Reader's Guide
This manual is compiled based on manufacturer's data. Due to equipment batch changes, the content may not fully match every physical unit. Please refer to the actual operation and on-device labels.
Copyright belongs to the manufacturer. For internal reference of purchasers only. If safety parameters in the manual differ from the physical unit, please suspend operation and consult our engineer for confirmation.
Disclaimer: This manual is for reference only. In case of any discrepancy, the physical unit shall prevail. Our company assumes no liability for any consequences arising from operations based on this manual.
YIHENG DZF series Vacuum oven instruction manual
The matters recorded here are extremely important and must be strictly observed
1. Safety Reminders
This product must be reliably grounded and kept away from sources of electromagnetic interference (neutral or centerline must not be used as ground).
Before use, please confirm that the voltage and frequency of the power supply match the product requirements.
Products should use independent power outlets, and ensure that plugs and sockets are properly grounded.
It is not allowed to unplug or plug in the power cord arbitrarily without turning off the power switch during operation.
It is not allowed to arbitrarily extend or cut the product's power cord.
Do not place flammable, explosive, volatile, or corrosive substances during drying or baking.
Do not fill inert gases during heating to avoid explosions or other accidents; violations are at your own risk.
Do not touch the doors, inspection windows, or surrounding surfaces of the product when operating at temperatures above 80°C to prevent burns.
Unauthorized repairs are prohibited; entrusted repairs must be carried out by professionals.
You must fully read and understand the product manual before proceeding.
The 304 stainless steel inner pot is not acid-resistant; please take corrosion prevention measures. Never use acidic media inside the box!
When unplugging the power cord, do not drag the power cord directly.
If any of the following situations occur, you must unplug the power cord of this product:
When replacing the fuse tube;
When a product malfunction is awaiting inspection and repair;
When the product is discontinued for a long time;
When moving products;
After the product is powered on, the upper deviation alarm function must be adjusted or confirmed.
Products should be placed on a hard and solid surface to keep them level.
A certain amount of space should be left around the product. (See Chapter 3, Section 1.4)
Products must be used under certain operating conditions. (See Chapter 3, Sections 1.1~1.6)
Do not open or close the product box door by gravity, as this may cause the door to fall off, damage the product, and cause injury.
2. Product Introduction
1. Appearance drawing
2. Overview of structural functions
The vacuum drying oven (hereinafter referred to as the vacuum chamber) is a benchtop structure. The vacuum chamber consists of four parts: the chamber body, the inner tank (chamber), the vacuum extraction system, and the temperature control system.
The cabinet is made of high-quality thin steel plates, with a surface powder coating and vibrant colors. The inner liner is made of general galvanized steel plate or 304 stainless steel plate. The inner chamber is square with semi-circular inner corners, and the space between the inner and outer boxes is filled with ultra-fine glass wool as insulation material; The door features a double-layer bulletproof glass inspection window in the middle, making it easy to observe the items being dried and processed inside; A thick tempered glass piece is installed on the inside of the box, and a long column-shaped door latch is used. This allows the spacing between the door latch and the thick tempered glass to be adjusted, so that after the door is closed, the rubber sealing ring can be tightly pressed to ensure no air leakage during vacuuming.
The vacuum extraction system consists of a vacuum pump (optional accessories), vacuum gauge, solenoid valve or vacuum valve, balance port, and other components. Depending on user requirements, optional drying filter tanks or air inlet valves (other gases can be input into the workroom). If a vacuum pump is optional, the pump's extraction rate must be ≥2L/s
The main component of the temperature control system is the thermometer. It is a microcomputer intelligent controller composed of a microcontroller and peripheral circuits for a dual-row LED or LCD display. Using a Pt100 platinum resistor as the temperature sensing element, the heating system is controlled by PID regulation. The temperature controller also features timed control, temperature control error correction, and deviation alarm protection. A 15-segment programmable LCD controller or a 30-segment variable temperature rise (cool) rate program controller can be configured according to user needs.
