YIHENG MP MPG-C Refrigeration and Heating Circulation Tank Instruction Manual

YIHENG MP MPG-C refrigeration and heating cycle tank manual, covering safety tips, product structure, temperature setting, timing operation, accuracy check, maintenance and Troubleshooting to ensure correct use of the equipment.
Product images

Tips for ensuring safety

The matters recorded here are extremely important and must be strictly observed.

1. Safety Reminders

! Danger (may result in serious property loss or personal injury)

  1. This product must be reliably grounded and kept away from sources of electromagnetic interference (neutral or centerline must not be used as ground).

  2. Before use, please confirm that the voltage and frequency of the power supply match the product requirements.

  3. Products should use independent power outlets, and ensure that plugs and sockets are properly grounded.

  4. It is not allowed to unplug or plug in the power cord arbitrarily without turning off the power switch during operation.

  5. It is not allowed to arbitrarily extend or cut the product's power cord.

  6. Repairs must not be performed without authorization; repairs commissioned by our company must be performed by professionals.

! Warning (may result in property damage or personal injury)

  1. You must fully read and understand the product manual before proceeding.

  2. When unplugging the power cord, do not drag the power cord directly.

  3. If any of the following situations occur, you must unplug the power cord of this product:
    3.1 When replacing the fuse tube;
    3.2 When a product malfunction is awaiting inspection and repair;
    3.3 When the product is out of use for a long time;
    3.4 When moving the product;

! Note (may affect service life and cause the product to malfunction)

  1. Products should be placed on a hard and solid surface to keep them level.

  2. A certain amount of space should be left around the product.

  3. Products must be used under certain operating conditions. (See Chapter 3, Sections 1.1~1.4)

  4. Products must be used in the correct settings. (See Chapter 3 for details)

2. Product Introduction

1. Appearance drawing

Refrigeration and heating circulation tank    
Figure 1: Refrigeration and heating circulation tank

2. Overview of structural functions

The refrigeration and heating circulation tank consists of three parts: the circulation controller, the enclosure, and the refrigeration system. The circulation controller consists of a thermostat, temperature sensor, heater, water level controller, and circulation pump. The cabinet is made of high-quality thin steel plates, with a surface powder coating and a stainless steel inner liner. The refrigeration system consists of compressors, condensers, evaporators, and other components.

This equipment relies on a balance of cooling and heating power to achieve precise temperature control, and the circulation pump ensures high temperature uniformity inside the bath. In addition to meeting its own constant temperature control needs, the circulation pump is also used to deliver the constant temperature bath liquid outward.

On the lower right side of the circulation controller, there is a connecting port for the inlet and outlet of the circulating bath liquid, used for external bath tank connections. When no external bath tank is installed, the bath liquid inlet and outlet are directly connected by a hose (see Figure 1 9). The thermostat is a continuous PID digital controller with a dual-row, four-bit display, composed of a CPU processor and peripheral circuits. It uses a Pt100 platinum resistor as the temperature sensing element, offering high control accuracy and good stability, as well as timer, over-temperature alarm, and liquid level protection alarm functions.

3. Product usage

1. Preparation before use

1.1 Products should operate normally under the following operating conditions: a. Ambient temperature: 5°C~35°C; Relative humidity: ≤85% b. Atmospheric pressure: (86~106)kPa c. Altitude not exceeding 2000 meters d. Power supply: (220±22)V (50±1)Hz e. Should be kept away from heat sources, with no corrosive gases, strong vibration sources, or strong electromagnetic fields.

1.2 The base for placing equipment must be flat and made of non-combustible materials.

1.3 Maintain at least 300mm of space around the equipment (front and rear ventilation grids) for good ventilation.

1.4 This equipment is equipped with a compressor. After transportation, do not operate the equipment immediately. Instead, leave it in the correct position and wait (1~2) days before starting it to ensure the compressor operates normally.

1.5 When pouring liquid into the bathtub, be careful not to pour the liquid into the inside of the instrument. It is recommended that the maximum filling level be 30mm below the edge of the bathtub or at least two-thirds of the total height to raise the float to its highest position.

