Bante901P-UK Portable multiparameter water quality measurement meter pH/Conductivity Meter
Measure pH/mV, conductivity and total dissolved solids (TDS) values, with a large backlit LCD screen, 1 to 5 point button calibrated.
Bante-
HANNA HI76483 Built-in temperature sensor dissolved Oxygen electrode
$ 836.00 -
JINGQI WD8216 Silver Tester
$ 481.00 -
HANNA HI97105 multiparameter photometer for seawater
$ 652.00 -
BANTE BI-680 industrial online Dissolved Oxygen Meter aquaculture water quality monitoring equipment
$ 455.00 -
TONGAO TR-1000H Water Quality Suspended Solids Tester SS Tester
$ 786.00
Bante901P Portable pH/Conductivity Meter
Multi-parameter portable pH meter, including 4 measurement functions, the meter is suitable for outdoor measurement of pH/mV, conductivity and total dissolved solids (TDS) value of liquids.
pH Mode Features
• Multi-parameter Water Quality Analyzer with large backlit LCD display.
• 1 to 5 point key calibration, automatic identification of pH standard buffer.
• Optional pH buffer standard (USA/NIST/DIN) or 2-point custom calibration.
• Automatic electrode slope display, prompting the user whether to replace the pH electrode.

Portable pH Meter
Conductivity/TDS mode functional characteristics
• 1 to 5 point key calibration, automatic identification of conductivity standard solution.
• Optional cell constant (K=0.1/1/10), TDS conversion factor, normalized temperature (20 or 25°C), linear and pure water temperature compensation modes.
common features
• Automatic temperature compensation to ensure accurate measurement over the full range.
• Automatic Calibration Reminder to remind users to calibrate the meter on a regular basis.
• Automatic end point locking, retaining stable measurement values for easy viewing and recording.

Portable pH Meter
• Automatic electrode diagnosis, you can check the zero offset, electrode slope and conductivity K value.
• The system menu can customize 12 functional parameters such as the number of calibration points, stability standard, data lock mode, and automatic shutdown.
• One key reset (Reset) automatically restores the instrument to the factory default settings.
• Expanded memory can store up to 500 sets of measurement data.
• The built-in clock automatically marks the date and time of stored data to comply with GLP regulations.
• USB communication interface to send stored data to PC.
• Multi-mode power supply scheme (battery, power adapter, computer's USB port) ensures the smooth application of the instrument.
Bante Bante901P-UK Portable Multi-Parameter Water Quality MeterSpecifications
| LIST | VALUE |
|---|---|
| pH Measurement Range | -2.000~20.000pH |
| pH measurement Accuracy | ±0.002pH |
| pH resolution | 0.1, 0.01, 0.001 pH, optional |
| pH calibrating point | 1 to 5 points |
| pH calibrating solution | USA, NIST, DIN, Custom |
| pH temperature compensation range | 0~100℃ |
| pH temperature compensation | Manual/Automatic |
| ORP Measurement Range | ±1999.9mV |
| mV measurement Accuracy | ±0.2mV |
| mV resolution | 0.1/1 mV, optional |
| Conductivity Measurement Range | 0~20mS/cm (max 200mS/cm) |
| EC measurement Accuracy | ±0.5%FS |
| EC resolution | 0.001, 0.01, 0.1, 1 |
| Conductivity calibrating point | 1 to 5 points |
| Conductivity calibrating fluid | 10µS/cm, 84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm |
| EC Temperature Compensation Range | 0~100℃, 32~212°F |
| EC Temperature Compensation Mode | Manual or automatic |
| EC Temperature Compensation Coefficient | 0.0~10.0%/℃ |
| EC Temperature Compensation Mode mode | Linear or pure water |
| Conductivity Detector cell constant | K = 0.1, 1, 10, custom |
| Standardized temperature | 20 ℃, 25 ℃, optional |
| TDS range | 0~10ppt (max 200ppt) |
| TDS measurement Accuracy | ±1%FS |
| TDS conversion factor | 0.1~ 1.0 (default 0.5) |
| Temperature Measurement Range | 0~105℃, 32~221℉ |
| Temperature measurement Accuracy | ±0.5℃ |
| Temperature Resolution | 0.1℃ |
| Temperature calibrating point | 1 point |
| Temperature unit | ° C, ° F, optional |
| Data stability condition | Fast or high Accuracy |
| Calibrated reminder | 1 to 31 days or closed |
| shutdown mode | Manual or automatic (10, 20, 30 minutes after no button operation) |
| data storage | 500 sets |
| Communication interface | USB |
| Connector | BNC, 6-pin MINI Connector |
| Power Supply | 3 AA batteries |
| Battery life | Approximately 150 hours (when the LCD backlight is turned off) |
| Dimensions | 170*85*30mm |
| instrument Weight | 300g |
| optional accessories | P11 type pH electrode, P11-LiCl type pH electrode, P11-HA type pH electrode, P11-NA type pH electrode |
| optional accessories | P12 type pH electrode, P13 type pH electrode, P15 type pH electrode, P16 type pH electrode, P18 type pH electrode, P19 type pH electrode, P21 type pH electrode, P22 type pH electrode, 501 type ORP electrode, 502 type ORP electrode |
| optional accessories | Conductivity Detector electrode of CON-0.1 type, Conductivity Detector electrode of CON-1 type, Conductivity Detector electrode of CON-10 type |
| optional accessories | TP-10K temperature probe, Conductivity Detector standard solution, pH standard buffer reagent (USA standard), pH standard buffer reagent (NIST standard) |
| Packaging Dimensions | 340*290*80mm |
| Encasement Weight | 1.5kg |
Bante Bante901P-UK Portable Multi-Parameter Water Quality Meter Packing list
Host X1, P11 type pH electrode (made in the UK) X1, K10 type Conductivity detector electrode (made in the UK) X1, TP-10K type temperature probe X1, pH standard buffer reagent (pH4.01/7.00/10.01) X1, Conductivity detector standard solution (84uS/cm, 1413uS/cm, 12.88mS/cm) X1, AA type battery X1, plastic carrying case X1, manual X1, certificate X1, warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveGB/T 693-2024
- ActiveGB/T 8884-2017
- ActiveJJF 1565-2016
- ActiveJJG 2059-2014
- Active 126-2009
- ActiveHJ 1147-2020
- Active 271-2018
- ActiveSL 144.2-2008
- ActiveGB/T 17650.2-2021
- ActiveGB/T 41079.2-2022
- ActiveGB/T 6908-2018
- ActiveGB/T 7977-2007
- ActiveGB/T 32697-2016
- ActiveISO 21318-2007
- ActiveISO 9944-1990
- ActiveHJ 2515-2012