Reader's Guide
This manual is compiled based on manufacturer's data. Due to equipment batch changes, the content may not fully match every physical unit. Please refer to the actual operation and on-device labels.
Copyright belongs to the manufacturer. For internal reference of purchasers only. If safety parameters in the manual differ from the physical unit, please suspend operation and consult our engineer for confirmation.
Disclaimer: This manual is for reference only. In case of any discrepancy, the physical unit shall prevail. Our company assumes no liability for any consequences arising from operations based on this manual.
YIHENG DKZ Thermostatic water bath instruction manual
Tips for ensuring safety
The matters recorded here are extremely important and must be strictly observed.
1. Safety Tips
Danger (May result in serious property loss or personal injury)
This product must be reliably grounded (neutral or centerline must not be used as ground).
Before use, please confirm that the voltage and frequency of the power supply match the product requirements.
Products should use independent power outlets, and ensure that plugs and sockets are properly grounded.
It is not allowed to unplug or plug in the power cord arbitrarily without turning off the power switch during operation.
It is not allowed to arbitrarily extend or cut the product's power cord.
When sound and light alarms occur, please first check whether the set temperature deviates from the normal range. If not, stop using the device and contact professional maintenance personnel for inspection.
Warning (Unauthorized repairs may result in property damage or personal injury; responsibility is borne by the driver.)
You must fully read and understand the product manual before proceeding.
When unplugging the power cord, do not drag the power cord directly.
The 304 stainless steel inner pot is not acid-resistant; please take corrosion prevention measures. Never use acidic media inside the box!
If any of the following situations occur, you must unplug the power cord of this product:
When replacing the fuse tube;
When a product malfunction is awaiting inspection and repair;
When the product is discontinued for a long time;
When moving products;
Do not press the power switch before adding water to avoid burning out the heating tube.
Note (Otherwise, it may affect the product's lifespan and prevent it from functioning properly.)
Products should be placed on a hard and solid surface to keep them level.
A certain amount of space should be left around the product.
Products must be used under certain operating conditions.
Do not remove the side panel of the temperature control device without the approval of the maintenance department, for safety.
2. Scope of Application
It is used by factories, mining enterprises, universities, and research institutions for bacterial and cell tissue cultivation and laboratory testing.
3. Technical Data
Table 1
| Model | DKZ-1 / DKZ-2 | DKZ-2B | DKZ-3 / DKZ-3B |
| Power supply voltage | AC220V±10% / 50Hz | ||
| Power consumption | 1250 W | 1650 W | |
| Temperature control range | Room temperature +5 ~ 99°C | ||
| Temperature control accuracy | ±0.1°C | ||
| Tracking alarms | 2.0°C | ||
| Oscillation frequency | 30 ~ 150 rpm | ||
| Amplitude | 30mm (40mm optional) | ||
| Studio Dimensions (W*D*H) | 438×310×250 mm | 618×310×250 mm | |
| Dimensions (W*D*H) | 643×350×353 mm | 823×350×355 mm | |
Note: The inner tank and outer shell of DKZ-2B and DKZ-3B are all made of stainless steel.
4. Features
Stainless steel inner pot is used.
High-precision microcomputer temperature controllers provide precise and reliable temperature control.
Operating temperature can be set arbitrarily, with an over-temperature alarm.
Stepless speed regulation is used, allowing direct drive of the shaker with low noise.
5. Usage Instructions
Add purified water to the oscillating water tank so that the liquid level exceeds half of the total height.
Press the power switch on the right side of the oscillating sink; a digital display will appear on the controller, indicating power is on.
Adjust the controller setting value and select the desired operating temperature. (See section 6 below)
Turn the speed control knob to the left to the minimum, then press the oscillation switch on the right side of the oscillating water tank, and adjust the oscillation speed knob so that the shaker's speed gradually reaches the desired vibration frequency.
To drain water, pull the drain plug from the left side of the casing out to the outside (about 300mm), then pull out the drain plug; water will flow out from the rubber tube.
6. Controller usage method
(1) (PV) Display — Displays measurement values and displays various prompts based on instrument status.
(2) (SV) Display — Displays set values and displays various parameter values based on the instrument status.
(3) AT (Operation Indicator Light) The controller lights up when operating, flashes during self-tuning, and turns off when stopped.
(4) HEAT (Heating Indicator Light) Lights up when there is heating output.
(5) ALM (Alarm Indicator Light) When there is an alarm output, it lights up and the buzzer sounds.
(6) COLD (Cooling Indicator Light) Lights up when there is a cooling output; (Note: This product does not have COLD cooling function.)
⑦ Used to adjust various parameter values or enter self-tuning mode.
⑧ Used to adjust internal parameter values or enter self-tuning mode.
(9) Shift key Used for shifting set values, internal parameters, and observing scheduled running times.
