YIHENG THZ-300/C thermostatic foster Shaker instruction manual

YIHENG THZ-300/C thermostatic foster Shaker manual, covering safety tips, structural functions, operation steps, timing appointments, temperature control calibrating, alarm settings, maintenance and troubleshooting.
THZ-300/300C model
Constant temperature incubation shaker
User manual

Table of Contents

  • 1. Safety Reminders ............ 3

  • 2. Product Overview (Appearance Drawing, Structural Functions) ............ 4

  • 3. Product Usage (Preparation, Power-Up, Functions, Alarms) ............ 5-8

  • 4. Product Maintenance and Key Points ............ 9

  • 5. Appendix (Technical Indicators, Parameter Table, Fault Handling, Wiring Diagrams) ............ 10-13

  • Packing list ............ 14

1. Safety Reminders

âš  Danger (may cause serious property damage or personal injury)

  • This product must be reliably grounded and kept away from sources of electromagnetic interference (neutral or centerline must not be used as ground).

  • Before use, please confirm that the voltage and frequency of the power supply match the product requirements.

  • Products should use independent power outlets, and ensure that plugs and sockets are properly grounded.

  • Do not unplug or plug in the power cord without turning off the power switch during operation.

  • It is not allowed to arbitrarily extend or cut the product's power cord.

  • Unauthorized repairs are prohibited; entrusted repairs must be carried out by professionals.

âš  Warning (may cause property damage or personal injury)

  • You must fully read and understand the product manual before proceeding.

  • The 304 stainless steel inner pot is not acid-resistant; please take corrosion prevention measures. Never use acidic media inside the box!

  • When unplugging the power cord, do not drag the power cord directly.

  • If any of the following situations occur, you must unplug the power cord of this product:
                       a) When replacing the fuse;                    b) When a product malfunction is awaiting inspection and repair;                    c) When the product is out of use for a long time;                    d) When moving the product.

  • When placing the product on the worktable, the four feet must be fixed to prevent movement caused by vibration, which could cause the product to fall and cause serious damage or personal injury.

📌 Note (may affect service life or cause abnormalities)

  • Products with model suffix "C" are equipped with a refrigeration system, and the tilt angle during handling must not exceed 45 degrees.

  • After products marked with "C" are transported and placed in place, they should be left to stand for (1~2) days before starting the machine, to ensure the refrigeration system operates normally and extends its lifespan.

  • Products should be placed on a hard and solid surface to keep them level.

  • There should be a certain gap around the product (recommended not less than 50cm).

  • Products must be used under specified conditions (ambient temperature, altitude, etc.).

  • Do not open or close the product box door by gravity, as this may cause the door to fall off and damage the product.

  • If the use is stopped for a long time, regular moisture removal should be performed.

2. Product Introduction

1. Appearance drawing

Exterior diagram            
Figure (1)

2. Overview of structural functions

The THZ series constant temperature incubator shaker adopts a benchtop structure, with an outer shell made of engineering plastic, inner chamber lined with stainless steel thin plates, a flip-top door with a built-in oscillation tray, and an LCD monitor controlling the oscillation speed and temperature.

  • Cabinet: The shell is molded from engineering plastic, with a motor-driven mechanical oscillator (circular trajectory) mounted at the bottom. The rear air duct panel of the studio is made of stainless steel thin plate, with the LCD display and operation buttons located on the front of the cabinet.

  • Box doors: Uses a large flip-top structure, made of molded glass or transparent plastic, making it easy to observe and place cultures.

  • Control: Microcomputer electronic controller uses PID adjustment for closed-loop control of temperature and speed. The LCD display displays multiple sets of data, making operation simple.

  • Cycle: The fan behind the air duct plate inside the workspace rotates, circulating hot and cold airflow to improve temperature uniformity.

  • Protection: Prevents multiple compressor starts in a short period (products with a tail "C"), with delayed start function; When the measured temperature exceeds the set temperature by 3°C, the buzzer alarms and cuts off heating; The door switch can cut off the power to heaters, fans, and turntables, ensuring safe operation.

3. Product usage

1. Preparation before use

  • Ambient temperature: 15°C~35°C, relative humidity not exceeding 85%;

  • Power supply: (220±22)V, (50±1)Hz;

  • Atmospheric pressure: (86~106) kPa;

  • The elevation should not exceed 2000 meters;

  • No strong vibration sources or strong electromagnetic fields nearby;

  • It should be placed on a stable, level, and sturdy countertop free of direct sunlight, with four legs fixed; The indoor area should be free of severe dust and corrosive gases;

  • There should be a gap of no less than 50cm around the product (front, back, left, right, top).

