Search
    Cart(0)
      总价
      前往结算
      您的购物车是空的 : )
      Home> High-Low Temperature Chamber> ZKHS> Compare
      Model selection comparator
      If you have any questions, please click below
      ZKHS GDJS-100D Temperature&Humidity Cycling Test Chamber
      56700.0 RMB
      ChinaChina
      Detail
      hzaoc GDJW-100D High-Low Temperature Cycling Test Chamber
      43700.0 RMB
      ChinaChina
      Detail
      Model comparison
      Type alternating alternating
      Min-TEMP -60℃ -60℃
      Max-TEMP 150℃ 150℃
      working Volume ≤100L ≤100L
      Parameter comparison
      Interior size 450*450*500mm 400*500*500mm
      Temperature range -60~150℃ -60~150℃
      Humidity control range 30~98%R.H
      Temperature fluctuation ≤±0.5℃ ≤±0.5℃
      Humidity Deviation +2、-3%R.H
      Heating rate -20→100℃≈35min 1.0~3.0℃/min
      Cooling rate 25~-40℃≈50min 0.7~1.0℃/min
      display (7.0 inch) large screen screen programmable touch temperature and humidity Controller
      Accuracy range Setting Accuracy: temperature ± 0.1 ℃, humidity ± 1% R.H
      Temperature transducer Platinum resistance PT100Ω/MV
      Heating system Fully independent system, nickel-chromium alloy electric heating heater
      Humidification system Built-in isolated, all stainless steel shallow surface evaporative humidifier
      Dehumidification system Adopt evaporator coil dewpoint temperature laminar flow contact dehumidification method
      Refrigeration system Original French "Taikang", fully enclosed air-cooled compression refrigeration method
      Circulatory system Temperature-resistant and low-noise air-conditioned motor, multi-leaf centrifugal wind turbine
      Enclosure Material A3 steel plate electrostatic spray treatment
      Chamber material SUS304不锈钢优质镜面钢板
      Thermal insulation material Polyurethane Rigid Foam, Ultrafine Glass Fiber Cotton
      Door frame insulation Double Layer High and Low Temperature Conditioning Silicone Rubber Door Seal
      Timer 0~9999 M, H adjustable 0~999h
      Standard configuration Multi-layer heating and defrosting with lighting Glass window 1 set, 2 test racks, 1 test lead hole φ50mm
      Safety Devices Leakage, short circuit, over temperature, water shortage, motor overheating, compressor overVoltage, overload, overcurrent protection
      Power AC220V ± 10% 50Hz AC380V? 10% 50Hz three-phase five-wire system 380V±10%
      Environment temperature 5~30℃
      Environment humidity ≤85%R.H
      Temperature Uniformity ≤+2℃ ≤±2℃

      Friendly reminder: This page is a comparison of ZKHS GDJS-100D,hzaoc GDJW-100D.NBCLab only provides you with the similarities and differences of the product in the model and specifications.

      For more information about ZKHS GDJS-100D,hzaoc GDJW-100Dsimilarities and differences, please refer to the NBCLab product details.