The electric heating system uses a resistance wire structure heater. Except for models 6021-1, 6022, 6031-1, 6032, 6050, 6051, 6050B, and 6053, which use electric heaters to heat the inner chamber, all other models have heaters installed inside shelves.
This series features high temperature control accuracy, minimal overshoot, minimal fluctuation, and deviation overtemperature protection. Additionally, an independent over-temperature protection system can be optionally configured according to user requirements.
3. Product usage
1. Preparation before use
Products should operate under the following usage conditions:
Ambient temperature: 5°C~40°C, relative humidity≤85%, atmospheric pressure: (86~106) KPa; The elevation does not exceed 2000 meters
No strong vibration sources or strong electromagnetic fields nearby;
It should be placed in a stable, level indoor area free of heavy dust, direct sunlight, and corrosive gases;
Leave enough space around the product, as shown in the upper right figure, and should not be placed under the fire alarm;
For product power supply, see technical specifications (Appendix 1);
When placing items in the studio, maintain a certain gap around the top and bottom, ensure the shelf is not bent or deformed by pressure, and ensure that dry items come into direct contact with the shelf (or through metal with good thermal conductivity).
2. Power on the machine
(1) (PV) Display Display Measurement Values Various prompts are displayed according to the instrument status.
(2) (SV) Display Display Set Values Displays various parameter values according to the instrument status
(3) AT (Operation Indicator Light): Lights up when the controller is running, flashes during self-tuning, and turns off when stopped;
(4) HEAT (Heating Indicator): Lights up when heating output is present.
(5) ALM (Alarm Indicator Light): Lights up when an alarm output is present, and the buzzer sounds.
(6) COLD (Cooling Indicator): Lights up when there is cooling output; (Note: This product does not have COLD cooling function.)
(7) Used to adjust various parameter values or enter self-tuning mode
(8) Used to adjust internal parameter values or enter self-tuning mode
(9) Shift key: Used to shift set values and internal parameters, and to observe the scheduled operating time;
(10) Function keys: Modify set values, retrieve parameter symbols, and confirm parameter modifications.
2.2 Vacuum Commissioning
2.2.1 For models 6020~6053, use the included vacuum connecting tube (inner diameter: φ16mm, wall thickness: 10mm) to connect the vacuum drying oven's exhaust pipe (outer diameter: φ16mm) and the vacuum pump (2XZ-2 model, inlet outer diameter: φ16mm; Pumping rate ≥2L/s) Connections are firm (recommended for middle string drying tanks).
2.2.2 Close the box door and tighten the door handle in place, close the balance port (so that the hole on the rubber plug and the hole on the balance port core are twisted 180°). First turn on the vacuum pump power switch, then open the vacuum valve (rotate clockwise 90°). The first use may be too tight for the vacuum valve, so you can rotate it forcefully. When the vacuum gauge reading reaches -0.1 MPa (the red arrow in Figure 2) and the vacuum pump needs to be shut off, the vacuum valve must be closed first to prevent the pump oil from flowing back into the workroom.
Figure 2: For specific unit conversions, see Appendix 6
2.3 Vacuum Box Commissioning
After vacuum adjustment is complete, proceed with the following steps:
2.3.1 Turn on the vacuum chamber power; the power indicator should turn on;
2.3.2 After 5 seconds of self-check, the thermostat automatically enters operating mode, where the PV screen displays the measured temperature and the SV screen displays the set temperature. At this time, the AT light is on. When PV < SV, the HEAT light should turn on, indicating the temperature has entered a heating state.
2.4 Check the temperature control accuracy
2.4.1 Place a 0.5°C gradient mercury thermometer (or a digital thermometer with 0.1°C resolution) into the product workroom; The mercury thermometer should be placed flat on the shelf (Note: The thermometer head must be placed close to the shelf surface to ensure good contact).
2.4.2 Set the SV temperature control value within the temperature control range. When the PV measurement reaches the set value, maintain the temperature for about (1~2) hours (the holding time varies by product model). Observe that the actual mercury thermometer reading and the PV display on the thermometer should differ by ≤±1°C.
