| maximum test force |
20kN |
| Accuracy class |
Level 0.5 |
| Measurement range |
0.4%~100%F.S |
| Test error of indication |
+/- 0.5% of the indication |
| Test force resolution |
500,000 code, the whole process is not divided into gears, and the whole process resolution remains unchanged |
| effective test width |
400mm |
| Cross beam travel (without fixtures installed) |
1150mm |
| Displacement error of indication |
+/- 0.5% of the indication |
| Displacement resolution |
0.04μm |
| Measurement range |
0.4%~100%F.S; |
| Deformation indication relative error |
+/- 0.5% of the indication |
| Stress control rate range |
0.005~5%FS/s |
| Stress control rate Accuracy |
When the rate is < 0.05% FS/s, it is within +/- 1% of the set value, and when the rate is ≥ 0.05% FS/s, it is within +/- 0.5% of the set value |
| Strain control rate range |
0.005~5%FS/s |
| Strain control rate Accuracy |
When the rate is < 0.05% FS/s, it is within +/- 1% of the set value, and when the rate is ≥ 0.05% FS/s, it is within +/- 0.5% of the set value |
| Displacement control rate range |
0.001~500mm/min |
| Displacement control rate Accuracy |
+/- 0.5% of the set value |
| Host size (width × depth × height) |
750*540*1750mm |
| Power Supply |
Single Phase, 220V +/- 10%, 50Hz; Power 1kW |
| Host Weight |
About 170kg |