Home> Library> Read
📖

Reader's Guide

This manual is compiled based on manufacturer's data. Due to equipment batch changes, the content may not fully match every physical unit. Please refer to the actual operation and on-device labels.

Copyright belongs to the manufacturer. For internal reference of purchasers only. If safety parameters in the manual differ from the physical unit, please suspend operation and consult our engineer for confirmation.

Disclaimer: This manual is for reference only. In case of any discrepancy, the physical unit shall prevail. Our company assumes no liability for any consequences arising from operations based on this manual.

YIHENG MP MPG-H heating circulation tank instruction manual

YIHENG MP MPG-H heating cycle tank instruction manual, covering structure, technical specifications, parameter settings, usage methods and troubleshooting, suitable for thermostatic control in Petrochemicals, medicine, life sciences and other fields.

1. Diagram

Schematic diagram of the constant temperature circulation tank    

2. Scope of use

The new generation MP(MPG)H series constant temperature circulation tank is developed by the company's years of production experience, incorporating German technology. This product features high quality, reliability, and stable performance, and is widely applicable in fields such as petroleum, chemical pharmaceuticals, life sciences, metrology, light industry, and scientific research.

3. Overview of the structure

The constant temperature circulation tank consists of two parts: the circulation controller and the casing body. The circulation controller consists of a temperature controller, temperature sensor, heater, low water level controller, and circulation pump. The cabinet is made of high-quality thin steel plates, with a surface powder coating and a stainless steel inner liner.

On the lower right side of the circulation controller, there is a bath liquid circulation inlet and outlet connection port for external bathtub connections. The temperature controller uses a CPU microprocessor for PID control, offering high control accuracy and good stability, as well as alarm functions such as timing, over-temperature alarms, and liquid level protection.

4. Technical Indicators

Specifications and modelsMP-5HMP-13HMP-19HMP-200HMPG-200H
Power supply voltage(220±22)V (50±1)Hz
Input Power (W)1000
Temperature control range (°C)(RT+5)~100(RT+5)~200
Temperature Resolution (°C)±0.1
Temperature fluctuation (°C)±0.1±0.2
Pump flow rate (L/min)4
Pump Head (cm)50
Filling Volume (L)6.710.922.510.914.5
Bathtub opening size (mm²)150×160240×170330×300240×170240×170
Dimensions (mm³)324×174×368328×266×368530×324×370352×324×370328×312×462

Note: Performance parameter test conditions are: ambient temperature 20°C, relative humidity ≤85%

5. Usage Instructions

1. Usage environment

a. Ambient temperature: 5°C~35°C;

b. Relative humidity: ≤85%;

c. Atmospheric pressure: 86 KPa ~ 106 KPa;

d. The elevation must not exceed 2000 meters;

e. Power supply: (220±22)V (50±1)Hz;

f. Stay away from heat sources, with no corrosive gases nearby, strong vibration sources, or strong electromagnetic fields.

2. The base for placing the equipment must be flat and made of non-combustible materials.

3. Maintain at least 300mm of space around the equipment (front and rear ventilation grids), ensure good ventilation, and ensure good grounding.

4. When pouring liquid into the bathtub, be careful not to pour it into the instrument. It is recommended to fill the maximum liquid level at least 30mm below the edge of the bathtub or at least two-thirds of the total height to raise the float to its highest position.

5. Add different bath liquids according to different temperature control ranges (strictly prohibited from overheating);

When the operating temperature is between 5°C~80°C, the bath liquid is purified water

When the temperature is above 80°C, the bath liquid is mineral oil (pay attention to the oil's ignition point to avoid fire!). )

6. Turn on the circulation controller power switch. At this point, the thermostat's PV screen displays the current liquid temperature, the SV screen shows the factory set temperature, and the circulation pump is running. When PV < SV, the HEAT heating light turns on and the instrument enters working mode.

