      Home> Coating Thickness Gauge> Qnix> Compare
      Model selection comparator
      If you have any questions, please click below
      Qnix QNix4200 Coating Thickness Gauge
      3740.0 RMB
      GermanyGermany
      Detail
      KETT LE-373 Film thickness meter
      In coming
      JapanJapan
      Detail
      Model comparison
      Principle Magnetic Magnetic
      Substrate Ferrous Ferrous Non-Ferrous
      Range 0 ~ 6000μm 0 ~ 2000m
      Precision ±3% ±3%
      Certificate Without Without
      Probe Integrated probe Split probe
      Application General General
      Accessories
      Parameter comparison
      Measurement principle magnetic Induction Magnetic Induction
      Measurement range 0~3000μm 0~2500μm
      margin of error ±3% <50μm±1μm ≥50μm<1000μm±2% ≥1000μm±3%
      resolution 0.1μm
      Minimum measurement surface Φ25mm
      Minimum radius of curvature Convex 5mm; Concave 25mm
      thinnest substrate 0.2mm
      probe Integrated probe Line probeTouch (LEP-J) or (LHP-J)
      statistical function Measurement times, MAX Maximum, MIN Minimum, Average, Standard deviation
      icon function
      Power Supply 2 × 1.5V AA battery 1.5V
      Standards DIN EN ISO 2808, ISO 2178, ASTM B 499, ASTM D 7091
      Measurement Certificate
      system language
      Display screen Graphic-LCD Digital LCD display (backlight can be self-selected switch)
      Interface USB、RS-232CComputer (USB) output or printer (RS-232C)
      supporting software
      Weight 105g
      Dimension 100*60*27mm

      Friendly reminder: This page is a comparison of Qnix QNix4200,KETT LE-373.NBCLab only provides you with the similarities and differences of the product in the model and specifications.

      For more information about Qnix QNix4200,KETT LE-373similarities and differences, please refer to the NBCLab product details.

      Focus on online sales of Lab Testing Instruments