KOSON KS-90 integrated floor Thickness Gauge, thickness measurement range 20~999mm
Test accuracy (mm) 20-200: +/- 1; 201-400: +/- 2; 401-600: +/- 3; 60 1 - 999 :≥ +/- 6, using the unique design of receiving probe and host in one, can be used for cast-in-place floor, wall, beam, column, wood and ceramic and other non-ferromagnetic thickness measurement.
KOSON-
LIULING CH-20-AT Percent Benchtop Rubber Thickness Gauge 0mm~ 20mm
$ 296.00 -
LIULING CH-1-NT Pressure sensitive adhesive tape desktop thickness gauge 0mm~ 1mm
$ 257.00 -
LIULING CH-1-B Latex Thickness Gauge 0~ 1mm
$ 242.00 -
LIULING CH-10-AT Percent Desktop Rubber Thickness Gauge 0~ 10mm
$ 211.00 -
WBT-02 Bottle Wall Film thickness Tester, digital display, bottle, box
$ 637.00
product description
KOSON KS-90 integrated floor Thickness Gauge is mainly used for non-destructive testing of concrete structures or thickness measurement of other non-ferromagnetic media, such as: measuring the non-ferromagnetic thickness of cast-in-place floors, walls, beams, columns, wood and ceramics Measurement: Adopt the unique design method of combining the receiving probe and the host into one, only need to test the transmitting probe on the corresponding surface (floor, etc.) to measure the non-ferromagnetic thickness value.
According to the standard
"Code for Construction Quality Acceptance of Concrete Structure Engineering"………………………… GB50204-2002
"Reinforcement Cover Thickness Measuring Instrument, Floor Thickness Measuring Instrument Measurement and Calibration Specifications"……………JJF1224-2009
Features
Adopt fast real-time positioning, display the position of the transmitting probe in real time, and realize fast positioning through direction indication;
Real-time display of measured values is used to automatically lock the true thickness;
2.8-inch high-resolution color LCD, the detection data can be clearly seen in any situation indoors and outdoors;
One-handed machine and operation design, easy to operate and quickly and automatically find the position of the transmitting probe;
Using USB mode data transmission, you can quickly upload the stored data to the computer through the USB cable;
Using functions such as data storage, viewing, and deletion in the data management mode, it can store more than 200,000 measuring point data;
It adopts the integrated design of the host and the receiving probe, which is small in size, light in weight, and easy to carry;
Four-wheel positioning method is adopted, and the outer wheel is equipped with high-strength wear-resistant tires, which is convenient for protecting the instrument from wear;
Equipped with professional and uniquely designed standard calibration test blocks;
The standard telescopic pole adopts professional and unique double 90-degree universal joints, which can be adjusted within 360°.
Pictures

KOSON KS-90 Integrated floor Thickness GaugeSpecifications
| LIST | VALUE |
|---|---|
| test mode | Ordinary thickness detection; compensated thickness detection |
| Display resolution | 320*240 |
| data transmission mode | USB |
| Host Weight | 0.6kg |
| Total Mass | 5.6Kg (Encasement total Mass) |
| Test Accuracy (mm) | 20-200:±1;201-400:±2;401-600:±3;601-999:≥±6 |
| Power Supply time | Host: not less than 26H; Probe: not less than 64H |
| Host size | 215×95×105mm |
| Telescopic rod size | 28×470mm |
KOSON KS-90 Integrated floor Thickness Gauge Packing list
Host x1, launch probe X1, walkie-talkie X1, telescopic rod X1, charger X1, USB data cable X1, software CD X1, instrument case X1, crayon X1, manual x1, warranty card x1, certificate x1
[Note] Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
- Thickness gauge selection guide: comprehensive analysis of thickness measurement tools from type to application
- Introduction of nondestructive thickness testing principle and instrument
- Correct use of Film Thickness Gauge
- Application of Ultrasonic Thickness Measurement Technology- "Selection of Thickness Gauges"
- Ultrasonic Instrument FAQ
- Small coating thickness test which coating Thickness Gauge measurement is more accurate?
- Why you need a calibrated coating Thickness Gauge before testing
- Principle of Magnetic Separation Thickness Gauge
- Measuring methods of coating Thickness Gauge
- Choice of Coatings
- ActiveJJF 1488-2014
- ActiveJJG (轻工)50.3-2000
- ActiveJJG (轻工)50.2-2000
- ActiveASTM F2251-2023
- ActiveQB/T 1055-2004
- ActiveGB/T 7125-2014
- AbolishHG/T 2041-1991


KS-90