3. Temperature and timer settings
3.1 In working mode, press the "SET" button once; the PV screen displays the character " "; press the or * key to make the SV screen display the required temperature value; (See Appendix 2 Function Extraction Process)
3.2 Press the "SET" key again; the PV screen will display the character " ". Press the or key to display the SV screen at the desired time value; (See Appendix 2 Function Extraction Process)
3.2.1 When ST is set to 0, the controller cancels the timing function and continues to run; When the ST setting is not zero, the controller has a timing function. When the controller's running time is up, the SV screen displays "END," the buzzer sounds, the controller stops working, any key is pressed to mute, and pressing the key and key for 4 seconds restarts simultaneously.
3.2.2 When the controller is in operating mode, simply press the shift button to display "TIME" on the PV screen, and the SV screen shows the operating time of the controller. Press the shift button again to return the controller to operating mode.
3.3 Timer Mode Press the "SET" button again to return to working mode and enter working mode
4. Setting up the upper deviation alarm
The setting of the upper deviation is reasonable, providing protection against system temperature control exceeding or losing control, and must be used during product operation.
4.1 When products leave the factory, AL=10~15 is generally set, i.e., the alarm temperature is: (SV+AL)°C
4.2 Press the "SET" key for about 4 seconds. When the PV screen displays the character " " (indicating the thermometer has entered the parameter menu), release it (indicating the thermometer has entered the parameter menu). However, you must unlock the electronic lock "Lk" before modifying the relevant parameters using the keys.
4.3 Unlocking Program: After entering the parameter menu, press the SET button several times. When the PV screen displays the character "", use the key to change the SV screen value from "0" to "Unlock Password", which will unlock the electronic lock (then press without a key, and after 1 minute, the thermostat will automatically return to working mode);
4.4 Press several "SET" keys; when the PV screen displays the character "" (AL), use the [ and " keys to set a reasonable deviation value (AL);
4.5 Press the "SET" key several times to bring up the character "", and use the key to change the "unlock password" to "0" to disable the electronic lock (Note: You can disable the electronic lock after all parameters that need to be changed).
4.6 After overheating, the buzzer intermittently alarms, and the over-temperature lamp AL1 stays on; press any key to mute it;
4.7 When setting from high temperature to low temperature, an alarm will also appear if the temperature exceeds the AL value. This is normal; press mute to proceed.
5. Methods to improve temperature control accuracy
5.1 After the product has been used for a period of time, the temperature control accuracy should be checked according to the method in 2.4
5.2 If a certain temperature point remains constant for 1-2 hours, the PV value (either high or low) still cannot reach the SV value, please adjust the corresponding instrument menu parameters. For example:
5.2.1 Inaccurate temperature control, below (or above the required temperature), causing static error, but the system remains stable: If PV < SV, adjust P (proportional band) to a smaller value (set 3~5 each adjustment), but generally not equal to 0, or adjust Ar (overshoot suppression) to a higher value (5~10 per adjustment). Generally, first modify Ar, then modify P if it cannot be satisfied, repeating until satisfied;
5.2.2 If the temperature overshoot is too high, adjust the value of Ar to a smaller value. If it cannot be met, then adjust the value to the larger P.
5.3 If the difference between the measured value of the mercury thermometer and the PV display exceeds ±1°C, the following method can be used to correct it:
5.3.1 Entering the Thermostat Parameter Menu (see section 4.2)
5.3.2 Unlocking Electronic Locks (see section 4.3)
5.3.3 Press the "SET" key several more times to find the " " symbol,
By PK = 4000 × (mercury meter value - PV measurement) / PV measurement After formula calculation, use the keys and s to modify the original factory PK value (Note: If one correction is inaccurate, it can be revised repeatedly until it matches);
5.3.4 Turning Off the Electronic Lock (see Clause 4.5)
6. Use of the optional "independent temperature limiting controller."
An independent temperature limiting controller is a separate protection system. When the temperature controller malfunctions and causes temperature loss and the temperature inside the workroom reaches the temperature limit set on the over-temperature dial, the independent temperature controller will automatically cut off heating and sound an alarm. When the temperature inside the workroom falls below the set temperature limit, the protection system is deactivated and the instrument resumes operation. This cycle repeats until the fault is resolved.