7. Set the desired temperature:

Press the SET button; at this point, the first position of the SV screen will be highlighted, while the others will flash. Use the key to repeatedly select the specific high-brightness positions displayed on the SV screen, and use the X or key to change the temperature value displayed on the original SV screen until it reaches the desired value. After setting, press the SET button once to display the PV display "". If the timer function is not used, set SV to 0. Press the SET button again to enter working mode. The PV screen displays the measured temperature, and the SV screen shows the set temperature;

(Note: If using the timer function, see Chapter 6 for details)

8. When the machine is first turned on or used for a period of time, or when the ambient temperature changes (seasonal variations), it is necessary to check whether the actual temperature inside the bathtub matches the measured temperature error with the instrument. (If it does not meet the requirements, please correct the error according to Chapter 6)

9. If an external bathtub is required, remove the rubber hose connecting the inlet and outlet of the circulating bath liquid, and use spare parts (or a self-provided rubber hose) to connect directly to the inlet and outlet of the external bathtub.

Note: When configuring external circulation, ensure that the inlet and outlet of the external bath tank are basically level with the inlet and outlet of the equipment.

10. After use, turn off the circulation controller power switch, unplug the power connector, unscrew the liquid outlet and drain liquid from the head (please check the bath temperature and be cautious!). to avoid burns) and dry the cabinet.

2. The order in which each function parameter is retrieved

6. Explanation of changes in various functional parameters to improve product performance

1. Thermometer panel description

Thermometer panel    

1. PV: Displays the measured value, unit: °C

2. SV: Displays the set value, unit: °C

3. Install deviation alarm indicator light

4. Cooling operation indicator light (this function is not available in this unit)

5. Function keys

6. Shift key/remaining time observation key

7. Add key

8. Reduce keys

9. Heating indicator light

10. Self-tuning/timer operation indicator light

Panel illustration    

3. Explanation of changes in each functional parameter

Since products undergo strict testing before leaving the factory, corrections are generally not required. However, after the first startup, after a period of use, or in harsh environments, if the external temperature does not match the factory temperature, or when the temperature changes before or after using the control temperature, the temperature display may not match the actual temperature inside the workroom or become unstable (overshoot). If the product exceeds the technical specifications, parameters can be modified to meet factory standards.

1) Timer function

● When T1 is set to 0, the instrument disables the timing function; The timing function only appears when T1 is not set to 0.

a) The timer range is (1~9999) minutes, which is countdown timer operation; During scheduled operation, changing the timing value is invalid; When the timer setting is complete, press SET to start the timer; When the countdown ends, the output is automatically cut off, the buzzer sounds, the PV window displays "End" or the shift button is pressed, and the SV window shows "End."

b) When the instrument is running on a scheduled timer, pressing the shift key will display the remaining timer in the SV window;

(Note: If the countdown ends and you need to use the timer again, press the SET button once.) If the power is suddenly cut off and the machine restarts during timer use, the timer must be reset and restarted before it becomes effective).

● Explanation of timer setting method:

a) Press the "SET" button once; the instrument enters the (A) menu and enters the set temperature value;

b) Press the "SET" key again; the PV window will display "", press the and keys to enter the set timer;

c) Press the "SET" key again to exit the instrument menu and enter working mode;

d) If the decimal point in the lower right corner or the AT light flashes, it indicates the instrument has entered countdown mode.

2) Explanation of control parameter changes

Press the SET key for more than 5 seconds (enter menu B). When the PV screen displays "AL1," release it. Press the SET button several more times to find the "LCK" prompt, press the key so that the SV screen shows 0 (i.e., unlock). Press the SET key several times to find the prompt for the control parameter you want to adjust, and press the or key to modify the parameter to the desired value. All control parameters can be adjusted at once. Then press the SET key to find the "LCK" prompt, press the key to make the SV screen display 2 (locked), press the SET button for more than 5 seconds to return to working mode. At this point, the instrument executes the newly modified parameters.

After holding down the function key "SET" for 10 seconds, release the PV screen when the measurement changes from "AL1" to "HL" (i.e., enter the C menu). As described above, use the SET key to change parameter symbols, and use shift and add/minus keys to modify parameter values. After modification, press "SET" again for 5 seconds to enter working mode.

(Note: After 30 seconds of pressing without a key, the system will automatically return to working mode, but any changed data will be considered invalid.)