(10) Function key — Modification of set values, retrieval of parameter symbols, and confirmation of parameter modifications.
1. Function retrieval process
2. Detailed explanations of each function
2.1 If "□□□" is displayed, it means the sensor is open-circuited or the input signal exceeds the measurement range.
3. Temperature and timer settings
3.1 In working mode, press the SET button once; the PV screen displays the character " ". Press , or press the key to show the SV screen at the required temperature value.
3.2 Press the SET button again; the PV screen will display the character " ". Press , or press the key to make the SV screen display the required time value. (See function extraction process)
3.2.1 When ST is set to 0, the controller cancels the timing function and continues to run; When the ST setting is not zero, the controller has a timing function. When the controller's running time is up, the SV screen displays "END," the buzzer sounds, the controller stops working, any key is pressed to mute, and pressing both the key and the key for 4 seconds restarts.
3.2.2 When the controller is in working mode, simply press the shift button to display "TIME" on the PV screen, and the SV screen shows the operating time of the controller. Press the shift button again to return the controller to operating mode.
3.3 Timer Mode Press the SET button again to return to working mode and enter working mode.
4. Setting up the upper deviation alarm
The setting of the upper deviation is reasonable, providing protection against system temperature control exceeding or losing control, and must be used during product operation.
4.1 When products leave the factory, AL=10~15 is generally set, i.e., the alarm temperature is: (SV+AL)°C.
4.2 Press and hold the SET key for about 4 seconds. When the PV screen displays the character " " (" (indicating the controller has entered the parameter menu), release (indicating the controller has entered the parameter menu), but you must unlock the electronic lock "Lk" before modifying the relevant parameters using the key .
4.3 Unlocking Procedure: After entering the parameter menu, press the SET button several times. When the PV screen displays the character " ", use the key to change the SV screen value from "0" to "Unlock Password", which will unlock the electronic lock (after that, pressing without a key will automatically return the controller to working mode after 1 minute).
4.4 Press several SET keys; when the PV screen displays the character " ", use the OR keys to set a reasonable deviation value (AL).
4.5 Press the SET key several times to bring up the character "", and use the key to change the "unlocking password" to "0" to disable the electronic lock (Note: You can disable the electronic lock after all parameters are modified).
4.6 After overheating, the buzzer intermittently alarms, and the SV window synchronously alternately displays "Set value/ ―――― "And the AL1 ultra-temperature lamp stays on continuously, with any key to silence it.
4.7 When setting from high temperature to low temperature, an alarm will also sound if the temperature exceeds the AL value. This is normal; press mute to proceed.
Since products undergo strict testing before leaving the factory, generally no modifications are needed. If the product is used in harsh environments, the external temperature does not match the instrument's ambient temperature, or if the temperature control range is used before and after the limit (since the test point differs from the factory test point, the factory test point is 37°C), it may cause a discrepancy between the temperature display value and the actual temperature inside the chamber. If the range exceeds the technical indicators, corrections can be made.
5. Function parameter table
Table 2
| Prompt | Name | Set the range | Explanation | Factory preset |
|---|---|---|---|---|
| AL/AL | Upper deviation alarm settings | 0~Full scale | When the temperature exceeds the (SV+AL) value, AL1 lights up, the buzzer sounds, the heating output is cut off, and the SV window alternately displays "Set value/----" | |
| CL/CL | Refrigeration control settings | 0~Full scale | When the temperature exceeds the (SV+CL) value, the COLD light turns on, the refrigeration contact is activated, and the compressor is started | This device does not have this feature |
| CT/CT | Cooling control delay | (0~3600) seconds | When the measured value reaches the alarm threshold, the alarm relay outputs after CT time | |
| P/P | Scale band | 1~ Full scale | It acts only on the heating side; the larger the P, the lower the system gain; Reducing P can improve system control accuracy and eliminate static errors | |
| I/I | Integration Time (Readjustment Time) | (0~3600) seconds | The time constant for the effect of integrals—the larger I, the weaker the integral, and the system is stable. | |
| d/d | Differential time (pre-tuned time) | (0~3600) seconds | The time constant for differential action: the larger d, the stronger the differential action, and it can overcome overshooting. I=0, d=0 is a half-proportional control, generally d is twice I, or equal to I | |
| AR/Ar | Overshoot suppression (ratio resetting) | (0~100)% | Changing the output power and increasing Ar can improve the heating rate, help eliminate static differences, reduce Ar, and prevent overshoot. | |
| T/T | Heating cycle | (1~300) seconds | The thyristor output is generally (2~3) seconds. For devices with large remaining power, increasing T can reduce the static error of PID control. | 3 |
| Pd/Pb | Zero Adjustment (Intercept) | -100~100 | When the instrument's zero error is large and the fullness error is small, adjust this value; generally, Pt100 rarely adjusts this value | |
| PK/PK | Full Scale Adjustment (Slope) | -(1000~1000) seconds | When the zero error of the instrument is small and the fullness error is large, adjust this value: PK = 4000x (specified value - actual displayed value) / actual display value. Generally, Pt100 adjusts this value first | |
| Fn | Fan control settings | 0~Full scale | When SP-Fn< = measured temperature < = SP + Fn, start the slow-speed fan to run; When measuring temperature <SP-Fn or >.>SP+Fn, start the fast fan to operate. | This device does not have this feature |
| L/LK | Combination lock | 0~255 | Enter the unlock password to change the above parameters | 0 |
7. Precautions for Use
Products undergo strict testing before leaving the factory and generally do not undergo corrections.