6.1 The temperature limit set value should be greater than or equal to (SV+AL)+(10~15)°C
6.2 See Figure 3. Use the + and — buttons on the panel dial to set the required temperature limit. Example: SV=130°C, AL=10, then 150°C should be set
4. Product maintenance and precautions
After each use, turn off the power and open the balance port. Wait until the vacuum level returns to zero before opening the door (if it cannot be opened, wait 5 minutes before opening; forcing it may damage the door handle).
During use, the principle of "open first, then close" is used for vacuum pumps, meaning the vacuum pump is opened first before the vacuum valve is opened during operation, and before finishing work, the vacuum valve is closed before the pump is closed. This prevents vacuum pump oil from backflowing into the chamber.
When removing items that have been dried, be very careful to avoid burns.
If you encounter dust or particles from the dryer during use, or if the dry material is moist with large moisture, it can cause oil contamination or emulsification, affecting normal vacuuming or increasing pump noise. Please replace the vacuum pump oil regularly according to usage conditions! Emulsification—Vacuum pump observation window visibly separates oil and water; Pollution - The vacuum pump observation window visibly shows the oil darkening and blackening, with a large amount of smoke at the outlet.
If the product is not used for a long time, both the inside and outside must be cleaned, the power cord removed, and the plastic dust cover covered.
If the storage environment is humid, regularly (about one month) power on and heated to remove moisture should be done.
Before reuse or when process requirements change. Temperature control accuracy verification should be carried out. (See Chapter 3, Clause 2.4)
Except for parameters such as SV, AL, Pb, Pk, etc., other control parameters require approval from our service center or be adjusted by professionals.
Aging and loss of elasticity of the door seal will cause the box to be not sealed. Generally, it should be replaced every six months, or if temperatures above 100°C are used for extended periods, the cycle should be shortened.
5. Appendix
1. Technical Specifications
This product is manufactured according to the enterprise standard Q/TIWY 7
| Technical parameters, specifications, specifications, and models | 6021 / 6021-1 | 6020 / 6022 | 6031 / 6031-1 | 6030A / 6032 | 6030B | 6050B | 6050 / 6053 | 6051 |
| Power supply voltage | AC220V / 50Hz | |||||||
| Input Power (W) | 400W / 700W | 500W / 800W | 250W | 600W | 1400W | |||
| Temperature control range | RT+10~200°C | RT+10~65°C | RT+10~200°C | |||||
| Temperature fluctuation | ±1.0°C | |||||||
| Achieve a vacuum degree | 133Pa | |||||||
| Shelf (layer) | 1 | 12 | 1 | 12 | 1 | 2 | 23 | 2 |
| Inner pot material | Cold-rolled sheet | Stainless steel | Cold-rolled sheet | Stainless steel | Cold-rolled sheet | |||
| The inner liner measures accordingly | 300mm×300mm×275mm | 320mm×320mm×300mm | 415mm×370mm×345mm | |||||
| External dimensions | 605mm×490mm×450mm | 630mm×510mm×490mm | 730mm×560mm×550mm | |||||
Note: According to user requirements, the temperature control range for 6050, 6053, 6030A, and 6020 can be expanded to RT+10~250°C
Notes: 1. RT stands for ambient temperature.
2. All technical parameters in the table are measured using a mercury thermometer with an accuracy of not less than ±0.5°C under ambient temperature of 25°C, relative humidity not exceeding 85%, and vacuum degree not less than 0.1MPa. The mercury head of the thermometer must be in good contact with the shelf surface inside the box (see 2.4.1).