2. Power on and operate

Operation panel diagram            
Figure (2) Operation Panel

Display and button description:

  • TIME: Time display window (running time/scheduled)

  • REV: Displays the measured speed

  • TEMP: Displays the measured temperature

  • RUN Operation Indicator Light: Lights up while the device is running

  • Heating lamp / cooling lamp / fan indicator light

  • TEMP SET: Displays the set temperature or parameter values

  • REV SET: Displays the set speed or parameter code

  • Function key (MODE): Modify settings/bring up parameters/confirm

  • Shift key/increase key/decrease key

  • Start/Stop button: Press and hold for more than 4 seconds to control operation/stop

Basic operation step-by-step diagram            
Figure (3) Basic operating steps

1) Timer function: Press the "MODE" button; when the time is set to 0, there is no timer function; When the setting time is not zero, you can use the "Increase", "Decrease", and "Shift" keys to set the required time. When the timer arrives, TIME displays "END" and beeps.

2) Speed setting: Then press the "MODE" button, the REV SET window value flashes, and use the adjust/shift keys to set the speed.

3) Temperature setting: Then press the "MODE" button; the TEMP SET window will flash to set the desired temperature. Press the "MODE" button again to return to the standard display.

4) Ambient temperature check: Press the "Shift" button for 4 seconds, then press again for another 4 seconds to return.

Note:After each parameter is modified, you need to press the "MODE" key to confirm; After all settings are complete, press the "START/STOP" button for about 4 seconds to start running.

3. Equipment scheduled startup, ambient temperature correction, temperature control accuracy calibration, etc

Function settings diagram            
Figure (4) Appointment/calibration function process

3.1 Scheduled Power Start-On Function

Long press the "MODE" button to display LK, enter password 3, press MODE to find KA, and adjust KA to 3 (scheduled startup mode) by upstream/shifting. In standard mode, press the MODE button, TIME flashes, set the scheduled time value, and confirm. After powering off and re-energizing, the scheduled countdown starts and runs automatically after zeroing.

3.2 Ambient temperature correction

Generally, no correction is needed. If there is a deviation, you can enter the A menu to adjust the Se value.

3.3 Checking Temperature Control Accuracy (Pb and PK Values)

Place a 0.1°C grading mercury thermometer in the workroom, maintain the temperature for 1 hour, compare the thermometer reading with the instrument reading, and the deviation should be within ±0.5°C. If exceeded, the PK value can be corrected:
       PK = 4000 × (mercury thermometer value - instrument reading) / instrument reading, modify the original factory PK value and repeatedly correct until it meets the requirements.

4. Instrument alarm function

  • If the measured temperature exceeds the set temperature by 3°C (factory AL=3), beep and cut off heating, and press any button to mute the sound.

  • Excessive oscillation load causes the motor to overload for more than 10 seconds, beeps and the oscillation stops; press any key to silence.

  • Automatic shutdown when scheduled ends, buzzer reminders, and any key is used to silence the sound.

  • If the PT100 is disconnected or short-circuited, causing the measurement temperature to exceed 60°C or fall below -1.0°C, the screen will display "------".

5. Independent temperature limit alarm settings

Set the required temperature limit temperature by using the + and - buttons on the panel overtemperature setting dial. For example, SV=37°C and AL=3, the temperature limit should be set to 45°C.

Independent temperature limit setting            
Set Button Diagram (5)

4. Product Maintenance and Key Points for Care

1. Key Maintenance Points

Storage conditions: Ambient temperature (-40~+55°C), relative humidity ≤95%, atmospheric pressure (80~106) kPa, altitude ≤ 2000 meters, no corrosive gases.

Moisture Removal Treatment: (1) Wipe the box dry after each use. (2) If not used for a long time, every month you should power on, heat to 40°C, run at 50r/min for 5 hours to remove moisture, and unplug the power cord afterward.

Defrosting (endnote "C"): When operating below ambient temperature for a long time or cooling performance deteriorates, set 40°C to allow the product to resume after 3~5 hours of operation.

Treatment without heating or shaking: First, check whether the door switch panel position is in place, as shown in the right diagram.

Door switch adjustment            
Figure (6) Door switch clipping plate

2. Key points to note

  • After the product has been stored for a period of time or been used intermittently for more than half a year, or after process parameters change, the temperature control accuracy should be verified.

  • Except for parameters such as temperature, rotational speed, time, and zero-point correction, all other parameter modifications should be standardized or adjusted by professionals.

