Search
    Cart(0)
      总价
      前往结算
      您的购物车是空的 : )
      Home> Gas detector> EYESKY> Compare
      Model selection comparator
      If you have any questions, please click below
      EYESKY ES10B11-O2 Fixed micro oxygen Detector
      In coming
      ChinaChina
      Detail
      Bosean BH-90M-O2 Gas Detector
      648.0 RMB
      ChinaChina
      Detail
      Model comparison
      Type Portable Portable
      Number of Gases Single gas Single gas
      Sensors
      Gases O2 O2
      Sampling Mode Diffusion Diffusion
      Parameter comparison
      Detect gas Oxygen O2
      Sensors principle principle of zirconia
      Sampling method Diffusion, circulation
      O2-range 0~5000PPm
      O2- resolution 0.01PPm
      O2- detection Accuracy ±3%F.S.
      Operation mode Infrared remote control for remote and close operation, easy to set up
      Repeatability ≤±1%
      Zero drift ≤ +/- 1% (F. S/year)
      Response time T90≤20s T90<60s
      recovery time ≤20s
      Explosion-proof sign Exd Ⅱ CT6 ExibIICT4Gb
      Protection level IP66
      Output signal Users can according to the actual site environment requirements, up to 2000 meters (single core 1mm2 Shielding cable) three-wire system 4-20mA current signal output, can be connected to the alarm Controller, PLC, DCS and other Control system (standard configuration)
      main material Shell: ADC12 aluminum alloy, strong, wear-resistant and corrosion-resistant (optional 316 stainless steel shell) Air chamber: high-strength wear-resistant and corrosion-resistant Aluminum profile, durable (optional 316 stainless steel air chamber)
      Installation method Wall-mounted, pipeline type
      Operating temperature -30~60℃
      Environmental humidity 0~95% RH non-condensing10%~95%RH
      Environmental humidity 0~95% RH non-condensing10%~95%RH
      Power Supply 24VDC (Normal Operating Voltage range 12~ 30VDC)
      power consumption 1.5W
      Operating Pressure ≤200Kpa
      Dimension 180*140*92mm
      Weight About 1.6Kg ≤ 172G (without portable accessories)
      execution standard GB15322.1-2003,GB3836.1-2010,GB3836.2-2010,GB3836.4-2010

      Friendly reminder: This page is a comparison of EYESKY ES10B11-O2,Bosean BH-90M-O2.NBCLab only provides you with the similarities and differences of the product in the model and specifications.

      For more information about EYESKY ES10B11-O2,Bosean BH-90M-O2similarities and differences, please refer to the NBCLab product details.

      Focus on online sales of Lab Testing Instruments