1.6 Different bath liquids should be added according to the temperature control range (strictly prohibited from overheating):

  • When the operating temperature is between 5°C~80°C, the bath liquid is purified water

  • When the temperature is above 80°C, the bath liquid is mineral oil (pay attention to the oil's ignition point to avoid fire!). )

  • When the operating temperature is below 5°C, please be aware of the risk of freezing! Recommended bath solution: ethanol or ethylene glycol.

1.7 If an external bathtub is needed, the rubber hose connecting the inlet and outlet of the circulating bath liquid should be removed, and a spare part (or self-provided) rubber hose should be used to directly connect it to the inlet and outlet of the external bathtub.

Note:When configuring external circulation, ensure that the inlet and outlet of the external bath tank are basically level with the inlet and outlet of the equipment.
If the equipment operates at low temperatures for a long time, ice may form on the evaporator and pipes. If temperature control is inaccurate, please turn off the compressor and heat to thaw it. The melting chamber may also seep due to the ice.

2. Power on the machine

2.1 Description of the thermostat panel

Thermometer panel diagram    
Figure 2: Thermostat panel view
1. PV: Displays measured values, unit: °C    2. SV: Displays the set value, unit: °C    3. COLD Cooling Operation Indicator Light    4. AL1 deviation alarm indicator light    5. SET function key    6. Shift key / remaining time observation key    7. Reduce keys    8. Add key    9. HEAT heating indicator light    10. AT self-tuning/timer work indicator light

2.2 When the circulation controller power switch is turned on, the thermometer's PV screen displays the current liquid temperature, the SV screen shows the factory set temperature, and the circulation pump is running. When PV < SV, the HEAT heating light turns on and the instrument enters working mode.

3. Temperature setting (see Figure 2)

Press the SET button; at this point, the first position of the SV screen will be highlighted, while the others will flash. Use â—€ â–¶ the key to repeatedly select the specific high-brightness positions displayed on the SV screen, and use the â–² or â–¼ keys to change the temperature value displayed on the original SV screen until it reaches the desired value. After setting, press the SET button again to display "T1" on the PV screen. If the timer function is not in use, set SV to 0. Press the SET button again to enter working mode. The PV screen shows the measured temperature, and the SV screen shows the set temperature;

If using temperatures below (RT+5)°C, the refrigeration power switch can be turned on. (RT refers to ambient temperature)

4. Scheduled settings

When T1 is set to 0, the instrument cancels the timing function; The timing function only appears when T1 is not set to 0.

  1. The timer range is (1~9999) minutes, and it operates as a countdown timer; During scheduled operation, changing the timing value is invalid; When the timer setting is complete, press SET to start the timer; When the countdown ends, output is automatically cut off, the buzzer sounds, the PV window displays "End" or pressing the shift key causes the SV window to show "End".

  2. When the instrument is running on a scheduled schedule, pressing the shift key will display the remaining timer in the SV window;

Note:When the countdown ends, to use the timer again, press the SET button. If the power is suddenly cut off and the machine restarts during timer use, the timer must be reset and restarted before it becomes effective.

3) Explanation of timer setting methods:

a) Press the "SET" button once; the instrument enters the (A) menu and enters the set temperature value;
b) Press the "SET" key again; the PV window will display "T1". Press â–² and â–¼ keys to enter the set timer value;
c) Press the "SET" key again to exit the (A) menu and enter working mode;
d) If the decimal point in the lower right corner or the AT light flashes, it indicates the instrument has entered countdown mode.

5. Check the accuracy of temperature control

If the machine is powered on for the first time, after a period of use, or when the ambient temperature changes (seasonal changes), it is necessary to verify whether the actual temperature in the studio (bathtub) matches the measured temperature error of the instrument. If it does not meet the requirements, please correct the error using the following methods.

To modify parameters, you must first unlock the electronic lock "LCK" before you can modify the relevant parameters; After modifying the parameters, the electronic lock must be turned off again to return to working mode. Only then does the instrument execute the newly modified parameters.

