      Home> STIP-scan> Bante> Bante Bante904P-UK Portable Multi-Parameter Water Quality Meter

      Bante904P-UK portable multiparameter Water Quality Meter conductivity dissolved oxygen

      Import electrode, while testing conductivity and dissolved oxygen, temperature, salinity, TDS, automatic temperature compensation, conductivity Measurement range 0~20mS/cm, dissolved oxygen range 0.00~20.00mg/L.

      Brand
      SKU: NB014748
      China Bante
      Stock
      Loading…
      Price
      $ 607.00
      Promotion
      Service
        Service
        The delivery and service By NBchao.Com
        VS
        Genuinegoods
        • Bante Bante904P-UK Portable Multi-Parameter Water Quality MeterOverview

        • Specifications
        • Comments
        • FAQs
        Bante904P-UK
        Portable Multi-Parameter Water Quality Meter
        • Bante904P-UK portable multiparameter Water Quality Meter conductivity dissolved oxygen
        • Bante904P-UK portable multiparameter Water Quality Meter conductivity dissolved oxygen Figure 1
        Import electrode, while testing conductivity and dissolved oxygen, temperature, salinity, TDS, automatic temperature compensation, conductivity Measurement range 0~20mS/cm, dissolved oxygen range 0.00~20.00mg/L.

        Bante904P Portable Conductivity/Dissolved Oxygen Meter

        Multi-parameter portable conductivity meter, including 5 measurement functions, the meter is suitable for measuring liquid conductivity, total dissolved solids (TDS), salinity, resistivity, dissolved oxygen concentration and saturation.

        Conductivity/TDS/salinity/resistivity mode features

        • Multi-parameter Water Quality Analyzer with large backlit LCD display.

        • 1 to 5 point key calibration, automatic identification of conductivity standard solution.

        • Selectable cell constant (K=0.1/1/10), TDS conversion factor, normalized temperature (20 or 25°C), linear and pure water temperature compensation, seawater and practical salinity measurement modes.

        conductivity meter


        Dissolved Oxygen Mode Features

        • 1 or 2 point pushbutton calibration using air saturated water or zero oxygen liquid.

        • Compensation for salinity and atmospheric pressure to improve the measurement accuracy of the instrument.

        common features

        • Automatic temperature compensation to ensure accurate measurement over the full range.

        • Automatic Calibration Reminder to remind users to calibrate the meter on a regular basis.

        • Automatic end point locking, retaining stable measurement values for easy viewing and recording.

        • Automatic electrode diagnosis, you can check the electrode slope value (K).

        conductivity meter


        • The system menu can customize 13 functional parameters such as the number of calibration points, stability standard, data lock mode, and automatic shutdown.

        • One key reset (Reset) automatically restores the instrument to the factory default settings.

        • Expanded memory can store up to 500 sets of measurement data.

        • The built-in clock automatically marks the date and time of stored data to comply with GLP regulations.

        • USB communication interface to send stored data to PC.

        • Multi-mode power supply scheme (battery, power adapter, computer's USB port) ensures the smooth application of the instrument.

        VS

        Bante Bante904P-UK Portable Multi-Parameter Water Quality MeterSpecifications

        LIST VALUE
        Conductivity range 0~20mS/cm (max. 200mS/cm)
        Measurement Accuracy of conductivity ±0.5%FS
        EC resolution 0.001, 0.01, 0.1, 1
        Conductivity calibrated point 1 to 5 point
        Conductivity calibrating fluid 10µS/cm, 84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm
        EC Temperature Compensation Range 0~100℃, 32~212°F
        EC Temperature Compensation Mode Manual or automatic
        EC Temperature Compensation Coefficient 0.0~10.0%/℃
        EC Temperature Compensation Mode mode Linear or pure water
        Conductivity Detector cell constant K = 0.1, 1, 10, custom
        Standardized temperature 20 ℃, 25 ℃, optional
        TDS range 0~10ppt (max 200ppt)
        TDS Measurement Accuracy ±1%FS
        TDS conversion factor 0.1~ 1.0 (default 0.5)
        Salinity range 0~10ppt (max 80ppt)
        Measurement Accuracy of salinity ±1%FS
        Measurement mode of salinity Utility salinity (psu) or seawater salinity (ppt)
        Resistivity range 0~10MΩ (up to 20MΩ)
        Measurement Accuracy of resistivity ±1%FS
        Resistivity resolution 0.01, 0.1, 1
        DO range 0.00~20.00mg/L
        DO Measurement Accuracy ±0.2mg/L
        DOS resolution 0.1, 0.01 mg/L, optional
        DO temperature compensation range 0~50℃
        DO temperature compensation automatic
        Pressure compensation range 60.0~112.5kPa, 450~850mmHg
        Salinity compensation range 0~50g/L
        DO% Measurement range 0.0~200.0%
        DO% Measurement Accuracy ±2.0%
        DO% resolution 0.1, 1%
        Measurement range 0~105℃, 32~221℉
        Measurement Accuracy ±0.5℃
        Temperature Resolution 0.1℃
        Temperature calibrating point 1 point
        Temperature unit ℃, ° F, optional
        Data stability condition Fast or high-precision
        Calibrated reminder 1 to 31 days or closed
        shutdown mode Manual or automatic (10, 20, 30 minutes after no button operation)
        Data storage 500 sets
        Communication interface USB
        Connector 6-Pin MINI Connector
        Power Supply 3 AA batteries
        Battery life Approximately 150 hours (when the LCD backlight is turned off)
        Dimension 170*85*30mm
        instrument Weight 300g
        Optional Accessories TP-10K temperature probe, dissolved Oxygen electrode film cap, dissolved Oxygen electrode filling solution, Conductivity Detector standard solution, CON-0.1 type Conductivity Detector electrode, CON-1 type Conductivity Detector electrode, CON-10 type Conductivity Detector electrode, DO100 type dissolved Oxygen electrode,
        Packaging Dimensions 430*310*100mm
        Encasement Weight 2kg

        Bante Bante904P-UK Portable Multi-Parameter Water Quality Meter Packing list

        Host X1, Con-1 type Conductivity detector electrode X1, DO100 type dissolved Oxygen electrode (made in the United States) X1, TP-10K type temperature probe X1, Conductivity detector standard solution (84uS/cm, 1413uS/cm, 12.88mS/cm) X1, dissolved Oxygen electrode filling solution X1, dissolved Oxygen electrode film cap X1, AA type battery X1, plastic carrying case X1, manual X1, certificate X1, warranty card X1

        [Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.

        FAQ

        [Q]
        Is the multi-function STIP-scan portable for outdoor testing?
        [A]
        Yes, this is the Bante904P.
        [Q]
        Is the water quality tester equipped with a temperature probe?
        [A]
        Yes, it comes standard.
        Bante