4. Give examples

1) Correction of measurement temperature and actual temperature error inside the workroom:

a) Place the glass mercury thermometer (using a mercury thermometer with ±0.1°C accuracy and a certified meter) into the workroom, with the mercury end positioned at the geometric center of the room; (The reading from the glass mercury thermometer is the actual temperature.)

b) When the device is powered on, after the temperature reaches the set temperature and stabilizes for about 1~2 hours (at constant temperature), compare the difference between the actual temperature and the PV display temperature. The difference (actual temperature minus PV display value) is the value of the temperature correction parameter that needs to be modified, i.e.:

SC = Raw SC value + (actual temperature - PV display value)

c) Calculate the SC value based on the above formula and enter it (if one adjustment is inaccurate, it can be repeated multiple times until it meets the standard).

d) If the SC temperature value is frequently changed during use, first set SC to 0 and return to working mode; Press the SET key for 10 seconds to enter the C menu. The PV screen will show "HL"; change HL to 1.000 and return to working mode; Within the control temperature range, select two test points, such as P₁ and P₂ points, (P₂> P₁), and the slope is:

HL = (glass P₂ - glass P₁) / (instrument P₂ - instrument P₁) (always keep three decimal places)

Set a new HL parameter value; If still inaccurate, adjust SC as in the previous three steps (repeat multiple times until the standard is met).

2) Suppression of temperature instability (excessive overshoot):

Method 1:

Enter menu B, decrease parameter P (proportional band), change it by about 0.5 each time, and change parameter I by about 500 seconds each time, and simultaneously change parameter D, D = I/4. Then enter menu C, decrease OUT (output power) by about 10% each time, and reheat and observe until overshoot meets the requirements.

Method 2:

When the required temperature is low, a secondary setting method can be used to reduce overshoot. For example: when the required temperature is 40°C, set it to 38°C first, and after the temperature stabilizes, set it to 40°C. This can reduce or even eliminate overshoot.

5. Tables of various function parameters

Menu B

PromptNameSet the rangeExplanationFactory preset
FIL1/AL1Upper deviation and water level alarm settings(-99.9~999.9)°CWhen the temperature exceeds the (SV+AL1) value or the liquid in the bathtub falls below the required threshold, the AL1 light turns on, the buzzer sounds, and the heating output is cut off
Cold/COLDLower deviation alarm settings(-99.9~999.9)°CWhen the temperature is below the (SV+COLD) value, the COLD light turns on
SC/SCCorrection of display value errors(-20~20)°CMeasure the actual temperature inside the box and compare it with the PV display temperature to correct display errors
ARU/ATUSelf-organizing command0,10:OFF 1:ON allows self-tuning of PID parameters0
P/PScale band(0~100)°CHeating ratio control: the higher the P ratio, the lower the system gain; Reducing P can improve system control accuracy and eliminate static differences, but if too low, it will cause system oscillation and instability
I/IIntegration Time (Readjustment Time)(1~4320)sThe time constant for the integration action—the larger I, the weaker the integral, and the system is stable
d/dDifferential Time (Pre-Tuned Overshoot)(0~1200)sThe time constant of the differential action: the larger the d, the stronger the differential action, and it can overcome overshoot. Generally, d is taken as (I/4)
「/THeating cycle(1~60)sThe thyristor output is (2~3) seconds. For devices with large remaining power, increasing T can reduce the static difference of PID control
LCE/LCKCombination lock0,1,20: Unlocking 1: Full lock 2: Locking all except SV2
HL/HLSlope0.500~1.500This ensures consistent temperature control accuracy throughout the entire measurement range
OUT⁻/OUT⁻Output power(0~100)%Adjustable heater output power

※ Products undergo strict testing before leaving the factory. When the studio's technical specifications meet requirements and work normally, generally no corrections are required.

Menu C

6. Pump flow regulation

Pump flow regulation diagram 1    
Pump flow regulation diagram 2    

The pump flow rate is preset before leaving the factory, but it can also be adjusted according to requirements.

· Use a screwdriver to hold down the screw (1) and rotate 360° counterclockwise.

· Use flat-nose pliers to rotate the control valve indicated in the diagram to the desired position.

· Tighten the screws

Example:

Internal circulation applications in bathtubs

A: 100% internal circulation of bathtubs (for larger bathtubs)

B. Smaller bathtubs have internal circulation

(Used for liquids with smooth surfaces and high flowability) External/Internal circulation applications

C. 40% external circulation flow, 60% internal circulation; (For larger bathtubs)

D: 80% external circulation flow, 20% internal circulation. (Used for some smaller bathtubs)

7. Self-tuning function

Generally, it is not recommended, so it will not be explained in detail.

