YIHENG THZ series thermostatic foster Shaker instruction manual
1. Safety Reminders
! Danger (may result in serious property loss or personal injury)
This product must be reliably grounded and kept away from sources of electromagnetic interference (neutral or centerline must not be used as ground).
Before use, please confirm that the voltage and frequency of the power supply match the product requirements.
Products should use independent power outlets, and ensure that plugs and sockets are properly grounded.
It is not allowed to unplug or plug in the power cord arbitrarily without turning off the power switch during operation.
It is not allowed to arbitrarily extend or cut the product's power cord.
Repairs must not be performed without authorization; repairs commissioned by our company must be performed by professionals.
! Warning (may result in property damage or personal injury)
You must fully read and understand the product manual before proceeding.
The 304 stainless steel inner pot is not acid-resistant; please take corrosion prevention measures. Never use acidic media inside the box!
When unplugging the power cord, do not drag the power cord directly.
If any of the following situations occur, you must unplug the power cord of this product:
When replacing the fuse tube;
When a product malfunction is awaiting inspection and repair;
When the product is discontinued for a long time;
When moving products;
When placing the product on the workbench, the four feet must be fixed to prevent movement caused by vibrations during long-term use, which could cause the product to fall and cause serious damage or personal injury.
! Note (may affect service life and cause the product to malfunction)
Products with model suffix "C" equipped with a refrigeration system must not exceed 45 degrees during handling to avoid damage to the refrigeration system.
After handling and placing products with the model stub "C", they should be left to stand for (1~2) days before starting the machine, to ensure the refrigeration system can operate normally and extend its lifespan.
Products should be placed on a hard and solid surface to keep them level.
A certain amount of space should be left around the product.
Products must be used under certain operating conditions.
Do not open or close the product box door by gravity, as this may cause the door to fall off, damage the product, and cause injury.
When the product is out of use for a long time, regular moisture removal should be performed to avoid damaging related components.
(5) Power outlet
2. Product Introduction
1. Appearance drawing
2. Overview of structural functions
The THZ series constant temperature incubator shaker adopts a benchtop structure, with a cabinet shell made of engineering plastic (Type 103B is metal), lined with stainless steel thin plates inside, and a flip-top door with a built-in oscillation tray. The vibration speed and temperature of the shaker are controlled by an LCD display controller
Cabinet —— Except for the 103B model, which has a metal shell, the rest of the casings are molded from engineering plastic. The bottom is equipped with a motor-driven mechanical oscillator with a circular oscillation trajectory. The rear air duct panel of the studio is made of stainless steel thin plate, and the LCD display control buttons are all located on the front of the cabinet.
Box door —— Adopts a large flip-top structure made of molded transparent plastic, making it extremely convenient to observe, remove, and place cultures.
Control — Using a microcomputer electronic controller and PID adjustment method for closed-loop control of temperature and speed, with an LCD display capable of displaying multiple sets of data and related text, simple operation.
Cycle — The fan behind the air duct plate inside the workspace rotates, circulating hot and cold airflow to improve temperature uniformity inside the workspace.
Protection — Prevents multiple compressor starts in a short period (for products with the model ending "C"); the controller has a delayed start function before starting the compressor.
When the measured temperature exceeds the set temperature by 3°C, the buzzer will emit an over-temperature alarm and simultaneously cut off the heater.
— When the box door opens, the buzzer emits an alarm, and the heater, fan, and turntable stop working, allowing the operator to safely pick up and place the culture.
3. Product usage
1. Preparation before use
1.2 Power supply: (220±22)V (50±1)Hz;
1.3 Atmospheric pressure: (86~106) KPa;
1.4 The elevation must not exceed 2000 meters;
1.5 No strong vibration sources or strong electromagnetic fields nearby;
1.6 It should be placed on a stable, level, sturdy surface away from direct sunlight, and the four legs of the product should be secured. The room should be free of severe dust and corrosive gases;
