YIHENG THZ-300/C thermostatic foster Shaker instruction manual
Table of Contents
1. Safety Reminders ............ 3
2. Product Overview (Appearance Drawing, Structural Functions) ............ 4
3. Product Usage (Preparation, Power-Up, Functions, Alarms) ............ 5-8
4. Product Maintenance and Key Points ............ 9
5. Appendix (Technical Indicators, Parameter Table, Fault Handling, Wiring Diagrams) ............ 10-13
Packing list ............ 14
1. Safety Reminders
âš Danger (may cause serious property damage or personal injury)
This product must be reliably grounded and kept away from sources of electromagnetic interference (neutral or centerline must not be used as ground).
Before use, please confirm that the voltage and frequency of the power supply match the product requirements.
Products should use independent power outlets, and ensure that plugs and sockets are properly grounded.
Do not unplug or plug in the power cord without turning off the power switch during operation.
It is not allowed to arbitrarily extend or cut the product's power cord.
Unauthorized repairs are prohibited; entrusted repairs must be carried out by professionals.
âš Warning (may cause property damage or personal injury)
You must fully read and understand the product manual before proceeding.
The 304 stainless steel inner pot is not acid-resistant; please take corrosion prevention measures. Never use acidic media inside the box!
When unplugging the power cord, do not drag the power cord directly.
If any of the following situations occur, you must unplug the power cord of this product:
a) When replacing the fuse; b) When a product malfunction is awaiting inspection and repair; c) When the product is out of use for a long time; d) When moving the product.When placing the product on the worktable, the four feet must be fixed to prevent movement caused by vibration, which could cause the product to fall and cause serious damage or personal injury.
📌 Note (may affect service life or cause abnormalities)
Products with model suffix "C" are equipped with a refrigeration system, and the tilt angle during handling must not exceed 45 degrees.
After products marked with "C" are transported and placed in place, they should be left to stand for (1~2) days before starting the machine, to ensure the refrigeration system operates normally and extends its lifespan.
Products should be placed on a hard and solid surface to keep them level.
There should be a certain gap around the product (recommended not less than 50cm).
Products must be used under specified conditions (ambient temperature, altitude, etc.).
Do not open or close the product box door by gravity, as this may cause the door to fall off and damage the product.
If the use is stopped for a long time, regular moisture removal should be performed.
2. Product Introduction
1. Appearance drawing
2. Overview of structural functions
The THZ series constant temperature incubator shaker adopts a benchtop structure, with an outer shell made of engineering plastic, inner chamber lined with stainless steel thin plates, a flip-top door with a built-in oscillation tray, and an LCD monitor controlling the oscillation speed and temperature.
Cabinet: The shell is molded from engineering plastic, with a motor-driven mechanical oscillator (circular trajectory) mounted at the bottom. The rear air duct panel of the studio is made of stainless steel thin plate, with the LCD display and operation buttons located on the front of the cabinet.
Box doors: Uses a large flip-top structure, made of molded glass or transparent plastic, making it easy to observe and place cultures.
Control: Microcomputer electronic controller uses PID adjustment for closed-loop control of temperature and speed. The LCD display displays multiple sets of data, making operation simple.
Cycle: The fan behind the air duct plate inside the workspace rotates, circulating hot and cold airflow to improve temperature uniformity.
Protection: Prevents multiple compressor starts in a short period (products with a tail "C"), with delayed start function; When the measured temperature exceeds the set temperature by 3°C, the buzzer alarms and cuts off heating; The door switch can cut off the power to heaters, fans, and turntables, ensuring safe operation.
3. Product usage
1. Preparation before use
Ambient temperature: 15°C~35°C, relative humidity not exceeding 85%;
Power supply: (220±22)V, (50±1)Hz;
Atmospheric pressure: (86~106) kPa;
The elevation should not exceed 2000 meters;
No strong vibration sources or strong electromagnetic fields nearby;
It should be placed on a stable, level, and sturdy countertop free of direct sunlight, with four legs fixed; The indoor area should be free of severe dust and corrosive gases;
There should be a gap of no less than 50cm around the product (front, back, left, right, top).
2. Power on and operate
Display and button description:
TIME: Time display window (running time/scheduled)
REV: Displays the measured speed
TEMP: Displays the measured temperature
RUN Operation Indicator Light: Lights up while the device is running
Heating lamp / cooling lamp / fan indicator light
TEMP SET: Displays the set temperature or parameter values
REV SET: Displays the set speed or parameter code
Function key (MODE): Modify settings/bring up parameters/confirm
Shift key/increase key/decrease key
Start/Stop button: Press and hold for more than 4 seconds to control operation/stop
1) Timer function: Press the "MODE" button; when the time is set to 0, there is no timer function; When the setting time is not zero, you can use the "Increase", "Decrease", and "Shift" keys to set the required time. When the timer arrives, TIME displays "END" and beeps.
