Search
    Cart(0)
      总价
      前往结算
      您的购物车是空的 : )
      Home> Thermal Shock Test Chamber> Apkjgroup> Compare
      Model selection comparator
      If you have any questions, please click below
      Apkjgroup 2AP-CJ-80A Thermal Shock Test Chamber
      103900.0 RMB
      ChinaChina
      Detail
      Factory NBC-80B Thermal Shock Test Chamber
      95000.0 RMB
      ChinaChina
      Detail
      Model comparison
      standard Cabinet Cabinet
      Cooling method Air-Cooled Air-Cooled
      Number of cabins Two-Chamber Two-Chamber
      With humidity function without without
      Volume ≤100L ≤100L
      Parameter comparison
      Temperature range -40 ℃~ + 150 ℃ (air-cooled)
      Temperature fluctuation ±0.5℃
      Temperature deviation ±2.0℃
      Heating rate + 50 ℃ → + 195 ℃ ≤ 40min; Note: The heating time is the performance of the high greenhouse running alone
      Cooling rate From room temperature~ -55 ℃ ≤ 60min
      heater U-shaped fin type 304 #seamless steel high-speed heating electric heater.
      Temperature control method
      power
      Weight About 550kg
      Inner Volume 80L
      Liner size 400*500*400mm
      Overall Dimensions 1400*1950*1850mm
      Low Temperature Impact Range -10 ℃~ -40 ℃ (optional)RT~-55°C
      High Temperature Impact Range 60 ℃~ 150 ℃ (arbitrarily adjustable)RT~+195°C
      Test environmental conditions The temperature of the environment around the machine is maintained between + 25~ + 30 ℃, Relative Humidity: ≤ 85%; air pressure: 86kPa~ 106kPa normal atmospheric pressure.
      High Temperature Impact Range 60 ℃~ 150 ℃ (arbitrarily adjustable)RT~+195°C
      Heating rate + 50 ℃ → + 195 ℃ ≤ 40min; Note: The heating time is the performance of the high greenhouse running alone
      Low Temperature Impact Range -10 ℃~ -40 ℃ (optional)RT~-55°C
      Cooling time + 20 ℃ → -55 ℃ ≤ about 90min; Note: The cooling time is the performance of the low greenhouse running alone
      Test method Place the product in the gondola test area, two boxes of mobile heat & cold shock, safe air pressure to drive the test object up and down Impact.
      Temperature recovery time ≤ 5min; the conversion temperature only needs ≤ 10s (the movement time of the hanging basket: within 3 seconds)
      Impact exposure time ≥30min
      Outer wall material The outer material can be selected from SUS304 high-quality stainless steel or fine powder paint treatment
      Liner material The inner plate material is high-quality SUS304 stainless steel, with good corrosion resistance, heat resistance, low temperature strength and mechanical properties; stamping, bending and other hot workability, no heat treatment hardening phenomenon (no magnetic, operating temperature -196 ℃~ 800 ℃)
      Box insulation material Rigid polyurethane foam
      Air conditioning channel Fan, heater, evaporator, damper, temperature transducer
      Specimen shelf 1 set, a total of two layers (with four hooks) bearing 25KG (please indicate when placing an order for larger bearing)
      Lead Hole (Optional) 1, located on the left side of the box, with a round hole with a size of φ50mm
      Test chamber door Single open, left hinge, right handle
      Circulating Fan High temperature/ring temperature exposure: 3φ, 400W, centrifugal; low temperature exposure: 3φ, 400W, centrifugal
      Control Panel Controller display, fault indicator, timer, USB or R232 interface, over-temperature protection setter, main power switch
      Machinery room Refrigeration unit, connected to drainage device, exhaust fan
      Distribution control cabinet Total power leakage circuit breaker, Controller, switchboard, cooling fan, high and low temperature Circulating Fan motor
      Heating system Nickel-chromium alloy heating wire heater
      Heater control method No touchpoint equal period pulse width modulation, SSR (solid state relay)
      Power cord holes and drain holes On the back of the box
      Refrigeration system France Taikang or Japan Hitachi fully enclosed low noise rotary compressor
      Evaporator finned tube heat exchanger
      Condenser Shell and tube heat exchanger
      Throttle device Thermal expansion valve, capillary
      Evaporative condenser Stainless steel brazed plate heat exchanger
      Refrigerator control method The PLC (programmable logic Controller) of the Control system automatically adjusts the operating conditions of the refrigerator, the compressor return air chill down circuit, the energy adjustment circuit, and the condensing pressure control valve according to the test conditions
      refrigerant R404a/R23 (ozone depletion index is 0)
      Controller model Programmable Controller, touch screen input, different high and low temperature Impact temperature points can be arbitrarily set according to customer requirements to meet different needs for testing
      display LCD liquid crystal display
      Operation mode procedural method
      Setup method All Chinese liquid crystal display (Chinese and English can be switched) LCD touch panel (resistant to 100.000 line drawing test)
      Program Capacity Up to 120 groups, each program is divided into three areas, three areas and 3 segments. Each command can reach 999 times; the program segment time is set to 0~999Hour59Min.
      Display resolution Temperature: 0.1 ℃; Time: 0.1min
      Electrical input thermocouple
      storage device USB and R232 communication interface
      Control method Anti-integral saturation PID, fuzzy algorithm; BTC equilibrate temperature control method
      Curve recording Real-time display of graphics during program execution Linear dispersion
      subsidiary function With power-off program memory, it automatically starts and continues to execute program functions after power-back, with date, time adjustment, reservation start, shutdown function and screen lock (LOCK) function.
      Environmental conditions Temperature: 5 ℃~ 35 ℃; Relative Humidity: ≤ 85%; air pressure: 86kPa~ 106kPa normal atmospheric pressure.
      Air source pressure ( 5 ~ 7 ) kg / cm2
      Air source connecting pipe diameter fo8mm
      Air source flow 0.1m3/min
      Maximum current consumption Current: 18A; Power: about 12.0KW [Voltage range: no more than +/- 10%]
      Storage environment When the equipment is not working, the temperature of the environment should be kept within 0 ℃~ + 45 ℃; when the environment temperature is lower than 0 ℃, the water remaining in the equipment should be discharged cleanly to prevent the water in the pipeline from freezing and damaging the pipeline.

      Friendly reminder: This page is a comparison of Apkjgroup 2AP-CJ-80A,Factory NBC-80B.NBCLab only provides you with the similarities and differences of the product in the model and specifications.

      For more information about Apkjgroup 2AP-CJ-80A,Factory NBC-80Bsimilarities and differences, please refer to the NBCLab product details.

      Focus on online sales of Lab Testing Instruments