![]() Model selection comparator
If you have any questions, please click below
|
![]() xiangmin XM-LQ001 Leakage marking Tester
In coming
ChinaDetail
|
![]() Factory NBC-4327 Leakage marking Tester
7170.0 RMB
ChinaDetail
|
|---|
| Electrode material | Test electrode - platinum, electrode Connector - silver (imported) | ||||
|---|---|---|---|---|---|
| Electrode size | (2Mm +/- 0.1mm) × (5mm +/- 0.1mm) × (40mm +/- 5mm), platinum electrode 12mm, 30 ° +/- 2 ° inclined surface; | ||||
| Electrode distance | 4.0Mm +/- 0.01mm, included angle 60 ° +/- 5 ° | ||||
| Electrode pressure | 1.00N±0.001N | ||||
| test solution resistance | Liquid A 0.1% NH 4Cl, 3.95 +/- 0.05 Ω m, Liquid B 1.98 +/- 0.05 Ω m | ||||
| Droplet Volume | 20 drops 0.380g~ 0.480g, 50 drops 0.997g~ 1.147g (slightly adjustable) | ||||
| Droplet height | 35Mm +/- 5mm (adjustable) | ||||
| Drop time | 30s +/- 0.1s (better than standard) (digital display, can be preset and adjusted), 50 drops time 24.5min +/- 2min | ||||
| Drop count | 1~ 9999 (digital display, preset) | ||||
| test wind speed | 0.2M/s (new standard) | ||||
| test Voltage | 100V~ 1000V (25V indexing, adjustable) | 100~600V | |||
| Power Supply Voltage drop | 8% at 1.0 A +/- 0.1A | ||||
| Trace judgment | 0.50A±10% , 2.00s±10% | ||||
| Dimension | width 600 * depth 500 * height 1000mm, exhaust aperture ø100mm | ||||
| Test Power Supply | 220V 0.6kVA 50-60Hz | ||||
Friendly reminder: This page is a comparison of xiangmin XM-LQ001,Factory NBC-4327.NBCLab only provides you with the similarities and differences of the product in the model and specifications.