2) Speed setting: Then press the "MODE" button, the REV SET window value flashes, and use the adjust/shift keys to set the speed.
3) Temperature setting: Then press the "MODE" button; the TEMP SET window will flash to set the desired temperature. Press the "MODE" button again to return to the standard display.
4) Ambient temperature check: Press the "Shift" button for 4 seconds, then press again for another 4 seconds to return.
Note:After each parameter is modified, you need to press the "MODE" key to confirm; After all settings are complete, press the "START/STOP" button for about 4 seconds to start running.
3. Equipment scheduled startup, ambient temperature correction, temperature control accuracy calibration, etc
3.1 Scheduled Power Start-On Function
Long press the "MODE" button to display LK, enter password 3, press MODE to find KA, and adjust KA to 3 (scheduled startup mode) by upstream/shifting. In standard mode, press the MODE button, TIME flashes, set the scheduled time value, and confirm. After powering off and re-energizing, the scheduled countdown starts and runs automatically after zeroing.
3.2 Ambient temperature correction
Generally, no correction is needed. If there is a deviation, you can enter the A menu to adjust the Se value.
3.3 Checking Temperature Control Accuracy (Pb and PK Values)
Place a 0.1°C grading mercury thermometer in the workroom, maintain the temperature for 1 hour, compare the thermometer reading with the instrument reading, and the deviation should be within ±0.5°C. If exceeded, the PK value can be corrected:
PK = 4000 × (mercury thermometer value - instrument reading) / instrument reading, modify the original factory PK value and repeatedly correct until it meets the requirements.
4. Instrument alarm function
If the measured temperature exceeds the set temperature by 3°C (factory AL=3), beep and cut off heating, and press any button to mute the sound.
Excessive oscillation load causes the motor to overload for more than 10 seconds, beeps and the oscillation stops; press any key to silence.
Automatic shutdown when scheduled ends, buzzer reminders, and any key is used to silence the sound.
If the PT100 is disconnected or short-circuited, causing the measurement temperature to exceed 60°C or fall below -1.0°C, the screen will display "------".
5. Independent temperature limit alarm settings
Set the required temperature limit temperature by using the + and - buttons on the panel overtemperature setting dial. For example, SV=37°C and AL=3, the temperature limit should be set to 45°C.
4. Product Maintenance and Key Points for Care
1. Key Maintenance Points
Storage conditions: Ambient temperature (-40~+55°C), relative humidity ≤95%, atmospheric pressure (80~106) kPa, altitude ≤ 2000 meters, no corrosive gases.
Moisture Removal Treatment: (1) Wipe the box dry after each use. (2) If not used for a long time, every month you should power on, heat to 40°C, run at 50r/min for 5 hours to remove moisture, and unplug the power cord afterward.
Defrosting (endnote "C"): When operating below ambient temperature for a long time or cooling performance deteriorates, set 40°C to allow the product to resume after 3~5 hours of operation.
Treatment without heating or shaking: First, check whether the door switch panel position is in place, as shown in the right diagram.
2. Key points to note
After the product has been stored for a period of time or been used intermittently for more than half a year, or after process parameters change, the temperature control accuracy should be verified.
Except for parameters such as temperature, rotational speed, time, and zero-point correction, all other parameter modifications should be standardized or adjusted by professionals.
For products marked with "C" in the tail, the handling angle should not exceed 45 degrees. After placing them in place, let them sit for 1~2 days before starting the machine.
When the equipment is placed on the platform, the edge must be equipped with protruding barriers to prevent displacement and falling.
After the equipment is placed on the ground, it must be leveled and must not tilt; Leave at least 50cm of space on the left and right sides and back. Especially for products with "C" in environments above 30°C, it is recommended to improve rear ventilation.
Stay away from sources of electromagnetic interference and ensure the grounding wire is in good condition.