1.7 There should be a gap of no less than 50cm around the product (front, back, left, right, top).
2. Power on and operate
2.1 The product operation panel diagram is as follows:
1) TIME: Time display window; When the controller is running, TIME displays the running time;
2) REV: Displays the measurement speed;
3) TEMP: Displays measurement temperature;
4) Heating lamp: lights up when heating output is available;
5) REV SET: Displays set speed or parameter code;
6) TEMP SET: Displays the set temperature or parameter values;
7) Fan indicator light: lights up when the fan is output;
8) Oscillation indicator light: lights up when oscillation output is present;
(1) MODE function key: Used to modify set values; Retrieve parameters and confirm parameter modifications;
Shift button: When in the set state, press this button to shift;
Decrease button: When set to the set state, press this key to decrease the number by one, then hold this key to continuously decrease the number;
Add button: When in the set state, press this key to increment the number by one, then hold this key to continuously increase the number;(5) START/STOP Start/Stop Key: Press and hold this button for more than 4 seconds to control program start/stop;
3 Operating Steps
2) Speed setting: Press the MODE button again; the "REV SET" window will flash, indicating the speed can be set as needed. Use the increase, decrease, and shift keys to set the required speed.
3) Temperature setting: Press the MODE button again; the "TEMP SET" window will flash, indicating the temperature can be set as needed. Use the increase, decrease, and shift keys to set the desired temperature. Press the MODE button again to return to standard display mode.
Note: (1) For each parameter modification, you must press the "MODE" key to confirm for the modification to be valid.
(2) After all parameters are set, press the "START/STOP" button and wait about 4 seconds before starting operation.
4. Alarm function
(2) When the oscillation load is too heavy and the motor is overloaded for more than 10 seconds, the instrument will emit a beep, the oscillation will stop, and pressing any key will silence the sound.
(3) When the temperature inside the chamber reaches the set temperature, it starts running at a scheduled timer. When the timer ends, it automatically stops and emits a beeping sound. Pressing any key can mute the sound.
(4) When the PT100 experiences faults such as broken wires or short circuits, causing the measurement temperature to exceed 60°C or fall below -1.0°C, the LCD screen displays "------";
5. Upper deviation alarm settings
The setting of the upper deviation is reasonable, providing protection against system temperature control exceeding or losing control, and must be used during product operation.
Example: If the product is set to AL=3.0 at the factory, the alarm temperature is: (set temperature value + AL value)°C
6. Check the accuracy of temperature control
The mercury thermometer head should be positioned at the geometric center of the effective workspace in the studio
6.2 Within the product's temperature control range, select any point. When the measured temperature equals the set value, maintain the temperature for about 1 hour. Observe that the difference between the actual measured temperature on the mercury thermometer and the measured value displayed on the thermometer should be ≤±0.5°C.
7. Methods to Improve Temperature Control Accuracy
7.2 Enter parameter settings, find the symbol ${ \ L } _ { 6 6 } \prod \bigl [ \int \ d z , \bigr ]$ symbol,
After calculating according to the formula PK = 4000 × ((instrument measurement value - mercury meter value) / mercury meter value), modify based on the original factory PK value (Note: If one correction is inaccurate, you may repeatedly correct until it matches);
4. Product Maintenance and Key Points for Care
1. Key Maintenance Points:
(1) Ambient temperature: (-40~+55°C);
(2) Relative humidity should not exceed 95%;
(3) Atmospheric pressure: (80 ~ 106) Kpa;
(4) Altitude not exceeding 2000 meters;
(5) No corrosive gases nearby.
2) Product moisture-wicking treatment
(1) After each use, wipe the moisture inside the box with a dry cloth to prevent rust.
(2) If the product is not used for a long time (over a month), regularly power it on, heat it to 40°C, run at 50r/min for 5 hours to dispel moisture and prevent rust. After use, unplug the power cord.
3) Defrosting (products with tail labels marked "C")
(1) When the product operates below ambient temperature for a long time or when cooling performance deteriorates (causing static difference), defrosting should be applied.
(2) Set the temperature to 40°C, let the product operate for (3~5) hours, then restore the operating temperature and resume the operating speed.
2. Key Points:
2) Except for parameters such as temperature, rotational speed, time, and zero-point correction, all other parameters must be modified with the approval of our company's service center or adjusted by professionals.
3) For low-temperature products marked with a "C" in the tailline, the tilt angle during handling should not exceed 45 degrees to avoid damage to the refrigeration system.
4) After handling and placing the tail-end products marked "C", let them sit for (1~2) days before starting the machine, to ensure the compressor in the refrigeration system operates normally and extends its lifespan.
5) When the equipment is placed on the platform for operation, protruding barriers must be installed along the platform edge to prevent the shaking box from displacing or falling after prolonged operation.
