Qiwei BND-4T Portable Coated -4 Viscometer Pure Copper/4mm
Qiwei BND-4T portable coating-4 Viscometer second measuring cup outflow time 0~150s, flow hole diameter 4mm, capacity 100ml, cup body material pure copper, applicable standard GB/T 1723, correction coefficient K = 1 ± 0.03.
QiWei-
TRUIT TR 1004/6 French standard cup, aperture φ6.0mm
$ 126.00 -
JINGKELIAN QND-4A Viscometer (Bakelite Cup) is suitable for measuring viscosity of Coating products under 150 seconds
$ 118.00 -
TRUIT TR 1005/8 Benchtop DIN Cup, T Cup, Outflow Aperture 8mm
$ 126.00 -
TQC VF2000 DIN cup, aluminum alloy Viscosity Cup 2mm
$ 408.00 -
TQC VF2185 Handheld ISO Viscosity Cup aluminum alloy/5mm
$ 415.00
Principle and Calculation
The viscosity measured by the Tu-4 viscometer is the conditional viscosity, which is the time for a certain amount of sample to flow out of a hole with a specified aperture at a certain temperature, expressed in seconds. Convert the obtained sample outflow time (s) into kinematic viscosity value (mm²/s) with the following formula:
When t<23s, t=0.154μ+11
When 23s≤t≤150s, t=0.223μ+6.0
In the formula:
t——outflow time, s;
μ—kinematic viscosity, mm²/s.
main technical characteristics
1. The Tu-4 viscometer is made of brass, with a capacity of about 100ml. Its geometric dimensions and inner wall roughness meet the requirements of "GB/T 1723 Method for Determination of Viscosity of Paint". The bottom is a stainless steel outflow hole.
2. Correction coefficient K: Multiply the viscosity measured by the calibrated viscometer by the K value, which is the conditional viscosity measured by the viscometer. The correction coefficient K should be within the range of 1±0.03, otherwise the viscometer should be replaced. The viscometer should be calibrated periodically according to the frequency of its use.
3. The difference between the measured value and the calculated value of the standard viscosity liquid with known kinematic viscosity flowing out of the Tu-4 viscometer should be within ±3% of the calculated value.
QiWei BND - 4T Tu -4 ViscoMeterSpecifications
| LIST | VALUE |
|---|---|
| outflow time | 0~150s |
| Flow hole diameter | 4mm |
| Capacity | 100ml |
| Cup body material | pure copper |
| Applicable standard | GB/T 1723 |
| Correction factor | K=1±0.03 |
QiWei BND - 4T Tu -4 ViscoMeter Packing list
Viscosity cup × 1, certificate × 1, manual × 1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- IncomingGB/T 17001.3-2025
- Active
- ActiveJB/T 3078-2015
- ReplacedASTM
- ActiveGB/T 1727-2021
- ActiveGB/T 17001.6-2022
- ActiveHG/T 5187-2017
- ActiveHG/T 3369-2003
- ActiveHG/T 3349-2003
- ActiveCY/T 205-2019
- ActiveGB/T 13217.4-2020
- ActiveJJG 743-2018