TONGAO TR 6900Z Desktop heavy metals STIP-scan
Detection of heavy metals content, including total chromium, hexavalent chromium, copper, zinc, nickel, iron and other parameters, built-in thermal printer, using digestion Colorimetric integrated tube Colorimetric.
TongAo-
HANNA HI93720-01 Photometer Reagent, Calcium Hardness 100 times
$ 189.00 -
JINGQI WD8802 residual chlorine analyzer
$ 481.00 -
HANNA HI97753 chlorides [CI] Portable Photometer
$ 540.00 -
HANNA HI7032L TDS calibrated solution 1382ppm (mg/L)/500mL
$ 117.00 -
BANTE 502 ORP electrode for measurement of weak Oxidation-reduction potential
$ 158.00
Introduction
The all-in-one heavy metal Detector can quickly measure total chromium, hexavalent chromium, copper, nickel, zinc, iron, etc. in water. It integrates mature technologies in the field of water quality analysis, making the entire analysis system more complete, and the operation is simpler and faster. More stable and reliable. Widely used in emergency monitoring, industrial wastewater, domestic (CN) sewage, sewage treatment, pharmaceutical and chemical industry, food, beverage, textile printing and dyeing, electroplating, papermaking, petroleum metallurgy, river management, scientific research units, colleges and universities and other fields.
operating principle
The integrated digestion colorimetric tube and the multi-channel water quality rapid analyzer can be used seamlessly, and are compatible with φ25mm colorimetric tubes.
● The detection of each parameter conforms to national standards, industry standards or international standards.
● Built-in two COD detection methods
Method 1: HJ/T399-2007 (national standard)
Method 2: Import detection method
● The standard detection parameters are as follows:
- COD (premade reagents)
- COD (powder pack reagent or custom reagent)
- Ammonia nitrogen (NH3-N)
- Total phosphorus (TP)
- Total Nitrogen ( TN )
- heavy metal
● Strong scalability in the later stage, including but not limited to the following parameters:
- Phosphate - Manganese method COD
- Nitrite - Nitrate Nitrogen
- Free/Total Chlorine - Fluoride
…
Principle
1. Digestion colorimetric integrated tube
The digestion and colorimetric integrated tube is used for operation, and the digestion of the sample and the reading of the result share a reaction tube, and there is no need to transfer the liquid to the cuvette for detection.
2. Display the result directly
A cold light source with a specific wavelength is used to measure the absorbance of the sample, and the instrument system converts the detection result directly into the sample concentration value through a specific algorithm.
3. Measure multiple times and remove outliers
Measure the water sample 10 times or more, and take the average value after removing outliers, so as to reduce the influence of dirt, scratches, fingerprints or defective glass tubes on the results.
4. Detection of abnormal prompts
Prompts for abnormal detection, such as prompts for blank comparison, prompts for results exceeding the upper limit of the range, etc.
TongAo TR 6900Z Desktop STIP-scanSpecifications
| LIST | VALUE |
|---|---|
| measurement error | ≤± 5% |
| Repeatability | ≤±5% |
| Optical inspection Stability | ≤ +/- 0.001A/20min (100,000 hours life) |
| Data storage | 30,000 records (including item, date, time, sample ID, etc.) |
| Linear dispersion quantity | 96 Linear dispersions (expandable) |
| printer | Built-in thermal printer |
| Display | 5-Inch high-definition color LCD |
| batch Volume | 16/25 water samples/batch |
| Colorimetric method | Digestion Colorimetric Tube |
| Sample compatibility | 16 mm (round), 25mm (round) |
| Communication interface | USB cable, LAN, etc |
| exception notification | Blank comparison prompt, overrange prompt, etc |
| Operating interface | Membrane push-button operation, Chinese/English (optional) |
| Power Supply | AC(220V±10%),50Hz |
| Power | 5W |
| Protection level | IP30 |
| Relative Humidity | ≤ 85% non-condensing |
| Ambient Temperature | 5~40℃ |
| Host size | 348*250*145mm |
| Instrument Weight | Main unit less than 2.6kg |
| Total chromium detection method | GB7466-87 |
| Measurement range of total chromium | 0~5mg/L |
| Total chromium detection lower limit | 0.01mg/L |
| Total chromium digestion environment | 125℃,30min |
| Hexavalent chromium detection method | GB7467-87 |
| Measurement range of hexavalent chromium | 0~5mg/L |
| Lower limit of hexavalent chromium detection | 0.01mg/L |
| Hexavalent chromium digestion environment | No need to digest |
| Copper detection method | Diacetaldehyde oxalyl dihydrazone |
| Measurement range | 0~50mg/L |
| Lower limit of copper detection | 0.01mg/L |
| Copper digestion environment | No need to digest |
| Nickel detection method | GB11910-89 |
| Measurement range | 0~4mg/L |
| Lower limit of nickel detection | 0.01mg/L |
| Nickel digestion environment | No need to digest |
| Zinc detection method | Xylenol Orange Spectrophotometry |
| Measurement range | 0~15mg/L |
| Zinc detection lower limit | 0.01 mg/L |
| Zinc digestion environment | 无需消解 |
| Iron detection method | HJ/T345-2007 |
| Measurement range | 0~50 mg/L |
| Lower limit of iron detection | 0.01mg/L |
| Iron digestion environment | No need to digest |
TongAo TR 6900Z Desktop STIP-scan Packing list
Determination host X1, digestion instrument X1, digestion shield X1, determination reagent X1, standard solution X1, digestion Colorimetric tube X1 set, Colorimetric tube frame X1, Colorimetric tube cleaning cloth X1, host power adapter X1, USB data cable X1, simple operation flow chart X1, instruction manual X1, certificate of conformity/warranty card X1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- ActiveGB 45671-2025
- ActiveGB/T 693-2024
- ActiveGB/T 16311-2024
- ActiveJJF 1565-2016
- ActiveGB/T 9735-2008
- AbolishGB/T 22930-2008
- ReplacedGB/T 23950-2009
- AbolishGB/T 6276.9-2010
- ActiveGB/T 30810-2014
- ActiveGB/T 30799-2014
- ActiveGB/T 40493-2021
- ActiveGB/T 41058-2021
- ActiveGB/T 17593.3-2006
- ActiveGB/T 36021-2018
- AbolishGB 20814-2006
- ActiveGB 20814-2014
- ActiveGB/T 5750.3-2023
- ActiveGB/T 41837-2022
- ActiveGB/T 32448-2015
- ActiveGB/T 33328-2016


TR 6900Z



![HANNA HI97753 chlorides [CI] Portable Photometer](http://img.nbchao.com/imgs/200/2024/03/29/1357160732693843.jpg)