2. Functional parameter tables
| Prompt | Name | Set the range | Explanation | Factory preset |
| RL/AL | Upper deviation alarm settings | 0~Full Scale, 0.0~Full Scale | When the temperature exceeds (SV+AL), AL1 lights up, the buzzer sounds, the heating output is cut off, and the SV window displays alternately in sync | |
| CL/CL | Refrigeration control settings | 0~Full Scale, 0.0~Full Scale | When the temperature exceeds the (SV+CL) value, the COLD light turns on, the refrigeration contacts are connected, and the compressor is started | / |
| CT/CT | Cooling control delay | (0~3600) seconds | When the measured value reaches the alarm threshold, the alarm relay output only after CT time | / |
| P/P | Scale band | 1~Full Scale, 1.0~Full Scale | It acts only on the heating side; the larger the P, the lower the system gain; Reducing P can improve system control accuracy and eliminate static errors | |
| I/I | Integration Time (Readjustment Time) | (0~3600) seconds | The time constant for the integration action—the larger I, the weaker the integral, and the system is stable. | |
| d/d | Differential time (pre-tuned time) | (0~3600) seconds | The time constant for differential action: the larger d, the stronger the differential action, and it can overcome overshoot. I=0, d=0 is controlled by half-proportionality, generally d is twice I, or equal to I | |
| AR/Ar | Overshoot suppression (ratio resetting) | (0~100)% | Changing the output power and increasing Ar can improve the heating rate, help eliminate static differences, reduce Ar, and prevent overshoot. | |
| T/T | Heating cycle | (1~300) seconds | The thyristor output is generally (2~3) seconds. For equipment with large remaining power, increasing T can reduce the static error of PID control. | 3 |
| Pd/Pb | Zero Adjustment (Intercept) | -100~100 -100.0~100.0 | When the zero error of the instrument is large and the fullness error is small, adjust this value; generally, Pt100 rarely adjusts this value | |
| PE/PK | Full Scale Adjustment (Slope) | -(1000~1000) seconds | When the instrument's zero position error is small and the fullness error is large, adjust this value: PK = 4000 × (specified value - actual displayed value) / actual display value. Generally, Pt100 adjusts this value first | |
| Fn | Fan control settings | 0~Full scale | When SP-Fn< = measured temperature < = SP + Fn, start the slow-speed fan to run; When measuring temperature SP+Fn, start the fast fan to operate. | / |
| L/LK | Combination lock | 0~255 | When LK=18, the above parameters can be changed | 0 |
3. Cause and handling of faults
| Fault phenomena | Speculation on the cause of the fault | Troubleshooting methods |
| No power supply when powered on (indicator light not on) | The external power outlet has no electricity | Check whether the wiring trips and whether the sockets are functioning properly |
| The power cord plug is not properly plugged in or is broken | Replug or repair the power cord | |
| Fuse not installed or breaking | Check that there are no short circuits in the vacuum chamber wiring and replace the fuse | |
| The instrument PV screen displays "☐☐☐" | Temperature sensor Pt100 is faulty | Check Pt100, replace (0°C is 100Ω, 0.3°C/Ω) |
| The temperature sensor wiring has come off | Rewire | |
| It does not heat up | Set the temperature low | Set temperature: SV≥RT+10°C |
| The instrument output circuit wiring has come loose | Rewire | |
| The thermostat has no output signal or is faulty | 3041 or BTA is faulty, swap it | |
| Electric heater faulty (open circuit, short circuit) | Swap | |
| Timing function enabled or settings are incorrect | St=0 or St=(heating + constant temperature). | |
| Temperature runaway, with the set temperature and the measured temperature error exceeding the limit | The thermostat output signal is out of control | 3041 or BTA is faulty, swap it |
| Does not meet usage conditions | SV≥RT+10°C | |
| Pt100 sensor has poor contact | Reduces contact resistance | |
| The relevant parameter settings are incorrect | Resetting relevant parameters (e.g., Ar, P, etc.) | |
| The measured temperature differs from the actual temperature inside the chamber | Not in a vacuum | Vacuum to reach the required vacuum level |
| The mercury temperature sensor does not touch the shelves | Replace | |
| Instrument or Pt100 parameter changes | Re-correct Pb and Pk parameters or swap Pt100 | |