(1) Unlocking procedure: Press the SET key for more than 5 seconds (enter menu B). When the PV screen displays "AL1," release it (indicating the thermometer has entered the parameter menu). Press the SET key several times to find the "LCK" prompt, press the â–¼ key to make the SV screen show 0 (i.e., unlock). Press the SET key several times to find the prompt for the control parameter you want to modify, and press the â–² or â–¼ key to display the parameter as the desired value. All control parameters can be adjusted in one go.

(2) Lockout procedure: Press the SET key to find the "LCK" prompt, press the â–² key to make the SV screen display as 2 (locked), then press the SET key for more than 5 seconds to return to working mode. At this point, the instrument executes the newly modified parameters.

(Note: You can disable the electronic lock after all the parameters that need to be modified.)

For example:

1) Correction of measurement temperature and actual temperature error inside the workroom:

a) Place the glass mercury thermometer (using a mercury thermometer with ±0.5°C accuracy and a measurement and authentication test) into the bathtub. The mercury end should be immersed in bath liquid and positioned at the geometric center of the bathtub; (The reading from the glass mercury thermometer is the actual temperature.)

b) When the machine is powered on, after the temperature reaches the set temperature and stabilizes for about 1~2 hours (at constant temperature), compare the difference between the actual temperature measured by the mercury thermometer and the PV display temperature. The difference (PV display value - actual temperature) is the value that needs to be modified for the temperature correction parameter, i.e., SC = original SC value + (PV display value - actual temperature)

c) Calculate the SC value based on the above formula and enter it (if one adjustment is inaccurate, it can be repeated multiple times until it meets the standard).

d) If the SV temperature value is frequently changed during use, first set SC to 0 and return to working mode; Press the SET key again for 10 seconds to enter the C menu. The PV screen will show "HL"; change HL to 1.000 and return to working mode; Within the control temperature range, select two test points, such as P1 and P2, (P2 > P1), and the slope is as follows:

HL = (glass P2 - glass P1) / (instrument P2 - instrument P1) (always keep three decimal places)

Set the new HL parameter value; If still inaccurate, adjust the SC according to the previous three steps (repeating multiple times until the standard is met).

2) Suppression of temperature instability (excessive overshoot):

Method 1: First, enter the C menu and reduce OUT (output power), decreasing by about 10% each time; Alternatively, enter menu B, increase the P (proportional band) variation, changing it by about 0.5 each time, and simultaneously change the parameter I, decreasing or increasing it by about 200 seconds each time, then simultaneously change the D parameter so that D = I/4, then reheat and observe until the overshoot meets the requirements.

Method 2: When the required temperature is low, to reduce temperature overshoot, a secondary setting method can be used. For example: when the required temperature is 40°C, it should first be set to 38°C to stabilize the temperature before setting it to 40°C. This can reduce or even eliminate overshoot.

6. Improves temperature control accuracy

1) Explanation of changes in each functional parameter

Since products undergo strict testing before leaving the factory, corrections generally do not need to be made. However, after the first startup, after a period of use, or due to harsh environments, inconsistent external temperature with the factory temperature, or changes before or after using the control temperature, the temperature display value may not match the actual temperature inside the workroom or become unstable (overshoot). If the technical specifications exceed the specified range, parameters can be modified to meet factory standards.

2) Explanation of control parameter changes

After unlocking the unlocking program; By modifying according to "7. Extraction Program for Functional Parameters" and "Appendix 3, Function Parameter Table," all control parameters can be adjusted in one go. You can disable the electronic lock after all the parameters that need to be modified.

Note:After 30 seconds of keyless pressing, it will automatically return to working mode, but any changed data will be considered invalid.

7. The order in which each function parameters are retrieved

Parameters bring up the sequence diagram

4. Product maintenance and precautions

  1. The device must be connected to a well-grounded power outlet! to ensure personal safety;

  2. When using temperatures above 80°C, carefully observe the distance between the liquid level and the edge of the bathtub to prevent thermal expansion of the bath oil during heating and avoid bath liquid overflow;

  3. When the bath liquid is water, pay attention to the temperature to prevent freezing or boiling, which could cause inaccurate temperature control;

  4. When replenishing bath liquid, the model of bath liquid injected into the bathtub must be the same; (Note that the liquid medium in the workspace must be filled with different bath solutions according to different operating temperatures.)