8. Notes

Please ensure that you have read and understood the safety instructions listed in this manual before installing or operating the equipment.

1. When opening the packaging box, check whether the equipment is intact and whether any accessories are missing. If there is any damage or missing items, contact our company immediately;

2. The device must be connected to a well-grounded power outlet! Ensure the equipment casing is connected to the ground wire;

3. This equipment does not include a refrigeration device; its minimum control temperature is limited by ambient temperature at (RT+5)°C (RT is ambient temperature); If the ambient temperature is too high and causes inaccurate temperature control, please use indoor air conditioning to improve the ambient temperature;

4. When using temperatures above 80°C, carefully observe the distance between the liquid level and the edge of the bathtub to prevent thermal expansion of the bath oil during heating and avoid bath liquid overflow;

5. Be careful when draining hot bath liquid! Before draining, check the bath temperature (for example: open the bathtub cover and keep it open for a while);

6. Securely connect the rubber tube and clamp it tightly to prevent detachment;

7. During continuous operation, parts of the bathtub cover and the connection to the circulation pump become very hot. Therefore, be very careful when touching these areas to avoid burns;

8. Pay attention to the liquid medium in the workspace; be sure to add different bath fluids according to different operating temperatures;

9. When filling the bath, the bath liquid injected into the bathtub must be of the same model;

10. Do not use flammable bath liquids! If you cause damage caused by improper use of the bath solution, you are responsible for the consequences!

11. Before moving your device, don't forget to empty the bath liquid;

12. Before any repair, maintenance, or mobile device use, turn off the device and disconnect the socket;

13. Equipment repairs should be carried out by qualified technicians, and please contact our after-sales service center before repairs;

14. If you need technical support or have your instrument shipped back for repair, please contact our after-sales service center. During transport:

Please clean the equipment first to avoid harming maintenance personnel;

Attached is a brief fault error description from the previous page;

During transportation, keep the equipment upright and upright;

When transporting the instrument back, please pack it carefully.

Our company is not responsible for any damage caused by improper packaging!

9. Fault Handling

Fault phenomenaSpeculation on the cause of the faultTroubleshooting methods
No power supply when powered on (indicator light not on)Power outlets have no voltage or poor contact with plug/socketReplug or repair
Power supply cable is broken or the power switch is faultyReplacement and repair
The fuse blowsIf the device still breaks after replacement, you must check the temperature controller, circulation pump, heater, and other components, and repair them before starting the machine
The instrument panel shows no displayThe thermostat's power transformer is faulty or the incoming terminal is detachedReplace the thermostat or reconnect the incoming wires
Instrument PV screen display----Pt100 sensor is faulty or instrument input circuit faultReplace the Pt100 or the thermostat
An error in the full-scale setting of the temperature controller causes the measurement value to go out of rangeReset
Temperature loss of controlThermostat fault or thyristor conductionReplace the thermostat or BTA16 thyristor
It does not heat upThe heater is brokenReplace
The HEAT light on the temperature controller does not turn on, indicating a malfunctionReplace the instruments
The HEAT light is on, and the trigger or thyristor is faultyCheck the optocoupler 3041 or BTA16
T1 (timer) ≠0, incorrect setting, stops heating when the timer endsReset T1 = (T heating + T constant temperature) or set T1 = 0
The water level alarm light is onThere is little bath liquid in the studioAdd bath liquid
Instrument failureReplace the temperature controller
The floating ball leaked waterReplace
Large temperature control errorsThe PT100 sensor has poor contact and large errorsReplace with Pt100
Corrected SC and HL parameters (large errors)Refer to this manual
Adjust PID and other parameters (static difference)Refer to this manual
Poor uniformity in the studioIf the bath liquid does not circulate, the circulation pump flow rate is insufficient or faultyAdjust or replace

10. Electrical wiring schematic

Electrical wiring schematic    

Packing list

Product Name: Constant Temperature Circulation Tank

Serial numberCategoryNameUnitQuantityNote
1DocumentsUser manualportion1
2DocumentsPacking listportion1
3DocumentsGuarantee letterportion1
4DocumentsWarranty cardportion1
5Spare partsFuseOnly2
6Spare partsComes with an external bathtub connection pipeRoot2φ10(Inner diameter)*500mm

The items listed on this order match the actual items inside the box

Packer: 2

Inspector: 1