| It cannot be vacuumed | The vacuum pump model or specifications are incorrect, or water is inflowing | A vacuum pump or oil change pump with a pumping rate of no less than 2L/s should be selected |
| Various connecting pipes (heads) are loose, and the inner diameter is too thin | Reselect a suitable inner diameter connecting tube and tighten the joint firmly | |
| The vacuum gauge was damaged | Replace | |
| The box door was not tightly closed | Use the included wrench to adjust the spacing of the door latch | |
| The rubber of the door seal has aged and lost its elasticity | Replace the door seal | |
| The vent valve and vacuum valve are in the wrong position | Adjust the position | |
| Air leakage (vacuum level drops from -0.1MPa to less than 0.085MPa within 24 hours) | Various connecting pipes or door sealing rings are leaking | Replace after inspection |
| Except for models 6050, 6051, and 6053, the heater "0" type sealing ring is either not sealed or not tightened | Use a hex wrench to tighten the heater base or replace the "0" type sealing ring | |
| The balance port is positioned incorrectly | Place it in the correct position | |
| Leakage at the vacuum valve area | Swap |
4. Wiring schematic
Appendix 5 Introduction to the Use of the YLGZ-21WG Programmable LCD Temperature Controller
1. Layout of the product control panel and thermometer panel (see Figure 1);
1.1 Indicator Light Description
TEMP zone (PV): displays measurement temperature;
SET Area (SV): Displays the set temperature;
Heating lamp: lights up when heating output is available;
Fan indicator light: lights up when the fan output is present (this unit does not have this function);
TIME: Time display window; Displays runtime or parameter values;
STEP area: Displays working time or parameters;
PROG Zone: Displays program workgroups or settings groups;
1.2 Key Description
SET key: Used to modify set values or enter internal parameter settings. Press and hold the setting button for more than 3 seconds while parameter setting mode to exit;
Shift button: Used to shift set values, internal parameters, and view ambient temperature;
Decrease key: Used to modify set values, various parameters, or start/stop auto-tune;
Add key: Used to modify set values, internal parameters, or view remaining periods.
RUN/STOP: Press for 3 seconds to run or stop the controller.
2. Examples of procedures
For example, now it needs to run at a constant temperature of 80°C for 120 minutes; Then run at 160°C at a constant temperature for 240 minutes; This program is run repeatedly from start to finish 10 times, set to automatically start when the controller is powered on;
2.1 Program Editing
2.1.1 In standard mode, press the SET button once, the PROG area value flashes, press the plus/minus key to select the setting group to modify, press the SET button once again, PROG flashes, STEP shows "1", you can set the parameters for the first segment of the group, the TIME area flashes, press plus/minus, and the shift key sets the period for the group to 1; press the SET key again, the SV area flashes, press the add/subtract/shift keys to set the temperature of the first segment of the group to 80°C. Press the SET key again; the second digit of the STEP area is 2, prompting you to enter the second segment of the program setting for that group. (See Figure 4)
| PROG 1 | STEP 2 | TIEM 02:00 |
| TEMP 80.0 °C | ||
| SET 80.0 °C | ||
| RUN ▲▲∅ | ||
2.1.2 The TIME area flashes, setting the time to 120 minutes; Press the SET button again; the SV area flashes, setting the temperature to 80°C. Press the SET key again; the second digit of the STEP area is 3, prompting you to enter the third segment of the program setting for that group. (See Figure 5)
| PROG 1 | STEP 3 | TIEM 00:01 |
| TEMP 82.3 °C | ||
| SET 160.0°C | ||
| RUN ▲▲⊗ | ||
2.1.3 The TIME area flashes, setting the time to 1 minute; Press the SET button again, the SV area flashes, setting the temperature to 160°C. Press the SET key again; the second digit of the STEP area is 4, prompting you to enter the fourth segment of the program setting for that group. (See Figure 6)
| PROG 1 | STEP 4 | TIEM 04:00 |
| TEMP 160.0 °C | ||
| SET 160.0 °C | ||
| RUN ▲▲ ≈ | ||
2.1.4 The TIME zone flashes, setting the time to 120 minutes; Press the SET button again; the SV area flashes, setting the temperature to 80°C. Press the SET key again; the second digit of the STEP area is 5, prompting you to enter the fifth segment of the program setting in the group. (See Figure 7)