6) After the equipment is installed, if the ground is uneven, it must be leveled and must not be tilted to the side; Leave at least 50 cm of space on the left and right sides and back of the equipment. Especially for products with a "C" in the tail (with refrigeration system), if the device is used at ambient temperatures above 30°C, it is recommended to strengthen ventilation measures behind the equipment and lower the ambient temperature to prevent compressor crashes due to overheating.
7) This equipment should be kept away from sources of electromagnetic interference and have good grounding wires;
5. Appendix
1. Technical Specifications
Table 1
| Item\Model | THZ-100B | THZ-100 | THZ-103B | THZ-100C |
| Input power | 450W | 400W | 450W | / |
| Temperature control range | (RT+5)°C~50°C | 4°C~50°C | ||
| Resolution | 0.1°C | |||
| Temperature fluctuation | ±0.5°C | |||
| Temperature uniformity | ±1.0°C | |||
| Timed range | 0~99h:59min | |||
| Pallet rotation speed | (40~300)r/min | |||
| Tray amplitude | 20mm | |||
| Pallet load | ≤10kg | |||
| Pallet size | 350×350 mm² | |||
| External dimensions | 500×610×480 mm³ | / | ||
RT - ambient temperature
2. Functional parameter tables
1) Press and hold the MODE key. When the LCD shows LK, enter the password and press the MODE button to enter the settings; A menu
| Prompt | Name | Set the range | Explanation | Factory setup |
| Fn | Fan control settings | 1.0~100.0 | No effect | |
| KA | Power-on mode | 0~3 | 1 KA=0: Power-on stops, must hold the start/stop button to start; 2 KA=1: Runs automatically; 3 KA=2: Operation starts from the last power outage | |
| SE | Ambient temperature correction | -19.9~80.0 | If the ambient temperature does not match the actual temperature, correct it to match the value. | |
| Pb | Zero position adjustment | -100.0~100.0 | When the zero error is large and the fullness error is small, adjust this value; generally, Pt100 is rarely adjusted. | |
| Pk | Full level adjustment | -1000~1000 seconds | When the zero error is small and the fullness error is large, this value can be adjusted. |
2) Press and hold the MODE key. When the LCD shows LK, enter the password and press the MODE key to enter the settings; Menu B
| Prompt | Name | Set the range | Explanation | Factory setup |
| AL | Alarm settings | 0~100.0°C | When the measured temperature exceeds the value of "set temperature + AL," the over-temperature alarm indicator lights up and cuts off the heating output | |
| CL | Refrigeration control settings | 0.0~100.0 | When the temperature exceeds the "set temperature + CL" and meets the compressor cooling delay time, the refrigeration indicator light turns on, the refrigeration contacts are connected, and the compressor is started | |
| Ct | Cooling control delay | 0~3600 (seconds) | Pressing the compressor twice in a row requires a delay; Ct=0 cancels the compressor function | |
| P | Scale band | 1.0~300.0 | Proportional action adjustment: the larger the P, the smaller the proportional effect, the lower the system gain, acting only on the heating side. | |
| I | Scoring time | 10~3600S | The time constant for the effect of integrals, the larger I, the weaker the integral. | |
| d | Differential time | 0~3600S | The time constant of differential action: the larger d, the stronger the differential action, and it can overcome overshoot (D=0 PI control) | |
| Ar | Overshoot suppression | 0~100% | Used to suppress overshoot (Ar determination: 1.5~2 times steady-state output duty cycle) | |
| t | Control cycle | 1~300S | The thyristor output is generally 2~3 seconds. For devices with large remaining power, increasing T can reduce the static difference of PID control. |
3) Press and hold the MODE button. When the LCD screen shows LK, enter the password, and press the MODE button to enter the speed control parameter setting menu as follows: C menu
| Prompt | Name | Set the range | Explanation | Factory setup |
| Pr | Speed regulation ratio | 2~999 | The smaller the Pr, the stronger the proportional effect; if too small, oscillations are more likely to occur | |
| lr | Speed adjustment points | 2~999 | To reach the set speed quickly, increase Ir; otherwise, decrease Ir | |
| rA | Motor starting voltage percentage | 0~50 | The voltage required to start the motor is a percentage of the rated voltage | |
| rP | Each pulse pulse is rapid | 1-2 | 1: 80W motor; 2: 40W motor | |
| Cc | Gear reduction ratio | 1.0-5.0 | The ratio of gear reduction | |
| rn | Maximum speed | 40-300 | Maximum speed setting |
Changes in each functional parameter can alter control effectiveness. If you don't press the "MODE" button for one minute, it will automatically return to standard mode, as some functional parameters may not have been changed.
