      Home> Conductivity Meter> Bante> Compare
      Model selection comparator
      If you have any questions, please click below
      Bante Bante950-DL Laboratory conductivity /TDS/salinity/resistivity meter
      2930.0 RMB
      ChinaChina
      Detail
      FZchina DDS-609 High-Precision Module Conductivity Meter
      5100.0 RMB
      ChinaChina
      Detail
      Model comparison
      Type Benchtop Benchtop
      Range Miscellaneous Miscellaneous
      Precision 0.5 0.5
      Resolution 0.001
      Temperature compensation Automatic Manual
      PH Without
      TDS With With
      Salinity with with
      Resistivity with with
      Accessories
      Parameter comparison
      Measurement parameters
      Instrument level Level 0.5
      Measuring range 0.05µS/cm~20mS/cm
      Measurement error
      resolution 0.001, 0.01, 0.1, 1
      Stability
      Temperature compensation range 0~100℃
      Electrode constant specification
      data storage 500 groups
      Power DC5V power adapter
      Data Interface USB
      Operating Temperature
      Operating Humidity
      Operating Conditions
      Dimensions 210*188*60mm
      Weight 1.5kg
      Conductivity Detector measurement Accuracy ±0.5% F.S
      Conductivity Detector calibrated point 1 to 5 point
      Calibrating fluid 10µS/cm, 84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm
      Temperature compensation method Manual/AutomaticManual/Automatic
      temperature compensation coefficient 0.0~10.0%/℃
      Temperature compensation mode Linear or pure water
      Conductivity Detector cell constant K = 0.1, 1, 10, custom
      Standardized temperature 20 ℃, 25 ℃, optional
      TDS range 0~10ppt (Max. 20ppt) 0.0000mg/L~2000.0g/L
      TDS measurement Accuracy ±1% F.S.
      TDS conversion factor 0.1~ 1.0 (default 0.5)
      Salinity range 0~10ppt (0.00~80.00)ppt
      Salinity measurement Accuracy ±1% F.S
      Salinity measurement mode Practical salinity (psu) or seawater salinity (ppt)
      Resistivity Measurement range 0 ~ 20MZ
      Resistivity measurement Accuracy ±1% F.S
      Resistivity display resolution 0.01, 0.1, 1
      Temperature Measurement range 0~105℃
      Temperature measurement Accuracy ±0.5℃
      Temperature display resolution 0.1℃
      Temperature calibrating point 1 point
      Temperature unit ℃/℉
      data lock Manual or automatic end point locking
      Data stability condition Fast or high precision
      Calibrated reminder 1 to 31 days or closed
      shutdown mode Manual/Automatic
      One-click reset function contain
      data storage 500 groups
      Data Interface USB
      Connector 6-Pin MINI Connector
      Display Customized LCD (130 × 110mm)
      Optional accessories TP-10K temperature probe, Conductivity Detector standard liquid, CON-0.1 type Conductivity Detector electrode, CON-1 type Conductivity Detector electrode, CON-10 type Conductivity Detector electrode,
      Packaging Dimensions 300*240*170mm
      Encasement Weight 1.7kg
      Temperature compensation method Manual/AutomaticManual/Automatic