| PROG | 1 | STEP | 5 | TIEM | 00:00 |
| TEMP | 158.0 °C | ||||
| SET | 160.0 °C | ||||
| RUN | ♦ | ||||
2.1.5 The TIME area flashes, setting the time to 0 minutes; This indicates that the program has run until it ends.
STEP sets 80°C, TIME sets to 1 minute, indicating the goal to raise the temperature from RT room temperature to 80°C as quickly as possible; (There is no requirement for heating rate, just to reach the temperature in the shortest possible time); After running for 1 minute, the program still stays in this segment because it waits for the temperature: when the controller is in the ramp heating section, the time is up. If the temperature setpoint > temperature measurement + waiting temperature appears, the controller automatically pauses until the temperature setpoint < (temperature measurement + waiting temperature) enters the next curve; When the controller is in the slope cooling section, the time is up. If the temperature setpoint < (temperature measurement - waiting temperature) occurs, the controller automatically pauses until the temperature set value > (temperature measurement - waiting temperature) before entering the next curve;
3. Program operation instructions:
For each parameter modification, you must press the "SET" key to confirm for the modification to be valid.
After setting all parameters, press the "RUN/STOP" button and wait about 3 seconds before starting operation.
In standard mode, press the "SET" key; the first position of the PROG area flashes, then press the increase or decrease key to select the desired number of groups; Users can set the required temperature and time in the PROG group according to their needs, so next time they need to operate the group, just pull it out and run it.
Appendix 6. Concept of Vacuum Degree and Unit Conversion
"Vacuum degree," as the name suggests, refers to the degree of vacuum in the area. The so-called "vacuum" refers to the state of gas in a given space where the pressure is below 101325 pascals (that is, a standard atmospheric pressure of about 101 KPa). In a vacuum, the degree of rareficity of a gas is usually expressed by its pressure value; clearly, the lower the pressure, the thinner the gas.
Common vacuum degree unit conversion tables:
| Conversion value\unit item | kilopascal (KPa) | Inches of mercury ("Hg") | millimeter of mercury (mmHg) | Standard atmospheric pressure (atm) | Shows allowable deviations |
| 1 bar | 100 | 29.53 | 750 | 0.987 | ±2.5% |
Packing list
Product Name: Vacuum Drying Oven
| Serial number | Category | Name | Unit | Quantity | Note |
| 1 | Documents | User manual | portion | 1 | |
| 2 | Documents | Packing list | portion | 1 | |
| 3 | Documents | Guarantee letter | portion | 1 | |
| 4 | Documents | Warranty card | portion | 1 | |
| 5 | Spare parts | Fuse | Only | 2 | |
| 6 | Accessories | Vacuum connecting tubes | Root | 1 | Inner diameter: Φ14mm, wall thickness: 8mm |
| 7 | Accessories | Shelf | Block | See Table 1 | |
| 8 | Accessories | Wrenches | Handle | 2 | One 8"-10" 14"-17" each |
The items listed on this order match the actual items inside the box
| Packer: 2 | Inspector: 1 |
- 1YIHENG DZF-6090/6210 Vacuum oven instruction manual
- 2YIHENG BPZ-LC/LCB series Vacuum oven instruction manual
- 3YIHENG BGZ series Vacuum oven instruction manual
- 4LINSHANG LS215 Ultrasonic Thickness Gauge Instruction Manual
- 5LINSHANG LS213 Ultrasonic Thickness Gauge Instruction Manual
- 6LINSHANG LS212 Ultrasonic Thickness Gauge Instruction Manual
- 7LINSHANG LS211 Ultrasonic Thickness Gauge Instruction Manual
- 8LINSHANG LS190 glossmeter instruction manual
- 9LINSHANG LS161 Solar Film Tester Instruction Manual
- 10LINSHANG LS163A Solar Film Tester Instruction Manual