When the operating temperature is set at 37°C, if the ambient temperature is above 32°C, it is best to place the oscillating sink indoors with air conditioning to keep the operating ambient temperature below 32°C.
Open the top lid, place the items to be cultivated into the tank, and close the top lid. If the top lid has been open for too long, the temperature inside the tank will fluctuate—this is normal.
After the setting is complete, the sink enters a heating state and the heating indicator light turns on. When the temperature inside the tank approaches the set temperature, the heating indicator light turns on and off, repeating multiple times to control the temperature to a constant temperature state. After maintaining a constant temperature for 90 minutes, the temperature inside the tank basically stabilizes.
Test Note: The temperature gauge used for testing inside the tank should be a mercury gauge with 0.1 accuracy, with the mercury end placed at the geometric center position inside the workspace, and the top cover should be properly covered.
8. Maintenance and servicing
The inside and outside of the vibrating sink should be kept clean at all times. When spraying paint the casing, do not use reactive chemical solutions to wipe it to prevent chemical reactions.
If the instrument will not be used for a long time, both the inside and outside should be dried and covered with a plastic film dust cover to prevent moisture from affecting the instrument's use during temperature control.
The instrument should not be used under high voltage, high current, strong magnetic fields, or corrosive gas environments to avoid interference, damage, and electric shock hazards.
9. Fault Handling
Table 3
| Fault phenomena | Speculation on the cause of the fault | Troubleshooting methods |
|---|---|---|
| No power supply | 1. No power outlet | 1. Replace the socket |
| 2. Plugs are not properly inserted or the wires are broken | 2. Plug in or connect the wires | |
| 3. Fuse open circuit | 3. Replace the fuse | |
| 4. The power switch is broken | 4. Replace | |
| The temperature inside the chamber does not rise | 1. Low set temperature | 1. Adjust the set temperature |
| 2. The electric heater is faulty | 2. Battery swap heater | |
| 3. Controller failure | 3. Replace the controller | |
| 4. Incorrect timing settings | 4. Reset | |
| The set temperature and the temperature inside the chamber have a large discrepancy between the set temperature | 1. Component errors | 1. Edit Pb, Pk |
| Temperature loss of control | 1. The temperature sensor is fixed and detached | 1. Fixed temperature sensor |
| 2. Controller faulty, thyristor fault | 2. Replace the controller and the thyristor | |
| Display ☐☐☐ | 1. Temperature sensor is faulty | 1. Replace |
| 2. Controller failure | 2. Replace | |
| No vibration | 1. Faulty speed control plate or motor | 1. Replace |
10. Wiring schematic
Packing list
Product Name: Electric Constant Temperature Swinging Water Tank
| Serial number | Category | Name | Unit | Quantity | Note |
|---|---|---|---|---|---|
| 1 | Documents | User manual | portion | 1 | |
| 2 | Documents | Packing list | portion | 1 | |
| 3 | Documents | Guarantee letter | portion | 1 | |
| 4 | Documents | Warranty card | portion | 1 | |
| 5 | Spare parts | Fuse | Only | 2 | |
| 6 | Spare parts | and released water to block the block | Only | 1 | |
| 7 | Spare parts | Oil pot | Bottle | 1 | |
| 8 | Spare parts | Power cord | Root | 1 |
The items listed on this order match the actual items inside the box
Packer: 2
Inspector: 1
- 1YIHENG HWS Water bath instruction manual
- 2YIHENG THZ series thermostatic foster Shaker instruction manual
- 3YIHENG DKZ-C Thermostatic water bath instruction manual
- 4YIHENG THZ-300/C thermostatic foster Shaker instruction manual
- 5YIHENG THZ series thermostatic Shaker instruction manual
- 6YIHENG BPZ-Series Vacuum oven instruction manual
- 7YIHENG DZF series Vacuum oven instruction manual
- 8YIHENG DZF-6090/6210 Vacuum oven instruction manual
- 9YIHENG BPZ-LC/LCB series Vacuum oven instruction manual
- 10YIHENG BGZ series Vacuum oven instruction manual