3. Cause and handling of faults
Table 2
| Serial number | Fault phenomena | Speculation on the cause of the fault | Troubleshooting methods |
| 1 | No power supply when powered on | Power outlets have no power or poor contact | Inspection and repair |
| Power supply line disconnected | Swap | ||
| The power switch is either not on or is broken | Turn (close) the switch on or off, swap the switch | ||
| Fuse is either not installed or blown out | Install the appropriate fuse tube, check the cause of the blowout, repair it, and then turn it on | ||
| 2 | The device doesn't work when powered on | The controller power is not turned on or the run button is not pressed after it is turned on | Follow the instructions |
| Components such as controllers or motors are faulty | Swap | ||
| 3 | The screen displays "____" | Sensor Pt100 is faulty | Repair: Pt100=0°C is 100Ω, and for every 1°C increase, it increases by about 0.3Ω |
| 4 | No temperature rise, or no temperature increase after a certain degree rise | Set temperature lower than RT (ambient temperature) | Reset |
| Check the timer settings | If the timer is insufficient, reset the time to sufficient or set the timer to 0 | ||
| The controller has no voltage output | Controller failure, replacement, repair | ||
| There is voltage across both ends of the heating tube | The heating tube connector may fall off or be damaged | ||
| The door switch failed | Adjust the buckle position or replace the switch if it is faulty | ||
| 5 | Displays large actual temperature errors | Accuracy exceeds the limit | Zero and fullness corrections require professional debugging |
| 6 | Significant static error or overshoot occurs | Setting parameters is unreasonable | Set the temperature value to <(RT+5)°C, or professional adjustment and modification of parameters such as P, I, D, and power attenuation |
| 7 | Temperature loss of control | The heater is broken | Swap |
| The controller is broken | Swap | ||
| The fan does not rotate, and the actual temperature inside the workroom is low | Repair and replacement | ||
| 8 | Loud noise (including cooling fans) | Product placement is unstable, and screws such as the fan support plate are loose | Repair |
| 9 | The oscillation is either not working or out of control | The optical breaker is faulty or the connecting wire has poor contact | Light replacement intermittent devices, etc., repair |
| The control board is broken | Switch control boards | ||
| The motor does not run | Replace or repair the motor if the motor is faulty or overloaded, or if the door switch is not in contact or is faulty | ||
| The door switch failed | Adjust the buckle position or replace the switch if it is faulty | ||
| 10 | It wouldn't spin | The mechanical transmission part is seized | Manual push test |
| The starting torque is too low | Modifications to related parameters require professional debugging |
4. Electrical wiring schematic
No changes will be notified
Packing list
Product Name: Constant Temperature Culture Shaking Machine
| Serial number | Category | Name | Unit | Quantity | Note |
| 1 | Documents | User manual | portion | 1 | |
| 2 | Documents | Packing list | portion | 1 | |
| 3 | Documents | Guarantee letter | portion | 1 | |
| 4 | Documents | Warranty card | portion | 1 | |
| 5 | Spare parts | Fuse core | Only | 2 | |
| 6 | Spare parts | Power cord | Root | 1 |
The items listed on this order match the actual items inside the box
Packer: 2
Inspector: 1
- 1YIHENG DKZ-C Thermostatic water bath instruction manual
- 2YIHENG THZ-300/C thermostatic foster Shaker instruction manual
- 3YIHENG THZ series thermostatic Shaker instruction manual
- 4YIHENG BPZ-Series Vacuum oven instruction manual
- 5YIHENG DZF series Vacuum oven instruction manual
- 6YIHENG DZF-6090/6210 Vacuum oven instruction manual
- 7YIHENG BPZ-LC/LCB series Vacuum oven instruction manual
- 8YIHENG BGZ series Vacuum oven instruction manual
- 9LINSHANG LS215 Ultrasonic Thickness Gauge Instruction Manual
- 10LINSHANG LS213 Ultrasonic Thickness Gauge Instruction Manual