JJF 1187-2008 specifies a calibration method for Thermal Imaging Camera with temperature measurement for equipment with a Temperature range from -20 ° C to 2000 ° C. Thermal Imaging Camera enables accurate measurement of the surface temperature of an object and its distribution by converting the thermal radiation from the surface of the object into a visible image. The specification relates to the error of indication for Thermal Imaging Camera, i.e. the difference between the instrument temperature indication and the true temperature of the blackbody radiation source. The specification also specifies corrective factors to be considered in calibrating, including emissivity, target distance, Environmental temperature and atmospheric effects. The specification cites relevant literature and requires the use of currently valid versions of the literature to ensure the accuracy and conformity of the calibrated.
| Status | Active | ||
|---|---|---|---|
| CCS | ICS | ||
| Release Date | 2008-01-31 00:00:00 | Implementation Date | 2008-04-30 00:00:00 |