  5. Do not use flammable bath liquids! If you cause damage caused by improper use of the bath solution, you are responsible for the consequences!

  6. When connecting rubber tubes, use clamps to tighten the connection to prevent it from falling off;

  7. With continuous operation, parts of the bathtub cover and the connection to the circulation pump can become very hot or cold. Therefore, be very careful when touching these areas to prevent burns or frostbite.

  8. Be careful when draining hot bath liquid! Before draining, check the temperature of the bath (for example: open the device cover and let it cool for a while);

  9. If the equipment operates at low temperatures for a long time, icing may form on the evaporator and pipes, causing inaccurate temperature control in the bathtub. In this case, please turn off the compressor and perform heating, frosting, and defrosting. Water leakage may occur due to the freezing and melting of the chamber.

  10. After use, turn off the refrigeration power switch and the circulation controller power switch, unplug the power connection cable, and unscrew the liquid outlet to drain liquid (please pay attention to the bath temperature and be cautious!). to avoid burns or frostbite) and dry the cabinet.

When shutting down, please turn off the refrigeration power switch first, then turn off the power to the circulation controller;
  1. Before moving the device, don't forget to empty the bath liquid; the device must not be tilted (not exceeding 30°) or placed upside down;

5. Appendix

1. Main Technical Specifications

Table 1

Specifications and modelsMP-10CMPG-10CMP-20CMPG-20CMP-30CMPG-30CMP-40CMPG-40CMP-50CMPG-50C
Input Power (W)Heating2000
Refrigeration(MP) 250(MPG) 300(MP) 300(MPG) 450(MP) 550(MPG) 900
Temperature control range-10°C~100°C-20°C~100°C-30°C~100°C-40°C~100°C-50°C~100°C
Temperature resolution0.1°C
Temperature fluctuation±0.2°C
Pump flow rate (L/min)410
Pump Head (cm)50
Filling Volume (L)(MP) 4.5 / (MPG) 13
Bathtub opening/depth (mm)(MP) 150×160/150 / (MPG) 240×170/200
Dimensions (mm³)(MP) 300×540×660 / (MPG) 380×640×860

Note: Performance parameter test conditions are: ambient temperature 20°C, relative humidity ≤85%, liquid temperature 37°C, liquid medium purified water.

2. Pump flow regulation

Pump regulation 1    Pump regulation 2

The pump flow rate is preset before leaving the factory, but it can also be adjusted according to requirements.

  • Use a screwdriver to hold down the screw. (1) Rotate 360° counterclockwise.

  • Use flat-nozzle pliers to turn the control valve indicated in the diagram to the desired position.

  • Tighten the screws

Example:

Internal circulation applications in bathtubs
A: 100% internal circulation of bathtubs (for larger bathtubs)
B. Smaller internal circulation tanks (used for smooth, highly fluid liquids)

External/internal circulation applications
C. 40% external circulation flow, 60% internal circulation; (For larger bathtubs)
D: 80% external circulation flow, 20% internal circulation. (Used for some smaller bathtubs)

3. Tables of various functional parameters

Press Set key for 5 seconds to enter Menu B:

Table 2

PromptNameSet the rangeExplanationFactory preset
AL1Upper deviation and water level alarm settings(-99.9~999.9)°CWhen the temperature exceeds the (SV+AL1) value or the liquid in the bathtub falls below the required threshold, the AL1 light turns on, the buzzer sounds, and the heating output is cut off
COLDCooling start(-99.9~999.9)°CWhen the temperature is below the (SV+COLD) value, the COLD light turns on
SCCorrection of display value errors(-20~20)°CMeasure the actual temperature inside the box and compare it with the PV display temperature to correct display errors
ATUSelf-organizing command0,10:OFF 1:ON allows self-tuning of PID parameters0
PScale band(0~100)°CHeating ratio control: the higher the P ratio, the lower the system gain; Reducing P can improve system control accuracy and eliminate static differences, but if too low, it will cause system oscillation and instability
IIntegration Time (Readjustment Time)(1~4320)sThe time constant for the integration action—the larger I, the weaker the integral, and the system is stable
dDifferential Time (Pre-Tuned Overshoot)(0~1200)sThe time constant of the differential action: the larger the d, the stronger the differential action, and it can overcome overshoot. Generally, d is taken as (I/4)
THeating cycle(1~60)sThe thyristor output is (2~3) seconds. For devices with large remaining power, increasing T can reduce the static difference of PID control
LCKCombination lock0,1,20: Unlocking 1: Full lock 2: Locking all except SV2

