electrode tip into the standard solution to be calibrated and gently shake 3~5 times. 
Zero position calibrationAcidic slopeAlkaline slope
3Observe the current value in the main display area; if it is stable and flickering, hold down the CAL key to confirm. After calibration is complete, it will jump to the slope calibration value or return to the main interface. 
4The main display area flashes, indicating that the current value differs too much from the calibration value. At this time, long-pressing the CAL key for calibration will trigger a 2-second error message. 

(5) ORP calibration

Before calibration, wash the ORP electrode tip in clean water, gently wipe dry with a clean tissue, then place the electrode tip into the standard solution to be calibrated and gently shake 3-5 times.

ORP calibration should use a 220mV ORP standard solution.

StepsOperationDisplay
1On the main interface, short press the MODE button to enter the menu, then use the MODE or CAL keys to select the corresponding menu item. Press and hold the CAL key to enter the interface.ORP calibration iconCalibration
2Use the MODE or CAL keys to select the secondary display area's calibration value of 220mV. Place the electrode tip into the standard solution and gently shake 3~5 times.ORP calibration
4Observe the current value in the main display area; if it is stable and flickering, hold down the CAL key to confirm. After calibration is complete, you will return to the main interface. 
5The main display area flashes, indicating that the current value differs too much from the calibration value. At this time, long-pressing the CAL key for calibration will trigger a 2-second error message. 

(6) Mode switching

After switching modes, please replace the electrode head corresponding to the mode. For example, in pH mode, replace the pH electrode head. For replacement methods, please refer to the electrode replacement section. If the measurement is inaccurate, recalibration with standard solution is required.

StepsOperationDisplay
1On the main interface, short press the MODE button to enter the menu, then use the MODE or CAL keys to select the corresponding menu item. Press and hold the CAL key to enter the interface.Mode switching
2Use the MODE or CAL keys to select the mode to switch to. Press and hold the CAL key to confirm. The device automatically returns to the main interface.PH Mode ORP Mode

(7) Temperature calibration

The temperature sensor is located near the electrode head. Since air conducts heat slowly, it is recommended to place the electrode head together with a standard thermometer in water and adjust it based on the standard thermometer reading.

StepsOperationDisplay
1On the main interface, short press the MODE button to enter the menu, then use the MODE or CAL keys to select the corresponding menu item. Press and hold the CAL key to enter the interface.Temperature
2Use the MODE or CAL keys to edit the temperature value of the secondary display area up and down; long-press the MODE button to move the edit position. Place the electrode tip into the solution at the appropriate temperature.Edit the temperature value
3When the main display area temperature is stable, press and hold the CAL key to confirm. The device automatically returns to the main interface. 

(8) Backlight Adjustment (Large Screen Version)

The pen-style meter on the large screen allows adjustment of backlight brightness, while the small screen does not have this feature.

Backlight brightness adjustment range is L1~L5, factory default is L3. L5 is the brightest, L1 is equivalent to the backlight going out.

StepsOperationDisplay
1On the main interface, short press the MODE button to enter the menu, then use the MODE or CAL keys to select the corresponding menu item. Press and hold the CAL key to enter the interface.Backlight
2Use the MODE or CAL keys to adjust brightness up or down. Press and hold the CAL key to confirm. 

4. Maintenance

(1) Battery Replacement (Small Screen Version)

  1. Release the battery cover counterclockwise (do not discard the waterproof O-ring).

  2. Monitor the positive and negative directions of the battery and replace it with four new LR44 button cells.

  3. Put the battery cover and gasket back in their original places, then tighten them clockwise.

Battery replacement Figure 1        
Battery replacement Figure 2        
Battery replacement Figure 3        
Battery replacement Figure 4        

(2) Battery Charging (Large Screen Version)

  1. Release the battery cover counterclockwise (do not discard the waterproof O-ring).

  2. Plug in the Type-C cable, use a 5V charger on the USB end, and wait for charging to complete.

  3. Put the battery cover and gasket back in their original places, then tighten them clockwise.

Battery charging diagram        

Note:

If you don't plan to use the pen meter for a month or longer, you should remove the battery. Do not leave batteries inside the device.

(3) Electrode head replacement

  1. Screw the tightening sleeve clockwise.

  2. Remove the old electrode head, pay attention to the installation direction of the electrode head, and install the new electrode head.

  3. Tighten the screw-tightening sleeve counterclockwise.

Electrode head replacement diagram        

(4) Electrode head maintenance

  1. When measuring with a pen gauge, the electrode tip should first be cleaned in distilled water (or deionized water), then absorbed with filter paper to prevent impurities from entering the test solution. One third of the electrode tip should be inserted into the test solution.

  2. When not in use, the pen gauge should be cleaned thoroughly and the head inserted into a protective cover. You can use purified water or tap water to keep the sponge inside the cover moist.

  3. Electrode heads should be avoided from prolonged immersion in distilled water or protein solutions, and should not come into contact with silicone grease.

  4. For pH/ORP heads that have been used for a long time, their glass film may become semi-transparent or contain deposits. In such cases, dilute hydrochloric acid can be used for washing and rinsing with water.

(5) Frequently Asked Questions

  1. If the device does not display when powered on, it may be due to incorrect battery orientation or insufficient battery power. Please replace or recharge the battery promptly.

  2. If accidental operation causes abnormal measurement values, please restore factory settings and recalibrate before use.

  3. If a low battery warning appears, replace the battery or recharge it promptly to avoid inaccurate measurements.

  4. Under normal circumstances, equipment does not require calibration. If inaccurate values are found, first compare with the standard solution. If there is a significant deviation from the standard solution, recalibration is required.

  5. The lifespan of the electrode head depends on the usage environment and maintenance, ranging from six months to one year.