  • For products marked with "C" in the tail, the handling angle should not exceed 45 degrees. After placing them in place, let them sit for 1~2 days before starting the machine.

  • When the equipment is placed on the platform, the edge must be equipped with protruding barriers to prevent displacement and falling.

  • After the equipment is placed on the ground, it must be leveled and must not tilt; Leave at least 50cm of space on the left and right sides and back. Especially for products with "C" in environments above 30°C, it is recommended to improve rear ventilation.

  • Stay away from sources of electromagnetic interference and ensure the grounding wire is in good condition.

5. Appendix

1. Technical Specifications (Table 1)

Item\ModelTHZ-300THZ-300C
Input power600W1600W
Temperature control range(RT+5)°C~50°C4°C~50°C
Resolution0.1°C
Temperature fluctuation±0.5°C
Temperature uniformity2.0°C
Timed range0~99h:59min
Pallet rotation speed(40~300) r/min
Tray amplitude20mm
Pallet load≤10kg
Pallet size450×450 mm²
External dimensions610×717×500 mm³925×590×500 mm³

Note: RT - ambient temperature

2. Functional Parameter Tables (Tables 2~4)

Table 2 A Menu (Long press MODE to enter password)

PromptNameSet the rangeExplanation
FnFan control1.0~100.0No effect
KAPower-on mode0~30: Stop, 1: Automatic operation, 2: Resume operation after last power cut, 3: Scheduled startup
SeAmbient temperature correction-19.9~80.0Corrected ambient temperature display deviation
PbZero position adjustment-100.0~100.0Adjust when the zero error is large
PkFull level adjustment-1000~1000 secondsUsed for correcting fullness errors

Table 3 Menu B

PromptNameSet the rangeExplanation
ALAlarm settings0~100.0°CIf the temperature exceeds the set + AL alarm, the heating is cut off
CLRefrigeration control0.0~100.0Trigger compressor delay after temperature exceeds the limit
CtCooling delay0~3600 secondsCompressor delayed start time
PScale band1.0~300.0Adjust the intensity of the action
IScoring time10~3600SEliminating static errors
dDifferential time0~3600SSuppression of overshoot
ArOvershoot suppression0~100%Suppression of startup overshoot
tControl cycle1~300SThyristor output cycle

Table 4 Speed Regulation Parameters (C Menu)

PromptNameSet the rangeExplanation
PrSpeed regulation ratio2~999The strength of proportional effects
IrSpeed adjustment points2~999Integration speed
rAMotor starting voltage0~50Starting voltage percentage
rnMaximum speed40-300Speed limit

Any change in each functional parameter may affect the effectiveness of control. If you do not press the "MODE" button for one minute, it will automatically return to standard mode.

3. Causes and Handling of Faults (Table 5)

Fault phenomenaSpeculation on the cause of the faultTroubleshooting methods
No power supply when powered onSocket has no power or poor contact; Power cord breakage; Switch not turned on or damaged; The fuse burns outInspection and repair; exchange; Open/replace; Replace the fuse and identify the cause
The device doesn't work when powered onNo run button pressed; The controller or motor is faultyFollow the instructions; Replace parts
The screen displays "------"The PT100 sensor is faultyCheck sensor resistance (100Ω at 0°C, about +0.3Ω per kWh)
Either it does not heat up or heats up slowlySet value lower than ambient values; The timer function sets TIME≠0; The controller has no output; Heating tube failure; The door switch failedReset the temperature; Cancel the timer; Replace the controller; Check the heating tube joints; Adjust the door switch
Large temperature display errorAccuracy exceeds the limitZero and fullness corrections
Temperature loss of controlHeater/relay/controller is faulty; The wind does not turnReplace the corresponding components
Loud noiseUnstable placement; Loose screwsLeveling and tightening
Oscillation is not working or out of controlThe optical interrupter is broken; Control board failure; The motor does not rotate; The door switch failedReplace optical interrupters, control boards, motors, and adjust door switches
It wouldn't spinmechanical jamming; The starting torque is too lowManual Push Examination; Increase rA

4. Electrical wiring schematic

Electrical wiring schematic            

No changes will be notified

Packing list

Product Name: Constant Temperature Culture Shaking Machine

Serial numberCategoryNameUnitQuantity
1DocumentsUser manualportion1
2DocumentsPacking listportion1
3DocumentsCertificate of Conformityportion1
4DocumentsGuarantee letterportion1
5DocumentsWarranty cardportion1
6Spare partsFuseOnly2
7Spare partsPower cordRoot1