5. Appendix
1. Technical Specifications (Table 1)
| Item\Model | THZ-300 | THZ-300C |
|---|---|---|
| Input power | 600W | 1600W |
| Temperature control range | (RT+5)°C~50°C | 4°C~50°C |
| Resolution | 0.1°C | |
| Temperature fluctuation | ±0.5°C | |
| Temperature uniformity | 2.0°C | |
| Timed range | 0~99h:59min | |
| Pallet rotation speed | (40~300) r/min | |
| Tray amplitude | 20mm | |
| Pallet load | ≤10kg | |
| Pallet size | 450×450 mm² | |
| External dimensions | 610×717×500 mm³ | 925×590×500 mm³ |
Note: RT - ambient temperature
2. Functional Parameter Tables (Tables 2~4)
Table 2 A Menu (Long press MODE to enter password)
| Prompt | Name | Set the range | Explanation |
|---|---|---|---|
| Fn | Fan control | 1.0~100.0 | No effect |
| KA | Power-on mode | 0~3 | 0: Stop, 1: Automatic operation, 2: Resume operation after last power cut, 3: Scheduled startup |
| Se | Ambient temperature correction | -19.9~80.0 | Corrected ambient temperature display deviation |
| Pb | Zero position adjustment | -100.0~100.0 | Adjust when the zero error is large |
| Pk | Full level adjustment | -1000~1000 seconds | Used for correcting fullness errors |
Table 3 Menu B
| Prompt | Name | Set the range | Explanation |
|---|---|---|---|
| AL | Alarm settings | 0~100.0°C | If the temperature exceeds the set + AL alarm, the heating is cut off |
| CL | Refrigeration control | 0.0~100.0 | Trigger compressor delay after temperature exceeds the limit |
| Ct | Cooling delay | 0~3600 seconds | Compressor delayed start time |
| P | Scale band | 1.0~300.0 | Adjust the intensity of the action |
| I | Scoring time | 10~3600S | Eliminating static errors |
| d | Differential time | 0~3600S | Suppression of overshoot |
| Ar | Overshoot suppression | 0~100% | Suppression of startup overshoot |
| t | Control cycle | 1~300S | Thyristor output cycle |
Table 4 Speed Regulation Parameters (C Menu)
| Prompt | Name | Set the range | Explanation |
|---|---|---|---|
| Pr | Speed regulation ratio | 2~999 | The strength of proportional effects |
| Ir | Speed adjustment points | 2~999 | Integration speed |
| rA | Motor starting voltage | 0~50 | Starting voltage percentage |
| rn | Maximum speed | 40-300 | Speed limit |
Any change in each functional parameter may affect the effectiveness of control. If you do not press the "MODE" button for one minute, it will automatically return to standard mode.
3. Causes and Handling of Faults (Table 5)
| Fault phenomena | Speculation on the cause of the fault | Troubleshooting methods |
|---|---|---|
| No power supply when powered on | Socket has no power or poor contact; Power cord breakage; Switch not turned on or damaged; The fuse burns out | Inspection and repair; exchange; Open/replace; Replace the fuse and identify the cause |
| The device doesn't work when powered on | No run button pressed; The controller or motor is faulty | Follow the instructions; Replace parts |
| The screen displays "------" | The PT100 sensor is faulty | Check sensor resistance (100Ω at 0°C, about +0.3Ω per kWh) |
| Either it does not heat up or heats up slowly | Set value lower than ambient values; The timer function sets TIME≠0; The controller has no output; Heating tube failure; The door switch failed | Reset the temperature; Cancel the timer; Replace the controller; Check the heating tube joints; Adjust the door switch |
| Large temperature display error | Accuracy exceeds the limit | Zero and fullness corrections |
| Temperature loss of control | Heater/relay/controller is faulty; The wind does not turn | Replace the corresponding components |
| Loud noise | Unstable placement; Loose screws | Leveling and tightening |
| Oscillation is not working or out of control | The optical interrupter is broken; Control board failure; The motor does not rotate; The door switch failed | Replace optical interrupters, control boards, motors, and adjust door switches |
| It wouldn't spin | mechanical jamming; The starting torque is too low | Manual Push Examination; Increase rA |
4. Electrical wiring schematic
No changes will be notified
Packing list
Product Name: Constant Temperature Culture Shaking Machine
| Serial number | Category | Name | Unit | Quantity |
|---|---|---|---|---|
| 1 | Documents | User manual | portion | 1 |
| 2 | Documents | Packing list | portion | 1 |
| 3 | Documents | Certificate of Conformity | portion | 1 |
| 4 | Documents | Guarantee letter | portion | 1 |
| 5 | Documents | Warranty card | portion | 1 |
| 6 | Spare parts | Fuse | Only | 2 |
| 7 | Spare parts | Power cord | Root | 1 |
The items listed on this order match the actual items inside the box
- 1YIHENG THZ series thermostatic Shaker instruction manual
- 2YIHENG BPZ-Series Vacuum oven instruction manual
- 3YIHENG DZF series Vacuum oven instruction manual
- 4YIHENG DZF-6090/6210 Vacuum oven instruction manual
- 5YIHENG BPZ-LC/LCB series Vacuum oven instruction manual
- 6YIHENG BGZ series Vacuum oven instruction manual
- 7LINSHANG LS215 Ultrasonic Thickness Gauge Instruction Manual
- 8LINSHANG LS213 Ultrasonic Thickness Gauge Instruction Manual
- 9LINSHANG LS212 Ultrasonic Thickness Gauge Instruction Manual
- 10LINSHANG LS211 Ultrasonic Thickness Gauge Instruction Manual