Search
    Cart(0)
      总价
      前往结算
      您的购物车是空的 : )
      Home> Humidity tTest Chamber> HUDA> Compare
      Model selection comparator
      If you have any questions, please click below
      HUDA QT-E702-1000K60 Programmable Humidity Chamber
      32250.0 RMB
      ChinaChina
      Detail
      kesionots KS-HW1000E(150℃) humidity chamber
      65200.0 RMB
      ChinaChina
      Detail
      Model comparison
      Min-TEMP -60℃ -60℃
      Max-TEMP >100℃ >100℃
      Min-RH ≤20%RH ≤20%RH
      Max-humidity 98% RH 98% RH
      Volume >800L >800L
      TEMP fluctuation ±0.5℃ ±0.5℃
      Humidity deviation ±2% ±2%
      Parameter comparison
      Interior Volume 1000L 1000L
      Temperature range -60~150℃ -50~150℃
      Temperature Resolution 0.01℃ ±0.1℃
      Temperature Fluctuation ≤±0.5℃
      Temperature Uniformity ±2℃ ±1℃
      Humidity control range 20~98%R.H 20~98%R.H.
      Humidity Deviation ±2%R.H
      Humidity stability ±2%R.H
      Humidity Uniformity ±3%R.H ±3.0%R.H
      Working time
      Power 380V 220V AC (+/- 10%), protective ground wire, ground resistance ≤ 4Ω;
      Interior size 1000*1000*1000mm 1000*1000*1000mm
      Dimensions 1280mm × 2080mm (including cover) × 1900mm 1500*1860*1570mm
      Shelf (std.)
      Temperature deviation ≤±2℃
      Humidity resolution 0.1%R.H ±0.1%R.H
      Temperature control Accuracy ±0.01℃ ±1.0℃
      Humidity Control Accuracy ±0.1%RH ±2.0%R.H.
      Heating rate ≧3℃/min ≈4.0℃/min
      Cooling time ≧1℃/min
      Refrigeration system 2 * 0.75HP imported French Taikang fully enclosed compressorMechanical binary refrigeration method
      refrigerant Environmentally friendly refrigerant R404/R23/freezing oil
      Refrigeration system 2 * 0.75HP imported French Taikang fully enclosed compressorMechanical binary refrigeration method
      Circulation 离心风机带强制循环
      Condenser finned tube heat exchangerStainless steel brazed plate heat exchanger
      Condensation method air cooling
      Condenser finned tube heat exchangerStainless steel brazed plate heat exchanger
      Humidification system Stainless steel humidifier
      Heating system Nickel chromium alloy stainless steel heating device
      Water supply system Front-mounted water supply tank, fully automatic water supply system
      Controller Taiwan, China Youyi Controller Korean "TEMI" "TH-900C" or Japanese "OYO" card to choose
      Setup method Chinese/English interface, touch input
      Operation mode Program method/fixed value method
      Set range Maximum Temperature range 5 ℃
      Display resolution Temperature: 0.01 ℃; Time: 0.1min
      Power-off memory function The power failure recovery mode can be set as: hot start/cold start/stop
      input Platinum resistance
      Storage function Standard 2G SD card recording test data Linear dispersion, can upload and download data, can expand the storage Capacity
      Safety Devices Overall equipment overtemperature, fan overheating protection, overall equipment underphase/reverse phase, refrigeration system overload, refrigeration unit overpressure, overall equipment Timer, leakage, operation instructions, automatic shutdown after fault alarm, etc Air burn prevention, high Voltage protection, overheating protection, overcurrent protection, Pottery and porcelain E-16 fast fuse, no fuse overload protection
      Software Operating Environment IBM PC compatible machine, CPU above PII, memory above 128M, Simplified Chinese Windows2000 or Simplified Chinese WindowsXP operating system
      Enclosure Material Cold rolled steel plate (1.0mm thick)
      Inner box material SUS304 #stainless steel plate (1.0mm thick)
      Thermal insulation material 100mm polyurethane rigid foam + 100mm high density fiberglass cotton High temperature resistant high density formic acid vinegar foam insulator material
      Weight 534kg
      Rated current 32A
      Rated power 16kw
      Power 380V 220V AC (+/- 10%), protective ground wire, ground resistance ≤ 4Ω;
      Test hole diameter 100mm
      program amount Maximum 120 groups, 1 program can be combined from 1 to 99 segments, Capacity: 1200 segments
      Communication interface 1. R232/485 communication 2. Can display Linear dispersion, data acquisition 3. Can be used as monitoring and remote Control system RS-232 standard communication interface device, which can be connected to a personal computer (PC) at the same time for multi-machine control and management

      Friendly reminder: This page is a comparison of HUDA QT-E702-1000K60,kesionots KS-HW1000E(150℃).NBCLab only provides you with the similarities and differences of the product in the model and specifications.

      For more information about HUDA QT-E702-1000K60,kesionots KS-HW1000E(150℃)similarities and differences, please refer to the NBCLab product details.

      Focus on online sales of Lab Testing Instruments