Press Set key 10 seconds to enter the C menu:

Table 3

PromptNameSet the rangeExplanationFactory preset
HLSlope0.500~1.500This ensures consistent temperature control accuracy throughout the entire measurement range
OUTOutput power0~100%Adjustable heater output power
※ Products undergo strict testing before leaving the factory. When the studio's technical specifications meet requirements and work normally, generally no corrections are required.

4. Causes and handling of faults

Table 4

Fault phenomenaSpeculation on the cause of the faultTroubleshooting methods
1. No power on (indicator light not on)Power outlets have no voltage or poor contact with plug/socketReplug or repair

Power supply cable is broken or the power switch is faultyReplacement and repair

The fuse blowsIf the device still breaks after replacement, you must check components such as the thermometer, circulation pump, heater, and refrigeration unit, and repair them before starting the machine
2. No instrument displayThe thermostat's power transformer is faulty or the incoming terminal is detachedReplace the thermostat or reconnect the incoming wires
3. Instrument PV screen display----Pt100 sensor is faulty or instrument input circuit faultReplace the Pt100 or the thermostat

The temperature controller has a full-scale setting error, causing the measurement value to go out of rangeReset
4. Temperature runawayThermostat malfunction or thyristor conduction, temperature loss beyond rangeReplace the thermostat or BTA16 thyristor
5. Does not heat upThe heater is brokenReplace

The thermostat's HEAT light is not on, indicating a malfunctionReplace the instruments

The HEAT light is on, and the trigger or thyristor is faultyCheck the optocoupler 3041 or BTA16

T1 (timer) ≠0, incorrect setting, stops heating when the timer endsReset T1=T heating + T constant temperature or set T1=0
6. No coolingThe refrigeration power switch is not turned onOpen

Compressor is faulty, protector is faulty, and the machine won't start when startedReplace the compressor

Overheating causes the compressor protector to startAfter several hours of downtime, it naturally recovers

Refrigerant leaks, insufficient cooling capacity, and blocked pipelinesRemove blockage and add refrigerant

Evaporator frost is severeHeat and dissolve
7. The water level alarm light is onThere is little bath liquid in the studioAdd bath liquid

Instrument failureReplace the thermostat

The floating ball leaked waterReplace
8. Large temperature control errorsThe PT100 sensor has poor contact and large errorsReplace with Pt100

Corrected SC and HL parameters (large errors)Refer to this manual

Adjust PID and other parameters (static difference)Refer to this manual

Imbalance between cooling and heating powerAdjust OUT
9. Low temperatures may run out of control or cause large errorsHeating or cooling systems are not workingReplace the thermostat or repair the refrigeration
10. Poor uniformity of the studioIf the circulation pump flow is insufficient or faulty, the bath liquid does not circulateAdjust or replace

5. Electrical wiring schematic

Wiring diagram 1    Wiring diagram 2

Packing list

Product Name: Refrigeration and Heating Circulation Tank

Serial numberCategoryNameUnitQuantityNote
1DocumentsUser manualportion1
2DocumentsPacking listportion1
3DocumentsGuarantee letterportion1
4DocumentsWarranty cardportion1
5Spare partsFuseOnly2 eachφ5×20 / φ6×30
6AccessoriesComes with an external bathtub connection pipeRoot2φ10 (Inner diameter) ×500mm

The items listed on this order match the actual items inside the box

Packer: ____________ Inspector: ____________